The Flower Upper Extremity Plating Set is intended to be used for fixation of fractures, fusions, or osteotomies of the clavicle, humerus, radius and ulna.
Device Story
The Flower Upper Extremity Plating Set is a metallic bone fixation system consisting of various anatomical plates (clavicle, humerus, olecranon, coronoid) and variable angle locking/non-locking screws. Used by orthopedic surgeons in clinical settings to stabilize bone fractures, fusions, or osteotomies. The device provides mechanical support to bone segments, facilitating healing through rigid fixation. The system includes general-purpose instruments and trials to assist the surgeon during implantation. The plates and screws are implanted during surgical procedures to provide structural stability to the affected upper extremity bones.
Clinical Evidence
Bench testing only. Performance was validated through engineering analysis and testing against standards including ASTM F67, ASTM F136, ISO 5832-2/3, ISO 7153-1, ISO 10993-1/5, and ISO 11137-1. Testing demonstrated that the device provides appropriate mechanical strength and functions as intended.
Technological Characteristics
Plates: pure titanium (ASTM F67). Screws: titanium alloy (ASTM F136). Components include clavicle, humerus, olecranon, and coronoid plates with variable angle locking/non-locking screws. Sterilization: radiation (ISO 11137-1).
Indications for Use
Indicated for fixation of fractures, fusions, or osteotomies of the clavicle, humerus, radius, and ulna in patients requiring orthopedic bone fixation.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Flower Orthopedics, Small and Medium Implant Set (K123562, K131657)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background of the seal.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 15, 2014
Flower Orthopedics Corporation Mr. Gary Barnett RA/RQ Manager 100 Witmer Road, Suite 280 Horsham, Pennsylvania 19044
Re: K142306 Trade/Device Name: Flower Upper Extremity Plating Set Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliance and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: August 15, 2014 Received: August 19, 2014
Dear Mr. Barnett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Mr. Gary Barnett
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known): K142306
Device Name: Flower Upper Extremity Plating Set
Indications for Use:
The Flower Upper Extremity Plating Set is intended to be used for fixation of fractures, fusions, or osteotomies of the clavicle, humerus, radius and ulna.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page __ of ___________________________________________________________________________________________________________________________________________________________________
{3}------------------------------------------------
# 510(k) SUMMARY
# Flower Orthopedics Corporation's Upper Extremity Plating Set
#### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Flower Orthopedics Corporation 100 Witmer Road. Suite 280 Horsham PA 19044
(215) 394-8903 Phone: Facsimile: (215) 394-8904
Contact Person: Gary Barnett Date Prepared: August 15, 2014
#### Name of Device
Flower Upper Extremity Plating Set
#### Common or Usual Name/Classification Name
The Flower Upper Extremity Plating Set consists of bone plates classified under product code HRS (21 C.F.R. 888.3030. Single/multiple component metallic bone fixation appliance and accessories: Class II) and bone fixation screws classified under product code HWC (21 C.F.R. 888.3040, Smooth or threaded metallic bone fixation fastener; Class II).
## Predicate Devices
Acumed Inc., Congruent Bone Plating System (K012655, K063640, K071715) Flower Orthopedics, Small and Medium Implant Set (K123562, K131657)
#### Intended Use / Indications for Use
The Flower Upper Extremity Plating Set is intended to be used for fixation of fractures, fusions, or osteotomies of the clavicle, humerus, radius and ulna.
## Device Description
The Flower Upper Extremity Plating Set consists of the following components and accessories: midshaft clavicle plates, distal clavicle plates, medial distal humerus plates, lateral distal humeral plates, posterior lateral distal humerus plates, olecranon plates, variable angle locking screws, variable angle non-locking screws. The plates are all made of pure titanium compliant with ASTM F67, the screws are made of titanium alloy compliant with ASTM F136. The system accepts locking and non-locking screws cleared via K123562 and K131657, the locking and non-locking screws included in this submission extend the length range of the previously cleared screws. The device is provided with general purpose instruments, including appropriately sized trials.
## Technological Characteristics
The Flower Upper Extremity Plating Set consists of the following components/configurations: Midshaft Clavicle Plates, with a width of 11 mm and lengths of 70/110 mm; Distal Clavicle Plates, with a width of 11 mm and lengths of 91/136 mm;
{4}------------------------------------------------
Medial Distal Humerus Plates, with a width of 11 mm and lengths of 78/140 mm; Lateral Distal Humerus Plates, with a width of 11 mm and lengths of 87/146 mm: Posterior Lateral Distal Humerus Plates, with a width of 11 mm and lengths of 100/148 mm; Olecranon Plates, with a width of 10 mm and lengths of 95/135 mm; Coronoid Plates, with a width of 21 mm, and 40mm lengths; Variable Angle Locking Screws: 2.7 mm, 3.0 mm, 3.5 mm, 4.0 mm dia. x 65 mm long: Variable Angle Non-Locking Screws: 2.7 mm. 3.0 mm. 3.5 mm dia. X 8 mm long: Variable Angle Non-Locking Screws: 2.7 mm. 3.0 mm dia. X 60 mm long Variable Angle Non-Locking Screws: 2.7 mm, 3.0 mm, 3.5 mm, 4.0 mm dia. X 65 mm long
# Performance Data
The Flower Upper Extremity Plating Set was tested (worse case) according to the following standards:
- . ASTM F67, Standard Specification for Unalloyed Titanium for Surgical Implant Applications
- ISO 5832-2, Implants for Surgery, Metallic Materials. Part 2: Unalloyed Titanium ●
- ASTM F136, Standard Specification for Wrought Titanium- 6 Aluminum- 4 Vanadium ELI Alloy for Surgical Implant Applications
- o ISO 5832-3, Implants for Surgery. Metallic materials. Part 3: Wrought titanium 6-aluminium 4vanadium allov:
- o ISO 7153-1, Surgical instruments – Metallic materials – Part 1: Stainless steel (ISO 7153-1:1991, including Amendment 1:1999); German version EN ISO 7153-1:2000;
- o ISO 10993-1. Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process:
- . ISO 10993-5, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity;
- ISO 11137-1, Sterilization of health care products Radiation Part 1: Requirements for ● development, validation and routine control of a sterilization process for medical devices. (Sterility)
In addition, an engineering analysis was performed to demonstrate that the subject upper extremity plates provide appropriate mechanical strength for the claimed intended use.
In all instances, the Flower Upper Extremity Plating Set functioned as intended and test results, as well as an engineering analysis, demonstrate substantial equivalence with the cited predicate devices.
## Substantial Equivalence
The Flower Upper Extremity Plating Set is substantially equivalent to the identified predicate devices.
The subject devices have the same intended uses/indications, technological characteristics, and principles of operation as the predicate devices. An engineering analysis was performed to demonstrate that the components in the Flower Upper Extremity Plating Set provide appropriate mechanical strength for the claimed intended use. Thus, the subject upper extremity plates are substantially equivalent and do not present any new risks.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.