← Product Code [IYN](/productcode/IYN) · K142160

# LOGIQ E9 DIAGNOSTIC ULTRASOUND SYSTEM (K142160)

_GE Healthcare · IYN · Oct 10, 2014 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K142160

## Device Facts

- **Applicant:** GE Healthcare
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Oct 10, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic and vascular).

## Device Story

LOGIQ E9 is a mobile, console-based diagnostic ultrasound system for clinical imaging and fluid flow analysis. It acquires, processes, and displays ultrasound data via various transducers. The system features a 19-inch LCD display, 10-inch touch screen, and specialized controls for physician operation in clinical settings. It supports B-mode, M-mode, PW/CW Doppler, Color/Power Doppler, and 3D/4D imaging. It incorporates shear wave elastography for tissue elasticity assessment. The system processes inputs to provide real-time visualization, measurements, and reporting, aiding clinical decision-making in diverse diagnostic applications. It benefits patients by providing non-invasive, real-time diagnostic imaging.

## Clinical Evidence

No clinical studies were required to support substantial equivalence. Evidence is based on bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal/electrical/electromagnetic/mechanical safety testing in compliance with standards like AAMI/ANSI ES60601-1 and IEC60601-2-37.

## Technological Characteristics

Mobile console (approx. 58x86x141 cm) with 19-inch LCD display and 10-inch touch screen. Employs ultrasonic pulsed Doppler/echo imaging. Supports B, M, PW/CW Doppler, Color/Power Doppler, 3D/4D, and shear wave elastography. Complies with AAMI/ANSI ES60601-1, IEC60601-1-2, IEC60601-2-37, NEMA UD 2/3, ISO 10993-1, and ISO 14971. Connectivity via DICOM.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- GE LOGIQ E9 Diagnostic Ultrasound System ([K123564](/device/K123564.md))
- SuperSonic Imagine, Aixplorer Diagnostic Ultrasound System ([K132274](/device/K132274.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized emblem with three abstract human profiles facing right, stacked on top of each other. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 10, 2014

GE Healthcare % Mr. Bryan Behn Regulatory Affairs Manager 9900 Innovation Drive WAUWATOSA WI 53226

Re: K142160

Trade/Device Name: LOGIQ E9 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: September 29, 2014 Received: September 30, 2014

Dear Mr. Behn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

This determination of substantial equivalence applies to the following transducers intended for use with the LOGIQ E9 Diagnostic Ultrasound System, as described in your premarket notification:

| Transducer Model Number |          |          |           |
|-------------------------|----------|----------|-----------|
| 3CRF                    | 3CRF-D   | 6S-D     | 6Tc       |
| 9L-D                    | 11L-D    | C1-5-D   | IC5-9-D   |
| L8-18i-D                | M4S-D    | M5S-D    | M6C-D     |
| ML6-15-D                | P2D      | P6D      | RAB2-5D   |
| RAB4-8-D                | RIC5-9-D | RNA5-9-D | RSP6-16-D |
| S1-5                    | S4-10    | S4-10-D  | C2-9-D    |
| RAB6-D                  | S1-5-D   | C1-6-D   | C2-9VN-D  |

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

for

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K142160

Device Name LOGIQ E9

Indications for Use (Describe)

The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic and vascular).

| Type of Use (Select one or both, as applicable)                                       |                                                                                      |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <div> <span>☑</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span>☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form GE LOGIQ E9 Ultrasound System

|                               | Mode of Operation |   |         |         |         |         |         |          |          |       |              |  |
|-------------------------------|-------------------|---|---------|---------|---------|---------|---------|----------|----------|-------|--------------|--|
| Clinical Application          | B                 | M | PW      | CW      | Color   | Color M | Power   | Combined | Harmonic | Coded | Other        |  |
| Anatomy/Region of Interest    |                   |   | Doppler | Doppler | Doppler | Doppler | Doppler | Modes*   | Imaging  | Pulse | (Notes)      |  |
| Ophthalmic                    |                   |   |         |         |         |         |         |          |          |       |              |  |
| Fetal/Obstetrics[7]           | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [5,6,9]      |  |
| Abdominal[1]                  | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [3,5,6,9,10] |  |
| Pediatric                     | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [3,5,6,9]    |  |
| Small Organ[2]                | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [3,5,6,9,10] |  |
| Neonatal Cephalic             | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [5,6,9]      |  |
| Adult Cephalic                | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [5,6,9]      |  |
| Cardiac Adult                 | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     |              |  |
| Cardiac Pediatric             | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     |              |  |
| Peripheral Vascular           | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [3,5,6,9]    |  |
| Musculo-skeletal Conventional | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [3,5,6,9]    |  |
| Musculo-skeletal Superficial  | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [3,5,6,9]    |  |
| Other[4]                      | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [3,5,6,9]    |  |
| Exam Type, Means of Access    |                   |   |         |         |         |         |         |          |          |       |              |  |
| Transesophageal               | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     |              |  |
| Transrectal                   | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [3,5,6,9]    |  |
| Transvaginal                  | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [3,5,6,9]    |  |
| Transurethral                 |                   |   |         |         |         |         |         |          |          |       |              |  |
| Intraoperative[8]             | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [3,5,6,9]    |  |
| Intraoperative Neurological   |                   |   |         |         |         |         |         |          |          |       |              |  |
| Intravascular                 |                   |   |         |         |         |         |         |          |          |       |              |  |
| Laparoscopic                  |                   |   |         |         |         |         |         |          |          |       |              |  |

#### Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, Gyn

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[9] Volume navigation

[10] Shear wave elastography

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription User (Per 21 CFR 801.109)

Page 2 of 31

{4}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form GE LOGIQ E9 with 3CRF Transducer

| Clinical Application          |   | Mode of Operation |               |               |                  |                    |                  |                   |                     |                |                  |
|-------------------------------|---|-------------------|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
|                               | B | M                 | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Anatomy/Region of Interest    |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Ophthalmic                    |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Fetal/Obstetrics[7]           | P | P                 | P             | P             | P                | P                  | P                | P                 | P                   | P              | [5,6,9]          |
| Abdominal[1]                  | P | P                 | P             | P             | P                | P                  | P                | P                 | P                   | P              | [5,6,9]          |
| Pediatric                     | P | P                 | P             | P             | P                | P                  | P                | P                 | P                   | P              | [5,6,9]          |
| Small Organ[2]                |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Neonatal Cephalic             |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Adult Cephalic                |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Cardiac Adult                 |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Cardiac Pediatric             |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Peripheral Vascular           |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Musculo-skeletal Conventional |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Musculo-skeletal Superficial  |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Other[4]                      | P | P                 | P             | P             | P                | P                  | P                | P                 | P                   | P              | [5,6,9]          |
| Exam Type, Means of Access    |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Transesophageal               |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Transrectal                   |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Transvaginal                  |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Transurethral                 |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Intraoperative[8]             |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Intraoperative Neurological   |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Intravascular                 |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |
| Laparoscopic                  |   |                   |               |               |                  |                    |                  |                   |                     |                |                  |

#### Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[9] Volume navigation

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription User (Per 21 CFR 801.109)

Page 3 of 31

{5}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form GE LOGIQ E9 with 3CRF-D Transducer

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|---------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |         |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
| Fetal/Obstetrics[7]                                | E                 | E | E             | E             | E                | E                  | E                | E                  | E                   | E              | [5,6,9]          |         |
| Abdominal[1]                                       | E                 | E | E             | E             | E                | E                  | E                | E                  | E                   | E              | [5,6,9]          |         |
| Pediatric                                          | E                 | E | E             | E             | E                | E                  | E                | E                  | E                   | E              | [5,6,9]          |         |
| Small Organ[2]                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
| Cardiac Adult                                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
| Cardiac Pediatric                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
| Peripheral Vascular                                |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
| Other[4]                                           | E                 | E | E             | E             | E                | E                  | E                | E                  | E                   | E              | E                | [5,6,9] |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
| Transurethral                                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
| Intraoperative[8]                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |         |

#### Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[9] Volume navigation

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription User (Per 21 CFR 801.109)

Page 4 of 31

{6}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form GE LOGIQ E9 with 6S-D Transducer

