← Product Code [DQX](/productcode/DQX) · K142254

# CPS Excel MediGuide Enabled Guidewire, CPS Excel MediGuide Torque clip, CPS Excel MediGuide Guidewire Connector (K142254)

_St Jude Medical · DQX · Nov 13, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K142254

## Device Facts

- **Applicant:** St Jude Medical
- **Product Code:** [DQX](/productcode/DQX.md)
- **Decision Date:** Nov 13, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The St. Jude Medical CPS Excel™ MediGuide Enabled™ Guidewire is intended for use with the MediGuide™ System to enable real-time tip positioning and navigation within the coronary and peripheral vasculature (such as to facilitate left heart lead implantation). The MediGuide system is intended for use as an adjunct to fluoroscopy.

## Device Story

Device consists of CPS Excel™ MediGuide Enabled™ Guidewire and connector accessory (DS2M033); used with MediGuide™ System. Guidewire inserts into connector; connector features PCB with magnetic shielding to enable electrical connectivity to MediGuide Technology catheter connect box. System provides real-time tip positioning and navigation within coronary and peripheral vasculature; functions as adjunct to fluoroscopy. Used by clinicians in clinical settings (e.g., cath lab) to assist in procedures like left heart lead implantation. Output visualized by clinician to guide navigation; reduces reliance on fluoroscopy; potentially benefits patient by minimizing radiation exposure.

## Clinical Evidence

Bench testing only. Verification and validation included bubble leak testing (ASTM F2096), seal strength peel testing (ASTM F88), visual inspection (ISO 10555-1:2009), dimensional testing (ISO 11070:1999), cable joint flex testing (ANSI/AAMI EC53:1995/2008), and label testing (BS EN-45502-1). Results demonstrated compliance with design specifications.

## Technological Characteristics

Guidewire connector accessory includes flexible insulated cable, molded housing, and internal PCB with magnetic shielding. Connectivity via ODU-compatible connector to MediGuide Technology. Standards: ASTM F2096 (leak), ASTM F88 (seal), ISO 10555-1 (visual), ISO 11070 (dimensional), ANSI/AAMI EC53 (flex), BS EN-45502-1 (label).

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- CPS Excel™ MediGuide Enabled™ Guidewire and accessories ([K120298](/device/K120298.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an image of three stylized human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 20, 2014

St. Jude Medical Zachary Price Regulatory Affairs Specialist 15900 Valley View Court Sylmar, CA 91342

Re: K142254

Trade/Device Name: CPS Excel MediGuide Enabled Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: November 13, 2014 Received: November 13, 2014

Dear Mr. Price:

This letter corrects our substantially equivalent letter of November 13, 2014.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/1/Picture/8 description: The image shows the name "Melissa A. Torres -S" in a clear, sans-serif font. The text is horizontally aligned and appears to be part of a document or label. The letters are uniformly sized and spaced, contributing to the legibility of the name.

For Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K142254

#### Device Name

CPS Excel™ MediGuide Enabled™ Guidewire Models DS2M027, DS2M028, DS2M029

Indications for Use (Describe)

The St. Jude Medical CPS Excel™ MediGuide Enabled™Guidewire is intended for use with the MediGuide™ System to enable real-time tip positioning and navigation within the coronary and peripheral vasculature (such as to facilitate left heart lead implantation). The MediGuide system is intended for use as an adjunct to fluoroscopy.

Type of Use (Select one or both, as applicable)

|  | <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)   </span> |
|--|-----------------------------------------------------------------------------------------------------|
|  | <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)   </span>  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for St. Jude Medical. The logo consists of a green square grid on the left side and the text "ST. JUDE MEDICAL" on the right side. The text is in a sans-serif font and is black.

