← Product Code [LFZ](/productcode/LFZ) · K142133

# ELECYS CMV IGM IMMUNOASSAY; ELECYS PRECICONTROL CMV IGM (K142133)

_Roche Diagnostics · LFZ · Oct 24, 2014 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K142133

## Device Facts

- **Applicant:** Roche Diagnostics
- **Product Code:** [LFZ](/productcode/LFZ.md)
- **Decision Date:** Oct 24, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3175
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K142133 · Oct 24, 2014 | ELECYS CMV IGM IMMUNOASSAY; ELECYS PRECICONTROL CMV IGM | Roche Diagnostics | Retrospective clinical samples | A retrospective study of 134 preselected clinical samples was used to evaluate the performance of the Elecsys CMV IgM immunoassay in a population confirmed positive for CMV IgM. | Retrospective study; Clinical performance evaluation |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Preselected Confirmed CMV IgM Positive Population; Retrospective study | Patients positive for CMV IgM as determined by comparator device consensus; Sample Size: 134; Number of Sites: 3 clinical sites and 1 internal site | Consensus of 3 FDA-cleared CMV IgM assays | Positive percent agreement |

## Indications for Use

Immunoassay for the in vitro qualitative detection of IgM antibodies to CMV in human serum, lithium-heparin plasma, K2-EDTA plasma, and K3-EDTA plasma. The test is intended as an aid in the diagnosis of recent or current CMV infection in individuals for which a CMV IgM test was ordered, including pregnant women. Performance characteristics have not been evaluated in immunocompromised or immunosuppressed individuals. This test is not intended for use in neonatal screening or for use at point of care facilities. This assay is not intended for use in screening blood and plasma donors. The electrochemiluminescence immunoassay "ECLIA" is intended for use on Elecsys immunoassay analyzers.

## Device Story

The Elecsys CMV IgM is a μ-capture electrochemiluminescence immunoassay (ECLIA) performed on Elecsys immunoassay analyzers. The device processes human serum or plasma samples; it uses biotinylated anti-h-IgM antibodies and ruthenium-labeled recombinant CMV antigen to capture anti-CMV IgM antibodies on streptavidin-coated microparticles. The analyzer magnetically captures these particles on an electrode; voltage application induces chemiluminescent emission measured by a photomultiplier. The system automatically calculates results as reactive, non-reactive, or indeterminate based on a cutoff index (COI) derived from a 2-point calibration and master curve. The output is used by clinicians to aid in the diagnosis of recent or current CMV infection. The device is intended for clinical laboratory use.

## Clinical Evidence

Clinical performance was evaluated in a prospective multi-center study (n=617) of patients suspected of CMV infection, including a subset of pregnant women (n=199), and a retrospective study (n=134) of preselected positive samples. Comparator was a consensus of 2-3 FDA-cleared devices. In the prospective suspected population, negative agreement was 97.2% (95% CI: 95.1–98.6) and positive agreement was 87.0% (95% CI: 66.4–97.2). In the pregnant population, negative agreement was 98.5% (95% CI: 95.5–99.7). The preselected positive population showed 100% positive agreement.

## Technological Characteristics

μ-capture immunoassay; streptavidin-coated microparticles; biotinylated monoclonal anti-h-IgM; ruthenium-labeled recombinant CMV antigen. Detection: electrochemiluminescence (ECLIA). Form factor: reagent rackpack for automated analyzers (Elecsys/cobas e). Connectivity: analyzer-integrated. Software: automated calculation of cutoff index based on 2-point calibration and master curve. Materials: human serum-based calibrators/controls.

## Regulatory Identification

Cytomegalovirus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to cytomegalovirus in serum. The identification aids in the diagnosis of diseases caused by cytomegaloviruses (principally cytomegalic inclusion disease) and provides epidemiological information on these diseases. Cytomegalic inclusion disease is a generalized infection of infants and is caused by intrauterine or early postnatal infection with the virus. The disease may cause severe congenital abnormalities, such as microcephaly (abnormal smallness of the head), motor disability, and mental retardation. Cytomegalovirus infection has also been associated with acquired hemolytic anemia, acute and chronic hepatitis, and an infectious mononucleosis-like syndrome.

## Predicate Devices

- Diamedix Is-CMV IgM Capture Test Kit ([K001767](/device/K001767.md))
- Elecsys PreciControl Rubella IgM ([K092322](/device/K092322.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM

A. 510(k) Number:
K142133

B. Purpose for Submission:
New device

C. Measurand:
IgM antibodies to CMV

D. Type of Test:
electrochemiluminescence immunoassay

E. Applicant:
Roche Diagnostics Inc.

F. Proprietary and Established Names:
Elecsys CMV IgM Immunoassay
Elecsys PreciControl CMV IgM

G. Regulatory Information:
1. Regulation section:
21 CFR 866.3175
2. Classification:
Class II
3. Product code:
LFZ

{1}

4. Panel:

Microbiology

H. Intended Use:

1. Intended use(s):

(1) Elecsys CMV IgM:

Immunoassay for the in vitro qualitative detection of IgM antibodies to CMV in human serum, lithium-heparin plasma, K2-EDTA plasma, and K3-EDTA plasma. The test is intended as an aid in the diagnosis of recent or current CMV infection in individuals for which a CMV IgM test was ordered, including pregnant women.

