The ORS drape is a single-use equipment cover intended for use during various surgeries where warm irrigation, slush and/or cold solution is required.
Device Story
Non-Sterile ORS Warming and Slush Drapes are single-use polyurethane film covers; designed to protect surgical fluid warmers and slush machines from contamination during surgical procedures. Drapes are supplied in bulk to secondary processors/kit packers who perform final packaging, labeling, and sterilization. Device is used in clinical settings (OR); operated by surgical staff. Output is a physical barrier/drape; ensures equipment remains sterile/protected during use. Benefits include maintenance of sterile field and protection of surgical equipment.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation testing confirmed dimensional requirements, integrity of sterile barrier (no holes/defects), attachment to slush machines, and packaging durability per ASTM D4169-09.
Technological Characteristics
Materials: Polyurethane film (polypropylene skirt on model ORS-188231N). Form factor: Various sizes/shapes to fit specific surgical fluid warmers and slush machines. Connectivity: None. Sterilization: Not sterile (intended for sterilization by secondary processor).
Indications for Use
Indicated for use as a single-use equipment cover during surgeries requiring warm irrigation, slush, or cold solutions. No specific patient population or contraindications described.
Regulatory Classification
Identification
A powered heating unit is a device intended for medical purposes that consists of an encased cabinet containing hot water and that is intended to heat and maintain hot packs at an elevated temperature.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and features the department's name around the perimeter. In the center of the logo is an abstract symbol of a human figure, represented by three overlapping profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 28, 2014
Ecolab. Incorporated Jennifer Willner Director, Regulatory Affairs - Healthcare 370 Wabasha Street North St. Paul, Minnesota 55102
Re: K142080
Trade/Device Name: Non-Sterile ORS Warming and Slush Drapes Regulatory Class: Unclassified Product Code: LHC Dated: July 30, 2014 Received: July 31, 2014
Dear Ms. Willner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Erin I. Keith -S
Erin Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K142080
Device Name
Non-Sterile ORS Warming and Slush Drapes.
Indications for Use (Describe)
The ORS drape is a single-use equipment cover intended for use during various surgeries where warm irrigation, slush and/or cold solution is required.
The following models are included: ORS-100N, ORS-300N, ORS-301N, ORS-320N, ORS-330N and ORS-188231N.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Sajjad H. Syed
Digitally signed by Saiiad H. Sved -S DN: c=US. o=U.S. Government. ou=HHS. ou=FDA. ou=People. cn=Saijad H. Syed -S. 0.9.2342.19200300.100.1.1=2000601742 Date: 2014.08.26 17:00:38 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K140280 – 510(K) SUMMARY
{4}------------------------------------------------
Non-Sterile ORS Warming and Slush Drapes
Image /page/4/Picture/1 description: The image shows the logo for Ecolab Inc. The text "Ecolab Inc." is at the top of the image in a small, dark font. Below that is the Ecolab logo in a large, blue font. The logo includes a stylized image of a four-bladed propeller.
Section 5: 510(k) Summary
# Non-Sterile ORS Warming and Slush Drapes
As required by 21 CFR 807.92. Date: July 30, 2014 Administrative Information Submitter: Ecolab, Inc. Establishment
Registration Number:
| Registration Number: | 1043582 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jennifer Willner, RAC<br>370 Wabasha Street North<br>St. Paul, MN 55102-1390<br>Director, Regulatory Affairs - Healthcare<br>651.250.4348 |
Device Identification
| Device Name: | Non-Sterile ORS Warming and Slush Drapes |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Equipment Cover |
| Device Classification Name: | Warmer, Irrigation Solution |
| Device Classification: | Unclassified |
| Classification Product Code: | LHC |
| Panel: | General and Plastic Surgery |
| Classification Regulation: | Pre-amendment |
| Performance Standards: | No Recognized Consensus Standards |
| Predicate Device: | ORS-1000LD [covers ORS-100, ORS-300, ORS-301,<br>ORS-188231] cleared on 06/27/2002 via K021288 and<br>ORS-3000LD [covers ORS-320, ORS-321, ORS-330]<br>cleared 11/13/2002 via K023282 |
{5}------------------------------------------------
Non-Sterile ORS Warming and Slush Drapes
Ecolab Inc.