|                                                    | Mode of Operation |   |            |            |               |                 |               |                 |                  |              |               |  |
|----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|-----------------|------------------|--------------|---------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse‡ | Other [Notes] |  |
| Ophthalmic                                         |                   |   |            |            |               |                 |               |                 |                  |              |               |  |
| Fetal/Obstetrics [7]                               | P                 | P | P          | P          | P             | P               | P             | P               | P                | P            |               |  |
| Abdominal [1]                                      | P                 | P | P          | P          | P             | P               | P             | P               | P                | P            |               |  |
| Pediatric                                          | P                 | P | P          | P          | P             | P               | P             | P               | P                | P            |               |  |
| Small Organ[2]                                     |                   |   |            |            |               |                 |               |                 |                  |              |               |  |
| Neonatal Cephalic                                  | P                 | P | P          | P          | P             | P               | P             | P               | P                | P            | P             |  |
| Adult Cephalic                                     |                   |   |            |            |               |                 |               |                 |                  |              |               |  |
| Cardiac Adult                                      |                   |   |            |            |               |                 |               |                 |                  |              |               |  |
| Cardiac Pediatric                                  | P                 | P | P          | P          | P             | P               | P             | P               | P                | P            | P             |  |
| Peripheral Vascular                                |                   |   |            |            |               |                 |               |                 |                  |              |               |  |
| Musculo-skeletal Conventional                      |                   |   |            |            |               |                 |               |                 |                  |              |               |  |
| Musculo-skeletal Superficial                       |                   |   |            |            |               |                 |               |                 |                  |              |               |  |
| Other [4]                                          |                   |   |            |            |               |                 |               |                 |                  |              |               |  |
| Exam Type, Means of Access                         |                   |   |            |            |               |                 |               |                 |                  |              |               |  |
| Transesophageal                                    |                   |   |            |            |               |                 |               |                 |                  |              |               |  |
| Transrectal                                        |                   |   |            |            |               |                 |               |                 |                  |              |               |  |
| Transvaginal                                       |                   |   |            |            |               |                 |               |                 |                  |              |               |  |
| Transurethral                                      |                   |   |            |            |               |                 |               |                 |                  |              |               |  |
| Intraoperative[8]                                  |                   |   |            |            |               |                 |               |                 |                  |              |               |  |
| Intraoperative Neurological                        |                   |   |            |            |               |                 |               |                 |                  |              |               |  |
| Intravascular                                      |                   |   |            |            |               |                 |               |                 |                  |              |               |  |
| Laparoscopic                                       |                   |   |            |            |               |                 |               |                 |                  |              |               |  |

#### Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[9] Volume navigation

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription User (Per 21 CFR 801.109)

Page 5 of 31

{7}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form GE LOGIQ E9 with 6Tc Transducer

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|-----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse• | Other<br>[Notes] |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Fetal/Obstetrics[7]                                |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Abdominal[1]                                       |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Pediatric                                          |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Small Organ[2]                                     |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Cardiac Adult                                      | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P               |                  |  |
| Cardiac Pediatric                                  |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Peripheral Vascular                                |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Other[4]                                           |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Transesophageal                                    | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P               |                  |  |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Transurethral                                      |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Intraoperative[8]                                  |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |  |

## Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[9] Volume navigation

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription User (Per 21 CFR 801.109)

Page 6 of 31

{8}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form GE LOGIQ E9 with 9L-D Transducer

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Fetal/Obstetrics[7]                                | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [5,6,9]          |  |
| Abdominal[1]                                       | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [3,5,6,9]        |  |
| Pediatric                                          | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [3,5,6,9]        |  |
| Small Organ[2]                                     | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [3,5,6,9,10]     |  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Cardiac Adult                                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Cardiac Pediatric                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Peripheral Vascular                                | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [3,5,6,9]        |  |
| Musculo-skeletal Conventional                      | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [3,5,6,9]        |  |
| Musculo-skeletal Superficial                       | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [3,5,6,9]        |  |
| Other[4]                                           | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [3,5,6,9]        |  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transurethral                                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative[8]                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |

## Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[9] Volume navigation

[10] Shear wave elastography

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription User (Per 21 CFR 801.109)

Page 7 of 31

{9}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form GE LOGIQ E9 with 11L-D Transducer

| Clinical Application          | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse° | Other<br>[Notes] |
|-------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|-----------------|------------------|
| Anatomy/Region of Interest    |   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Ophthalmic                    |   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Fetal/Obstetrics[7]           |   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Abdominal[1]                  | P | P | P             |               | P                | P                  | P                | P                  | P                   | P               | [3,5,6]          |
| Pediatric                     | P | P | P             |               | P                | P                  | P                | P                  | P                   | P               | [3,5,6]          |
| Small Organ[2]                | P | P | P             |               | P                | P                  | P                | P                  | P                   | P               | [3,5,6]          |
| Neonatal Cephalic             |   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Adult Cephalic                |   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Cardiac Adult                 |   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Cardiac Pediatric             |   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Peripheral Vascular           | P | P | P             |               | P                | P                  | P                | P                  | P                   | P               | [3,5,6]          |
| Musculo-skeletal Conventional | P | P | P             |               | P                | P                  | P                | P                  | P                   | P               | [3,5,6]          |
| Musculo-skeletal Superficial  | P | P | P             |               | P                | P                  | P                | P                  | P                   | P               | [3,5,6]          |
| Other[4]                      |   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Exam Type, Means of Access    |   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Transesophageal               |   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Transrectal                   |   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Transvaginal                  |   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Transurethral                 |   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Intraoperative[8]             |   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Intraoperative Neurological   |   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Intravascular                 |   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Laparoscopic                  |   |   |               |               |                  |                    |                  |                    |                     |                 |                  |

#### Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[9] Volume navigation

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription User (Per 21 CFR 801.109)

Page 8 of 31

{10}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form GE LOGIQ E9 with C1-5-D Transducer

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                 |                  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|-----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse* | Other<br>[Notes] |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Fetal/Obstetrics[7]                                | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P               | [5,6,9]          |
| Abdominal[1]                                       | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P               | [5,6,9]          |
| Pediatric                                          | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P               | [5,6,9]          |
| Small Organ[2]                                     |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Cardiac Adult                                      |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Cardiac Pediatric                                  |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Peripheral Vascular                                | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P               | [5,6,9]          |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Other[4]                                           | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P               | [5,6,9]          |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Transurethral                                      |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Intraoperative[8]                                  |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |

#### Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[9] Volume navigation

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription User (Per 21 CFR 801.109)

Page 9 of 31

{11}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form GE LOGIQ E9 with IC5-9-D Transducer

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                 |                  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|-----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse* | Other<br>[Notes] |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Fetal/Obstetrics[7]                                | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P               | [5,6,9]          |
| Abdominal[1]                                       |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Pediatric                                          |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Small Organ[2]                                     |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Cardiac Adult                                      |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Cardiac Pediatric                                  |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Peripheral Vascular                                |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Other[4]                                           | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P               | [3,5,6,9]        |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Transrectal                                        | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P               | [3,5,6,9]        |
| Transvaginal                                       | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P               | [3,5,6,9]        |
| Transurethral                                      |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Intraoperative[8]                                  |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                    |                     |                 |                  |

## Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, Gyn

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[9] Volume navigation

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription User (Per 21 CFR 801.109)

Page10 of 31

{12}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form GE LOGIQ E9 with L8-18i-D Transducer

| Clinical Application          | Anatomy/Region of Interest | Mode of Operation |   |            |            |               |                 |               |                 |                  |             |               |
|-------------------------------|----------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|-----------------|------------------|-------------|---------------|
|                               |                            | B                 | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other [Notes] |
| Ophthalmic                    |                            |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Fetal/Obstetrics[7]           |                            |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Abdominal[1]                  |                            |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Pediatric                     | P                          | P                 | P |            | P          | P             | P               | P             | P               | P                | [3,5,6,9]   |               |
| Small Organ[2]                | P                          | P                 | P |            | P          | P             | P               | P             | P               | P                | [3,5,6,9]   |               |
| Neonatal Cephalic             | P                          | P                 | P |            | P          | P             | P               | P             | P               | P                | [5,6,9]     |               |
| Adult Cephalic                |                            |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Cardiac Adult                 |                            |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Cardiac Pediatric             |                            |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Peripheral Vascular           | P                          | P                 | P |            | P          | P             | P               | P             | P               | P                | [3,5,6,9]   |               |
| Musculo-skeletal Conventional | P                          | P                 | P |            | P          | P             | P               | P             | P               | P                | [3,5,6,9]   |               |
| Musculo-skeletal Superficial  | P                          | P                 | P |            | P          | P             | P               | P             | P               | P                | [3,5,6,9]   |               |
| Other[4]                      |                            |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Exam Type, Means of Access    |                            |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Transesophageal               |                            |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Transrectal                   |                            |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Transvaginal                  |                            |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Transurethral                 |                            |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Intraoperative[8]             | P                          | P                 | P |            | P          | P             | P               | P             | P               | P                | [3,5,6,9]   |               |
| Intraoperative Neurological   |                            |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Intravascular                 |                            |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Laparoscopic                  |                            |                   |   |            |            |               |                 |               |                 |                  |             |               |

#### Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[9] Volume navigation

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription User (Per 21 CFR 801.109)

Page 11 of 31

{13}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form GE LOGIQ E9 with M4S-D Transducer

| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
|----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
|                                                    |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Fetal/Obstetrics[7]                                | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [5,6]            |
| Abdominal[1]                                       | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [5,6]            |
| Pediatric                                          | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [5,6]            |
| Small Organ[2]                                     |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Neonatal Cephalic                                  |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Adult Cephalic                                     | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              |                  |
| Cardiac Adult                                      | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              |                  |
| Cardiac Pediatric                                  |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Peripheral Vascular                                |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Musculo-skeletal Conventional                      |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Musculo-skeletal Superficial                       |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Other[4]                                           |   |   |               |               |                  |                    |…

---

**Source:** [https://fda-staging.innolitics.com/device/K142160](https://fda-staging.innolitics.com/device/K142160)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com