K142216 · Clinical Laserthermia Systems, AB · GEX · Apr 21, 2015 · General, Plastic Surgery
Device Facts
Record ID
K142216
Device Name
Tranberg CLS Thermal Therapy System
Applicant
Clinical Laserthermia Systems, AB
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 21, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Tranberg 65 Thermal Therapy System is indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in areas of surgery including: gastroenterology, general surgery, plastic surgery, genitourinary (urology), gynecology (GYN), neurosurgery, otolaryngology (ENT) head and neck, orthopedics, ophthalmology, pulmonology, and thoracic surgery.
Device Story
Tranberg 65 Thermal Therapy System is a mobile laser unit (1064 nm) for soft tissue ablation, vaporization, excision, incision, and coagulation. System components include mobile laser, temperature sensor, and applicator kit. Laser energy is delivered via percutaneous or surgical procedure; tissue absorbs energy, causing heating and lesion formation. A tissue temperature feedback system controls heating and lesion size. Used in clinical settings by physicians. Output is controlled laser energy; healthcare providers use the system to perform thermal therapy. Benefits include controlled tissue destruction for surgical applications.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including engineering comparative temperature testing, verification and validation against product requirements, software testing, usability engineering (ISO 62366), electromagnetic compatibility (IEC 60601-1-2), electrical safety (IEC 60601-2-22), and sterilization validation (ISO 11135-1).
Technological Characteristics
Class IV diode laser (1064 nm); 1W-25W output power; TEC cooling; SMA 905 output port; 635 nm aiming beam. Includes integrated tissue temperature feedback system. Power source 100-240 V AC. Dimensions 540x450x180mm; weight 18kg. Sterilization via Ethylene Oxide (ISO 11135-1).
Indications for Use
Indicated for surgical soft tissue ablation, vaporization, excision, incision, and coagulation across multiple specialties including gastroenterology, general surgery, plastic surgery, urology, gynecology, neurosurgery, ENT, orthopedics, ophthalmology, pulmonology, and thoracic surgery.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is a symbol of medicine and healing. The logo also includes the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 21, 2015
Clinical Laserthermia Systems AB % Mr. David Makanani OMEDtech, LLC 1725 Signal Ridge Drive, Suite 150 Edmond, Oklahoma 73013
Re: K142216
Trade/Device Name: Tranberg 65 Thermal Therapy System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: March 20, 2015 Received: March 24, 2015
Dear Mr. Makanani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
K142216
# Indication for Use
510(k) Number: K142216
Device Name: Tranberg CLS Thermal Therapy System
Intended Use/Indications for Use:
"The Tranberg 65 Thermal Therapy System is indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in areas of surgery including: gastroenterology, general surgery, plastic surgery, genitourinary (urology), gynecology (GYN), neurosurgery, otolaryngology (ENT) head and neck, orthopedics, ophthalmology, pulmonology, and thoracic surgery."
Prescription Use: X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use: NO (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off
510(k)
{3}------------------------------------------------
## 510(K) SUMMARY
| Date | April 17, 2015 |
|------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTER: | Lars-Erik Eriksson, CEO<br>Clinical Laserthermia Systems, AB<br>Scheelevagen 2<br>Lund, Sweden 22381 |
| CONTACT PERSON: | David Makanani, CEO<br>OMEDtech, L.L.C.<br>1725 Signal Ridge Drive, Suite 150<br>Edmond, Oklahoma 73013<br>Tel: (405) 826-0713<br>Email: dmakanani@omedtech.com |
| DEVICE NAME:<br>Classification<br>Trade Name<br>Common Name<br>Classification<br>Product Code<br>Review Panel<br>PREDICATE DEVICE: | Class II<br>TRANBERGCLS Thermal Therapy System<br>TRANBERGCLS Thermal Therapy System<br>21 CFR 878.4810<br>GEX - Powered Laser Surgical Instrument<br>General and Plastic Surgery<br>K092197: BioTex, Inc.; PhoTex30 Diode Laser. |
| INTENDED USE: | The Tranberg CLS Thermal Therapy System is indicated<br>for use in surgical applications requiring the ablation,<br>vaporization, excision, incision, and coagulation of soft<br>tissue in areas of surgery including: gastroenterology,<br>general surgery, plastic surgery, genitourinary<br>(urology), gynecology (GYN), neurosurgery,<br>otolaryngology (ENT) head and neck, orthopedics,<br>ophthalmology, pulmonology, and thoracic surgery. |
#### DEVICE DESCRIPTION:
The TRANBERGCSS|Thermal Therapy System consists of three parts:
- TRANBERGCSS|Mobile Laser
{4}------------------------------------------------
- TRANBERGCLS|Temperature Sensor ●
- Applicator Kit (The Applicator kit is not included)
The mobile laser unit is provided with a laser generator operating at the wavelength 1064 nm. The generated laser light is locally applied by means of a single use applicator kit through a less invasive surgical or percutaneous procedure. The energy within the laser light is absorbed by the tissue resulting in increased tissue temperature. Tissue heating and lesion formation is controlled by a tissue temperature feedback system integrated into the TRANBERGCS | Thermal Therapy System.
For a detailed description of the function and the usage of the laser module and its accessories, view the IFU.
### TECHNOLOGICAL CHARACTERISTIC AND SUBSTANTIAL EQUIVALENCE:
Substantial equivalence of the TRANBERGSSSSSThermal Therapy System is claimed to the PhoTex 30 Diode Laser Series, cleared under K092197.