#### Device Description
The Non-Sterile ORS Warming and Slush Drapes are single-use equipment covers intended for use during various surgeries where warm irrigation, slush and/or cold solution is required. These equipment drapes consist of a polyurethane film base material manufactured to protect ORS surgical fluid warmers and slush machines from contamination during various procedures throughout the clinical setting.
These equipment drapes come in a variety of sizes and shapes to allow the device to properly fit ORS surgical fluid warmers and slush machines.
#### Special 510(k) Discussion
This Special 510(k) submission requests clearance for the manufacture and distribution of non-sterile ORS drapes from an identical subset of cleared sterile ORS devices. The nonsterile drapes will be sold in bulk packaging to other secondary processors, including kit packers, who will then package, label and sterilize the drapes prior to distribution to end users. The fundamental scientific technology of these equipment drapes remains unchanged. The Substantial Equivalence Table (Table 5-1) is provided below.
| Property or<br>Characteristic | Proposed Device | Predicate Device<br>(Warming Drapes) | Predicate Device<br>(Slush Drapes) |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| 510(k) No. | This 510(k)<br>Submission | K021288 | K023282 |
| Device Name | Non-Sterile ORS<br>Warming and<br>Slush Drapes<br>(includes: ORS-<br>100N, ORS-300N,<br>ORS-301N, ORS-<br>320N, ORS-321N,<br>ORS-330N and<br>ORS-188231N) | ORS-1000LD<br>(includes: ORS-100,<br>ORS-300, ORS-301,<br>and ORS-188231) | ORS-3000LD<br>(includes: ORS-320,<br>ORS-321, and ORS-<br>330) |
| Indications for<br>Use | The ORS drape is<br>a single-use<br>equipment cover<br>intended for use<br>during various<br>surgeries where | The ORS-1000LD<br>Leak Detection<br>Drape is an<br>equipment cover for<br>the ORS-2000LD<br>Solution Warmer. | The ORS-3000LD is<br>an equipment cover<br>for the ORS-<br>1075LD Hush-<br>Slush® machine.<br>This is a single use |
Table 5-1: Substantial Equivalence
{6}------------------------------------------------
| Property or<br>Characteristic | Proposed Device | Predicate Device<br>(Warming Drapes) | Predicate Device<br>(Slush Drapes) |
|-------------------------------|------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| | warm irrigation,<br>slush and/or cold<br>solution is<br>required. | This is a single use<br>product supplied<br>sterile. This device<br>is intended for use<br>during various<br>surgeries where<br>warm irrigation<br>solution is required. | product supplied<br>sterile. This device<br>is intended for use<br>during various<br>surgeries where<br>slush and/or cold<br>solution is required. |
| Conditions of<br>Use | Rx Only, Single<br>Use, Disposable | Identical | Identical |
| Materials | Polyurethane Film<br>Polypropylene<br>skirt for ORS-<br>188231N only | Identical | Identical |
| Principle of<br>Operation | Covers surgical<br>solution warmers<br>and/or slush<br>machines | Covers surgical<br>solution warmers | Covers surgical<br>solution warmers<br>and/or slush<br>machines |
| Packaging | Bulk packaged in<br>poly bag in<br>quantities up to 24 | Individually<br>packaged in<br>poly/Tyvek peel<br>pouches | Individually<br>packaged in<br>poly/Tyvek peel<br>pouches |
| Sterilized | No; intended to be<br>sterilized before<br>distribution to end<br>user | Yes; provided in<br>sterile condition via<br>EO at SAL 10-6 | Yes; provided in<br>sterile condition via<br>EO at SAL 10-6 |
#### Statement of Equivalence
The Non-Sterile ORS Warming and Slush Drapes perform as intended using the identical principles of operation as the predicate device(s). Differences between the Non-Sterile ORS Warming and Slush Drapes and the sterile versions do not raise any new questions of safety or efficacy when secondary processors finish the product as instructed. Based on the risk analysis, review of the product labeling, and successful performance and safety testing, the Non-Sterile ORS Warming and Slush Drapes are substantially equivalent to the legally marketed Sterile ORS Warming and Slush Drapes (K021288 and K023282). The fundamental scientific technology of the device remains unchanged.
