Talos-C(HA) Cervical Intervertebral Body Fusion Devices

K142345 · Meditech Spine, LLC · ODP · Feb 22, 2015 · Orthopedic

Device Facts

Record IDK142345
Device NameTalos-C(HA) Cervical Intervertebral Body Fusion Devices
ApplicantMeditech Spine, LLC
Product CodeODP · Orthopedic
Decision DateFeb 22, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Talos®-C (HA) Cervical Intervertebral Body Fusion Devices are intervertebral body devices intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the cervical spine at one level from C2-T1. Cervical degenerative disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on the posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by history and radiographic studies. Talos®-C (HA) Cervical IBF Devices are intended to be used with autologous bone graft to facilitate fusion. Non-operative treatment prior to treatment with Talos®-C (HA) Cervical Intervertebral Body Fusion Devices is six (6) weeks. Talos®-C (HA) Cervical IBF Devices are to be implanted via an open anterior approach. Talos®-C (HA) Cervical IBF Devices are also to be used with supplemental fixation.

Device Story

Talos-C (HA) Cervical IBF Device; rectangular intervertebral body fusion implant; made of hydroxyapatite-impregnated polyetheretherketone (HA PEEK); features ridged teeth for pullout resistance; includes tantalum markers for radiographic positioning. Implanted by surgeons via open anterior approach to cervical spine (C2-T1); used with autologous bone graft to facilitate fusion; requires supplemental fixation. Device accommodates anatomical variations via multiple sizes and lordotic/flat configurations. Benefits patients by providing structural support during spinal fusion for degenerative disc disease.

Clinical Evidence

Bench testing only. No clinical data provided. Mechanical equivalency established via static and dynamic compression, shear, and torsion testing per ASTM 2077 and ASTM 2267, plus subsidence testing. Simulated aging performed followed by additional dynamic mechanical testing.

Technological Characteristics

Material: Hydroxyapatite-impregnated polyetheretherketone (HA PEEK). Form factor: Rectangular, ridged teeth, curved lateral walls, rounded edges. Markers: Tantalum. Standards: ASTM 2077 (static/dynamic compression, shear, torsion), ASTM 2267 (subsidence). Connectivity: None. Energy source: None. Sterilization: Not specified.

Indications for Use

Indicated for skeletally mature patients with Degenerative Disc Disease (DDD) of the cervical spine at one level from C2-T1, defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation causing nerve root/spinal cord compression. Requires 6 weeks of failed non-operative treatment. Must be used with autologous bone graft and supplemental fixation via open anterior approach.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 22, 2015 Meditech Spine, LLC Mr. Jason Gromek Director of Engineering 1447 Peachtree Street, Northeast, Suite 440 Atlanta, Georgia 30309 Re: K142345 Trade/Device Name: Talos -C (HA) Cervical Intervertebral Body Fusion Devices Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP Dated: January 20, 2015 Received: January 22, 2015 Dear Mr. Gromek: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Lori A. Wiggins -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## meditech spir 1447 Peachtree St NE Suite 440, Atlanta GA 30309 | p (678) 974-5287 | f (404) 759-2104 | support@meditechspine.com ## Indications for Use Form 510(k) Number (if known): K142345 Device Name: Talos®-C (HA) Cervical Intervertebral Body Fusion Devices #### Indications for Use: Talos®-C (HA) Cervical Intervertebral Body Fusion Devices are intervertebral body devices intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the cervical spine at one level from C2-T1. Cervical degenerative disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on the posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by history and radiographic studies. Talos®-C (HA) Cervical IBF Devices are intended to be used with autologous bone graft to facilitate fusion. Non-operative treatment prior to treatment with Talos®-C (HA) Cervical Intervertebral Body Fusion Devices is six (6) weeks. Talos®-C (HA) Cervical IBF Devices are to be implanted via an open anterior approach. Talos®-C (HA) Cervical IBF Devices are also to be used with supplemental fixation. Prescription Use (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ # meditech spine, IIc 1447 Peachtree St NE Suite 440, Atlanta GA 30309 | p (678) 974-5287 | f (404) 759-2104 | support@meditechspine.com ## 510(k) Summary As required by section 807.92(c) Meditech Spine, LLC is requesting marketing clearance for the Talos®-C (HA) Cervical Intervertebral Body Fusion Devices. Summary prepared on the 18th of February, 2015. | | A. Sponsor/Manufacturer: | Meditech Spine, LLC | |--|--------------------------|---------------------------------------| | | Registration Number: | 3009405289 | | | | Jason Gromek, Director of Engineering | | | | 1447 Peachtree St NE Suite 440 | | | | Atlanta, GA 30309 | | | | 678-974-5287 Phone | | | | 404-759-2104 Fax | - B. Trade Name: Talos®-C (HA) Cervical Intervertebral Body Fusion Devices Common Name: Spinal Implant Classification Name: Intervertebral body fusion device (21 CFR 888.3080 Class II, Product Code ODP) - Predicate Device: K 122850: Meditech Advisors, LLC Talos®-C Cervical ﻥ Intervertebral Body Fusion Devices ### D. Device Description: The Talos® -C (HA) Cervical Intervertebral Body Devices (Talos®-C (HA) Cervical IBF Devices) are made of the polymer, hydroxyapatite impregnated polyetheretherketone (HA PEEK). The devices are open devices with ridged teeth on superior and inferior ends to resist implant pullout. The Talos®-C (HA) Cervical IBF Devices are rectangular devices and have curved lateral walls and rounded edges. The implants are available in a range of sizes as well as flat and lordotic angled implants to accommodate variations in patient's anatomy. In addition, tantalum markers at the opposite ends are offered which allows the Talos®-C (HA) Cervical IBF Device radiological confirmation for proper positioning. #### ت Intended Use: Talos®-C (HA) Cervical Intervertebral Body Fusion Devices are intervertebral body devices intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the cervical spine at one level from C2-T1. Cervical degenerative disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on the posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by history and radiographic studies. Talos®-C (HA) Cervical IBF Devices are intended to be used with autologous bone graft to facilitate fusion. Non-operative treatment prior to treatment with Talos®-C (HA) Cervical Intervertebral Body Fusion Devices is six (6) weeks. {4}------------------------------------------------ # meditech spine, IIc 1447 Peachtree St NE Suite 440, Atlanta GA 30309 | p (678) 974-5287 | f (404) 759-2104 | support@meditechspine.com Talos®-C (HA) Cervical IBF Devices are to be implanted via an open anterior approach. Talos® -C (HA) Cervical IBF Devices are also to be used with supplemental fixation. - F. Technological Characteristics: The technological characteristics of the Talos®C (HA) Cervical IBF Devices are the same as the predicate device, except for the material. The new material is hydroxyapatite filled polyetheretherketone (HA PEEK). HA is a naturally occurring mineral in bone and is widely used in the orthopedic field. - G. Non-clinical Testing: Tests according to ASTM 2077 and ASTM 2267 were performed on the Talos®-C (HA) Cervical IBF Devices to establish equivalency to the predicate device. The tests include static compression, compression shear and torsion tests, dynamic compression, compression shear and torsion tests, as well as subsidence testing. Simulated aging was performed on Talos®-C (HA) IBF Devices, which then were subjected to additional dynamic compression and torsion tests in accordance with ASTM 2077. The Talos®-C (HA) is equal in mechanical function and properties to the predicate device, establishing equivalency in safety and effectiveness. - H. Conclusion: The testing completed as well as a comparison of the technological characteristics have demonstrated that the Talos®-C (HA) Cervical IBF Devices are substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...