Performance characteristics have not been evaluated in immunocompromised or immunosuppressed individuals. This test is not intended for use in neonatal screening or for use at point of care facilities. This assay is not intended for use in screening blood and plasma donors.

The electrochemiluminescence immunoassay “ECLIA” is intended for use on Elecsys immunoassay analyzers.

(2) Elecsys PreciControl CMV IgM:

PreciControl CMV IgM is used for quality control of the Elecsys CMV IgM immunoassay on the Elecsys and cobas e immunoassay analyzers.

2. Indication(s) for use:

Same as intended use

3. Special conditions for use statement(s):

Not applicable

4. Special instrument requirements:

Elecsys immunoassay analyzers

I. Device Description:

(1) Elecsys CMV IgM:

The Elecsys CMV IgM is a μ-capture immunoassay with streptavidin microparticles, biotinylated recombinant CMV-specific antigen labeled with a ruthenium complex and

2

{2}

electrochemiluminescence detection. The results are determined using a calibration curve which is instrument-specifically generated by a 2-point calibration and a master curve provided via the reagent bar code. The test system contains the human serum-based calibrators intended for use with the system.

The CMV IgM assay begins with an automatic 1:20 predilution of sample with Elecsys Diluent Universal and the addition of biotinylated monoclonal anti-h-IgM-specific antibodies.

During the second incubation, CMV-specific recombinant antigen labeled with a ruthenium complex and streptavidin-coated microparticles are added. Anti-CMV IgM antibodies present in the sample react with the ruthenium-labeled CMV-specific recombinant antigen. The complex becomes bound to the solid phase via interaction of biotin and streptavidin.

The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode induces chemiluminescent emission which is measured by a photomultiplier.

The analyzer automatically calculates the cutoff based on the measurement of Cal1 and Cal2. The result of the samples is given either as reactive or non-reactive as well as in the form of a cutoff index (signal samples/cutoff). Samples with a cutoff index (COI) &lt; 0.7 are non-reactive. Samples with a COI between 0.7 and &lt; 1.0 are considered indeterminate (border). Samples with a COI ≥ 1.0 are considered reactive.

(2) Elecsys PreciControl CMV IgM:

The Elecsys PreciControl CMV IgM contains liquid control serum based on human serum. The controls are used for monitoring the accuracy of the Elecsys CMV IgM immunoassay. The reagents and calibrators are packaged together in the Elecsys CMV IgM assay kit, while the associated Elecsys PreciControl CMV IgM is packaged separately.

The following reagents are provided in the Elecsys CMV IgM assay kit:

1. The reagent rackpack consists of reagents M, R1, and R2 and is labeled as CMVIGM:

-M: Streptavidin-coated microparticles (transparent cap), 1 bottle, 6.5 mL: Streptavidin-coated microparticles 0.72 mg/mL; preservative.

-R1: Anti-h-IgM-Ab-biotin (gray cap), 1 bottle, 9 mL: Biotinylated monoclonal anti-h-IgM antibody (mouse) &gt; 500 µg/L, MES buffer 50 mmol/L, pH 6.5; preservative.

-R2: CMV-Ag-Ru(bpy) (black cap), 1 bottle, 9 mL: CMV-specific antigen (recombinant, E. coli) labeled with ruthenium complex &gt; 50 µg/L; MES buffer 50 mmol/L, pH 5.5; preservative.

{3}

2. CMVIGM Cal1: Negative calibrator 1 (white cap), 2 bottles of 1.0 mL each: Human serum, non-reactive for anti-CMV IgM; preservative.
3. CMVIGM Cal2: Positive calibrator 2 (black cap), 2 bottles of 1.0 mL each: Human serum reactive for anti-CMV IgM; buffer; bovine albumin; preservative.

The following are the materials that are required but not provided:

1. PreciControl CMV IgM, 8 x 1.0 mL each of PreciControl CMV IgM 1 and 2
2. Diluent Universal, 2 x 36 mL sample diluent
3. CalSet Vials, 2 x 56 empty snap-cap bottles
4. General laboratory equipment
5. Elecsys 2010 analyzer
6. Accessories for Elecsys 2010 analyzer

J. Substantial Equivalence Information:

1. Predicate device name(s):

Diamedix Is-CMV Capture Test System

2. Predicate 510(k) number(s):

K001767

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Indications for Use | Immunoassay for the in vitro qualitative detection of IgM antibodies to CMV in human serum, lithium-heparin plasma, K2-EDTA plasma, and K3-EDTA plasma. The test is intended as an aid in the diagnosis of recent or current CMV infection in individuals for which a CMV IgM test was ordered, including pregnant women. Performance characteristics have not been evaluated in immunocompromised or immunosuppressed | The Diamedix Is-CMV IgM Capture Test Kit is a capture enzyme immunoassay (EIA) for the qualitative detection of IgM antibodies to CMV in human serum as an aid in the diagnosis of recent or current infection with CMV. These reagents can be used either manually or in conjunction with the MAGO Plus Automated EIA Processor. This product has not been cleared/approved by the FDA for blood/plasma  |

