MERITS TRANSFORMER POWER SCOOTER, S941

K142162 · Merits Health Products Co., Ltd. · INI · May 1, 2015 · Physical Medicine

Device Facts

Record IDK142162
Device NameMERITS TRANSFORMER POWER SCOOTER, S941
ApplicantMerits Health Products Co., Ltd.
Product CodeINI · Physical Medicine
Decision DateMay 1, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3800
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Merits Transformer Power Scooter, S941, is to provide mobility to adult persons limited to a sitting position and capable to operate a few simple controls.

Device Story

Battery-powered, rear transaxle-driven motorized scooter; provides indoor/outdoor mobility for adults with limited mobility. User operates via tiller-mounted throttle lever and control buttons. Dynamic Rhino-2 120amp controller manages motor/brake system; throttle activation releases solenoid electromechanical brakes for movement; throttle release triggers automatic braking. Powered by two 12 VDC 50ah batteries; charged via off-board charger. Welded steel frame construction. Used in home or community settings on dry, level surfaces (concrete, blacktop, asphalt). Provides patient benefit by restoring independent mobility.

Clinical Evidence

Bench testing only. Compliance with ISO 7176 series standards for static/dynamic stability, braking effectiveness, energy consumption, dimensions/maneuvering, speed/acceleration, strength, climatic testing, obstacle climbing, and electromagnetic compatibility (ISO 7176-21). Upholstery ignition resistance tested per ISO 7176-16.

Technological Characteristics

Welded steel frame; Dynamic Rhino-2 120amp controller; two 12 VDC 50ah batteries; solenoid electromechanical brakes. Dimensions/performance tested per ISO 7176-1 through 7176-16 and 7176-21. Upholstery meets ISO 7176-16:2012. Off-board 5A charger with 3-pin microphone connector.

Indications for Use

Indicated for adult persons with limited mobility who are restricted to a sitting position and possess the capability to operate simple controls.

Regulatory Classification

Identification

A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, which is a traditional symbol of medicine. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 1, 2015 Merits Health Products Company, LTD Vincent Chen Manager, R&D Department No. 18, Jingke Road, Nantun District, Taichung City, Taiwan 40852 Re: K142162 > Trade/Device Name: Merits Transformer Power Scooter, S941 Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: March 25, 2015 Received: March 31, 2015 Dear Mr. Chen, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Felipe Aquel - S for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K142162 Device Name Merits Transformer Power Scooter, S941 Indications for Use (Describe) The Merits Transformer Power Scooter, S941, is to provide mobility to adult persons limited to a sitting position and capable to operate a few simple controls. | | Type of Use (Select one or both, as applicable) | | |--|-------------------------------------------------|--| |--|-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) SUMMARY Merits Health Products Co., LTD. 510(k) Premarket Notification #### Submitter: Merits Health Products Co., LTD. No.18, Jingke Rd., Nantun Dist. Taichung City 40852, Taiwan (R.O.C.) ### Contact Person: Vincent Chen Merits Health Products Co., LTD. No.18, Jingke Rd., Nantun Dist., Taichung City 40852, Taiwan (R.O.C.) Phone: +886-4-23594985 ext.200 Fax: +886-4-23594992 ### Date Prepared: July 30, 2014 # Proprietary Name: Merits Transformer Power Scooter, S941 ## Common name: Four Wheel Power Scooter ### Classification name: Physical Medicine / Motorized Three - Wheeled Vehicle #### Product Code: INI ### Comparison to Predicate Devices: This submission indicates the Substantial Equivalence of the Merits Transformer Power Scooter, S941, with the predicate Merits Travel Ease Electric Scooter, Pioneer 4 (K011753). S941 has the same intended uses and similar indications, technological characteristics and principles of operation with predicate device. {4}------------------------------------------------ ## Device Description The Merits Transformer Power Scooter, S941, is battery powered, rear transaxle driven and is controlled by the Dynamic Rhino-2 120amp controller. The user interface is a throttle lever and control buttons on the control panel. S941 is powered by two 12 VDC 50ah (22NF) batteries. The batteries are charged by 5A off-board charger connect with 3-pin Microphone Connector to charging socket on Tiller. The approximate driving range on fully charged batteries is up to 40km (25mi). The scooter frame is a welded steel construction and includes 2 rear wheels with drive units (including motor, gear box, and brake), batteries, 2 front wheels and a tiller to control its turning direction. When the user activates the throttle lever, the controller receives a signal to release the brakes. With the brakes released, the scooter is allowed to move forwards or backwards. When the user releases the throttle lever, the scooter slows to a stop and the brakes are automatically re-engaged. The solenoid electromechanical brakes allow the user stop by letting go of the throttle lever. The upholstery of the device complies with ISO 7176-16:2012 Resistance to ignition of postural support devices. The device can be operated on dry, level surfaces composed of concrete, blacktop, or asphalt under normal driving conditions. The Merits Transformer Power Scooter, S941, is substantially equivalent to the Merits Travel Ease Electric Scooter, Pioneer 4 (K011753). Both products are battery power, motorized vehicles designed for use as personal power mobility aids. Performance characteristics and drive mechanisms are similar and all have the same intended function and use which is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair. Additional, they are all constructed from the same basic materials, have the same basic operational principles and all use DC batteries as their source of power. Although there are some minor differences between S941 and its predicate device. But they raise no new issues of safety or effectiveness. Performance data demonstrate that S941 is safe. The non-clinical testing and the predicate comparisons demonstrate that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. ## Intended Use The Merits Transformer Power Scooter, S941, is to provide mobility to adult persons limited to a sitting position and capable to operate a few simple controls. {5}------------------------------------------------ # Discussions of Non-clinical Tests Performed for Determinations of Substantial equivalence are as follows: - ISO 7176-1:1999 Determination of Static Stability - ISO 7176-2:2001 Determination of Dynamic Stability of electric wheelchairs - ISO 7176-3:2012 Determination of effectiveness of brakes - Energy consumption of electric wheelchairs and scooters for ISO 7176-4:2008 determination of theoretical distance range - Determination of overall dimensions, mass and maneuvering ISO 7176-5:2008 space - Determination of maximum speed, acceleration and ISO 7176-6:2001 deceleration of electric wheelchairs - ISO 7176-7:1998 Method of Measurement of Seating and Wheel Dimensions - Requirements and test methods for static, impact and fatigue ISO 7176-8:1998 strengths - ISO 7176-9:2009 Climatic tests for wheelchairs - Determination of obstacle-climbing ability of electrically power ISO 7176-10:2008 wheelchairs - ISO 7176-11:2012 Test dummies - ISO 7176-13:1989 Determination of coefficient of friction of test surfaces - Power and control svstems for electrically powered ISO 7176-14:2008 wheelchairs and scooter- Requirements and test methods. - Requirements for Information Disclosure, Documentation and ISO 7176-15:1996 Labeling - ISO 7176-16:2012 Resistance to ignition of postural support devices - Requirements and test methods for electromagnetic ISO 7176-21:2009 compatibility of electrically powered wheelchairs and scooters, and battery chargers. - Medical devices -- Application of risk management to medical ISO 14971:2007 devices The results of the testing confirm that the device meets specifications and is substantially equivalent to the predicate device. ## Conclusions According to comparison table, the differences on weight capacity, speed, battery size and dimensions do not deleteriously affect the safety and effectiveness of the device. So based on the design, performance specifications and testing and intended use, the Merits Transformer Power Scooter, S941, is substantially equivalent to the Merits Travel Ease Electric Scooter, Pioneer 4 (K011753).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...