K142173 · Ecolab, Inc. · LHC · Dec 30, 2014 · Physical Medicine
Device Facts
Record ID
K142173
Device Name
NON-STERILE ORS-400N SCOPE PILLOW WARMER DRAPE
Applicant
Ecolab, Inc.
Product Code
LHC · Physical Medicine
Decision Date
Dec 30, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5950
Device Class
Class 1
Indications for Use
The ORS pillow drape is a single-use disposable device intended for use as an accessory for ORS Solution Warmers which is designed to hold the optical end of various endoscopes above the warm solution to prevent fogging or wetting of the scope eyepiece.
Device Story
The ORS Scope Pillow Warmer Drape is a single-use, disposable equipment cover and endoscope holder accessory for ORS surgical fluid warmers. It consists of a polyurethane film drape with an integrated polyurethane foam pillow. The device is used in clinical settings by surgical staff to protect fluid warmers from contamination and to provide stable positioning for endoscopes, preventing them from rolling or tipping while resting in the warm solution. The non-sterile version (ORS-400N) is supplied to secondary processors who package, label, and sterilize the device before clinical use. By keeping the scope eyepiece above the fluid, the device prevents fogging and wetting, ensuring clear visualization for the surgeon during procedures.
Clinical Evidence
Bench testing only. Performance testing included dimensional verification, drape integrity (sterile barrier maintenance), and simulated distribution testing per ASTM D4169-09. All tests passed.
Technological Characteristics
Materials: Polyurethane film and polyurethane foam. Form factor: 44 in. x 66 in. drape with integrated foam pillow. Connectivity: None. Sterilization: Provided non-sterile; intended for sterilization by secondary processors. Energy source: None (passive).
Indications for Use
Indicated for use as an accessory for ORS Solution Warmers to hold endoscope optical ends above warm solutions to prevent fogging/wetting during surgical procedures.
Regulatory Classification
Identification
A powered heating unit is a device intended for medical purposes that consists of an encased cabinet containing hot water and that is intended to heat and maintain hot packs at an elevated temperature.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to create a sense of depth and connection. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 30, 2014
Ecolab, Inc. Ms. Jennifer Willner Senior Director, Regulatory Affairs 370 Wabasha Street North St. Paul, MN 55102
Re: K142173
Trade/Device Name: ORS Scope Pillow Warmer Drape Regulation Number: Unclassified Regulation Name: Equipment Cover Regulatory Class: Unclassified Product Code: LHC Dated: December 3, 2014 Received: December 4, 2014
Dear Ms. Willner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Willner
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
of Surveillance and Biometrics/Division of Postmarket Surveillance.
Sincerely yours,
Erin I. Keith -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K142173
Device Name
ORS Scope Pillow Warmer Drape
Indications for Use (Describe)
The ORS pillow drape is a single-use disposable device intended for use as an accessory for ORS Solution Warmers which is designed to hold the optical end of various endoscopes above the warm solution to prevent fogging or wetting of the scope eyepiece.
The following model(s) are included: ORS-400 and ORS-400N
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-weight: normal;"> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span> | <span> <span style="font-family: Symbol;"> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-weight: normal;"> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span> | <span> <span style="font-family: Symbol;"> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "Ecolab Inc." The text is in a simple, sans-serif font and is horizontally aligned. The background of the image is plain white, which makes the text stand out clearly.
a u
#### Section 5: 510(k) Summary – K142173
## ORS Scope Pillow Warmer Drape
As required by 21 CFR 807.92.
| Date:<br>Administrative Information | August 6, 2014 |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Ecolab, Inc. |
| Establishment<br>Registration Number: | 1043582 |
| Contact Person: | Jennifer Willner, RAC<br>370 Wabasha Street North<br>St. Paul, MN 55102-1390<br>Sr. Director, Regulatory Affairs - Healthcare<br>651.250.4348 |
| Device Identification | |
| Device Name: | ORS Scope Pillow Warmer Drape |
| Common Name: | Equipment Cover |
| Device Classification Name: | Warmer, Irrigation Solution |
| Device Classification: | Unclassified |
| Classification Product Code: | LHC |
| Panel: | General and Plastic Surgery |
| Classification Regulation: | Pre-Amendment |
| Performance Standards: | No Recognized Consensus Standards |
| Predicate Device: | ORS Warming Drape cleared on 06/27/2002 via<br>K021288 and ORS Endoscope Holder cleared on<br>08/30/2005 via K051979 |
{4}------------------------------------------------
#### Device Description
The ORS-400N Scope Pillow™ Warmer Drape is a single-use disposable device provided non-sterile to secondary processors, who then package, label and sterilize the device. The Scope Pillow Warmer Drape is an accessory/equipment drape for use with the ORS-2057 Irrigation Solution Warming System which is designed to both protect ORS surgical fluid warmers from contamination during various procedures as well as hold the optical end of various endoscopes above the warm solution to prevent fogging or wetting of the scope eyepiece.