{4}

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | individuals. This test is not intended for use in neonatal screening or for use at point of care facilities. This assay is not intended for use in screening blood and plasma donors. | donor screening.  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Detection Technology | Electrochemiluminescent Immunoassay | Enzyme-linked immuno-adsorbent assay  |
|  Sample type | Human serum, lithium-heparin plasma, K2-EDTA plasma, and K3-EDTA plasma | Human serum  |

# K. Standard/Guidance Document Referenced (if applicable):

- CLSI EP5-A2: Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline - 2nd Edition
- CLSI EP7-A2: Interference Testing in Clinical Chemistry; Approved Guideline - 2nd Edition
- CLSI EP15-A2: User Verification of Performance for Precision and Trueness; Approved Guideline - 2nd Edition

# L. Test Principle:

$\mu$ -Capture test principle. Total duration of assay: 18 minutes.

- 1st incubation:  $10~\mu \mathrm{L}$  of sample are automatically prediluted 1:20 with Diluent Universal. Biotinylated monoclonal anti-h-IgM-specific antibodies are added.
- 2nd incubation: CMV-specific recombinant antigen labeled with a ruthenium complex (Tris(2,2'-bipyridyl)ruthenium(II)-complex (Ru(bpy)) and streptavidin-coated microparticles are added. Anti-CMV IgM antibodies present in the sample react with the ruthenium-labeled CMV-specific recombinant antigen. The complex becomes bound to the solid phase via interaction of biotin and streptavidin.
- The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier.
- Results are determined automatically by the software by comparing the electrochemiluminescence signal obtained from the reaction product of the sample with the signal of the cutoff value previously obtained by calibration.

{5}

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Precision:

The following precision results were obtained with the Elecsys CMV IgM assay in human serum samples and quality control materials. Within-run precision (repeatability) and total imprecision (intermediate precision) were determined according to CLSI Guideline EP5-A2. All results met the pre-defined acceptance criteria for repeatability and intermediate precision.

Summary of Precision Results for Elecsys CMV IgM

|  Sample | Mean (COI) | Repeatability |   | Intermediate Precision  |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  SD | CV (%) (UCL95%) | SD | CV (%) (UCL95%)  |
|  Control 01 | 0.171 | 0.003 | 1.8 (2.9) | 0.005 | 2.9 (4.4)  |
|  Control 01 | 0.220 | 0.002 | 0.9 (1.7) | 0.007 | 3.2 (4.6)  |
|  Control 02 | 1.98 | 0.040 | 2.0 (2.4) | 0.105 | 5.3 (6.7)  |
|  Control 02 | 2.03 | 0.032 | 1.6 (2.1) | 0.085 | 4.2 (5.6)  |
|  HS B | 0.175 | 0.003 | 1.7 (2.8) | 0.004 | 2.3 (3.9)  |
|  HS A | 0.200 | 0.002 | 1.0 (1.7) | 0.004 | 2.0 (3.7)  |
|  HS 07 | 0.249 | 0.004 | 1.6 (2.4) | 0.007 | 2.8 (3.7)  |
|  HS 04 | 0.841 | 0.010 | 1.2 (1.6) | 0.026 | 3.1 (4.1)  |
|  HS 05 | 0.913 | 0.013 | 1.4 (1.8) | 0.027 | 3.0 (4.0)  |
|  HS 01 | 0.957 | 0.010 | 1.0 (1.2) | 0.031 | 3.2 (4.4)  |
|  HS 06 | 1.13 | 0.018 | 1.6 (2.1) | 0.040 | 3.5 (4.6)  |
|  HS 02 | 1.21 | 0.016 | 1.3 (1.7) | 0.042 | 3.5 (4.8)  |
|  HS D | 1.22 | 0.022 | 1.8 (2.2) | 0.038 | 3.1 (3.7)  |
|  HS C | 1.62 | 0.024 | 1.5 (1.8) | 0.055 | 3.4 (4.3)  |
|  HS F | 4.16 | 0.067 | 1.6 (1.9) | 0.141 | 3.4 (4.3)  |
|  HS 03 | 5.07 | 0.066 | 1.3 (1.6) | 0.157 | 3.1 (4.1)  |
|  HS E | 13.9 | 0.236 | 1.7 (2.1) | 0.445 | 3.2 (3.9)  |

Reproducibility:

The reproducibility was assessed by testing the Elecsys CMV IgM assay on the Elecsys 2010 analyzer at three sites. Imprecision testing was conducted using three replicates in two runs per day for five days consistent with requirements in CLSI EP5-A2 and EP15-A2 with four serum pools and two PreciControl CMV IgM materials. One reagent lot was used for all of the testing at all three sites. The sample panels evaluated included samples close to the cutoff, reactive and non-reactive. All

{6}

sample panels demonstrated a CV % less than 10 % and the results are presented in the tables below.