The CLS device is verified and validated to have the same performance as the predicate device when used together with the Applicator kit cleared under K053087
The following table provides more detailed information regarding the basis for the determination of substantial equivalence:
| Parameter | | |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product name | TRANBERGCLS Thermal Therapy System | PhoTex 30 Diode Laser Series |
| Manufacturer | Clinical LaserThermia Systems CLS, Sweden | BioTex, US |
| Intended use /<br>Indications for use | "The Tranberg <span style="font-variant: small-caps;">CLS</span> Thermal Therapy<br>System is indicated for use in surgical<br>applications requiring the ablation,<br>vaporization, excision, incision, and<br>coagulation of soft tissue in areas of<br>surgery including: gastroenterology,<br>general surgery, plastic surgery,<br>genitourinary (urology),<br>gynecology (GYN), neurosurgery,<br>otolaryngology (ENT) head and neck,<br>orthopedics, ophthalmology,<br>pulmonology, and thoracic surgery." | "The PhoTex3 Diode Laser Series is<br>indicated for use in surgical applications<br>requiring the ablation, vaporization,<br>excision, incision, and coagulation of soft<br>tissue in areas of surgery including:<br>gastroenterology, general surgery, plastic<br>surgery, genitourinary (urology),<br>gynecology (GYN), neurosurgery,<br>otolaryngology (ENT) head and neck,<br>orthopedics, ophthalmology, pulmonology,<br>and thoracic surgery." |
| Device Regulatory<br>Classification | FDA 878.4810 | FDA 878.4810 |
| Product code | GEX | GEX |
| Device class | 2 | 2 |
| 510(k) No | To be obtained | K092197 |
| Diode laser<br>generator | | |
| Wavelength | 1064nm<br>Adapted to indication for use of the laser<br>applicator / hand piece | 980nm, 810nm or 940 nm |
| Output power | 1W - 25W at output port | 3W – 30W at output port |
| Output power<br>accuracy | +/- 10% of selected value | +/- 20% of selected value |
| Mode of operation | Continuous wave or controlled by tissue<br>temperature monitored by a temperature<br>sensor | Continuous wave (CW), pulsed, or external<br>modulation modes. |
| Output power<br>increments | 1W | 0.5 W |
| Cooling | TEC | TEC |
| Channel(s) | 1 | 1 |
| Output port | SMA 905 | SMA 905 |
| Aiming wavelength | 635 nm | 650 nm |
| Laser type<br>IEC60825-1 | Class IV | Class IV |
| General technical<br>characteristics | | |
| Power source | 100-240 V AC / 50-60 Hz | 100-240 V AC / 50-60 Hz |
| Operating<br>temperature range | 15°C to 28°C | 10-35 °C |
| Average dimensions | 540, 450, 180mm (width, depth, height) | 16.0"x12,5"x8,0" (406x318x203) |
| Weight | 18 Kg | 20 lbs (9,1kg) |
| Foot switch<br>operation | On/Off | On/Off |
| Emergency switch | Yes | Yes |
| Key activation of<br>laser output | Yes | Yes |
| Remote Interlock | Yes | Yes |
| Power ON/OFF<br>Visual Indicator | Yes | Yes |
| Laser Emission<br>Indicator | Yes | Yes |
| Internal Laser Power<br>Monitor | Yes | Yes |
| Manual Reset | Yes | Yes |
| Fiber Insertion<br>Interlock | Yes | Yes |
| Laser Emission<br>Energy Monitoring | Yes | Yes |
| Audio Warning<br>Signal Level | Fixed at HIGH | HIGH, MEDIUM, LOW, and OFF |
| Safety classification<br>FDA | Class II | Class II |
| Pump for cooling<br>liquid for applicator | Yes | Yes |
| Temperature<br>sensors included | Yes | No |
| Applicator kit<br>(Laser fiber and<br>Trochar) | | |
| Interface | Compatible with fiber optic delivery<br>accessory with a standard SMA905<br>connector having a core fiber diameter of<br>400 or 600 microns and a numerical<br>aperture of at least 0.37. | Compatible with fiber optic delivery<br>accessory with a standard SMA905<br>connector having a core fiber diameter of<br>400 or 600 microns and a numerical<br>aperture of at least 0.37. |
| Performance | The CLS device is verified and validated<br>to have the same performance when used<br>together with the Applicator kit cleared<br>under K053087 | The BioTex device is verified and validated<br>together with the Applicator kit cleared<br>under K053087 |
{5}------------------------------------------------
{6}------------------------------------------------
### PERFORMANCE TESTING - (NON-CLINICAL) BENCH:
The TRANBERG®|Thermal Therapy System has been determined through engineering bench testing to support substantial equivalence with this device and the predicates. This testing showed the TRANBERGSS|Thermal Therapy System to meet applicable ISO, IEC and FDA safety and performance standards,
Non-clinical bench performance testing completed:
- Engineering comparative temperature testing
- . Engineering Verification and Validation Testing to the Product Requirement Specification
{7}------------------------------------------------
- Software testing ●
- Usability Engineering Testing - ISO 62366
- Electromagnetic Compatibility – IEC 60601-1-2 Collateral Standard
- Electrical Safety for Laser Equipment — IEC 60601-2-22 Particular Standard
- Medical Device Sterilization of Health Care Products – ISO 11135-1; Ethylene Oxide
#### PERFORMANCE TESTING - CLINICAL:
There are no clinical data submitted with this Notification.
#### CONCLUSION:
Based on the results of non-clinical testing, the TRANBERGSS|Thermal Therapy System performs safely, as intended, and the comparative discussion of intended use, principle of operation, and technological characteristics, it is determined that the TRANBERGSSS|Thermal Therapy System is substantially equivalent to predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.