### 510(K) SUMMARY

### 510(k) Summary

The purpose of this special 510(K) is to gain clearance for the CPS Excel™ MediGuide Enabled™ Guidewire Models DS2M027, DS2M028, DS2M029 and accessories. There are no design changes to the Guidewire Models DS2M027, and DS2M029. The design features of the CPS Excel™ MediGuide Enabled™ Guidewire Connector accessory DS2M033 have only undergone minimal design changes to connect directly to the MediGuide technology as compared to the predicate, MediGuide Enabled™ Guidewire Connector DS2M032 (K120298).

| Submitter:                                                                | St. Jude Medical, CRMD                                                               |
|---------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| Address:                                                                  | 15900 Valley View Ct<br>Sylmar, CA 91342                                             |
| Phone:                                                                    | 818 493 2903                                                                         |
| Fax:                                                                      | 818 493 3615                                                                         |
| Contact Person:                                                           | Zachary Price                                                                        |
| Trade Name/Proprietary<br>Name:                                           | CPS Excel™ MediGuide Enabled™ Guidewire and accessories                              |
| Common Name:                                                              | Catheter, Guidewire<br>Model Numbers: DS2M027, DS2M028, DS2M029, DS2M030,<br>DS2M033 |
| Legally marketed device<br>to which your firm is<br>claiming equivalence: | CPS Excel™ MediGuide Enabled™ Guidewire and accessories<br>(K120298)                 |

### Device Description:

The device description of the CPS Excel™ MediGuide Enabled™ Guidewire Connector accessory DS2M033 is as follows.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for St. Jude Medical. On the left is a green square made up of smaller squares, with one square tilted. To the right of the square is the text "ST. JUDE MEDICAL" in a sans-serif font. The text is black.

The Connector accessory is a flexible, insulated cable that is used to connect the CPS Excel™ MediGuide Enabled " guidewire to the MediGuide Technology. The guidewire is inserted into the connector until it bottoms out. Once the guidewire cannot be pushed any further, the nut on the connector end is tightened to fixate the guidewire in the connector. Inside the molded connector housing is a printed circuit board (PCB) which is wrapped with magnetic shielding. The PCB has contact slots that connect with the electrical contacts of the guidewire for electrical functionality. The other end of the MediGuide guidewire connector is a ODU-compatible connector that connects to the MediGuide Technology catheter connect box.

## The indication for use is as follows:

The St. Jude Medical CPS Excel™ MediGuide Enabled™ Guidewire is intended for use with the MediGuide "" System to enable real-time tip positioning and navigation within the coronary and peripheral vasculature (such as to facilitate left heart lead implantation). The MediGuide system is intended for use as an adjunct to fluoroscopy.

## Technological Characteristics of the Device Compared to the Predicate Device:

The device has the same technological characteristics as the currently marketed CPS Excel™ MediGuide Enabled™ Guidewire connector, with only minimal changes to the cable length, connector type, connector construction to accommodate change to connect directly to the MediGuide Technology

### Non-clinical Test Summary:

The risk analysis method used to assess the impact of the modification of the CPS Excel™ MediGuide Enabled™ Guidewire Connector accessory DS2M033 documents the investigation of hazards and the mitigation of risks associated with its use and reports the results of the investigation. The risk analysis method used to assess the impact of the modifications was a Failure Modes and Effects Analysis (FMEA/FMECA). It was determined that overall residual risk is acceptable.

Verification and validation testing of the Connector accessory included the following testing and associated guiding standards: (1) bubble leak testing (ASTM F2096), (2) peel testing for seal strength (ASTM F88), (3) visual inspection testing (ISO 10555-1:2009), (4) dimensional testing (ISO 11070:1999), (5) cable joint flex testing (ANSI/AAMI EC53:1995/"2008), and (6) label testing (BS EN-45502-1). Successful completion of all verification and validation activities, (as shown in Table 4) demonstrated that the candidate devices meet their predetermined design and performance specifications and that the products are substantially equivalent to the predicate devices.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for St. Jude Medical. The logo consists of a green square grid on the left, with one square tilted. To the right of the grid is the text "St. Jude MEDICAL" in a simple, sans-serif font.

# Conclusion (Statement of Equivalence):

The results of the verification and validation tests and the risk analysis have demonstrated the CPS Excel™ MediGuide Enabled™ Guidewire and accessories functions in accordance with product specifications that were previously cleared for the predicate device, , the CPS Excel™ MediGuide Enabled™ Guidewire connector DS2M032 cleared in 510(k) K120298.

Therefore, St. Jude Medical considers the CPS Excel™ MediGuide Enabled™ Guidewire and accessories to be substantially equivalent to the legally marketed predicate device, the CPS Excel™ MediGuide Enabled™ Guidewire connector DS2M032 cleared in 510(k) K120298.

---

**Source:** [https://fda-staging.innolitics.com/device/K142254](https://fda-staging.innolitics.com/device/K142254)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