|  Elecsys CMV IgM between-site reproducibility  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  N = 90 |   | Repeatability |   | Between-day  |   |
|  Sample | Mean COI | SD COI | CV % | SD COI | CV %  |
|  HSPm) 01, Near cutoff | 0.839 | 0.012 | 1.4 | 0.000 | 0.0  |
|  HSP 02, Near cutoff | 1.09 | 0.016 | 1.5 | 0.000 | 0.0  |
|  Elecsys CMV IgM between-site reproducibility  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  N = 90 |   | Repeatability |   | Between-day  |   |
|  Sample | Mean COI | SD COI | CV % | SD COI | CV %  |
|  HSP 03, Reactive | 5.16 | 0.126 | 2.4 | 0.027 | 0.5  |
|  HSP 04, Non reactive | 0.242 | 0.003 | 1.3 | 0.002 | 0.7  |
|  PCn) CMV IgM 1 | 0.207 | 0.003 | 1.6 | 0.003 | 1.4  |
|  PC CMV IgM 2 | 1.85 | 0.042 | 2.3 | 0.000 | 0.0  |

m) HSP = Human serum pool
n) PC = PreciControl

|  Elecsys CMV IgM between-site reproducibility  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  N = 90 |   | Between-site |   | Reproducibility  |   |
|  Sample | Mean COI | SD COI | CV % | SD COI | CV %  |
|  HSP 01, Near cutoff | 0.839 | 0.039 | 4.6 | 0.043 | 5.1  |
|  HSP 02, Near cutoff | 1.09 | 0.047 | 4.3 | 0.052 | 4.8  |
|  HSP 03, Reactive | 5.16 | 0.058 | 1.1 | 0.186 | 3.6  |
|  HSP 04, Non reactive | 0.242 | 0.023 | 9.6 | 0.024 | 9.8  |
|  PC CMV IgM 1 | 0.207 | 0.019 | 9.2 | 0.020 | 9.5  |
|  PC CMV IgM 2 | 1.85 | 0.042 | 2.3 | 0.079 | 4.2  |

b. Linearity/assay reportable range:

Not Applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Reagent Stability:

Four studies were conducted to evaluate the stability of the Elecsys CMV IgM assay

{7}

reagent: reagent stability (unopened) at $2 - 8^{\circ}\mathrm{C}$, reagent stability (opened) at $2 - 8^{\circ}\mathrm{C}$, on-board reagent stability, and reagent stability stored alternately on-board the analyzer and refrigerated.

1. Reagent Stability (Unopened) at $2 - 8^{\circ}\mathrm{C}$: Shelf-life reagent stability for the Elecsys CMV IgM assay was tested on one Elecsys 2010 immunoassay analyzer. Three lots of reagent were evaluated using PreciControl CMV IgM (Levels 1 and 2). A fresh kit was placed on the analyzer and calibrated. Reference values for the PreciControl CMV IgM Levels 1 and 2 were then determined. After measurement, the kit was closed and stored refrigerated $(2-8^{\circ}\mathrm{C})$. The on-test reagent was stored at $2-8^{\circ}\mathrm{C}$ and PC CMV IgM materials tested in duplicate at several time points and the acceptance criteria were met at each time point. The Elecsys CMV IgM reagent was stable unopened at $2 - 8^{\circ}\mathrm{C}$ up to 18 months, which supports the claimed shelf life of 15 months.

2. Reagent Stability (Opened) at $2 - 8^{\circ}\mathrm{C}$: Reagent stability after first opening for the Elecsys CMV IgM assay was tested on an Elecsys 2010 immunoassay analyzer with duplicate measurements. A fresh kit was placed on the analyzer and calibrated. Reference values for samples on-test were then determined. After measurement, the kit was removed from the analyzer and kept at $2-8^{\circ}\mathrm{C}$ for up to 13 weeks (92 days). Calibration was repeated and all samples were placed on the analyzer and re-measured. The acceptance criteria were met for reagent stability after opening at $2 - 8^{\circ}\mathrm{C}$ for 13 weeks, supporting the claim of 12 weeks.

3. On-Board Reagent Stability: On-board reagent stability for the Elecsys CMV IgM assay was tested on one Elecsys 2010 immunoassay analyzer. A fresh kit was placed on the analyzer and calibrated. Reference values for the samples tested were then determined. After measurement, the kits were closed and stored at $20^{\circ}\mathrm{C} \pm 3^{\circ}\mathrm{C}$ to simulate on-board conditions. At specific time points, the kits were placed on the analyzer, calibrated, and the original test samples re-measured. Samples were measured in duplicate with one reagent lot in one run per day. Acceptance criteria were met at each time point for 34 days, supporting the on-board reagent stability claim of 2 weeks.

4. On-Board/Refrigerated Reagent Stability: On-board/refrigerated reagent stability for the Elecsys CMV IgM assay was tested on one Elecsys 2010 immunoassay analyzer by comparing the stressed samples with the reference samples. A fresh kit was placed on the analyzer and calibrated. Reference values for the samples tested were then determined. After measurement, the kit was closed and kept $5 \times 4$ hours on-board $(20^{\circ}\mathrm{C} \pm 3^{\circ}\mathrm{C})$ per week; the rest of the time it was stored refrigerated $(2 - 8^{\circ}\mathrm{C})$. The acceptance criteria were met, supporting reagent stability up to 6 weeks when stored alternately in the refrigerator and on the analyzer.

