PREMIUM, SHELTA Implant Systems
Device Facts
| Record ID | K142242 |
|---|---|
| Device Name | PREMIUM, SHELTA Implant Systems |
| Applicant | Sweden & Martina S.P.A. |
| Product Code | DZE · Dental |
| Decision Date | Feb 2, 2015 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3640 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
PREMIUM Implant Systems-SHELTA Implant Systems (Premium TG, Premium TG, Premium SP, Shelta and Shelta SL) are intended for both one- and two-stage surgical procedures. PREMIUM Implant Systems-SHELTA Implant Systems are intended for immediate placement and function on single tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function. Multiple tooth applications may be splinted with a bar. PREMIUM-SHELTA Abutments are intended to be used in conjunction with a PREMIUM-SHELTA Implants Systems in fully edentulous or partially edentulous maxillary and/or mandibular arches. The PREMIUM-SHELTA Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single tooth prostheses, in the mandible or maxilla. The prosthesis can be cemented, screw retained or friction fit to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant. PREMIUM-SHELTA Abutments are compatible with PREMIUM-SHELTA Implants Systems.
Device Story
Endosseous dental implant system comprising implants (Premium TG, Premium SP, Shelta, Shelta SL) and compatible abutments. Used by dental surgeons in clinical settings to restore chewing function in edentulous or partially edentulous patients. Implants surgically placed into maxillary or mandibular bone; abutments attached to implants to support prosthetic devices (cemented, screw-retained, or friction-fit). System facilitates single or multiple tooth replacement; supports immediate loading when primary stability is sufficient. Benefits include restoration of oral function and aesthetics.
Clinical Evidence
Bench testing only.
Technological Characteristics
Endosseous dental implants and abutments. Materials include titanium/titanium alloys. System includes various implant geometries (TG, SP, Shelta, SL). Designed for surgical implantation into alveolar bone. Non-software, mechanical device.
Indications for Use
Indicated for partially or fully edentulous patients requiring single or multiple tooth replacement in the maxilla or mandible. Supports one- or two-stage surgical procedures; allows immediate placement and function if primary stability is achieved.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.