The drape material is the identical polyurethane film that is used on all standard ORS Warmer Drapes (K021288). A soft polyurethane foam pillow is integrated within the polyurethane drape and has no direct or indirect patient contact. The pillow (endoscope holder) is designed to provide stable positioning of scope(s) during warming and prevents one or more scopes from rolling or tipping while the scope rests in the fluid warmer during the procedure (Figure 5-1).
Image /page/4/Picture/5 description: In the image, a medical instrument is placed on a table covered with a clear plastic sheet. The instrument has a long, slender metal shaft with a black handle at one end. The table is draped with blue covers, and the background is a light gray wall with horizontal lines.
Figure 5-1 Scope Pillow Warmer Drape
The ORS-400 and ORS-400N model drapes are identical with the exception of how they are packaged and labeled. The ORS-400 model consists of drapes individually packaged in labeled peel pouches or header bags which are stacked inside a double poly-lined corrugate case with Instructions for Use (IFU) insert and sterilized by an Ecolab subcontractor before it is released for sale. The ORS-400N model consists of the identical drapes which are individually wrapped in a poly tie secured by a product ID label and stacked inside a double poly-lined corrugate case with IFU insert that does not get sterilized before it is released for sale to kit packer customers. The IFU insert differs between the two models in that the ORS-400N contains sterilization instructions for kit packer customers to evaluate their sterilization process as it applies to placing the product in their commercial procedure kits.
NOTE: Both device models are designed and intended to be used by clinical staff in sterile condition.
## 510(k) Discussion
This 510(k) submission requests clearance for the manufacture and distribution of Non-Sterile ORS-400N Scope Pillow Warmer Drape (equipment cover with integrated pillow/endoscope holder). The non-sterile drape will be sold in bulk packaging to
{5}------------------------------------------------
secondary processors who will then package, label and sterilize the drapes as required under the regulation prior to distribution to end users. The non-sterile drape has the identical polyurethane drape material, and identical drape function as the predicate drape (K021288). The pillow (endoscope holder) provides the same holding function as the predicate Endoscope Holder (K051979). Therefore, the fundamental scientific technology of this equipment drape remains unchanged. The Substantial Equivalence Table (Table 5-1) is provided below.
| Property or<br>Characteristic | Proposed Device -<br>Non-Sterile ORS-<br>400N Scope Pillow™<br>Warmer Drape | Predicate Device -<br>ORS-1000LD<br>Equipment Cover | Predicate Device -<br>ORS Endoscope<br>Holder |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) No. | This 510(k)<br>Submission | K021288 | K051979 |
| Device Name | ORS-400N Scope<br>Pillow Warmer<br>Drape | ORS-1000LD<br>Warming Drape | ORS Model 6000<br>Endoscope Holder |
| Indications for<br>Use | The ORS pillow<br>drape is a single-use<br>disposable device<br>intended for use as an<br>accessory for ORS<br>Solution Warmers<br>which is designed to<br>hold the optical end<br>of various endoscopes<br>above the warm<br>solution to prevent<br>fogging or wetting of<br>the scope eyepiece. | The ORS-1000LD<br>Leak Detection Drape<br>is an equipment cover<br>for the ORS-2000LD<br>Solution Warmer.<br>This is a single use<br>product supplied<br>sterile. This device is<br>intended for use<br>during various<br>surgeries where warm<br>irrigation solution is<br>required. | The ORS 6000<br>Endoscope Holder is<br>a single use<br>disposable device<br>used as an accessory<br>for ORS Solution<br>Warmers. The ORS<br>6000 Endoscope<br>Holder is designed to<br>hold the optical end<br>of various endoscopes<br>above the warm<br>solution to prevent<br>fogging or wetting of<br>the scope eyepiece. |
| Conditions of<br>Use | Rx Only, Single Use,<br>Disposable | Identical | Identical |
| Materials | Polyurethane Film<br>Polyurethane Foam | Identical Film | Thermoplastic<br>polyester resins |
| Principle of<br>Operation | Covers surgical<br>solution warmers and<br>endoscope holder | Covers surgical<br>solution warmers | Endoscope Holder |