**Sample Stability:**

Four studies were conducted to evaluate sample stability: sample stability at $2 - 8^{\circ}\mathrm{C}$,

{8}

sample stability at 20 – 25°C, sample stability at -20°C, and sample stability through multiple freeze/thaw cycles.

1. Sample Stability at 2 – 8°C: Reference and on-test materials were tested in triplicate on the Elecsys 2010 analyzer. Samples were measured directly after collection (reference) and then stored at 2-8°C for 35 days. These samples were tested after storage at 2-8°C at multiple time points up to 35 days. The on-test recovery was calculated as a percent of the reference value. The acceptance criteria were met, supporting sample stability up to 4 weeks when stored at 2 – 8°C.

2. Sample Stability at 20 – 25°C: Reference and on-test material (human serum and plasma samples) was tested in triplicate and the controls were tested in duplicate on the Elecsys 2010 analyzer. The samples were measured directly after collection (reference) and then stored at 20-25°C for up to 10 days. The on-test recovery was calculated as a percent of or difference from the reference value. The acceptance criteria were met, supporting claimed sample stability up to 7 days at 20 – 25°C.

3. Sample Stability at -20°C: Reference and on-test material was tested in triplicate on the Elecsys 2010 analyzer. The samples were measured immediately after collection (reference) and then stored frozen at -20°C for approximately 4 months (119 days). The samples were re-measured after 119 days of frozen storage at -20°C. The on-test recovery was calculated as a percent of or difference from the reference value. The acceptance criteria were met, supporting claimed sample stability for 3 months at -20°C.

4. Sample Stability Through Multiple Freeze/Thaw Cycles: Reference and on-test material was tested in triplicate on the Elecsys 2010 analyzer. The samples were measured immediately after collection (reference) and aliquots were frozen/thawed 6 times. A sample measurement was taken after each freeze/thaw cycle and the results compared to the reference result. The acceptance criteria were met, supporting claimed sample stability through 5 freeze/thaw cycles.

## Calibration Stability:

Two studies were conducted to evaluate calibration stability: lot calibration stability and on-board calibration stability.

1. Lot Calibration Stability: The lot calibration stability was determined by comparing the calibration for two kits of the same lot. On Day 1, the first reagent kit was opened and calibrated, and samples were measured at Days 1 (reference), 29, and 36 on one Elecsys 2010 analyzer. All samples were measured in duplicate with one reagent in one run per day on Days 1, 29, and 36. The acceptance criteria were met for lot calibration stability of 28 days on the Elecsys 2010 analyzer.

9

{9}

2. On-Board Calibration Stability: On-board calibration stability for the Elecsys CMV IgM test system was tested on one Elecsys 2010 analyzer. One reagent kit was opened and samples were measured on Day 1 (reference). The same samples were then retested after 15 days with reagent bottles stored at $20 \pm 3^{\circ}\mathrm{C}$ (on-board condition) using the calibration from Day 1. Recovery was calculated based on the initial (reference) values. All samples were tested in duplicate with one reagent lot on one Elecsys 2010 analyzer in one run per day on two days. The acceptance criteria were met for on-board calibration stability of 8 days on the Elecsys 2010 analyzer.

## Calibrator Stability:

Two studies were conducted to evaluate calibrator stability: calibrator stability at $2 - 8^{\circ}\mathrm{C}$ and open vial calibrator stability.

1. Calibrator Stability at $2 - 8^{\circ}\mathrm{C}$: Reference and on-test materials were tested in duplicate with one reagent lot in one run per day on one Elecsys 2010 analyzer. The on-test material was opened and stored at $2 - 8^{\circ}\mathrm{C}$ for 8 weeks. It was measured at Day 1 (reference) and again at 8 and 10 weeks. The on-test recovery was calculated as the signal (counts) of the reference value. The acceptance criteria were met for calibrator stability at $2 - 8^{\circ}\mathrm{C}$ for 8 weeks.

2. Open Vial Calibrator Stability: Reference and open vial, on-test materials were tested in duplicate with one reagent lot on one Elecsys 2010 analyzer. The on-test material was opened and stored at $25^{\circ}\mathrm{C}$ for 6 hours to simulate on-board conditions. Open vial (on-test) material was also stored at $32^{\circ}\mathrm{C}$ for 6 hours on one Elecsys 2010 analyzer to simulate on-board conditions. Every hour, each calibrator was tested and the on-test recovery of signal (counts) was calculated as a percent of the reference value. The acceptance criteria were met for open vial calibrator stability for 5 hours on the Elecsys 2010 analyzer.

## Preci-Control Value Assignment:

The Elecsys CMV IgM calibrators and Elecsys PreciControl CMV IgM values are assigned through at least six independent series of analysis performed on at least three analyzers. All samples are tested in duplicate. The sample recovery (COI) is calculated as the median of each sample as reference to the target value.

## Stability of PreciControl CMV IgM:

Three studies were conducted to evaluate the stability of the Elecsys PreciControl CMV IgM materials: control stability at $2 - 8^{\circ}\mathrm{C}$, open vial control stability, and shelf-life stability.