{7}------------------------------------------------
# Ecolab Inc.
| Performance Data Summary | | | |
|-------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Table 5-2: Performance Data Summary of the Non-Sterile ORS Warming and Slush Drapes | | | |
| Requirement | Specification | Method | Result |
| Functional<br>Performance<br>Requirements<br>(of selected<br>drape types) | Dimensional Requirements:<br>ORS-301N (66 in. x 52 in.)<br>ORS-321N (66 in x 52 in.)<br>ORS-330N (66 in. x 44 in.)<br>ORS-188231N, Skirted<br>(44 in. x 44 in. x 36 in.) | Representative drapes were chosen to<br>cover all drape types. Drapes selected:<br>ORS-301N (largest size), ORS-<br>188231N (only skirted drape), ORS-<br>321N (largest size Disc drape), ORS-<br>330N (largest Plate drape)<br>Acceptable results following visual<br>inspection during V&V testing; VVR-<br>14-0002 V&V Summary Report,<br>Project 8 Track | Pass |
| | Drape remains intact (free<br>from holes or other defects<br>that would compromise the<br>sterile barrier) | Acceptable results following visual<br>inspection during V&V testing:<br>VVR-14-0002 V&V Summary<br>Report, Project 8 Track | Pass |
| | Disc/Plate must stay attached<br>to drape during use | Acceptable results following V&V<br>testing (based on sterile product<br>testing); VVR-14-0002 V&V<br>Summary Report, Project 8 Track | Pass |
| | Drape/Device must attach to<br>Slush Machine | Acceptable results following V&V<br>testing (based on sterile product<br>testing); VVR-14-0002 V&V<br>Summary Report, Project 8 Track | Pass |
| Packaging | Packaging Configuration: 24<br>per case, double poly bagged | Packaging Configuration per:<br>ORS-301N_DWG<br>ORS-321N_DWG<br>ORS-330N_DWG<br>ORS-188231N_DWG; acceptable<br>results following visual inspection<br>during V&V testing; documented in<br>V&V Summary Report VVR-14-0002 | Pass |
| | Product must be received by<br>customer with folds intact | Acceptable results following visual<br>inspection during V&V testing:<br>documented in V&V Summary Report<br>VVR-14-0002 | Pass |
| | Simulated Distribution Test | ASTM D4169-09 (Distribution Cycle<br>2. Assurance Level 1) Simulated<br>Distribution Test (PKG 001F) | Pass |
| Requirement | Specification | Method | Result |
| Labeling | Inspection for Drape<br>Damage | documented in Packaging Engineering<br>Report # REPT-18430<br>Acceptable results following visual<br>inspection during V&V testing;<br>documented in V&V summary report<br>VVR-14-0002 | Pass |
| | Master carton label is present<br>and per specification | Master label specifications:<br>ORS-301NMASTER<br>ORS-321NMASTER<br>ORS-330NMASTER<br>ORS-188231NMASTER<br>Acceptable results following visual<br>inspection during V&V testing;<br>documented in V&V summary report<br>VVR-14-0002 | Pass |
| | Ink (non-smudge/smear) | Acceptable results following visual<br>inspection during V&V testing;<br>documented in V&V summary report<br>VVR-14-0002 | Pass |
| | Insert Sheet/IFU is present<br>and per specification | Insert Sheet/IFU specifications:<br>1) ORS-WARMER DRAPE_2014-<br>5/NEW INSERT SHEET (ORS-<br>100N, ORS-300N, ORS-301N,<br>ORS-188231N)<br>2) ORS-320N/ORS-321N_2014-<br>5/NEW SLUSH/WARMER<br>DISC-DRAPE INSERT SHEET<br>3) ORS-330N_2014-5/NEW<br>SLUSH/WARMER PLATE<br>DRAPE INSERT SHEET | Pass |
| | Product Identification label<br>includes product code and<br>Ecolab Logo | Product Identification label<br>specification:<br>ORS-301NMISC<br>ORS-188231NMISC<br>ORS-321NMISC<br>ORS-330NMISC | Pass |
{8}------------------------------------------------
## Non-Sterile ORS Warming and Slush Drapes
Ecolab Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.