| Property or<br>Characteristic | Proposed Device –<br>Non-Sterile ORS-<br>400N Scope Pillow™<br>Warmer Drape | Predicate Device –<br>ORS-1000LD<br>Equipment Cover | Predicate Device –<br>ORS Endoscope<br>Holder |
| Packaging | Bulk packaged in<br>poly bag in quantities<br>up to 10 | Individually packaged<br>in poly/Tyvek peel<br>pouches | Individually packaged<br>in poly/Tyvek peel<br>pouches |
| Sterilized | No; intended to be<br>sterilized before<br>distribution to end<br>user | Yes; provided in<br>sterile condition via<br>EO at SAL 10-6 | Yes; provided in<br>sterile condition via<br>EO at SAL 10-6 |
#### Table 5-1: Substantial Equivalence
{6}------------------------------------------------
# Performance Data Summary
| | Table 5-2: Performance Data Summary of the Non-Sterile ORS-400N Scope Pillow | | |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Warmer Drape | | | |
| | | | |
| Requirement | Specification | Method | Result |
| Functional<br>Performance<br>Requirements | Dimensional Requirements:<br>ORS-400N (44 in. x 66 in.) | Acceptable results following<br>visual inspection during V&V<br>testing; VVR-14-0002 V&V<br>Summary Report, Project 8<br>Track | Pass |
| | Drape remains intact (free<br>from holes or other defects<br>that would compromise the<br>sterile barrier) | Acceptable results following<br>visual inspection during V&V<br>testing; VVR-14-0002 V&V<br>Summary Report, Project 8<br>Track | Pass |
| Packaging | Packaging Configuration:<br>10 per case, double poly<br>bagged | Packaging Configuration per:<br>ORS-400N_DWG;<br>Acceptable results following<br>visual inspection during V&V<br>testing; VVR-14-0002 V&V<br>Summary Report, Project 8<br>Track | Pass |
| | Product must be received by<br>customer with folds intact | Acceptable results following<br>visual inspection during V&V<br>testing; documented in V&V<br>summary report VVR-14-0002 | Pass |
| | Simulated Distribution Test | ASTM D4169-09 (Distribution<br>Cycle 2, Assurance Level 1)<br>Simulated Distribution Test | Pass |
{7}------------------------------------------------
| Requirement | Specification | Method | Result |
|-------------|--------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Labeling | Inspection for Drape<br>Damage | (PKG 001F) documented in<br>Packaging Engineering Report<br># REPT-18430<br>Acceptable results following<br>visual inspection during V&V<br>testing; documented in the<br>V&V summary report VVR-<br>14-0002 | Pass |
| | Master carton label is<br>present and per specification | Master label specifications:<br>ORS-400NMASTER<br>Acceptable results following<br>visual inspection during V&V<br>testing; documented in the<br>V&V summary report VVR-<br>14-0002. | Pass |
| | Ink (non-smudge/smear) | Acceptable results following<br>visual inspection during V&V<br>testing; documented in V&V<br>summary report VVR-14-0002 | Pass |
| | Insert Sheet/IFU is present<br>and per specification | Insert Sheet/IFU<br>specifications:<br>ORS-400N-INSERT SHEET | Pass |
| | Product Identification label<br>includes product code and<br>Ecolab Logo | Product Identification label<br>specification:<br>ORS-400NMISC | Pass |
## Statement of Equivalence
The Non-Sterile ORS Scope Pillow Warmer Drape performs as intended using the identical principles of operation as the predicate devices. Specifically, it is designed to protect ORS surgical solution warmers from contamination during various procedures as well as hold the optical end of various endoscopes above the warm solution to prevent fogging or wetting of the scope eyepiece. Differences between the Non-Sterile ORS Scope Pillow Warmer Drape and the sterile version (packaging and labeling), as well as the predicate devices (two separate components used together), do not raise any new questions of safety or efficacy when secondary processors sterilize the product as instructed. Based on the risk analysis, review of the product labeling, and successful performance testing (Table 5-2), the ORS Scope Pillow Warmer Drape is substantially equivalent to the legally marketed ORS Warming Drape (K021288) and ORS Endoscope Holder (K051979). The fundamental scientific technology of the device remains unchanged.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.