1. Control Stability at $2 - 8^{\circ}\mathrm{C}$: Stressed and unstressed samples of PreciControl CMV IgM were tested in duplicate with one reagent lot in one run per day on one Elecsys 2010 analyzer. All samples met the acceptance criteria, supporting the claimed PreciControl CMV IgM stability for 8 weeks when stored at $2 - 8^{\circ}\mathrm{C}$.

10

{10}

2. Control Open Vial Stability: Reference and open vial, on-test controls were tested in singlicate with one reagent lot on one Elecsys 2010 analyzer. The on-test material was opened, kept at 25°C on the Elecsys 2010 to simulate on-board conditions, and tested every hour for 6 hours. The on-test recovery of signal (counts) was calculated as a percent of the reference value. The acceptance criteria were met, supporting the claimed PreciControl CMV IgM open vial stability of 5 hours on the Elecsys 2010 analyzer.

3. Control Shelf-Life Stability: The on-test material was stored at 2-8°C and tested in duplicate on the Elecsys 2010 analyzer at several time points during and beyond the shelf life time of 18 months. Stability was measured by comparing the measurements of the stressed controls with the measurements of the unstressed controls (reference). The acceptance criteria were met at each time point, supporting the claimed PreciControl CMV IgM shelf-life stability of 15 months.

d. Detection limit:

Not Applicable

e. Analytical specificity:

Cross-Reactivity:

A study was conducted to evaluate the Elecsys CMV IgM for potential cross-reactivity using samples from individuals with antibodies to various medical conditions. Specimens (n = 205) were tested in duplicate with the Elecsys CMV IgM assay and a comparator CMV IgM assay. The first result was used for the performance evaluation. The presence of the potential cross-reactants in the samples tested was confirmed using FDA-cleared devices or laboratory methods. Potential cross-reactivity with autoimmune markers and antibodies against influenza vaccination could not be ruled out from the study.

11

{11}

Cross-Reactivity Test Results

|  Potential Cross-Reacting antibodies and disease conditions | Comparator CMV IgM Results (Negative) |   |   | Comparator CMV IgM Results (Positive)  |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  Elecsys CMV IgM Non-Reactive | Elecsys CMV IgM Indeterminate (Border) | Elecsys CMV IgM Reactive | Elecsys CMV IgM Non-Reactive | Elecsys CMV IgM Indeterminate (Border) | Elecsys CMV IgM Reactive  |
|  Autoimmune | 5 | 0 | 2 | 0 | 0 | 0  |
|  EBV | 14 | 0 | 0 | 0 | 0 | 0  |
|  E. coli | 5 | 0 | 0 | 0 | 0 | 0  |
|  HCV | 27 | 0 | 0 | 0 | 0 | 0  |
|  HSV | 4 | 0 | 0 | 1 | 0 | 0  |
|  HTLV | 43 | 0 | 0 | 0 | 0 | 0  |
|  Rubella | 10 | 0 | 0 | 0 | 0 | 0  |
|  HAV | 10 | 0 | 0 | 0 | 0 | 0  |
|  HBV | 18 | 0 | 0 | 1 | 0 | 0  |
|  HIV | 20 | 0 | 0 | 1 | 0 | 0  |
|  Influenza Vaccine | 9 | 0 | 1 | 0 | 0 | 0  |
|  Treponema palladium | 4 | 0 | 0 | 0 | 0 | 0  |
|  Toxoplasma gondii | 9 | 0 | 0 | 0 | 0 | 0  |

12

{12}

# Interferences:

To evaluate the effect of elevated levels of hemoglobin, bilirubin, Intralipid, and biotin on the CMV IgM assay, eight CMV IgM samples (negative, near cutoff and positive) were spiked with the potential interferents. Each interferent was evaluated at 11 numerical values. All samples were tested in singlicate on the Elecsys 2010 analyzer. Acceptance criteria for hemoglobin, bilirubin, Intralipid, and biotin: samples with  $\mathrm{COI} &lt; 0.7$  to be found non-reactive, and samples with  $\mathrm{COI} \geq 0.7$  to give recovery of  $\pm 15\%$ .

To evaluate the effect of rheumatoid factor (RF), 34 samples were spiked with different concentrations of RF and measured in singlicate on the Elecsys 2010 analyzer as well as a comparator device. The acceptance criterion for RF was agreement  $&gt;90\%$  against the comparator device.

The results of the interference studies are presented below:

|  Interferent tested | No interference up to (Concentration)  |
| --- | --- |
|  Hemoglobin | < 0.621 mmol/L or < 1.0 g/dL  |
|  Bilirubin | < 342 μmol/L or < 20 mg/dL  |
|  Intralipid | < 1500 mg/dL  |
|  Biotin | < 410 nmol/L or < 100 ng/mL  |
|  Rheumatoid factor | < 2000 IU/mL  |

# f. Assay cut-off:

The cut-off for the Elecsys CMV IgM assay was established with in-house studies by characterizing samples using several commercially available CMV IgG and CMV IgM assays. The specificity of the assay was evaluated using 931 samples from a low prevalence cohort. The assay cutoff was optimized to increase sensitivity by measuring 152 samples from individuals at different stages of primary CMV infection. Validation of the assay cutoff was performed by external clinical studies.

