# EXPEDIUM VERSE SPINE SYSTEM (K142185)

_Depuy Spine, Inc. · NKB · Nov 19, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K142185

## Device Facts

- **Applicant:** Depuy Spine, Inc.
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Nov 19, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The EXPEDIUM Verse System is intended to provide immobilization of spinal segments in sketally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities of the thoracic, lumbar and sacral spine. The EXPEDIUM Verse System is intended for noncervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients. When used in a posterior percutaneous approach with MIS instrumentation, the EXPEDIUM Verse System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients. When used for posterior non-cervical pedicle screw fixation in pediatric patients, the EXPEDIUM Verse System metallic implants are indicated as an adjunct to treat adolescent idiopathic scoliosis. The EXPEDIUM Verse system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

## Device Story

EXPEDIUM Verse Spine System; pedicle screw spinal system for thoracic, lumbar, and sacral stabilization. Features intraoperative polyaxial-to-monoaxial conversion; facilitates rod capture; provides reduction mechanism. Used in OR by surgeons; compatible with EXPEDIUM 5.5 rods, hooks, and mono screws. Implanted via open or posterior percutaneous MIS approach. Provides rigid fixation to support spinal fusion; benefits patients by correcting deformities and stabilizing unstable spinal segments.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing (mechanical testing) using ASTM 1717 standard test methods for spinal implant constructs in a vertebrectomy model, including static compression, dynamic compression, and static torsion.

## Technological Characteristics

Materials: Titanium alloy (Ti64AIV) and Cobalt chromium alloy (CoCr). Design: Polyaxial to monoaxial conversion mechanism; reduced profile thoracolumbar implant. Testing: ASTM 1717. Sterilization: Not specified.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- EXPEDIUM Spine System ([K111136](/device/K111136.md))
- VIPER Spine System ([K111136](/device/K111136.md))
- VIPER 2 Spine System ([K111136](/device/K111136.md))
- Merlin Spine System ([K033901](/device/K033901.md))
- EXPEDIUM Spine System ([K070387](/device/K070387.md))

## Reference Devices

- MOUNTAINEER LAMINOPLASTY SYSTEM ([K091994](/device/K091994.md))
- NAVIGATED CD HORIZON SOLERA SCREWDRIVERS, TAPS, ILIAC TAPS, LEGACY TAPS ([K124004](/device/K124004.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of three faces in profile, stacked on top of each other, representing the department's focus on people and their well-being.

November 19, 2014

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

DePuy Synthes Spine, a Johnson & Johnson Company Catherine Kilshaw, M.S. Senior Regulatory Affairs Associate 325 Paramount Drive Raynham, Massachusetts 02767

Re: K142185

Trade/Device Name: DePuy Synthes Spine EXPEDIUM® Verse Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MNI, MNH, KWP, KWQ Dated: August 21, 2014 Received: August 22, 2014

Dear Ms. Kilshaw:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K142185

Device Name

DePuy Synthes Spine EXPEDIUM® Verse Spine System

#### Indications for Use (Describe)

The EXPEDIUM Verse System is intended to provide immobilization of spinal segments in sketally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities of the thoracic, lumbar and sacral spine.

The EXPEDIUM Verse System is intended for noncervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

When used in a posterior percutaneous approach with MIS instrumentation, the EXPEDIUM Verse System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the EXPEDIUM Verse System metallic implants are indicated as an adjunct to treat adolescent idiopathic scoliosis. The EXPEDIUM Verse system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

| Type of Use (Select one or both, as applicable)                                       |                                                                                      |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <div> <span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{3}------------------------------------------------

## 510(k) Summary

Date Prepared: November 18, 2014

| Submitter:                       | DePuy Synthes Spine<br>325 Paramount Drive<br>Raynham, MA 02767                                                                                                                                                                               |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                  | Catherine Kilshaw<br>Sr. Regulatory Affairs Associate<br>Telephone: (508) 880-8412<br>Fax: (508) 828-3797                                                                                                                                     |
| Trade Name:                      | DePuy Synthes Spine EXPEDIUM® Verse Spine System                                                                                                                                                                                              |
| Device Class:                    | III                                                                                                                                                                                                                                           |
| Product Code:                    | NKB OSH KWP KWQ MNH MNI                                                                                                                                                                                                                       |
| Common Name:                     | Pedicle Screw Spine System                                                                                                                                                                                                                    |
| Classification Name:             | Pedicle Screw Spinal System                                                                                                                                                                                                                   |
| Regulation Number:               | 21 CFR 888.3070                                                                                                                                                                                                                               |
| Primary<br>Predicate Device:     | K111136 EXPEDIUM® Spine System, VIPER® Spine System, VIPER® 2<br>Spine System                                                                                                                                                                 |
| Additional Predicate<br>Devices: | K033901 Merlin Spine System (later re-named EXPEDIUM Spine System)<br>K070387 EXPEDIUM Spine System                                                                                                                                           |
| Reference Devices:               | K091994 MOUNTAINEER LAMINOPLASTY SYSTEM<br>K124004 NAVIGATED CD HORIZON SOLERA SCREWDRIVERS, TAPS, ILIAC TAPS,<br>LEGACY TAPS                                                                                                                 |
| Device Description:              | The EXPEDIUM® Verse System is designed to provide intraoperative polyaxial to<br>monoaxial conversion. It facilitates easier rod capture and provides a powerful<br>and precise reduction mechanism. Verse is a reduced profile thoracolumbar |

{4}------------------------------------------------

implant for use for with wide range of patient statures. EXPEDIUM Verse is a self-contained, efficient, and intuitive instrument system that is compatible with EXPEDIUM 5.5 rods, hooks and mono screws to enhance versatility.

Indications: The EXPEDIUM Verse System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

> The EXPEDIUM Verse System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

> When used in a posterior percutaneous approach with MIS instrumentation, the EXPEDIUM Verse System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the EXPEDIUM Verse System metallic implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The EXPEDIUM Verse system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

| Materials:        | Titanium alloy Ti64AIV, Cobalt chromium alloy CoCr                                                                                                                                                                                         |  |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Comparison to     |                                                                                                                                                                                                                                            |  |
| Predicate Device: | The substantial equivalence of the subject device to the predicate identified<br>above is based upon the similarity of intended use, design (fundamental<br>scientific technology) materials, performance, sterility and biocompatibility. |  |
| Non-clinical Test |                                                                                                                                                                                                                                            |  |
| Summary:          | Mechanical testing (using the ASTM 1717 Standard Test Methods for Spinal<br>Implant Constructs in a Vertebrectomy Model standard) was provided in order                                                                                    |  |

{5}------------------------------------------------

|               | to provide data to support a substantial equivalence determination. These tests<br>were performed to characterize the properties and functionality of the screw, as<br>well as to allow comparison with established acceptance criteria.                                                  |
|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|               | The following mechanical tests were conducted on the screw:                                                                                                                                                                                                                               |
|               | ●<br>Static Compression                                                                                                                                                                                                                                                                   |
|               | Dynamic Compression<br>●                                                                                                                                                                                                                                                                  |
|               | ●<br>Static Torsion                                                                                                                                                                                                                                                                       |
| Clinical Test |                                                                                                                                                                                                                                                                                           |
| Summary:      | N/A                                                                                                                                                                                                                                                                                       |
| Conclusion:   | Based upon the predicate comparison, the intended use, similar technological<br>characteristics and the results of the static compression testing, dynamic<br>compression testing and static torsion testing, the proposed device is<br>substantially equivalent to the predicate device. |

---

**Source:** [https://fda-staging.innolitics.com/device/K142185](https://fda-staging.innolitics.com/device/K142185)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).