The classification of samples, based on the cut-off establishment, verification, and validation, is as follows:

Sample results  $&lt; 0.7$  COI = Non-Reactive Sample
Sample results  $\geq 0.7$  to  $&lt; 1.0$  COI = Indeterminate (Border) Sample
Sample results  $\geq 1.0$  COI = Reactive Sample

# g. High Dose Hook-Effect:

Testing with the Elecsys CMV IgM assay demonstrated no high dose hook effect. Five highly positive, high-titer CMV IgM human samples were diluted with negative human serum in a dilution series with 11 steps. Each dilution was tested in triplicate

{13}

with one reagent lot in one run on one Elecsys 2010 analyzer. No high dose hook effect was observed up to 32.9 COI.

# h. IgM Specificity:

To confirm that the Elecsys CMV IgM assay specifically detects IgM-class antibodies, 11 samples with moderate to high levels of CMV IgM antibodies were selected for testing. These samples were treated with dithiothreitol (DTT) to destroy the IgM and then were retested using the Elecsys CMV IgM assay. Ten samples were rendered non-reactive (negative) following treatment with DTT, and one sample had an  $82\%$  decrease in COI value, confirming the specificity of the Elecsys CMV IgM assay for detecting IgM-class antibodies. The Elecsys CMV IgM assay is designed to specifically bind the anti-CMV IgM antibodies due to the  $\mu$ -capture design and antigens minimizing interference from CMV IgG antibodies.

# 2. Comparison studies:

# a. Method comparison between Analyzers:

The equivalence of the Elecsys CMV IgM assay on the Elecsys 2010 and MODULAR ANALYTICS E170 immunoassay analyzers was evaluated by a method comparison study. Native serum samples were tested on one Elecsys 2010 analyzer and one MODULAR ANALYTICS E170 analyzer. Positive and negative agreement of the results between the two platforms was calculated and demonstrated equivalence between the analyzer platforms for the determination of IgM antibodies to CMV using the Elecsys CMV IgM assay:

Concordance Rates:

Negative Percent Agreement (NPA) = 97.98 %

Positive Percent Agreement (PPA) = 99.22%

# b. Matrix comparison:

Studies were conducted to evaluate the suitability of the following types of samples to be used with the Elecsys CMV IgM immunoassay: Li-heparin plasma,  $\mathrm{K}_2$ -EDTA plasma,  $\mathrm{K}_3$ -EDTA plasma, and serum separator tubes. Samples were collected into matched serum and plasma collection tubes and assayed in duplicate. These studies were conducted using negative (non-reactive), near cutoff (indeterminate/border) and positive (reactive) samples. The results support the use of the following sample types: Li-heparin plasma,  $\mathrm{K}_2$ -EDTA plasma,  $\mathrm{K}_3$ -EDTA plasma, and serum separator tubes. Acceptance specifications are tabulated below.

|  Sample Matrix | Mean COI | Percent of samples showing differences in recovery relative to serum (COI) in non-reactive specimens  |   |   |
| --- | --- | --- | --- | --- |
|   |   |  < 0.07 COI | 0.07 – 0.1 COI | > 0.1 COI  |
|  Li-Heparin Plasma | 0.196 | 93 % | 7 % | 0 %  |
|  K2-EDTA Plasma | 0.280 | 100 % | 0 % | 0 %  |
|  K3-EDTA Plasma | 0.195 | 97 % | 3 % | 0 %  |
|  Serum Separator Tubes | 0.265 | 100 % | 0 % | 0 %  |

{14}

|  Sample Matrix | Mean COI | Percent of samples showing differences in recovery relative to serum (COI) in reactive specimens  |   |   |
| --- | --- | --- | --- | --- |
|   |   |  < 10 % | 10 – 20 % | > 20 %  |
|  Li-Heparin Plasma | 1.51 | 68 % | 32 % | 0 %  |
|  K2-EDTA Plasma | 2.12 | 100 % | 0 % | 0 %  |
|  K3-EDTA Plasma | 1.72 | 79 % | 21 % | 0 %  |
|  Serum Separator Tubes | 2.27 | 90 % | 10 % | 0 %  |

3. Clinical studies:

a. Clinical Sensitivity:

Not Applicable

b. Clinical specificity:

Not Applicable

c. Other clinical supportive data (when a. and b. are not applicable):

A multi-center study was conducted in the US to evaluate the performance of the Elecsys CMV IgM immunoassay in comparison to a consensus result derived from 2 or 3 FDA-cleared devices (depending on the availability of sample volume for testing). A two-out-of-three approach was used to determine the consensus result for samples where three FDA-cleared devices could be tested. Where only two FDA-cleared devices could be tested, results for which the FDA-cleared devices disagreed were omitted from analysis. All equivocal results obtained using the Elecsys CMV IgM assay were treated as not in favor of the Elecsys CMV IgM assay.

The main study was a prospective study consisting of 617 samples, all of which were from a population of patients suspected of CMV infection for whom a CMV IgM test was ordered. A subset of this population (n = 199) were pregnant women. In addition, a retrospective study was conducted, consisting of 134 samples that were preselected from a population of patients positive for CMV IgM as determined by the comparator device consensus of 3 FDA-cleared CMV IgM assays. Testing of specimens was done at three clinical sites and one internal site. The results of these studies are presented in the following tables.

{15}

16

# Prospectively Collected Suspected Infection Population

|   |  | Comparator Device Consensus
CMV IgM Result  |   |   |   |
| --- | --- | --- | --- | --- | --- |
|   |  | Positive | Equivocal | Negative | Total  |
|  Elecsys | Reactive | 20 | 0 | 6 | 26  |
|  CMV | Indeterminate | 0 | 0 | 5 | 5  |
|  IgM | (Border) | 0 | 3¹ | 384 | 387  |
|  Result | Non-Reactive | 20 | 3 | 395 | 418  |
|  Agreement | Numerator/ | Percent Agreement |   | 95% Confidence  |   |
|  Classification | Denominator | (%) |   | Interval  |   |
|  Negative Agreement | 384/395 | 97.2 |   | 95.1 – 98.6  |   |
|  Positive Agreement | 20/23 | 87.0 |   | 66.4 – 97.2  |   |

# Prospectively Collected Pregnant Population

|   |  | Comparator Device Consensus
CMV IgM Result  |   |   |   |
| --- | --- | --- | --- | --- | --- |
|   |  | Positive | Equivocal | Negative | Total  |
|  Elecsys | Reactive | 0 | 0 | 2 | 2  |
|  CMV | Indeterminate | 1 | 0 | 1 | 2  |
|  IgM | (Border) | 0 | 5 | 190 | 195  |
|  Result | Non-Reactive | 1 | 5 | 193 | 199  |
|  Agreement | Numerator/ | Percent Agreement |   | 95% Confidence  |   |
|  Classification | Denominator | (%) |   | Interval  |   |
|  Negative Agreement | 190/193 | 98.5 |   | 95.5 – 99.7  |   |
|  Positive Agreement | 0/6 | 0.00 |   | 0.00 – 45.9  |   |

# Preselected Confirmed CMV IgM Positive Population

|   |  | Comparator Device Consensus
CMV IgM Result  |   |   |   |
| --- | --- | --- | --- | --- | --- |
|   |  | Positive | Equivocal | Negative | Total  |
|  Elecsys | Reactive | 134 | 0 | 0 | 134  |
|  CMV | Indeterminate | 0 | 0 | 0 | 0  |
|  IgM | (Border) | 0 | 0 | 0 | 0  |
|  Result | Non-Reactive | 0 | 0 | 0 | 0  |
|   | Total | 134 | 0 | 0 | 134  |

¹ These samples were Equivocal according to the consensus result and Non-reactive (negative) by the Elecsys CMV IgM assay. As such, they were treated as Elecsys CMV IgM Reactive (positive) in the analysis (i.e., not in favor of the Elecsys CMV IgM assay).

{16}

|  Agreement Classification | Numerator/ Denominator | Percent Agreement (%) | 95% Confidence Interval  |
| --- | --- | --- | --- |
|  Negative Agreement | N/A  |   |   |
|  Positive Agreement | 134/134 | 100.0 | 97.3 – 100.0  |

# 4. Clinical cut-off:

Not Applicable

# 5. Expected values/Reference range:

The observed expected values in the prospectively collected US clinical study population using the Elecsys CMV IgM immunoassay were as follows:

|   |   | Elecsys CMV IgM Result  |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Age Group (years) | Sex | Reactive N (%) | Indeterminate (Border) N (%) | Non-Reactive N (%) | Total (N)  |
|  18 – 19 | Female | 0 (0.00) | 0 (0.00) | 2 (100.00) | 2  |
|   |  Male | 0 (0.00) | 0 (0.00) | 5 (100.00) | 5  |
|  20 – 29 | Female | 2 (1.96) | 1 (0.98) | 99 (97.06) | 102  |
|   |  Male | 0 (0.00) | 2 (4.44) | 43 (95.56) | 45  |
|  30 – 39 | Female | 2 (1.23) | 2 (1.23) | 159 (97.55) | 163  |
|   |  Male | 1 (1.30) | 2 (2.60) | 74 (96.10) | 77  |
|  40 – 49 | Female | 1 (3.03) | 0 (0.00) | 32 (96.97) | 33  |
|   |  Male | 0 (0.00) | 0 (0.00) | 70 (100.00) | 70  |
|  50 – 59 | Female | 2 (12.50) | 0 (0.00) | 14 (87.50) | 16  |
|   |  Male | 1 (1.67) | 0 (0.00) | 59 (98.33) | 60  |
|  60 – 69 | Female | 0 (0.00) | 0 (0.00) | 6 (100.00) | 6  |
|   |  Male | 1 (5.56) | 0 (0.00) | 17 (94.44) | 18  |
|  70 – 79 | Male | 0 (0.00) | 0 (0.00) | 2 (100.00) | 2  |
|  Totals | Female | 7 (2.17) | 3 (0.93) | 312 (96.89) | 322  |
|   |  Male | 3 (1.08) | 4 (1.44) | 270 (97.47) | 277  |
|   |  Overall | 10 (1.67) | 7 (1.17) | 582 (97.16) | 599  |

# N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

# O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K142133](https://fda-staging.innolitics.com/device/K142133)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
