The Emprint™ Procedure Planning Application is a stand-alone software product that allows physicians to visualize and compare CT imaging data. The display, annotation, and volume rendering of medical images aids in the intervention planning for videoassisted thoracoscopic surgery (VATS) and ablation procedures using the Emprint Ablation System. The software is not intended for diagnosis.
Device Story
Emprint Procedure Planning Application is stand-alone software on Windows-based workstation; imports DICOM CT image sets. Physician-operated; provides 3D rendering, annotation, and measurement tools. Enables pre-procedure planning and post-procedure review for VATS and thermal ablation. Overlays anticipated thermal ablation zones based on Emprint Ablation System settings (K133821). Output aids clinical decision-making by visualizing anatomy and ablation zones; does not control medical devices or provide diagnosis. Benefits include improved procedural planning and visualization for interventional thermal ablation.
Clinical Evidence
Bench testing only. System performance evaluated via verification and validation test plans to confirm all software functions met design specifications.
Technological Characteristics
Stand-alone software on Windows-based computer workstation. Imports DICOM CT data. Features: axial/coronal/sagittal views, 3D volume rendering, annotation, measurement, and ablation zone overlay. No direct patient contact.
Indications for Use
Indicated for physicians to visualize and compare CT imaging data to aid in intervention planning for video-assisted thoracoscopic surgery (VATS) and ablation procedures using the Emprint Ablation System. Not intended for diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three heads facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 17, 2014
Covidien LLC % Ms. Heather V. Nigro Global Senior Director, Regulatory Affairs 15 Hampshire Street MANSFIELD MA 02048
Re: K142048
Trade/Device Name: Emprint™ Procedure Planning Application Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 29, 2014 Received: July 30, 2014
Dear Ms. Nigro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Smh.f)
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142048
Device Name Emprint Procedure Planning Application
#### Indications for Use (Describe)
The Emprint™ Procedure Planning Application is a stand-alone software product that allows physicians to visualize and compare CT imaging data. The display, annotation, and volume rendering of medical images aids intervention planning for video-assisted thoracoscopic surgery (VATS) and ablation procedures using the Emprint Ablation System. The software is not intended for diagnosis.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Covidien. The logo consists of a blue square with a white rectangle in the center and a blue vertical line running through it. To the right of the square is the word "COVIDIEN" in blue, sans-serif font.
## 510(k) Summary
Date summary prepared: August 19, 2014
### 510(k) Submitter/Holder
Covidien, Inc. 15 Hampshire Street Mansfield, MA 02048
### Contact
Heather V. Nigro Global Senior Director, Regulatory Affairs Telephone: 508-452-1936 Fax: 508-452-1941 Email: heather.nigro@covidien.com
### Name of Device
| Trade Name: | Emprint™ Procedure Planning Application |
|----------------------|------------------------------------------------------------------------------|
| Common Name: | Radiological Image Processing System |
| Classification Name: | System, Image Processing, Radiological (21 CFR §892.2050,<br>Class II, LLZ). |
### Purpose of Submission
The purpose of this submission is to gain clearance for a new procedure planning application.
## Predicate Devices
Emprint™ Procedure Planning Application described in this submission is substantially equivalent to the following commercially available predicate devices:
| Trade Name: | LiverAnalysis/LiverViewer Software |
|---------------------|-----------------------------------------|
| Device Common Name: | Picture Archiving Communications System |
| 510(k) Number: | K051528 |
| Manufacturer: | MeVis |
| Trade Name: | Table |
| Device Common Name: | Radiological Image Processing System |
| 510(k) Number: | K140093 |
### Device Description
The Emprint™ Procedure Planning Application is a stand-alone software product that is intended to be used to view and compare CT image sets. The system is composed of
{4}------------------------------------------------
image review and planning software and a Windows-based computer. The system includes tools that provide 3-D rendering of the image sets, pre-procedure planning of interventional thermal ablation procedures, and post-procedure review of interventional thermal ablation procedures. The Emprint™ Procedure Planning Application does not prescribe therapy and is not intended for the diagnosis or treatment of any disease. It does not control or alter the functions or parameters of any medical device and has no direct patient contact.
## Intended Use
The Emprint™ Procedure Planning Application is a stand-alone software product that allows physicians to visualize and compare CT imaging data. The display, annotation, and volume rendering of medical images aids in the intervention planning for videoassisted thoracoscopic surgery (VATS) and ablation procedures using the Emprint Ablation System. The software is not intended for diagnosis.
## Technological Characteristics
The Emprint™ Procedure Planning Application is a stand-alone software product installed on a Windows-based computer workstation. The system imports DICOM data from CT scanners and can display the images in standard axial, coronal, or sagittal views and render the images in 3-D views. The substantial equivalence of the Emprint™ Procedure Planning Application to the predicates is shown by similarity in intended use, indications for use, materials, and performance.
# Principals of Operation
The Emprint™ Procedure Planning Application is composed of software running on a standard Windows-based computer that allows the user to import multiple DICOM compatible CT image sets, render them into 3-D and compare them. The software also provides tools to mark and measure anatomical features and to overlay anticipated thermal ablation zones as defined by the ablation tables associated with the Emprint Ablation System settings for a given type of tissue (provided in K133821 labeling and also included as an Attachment to this 510k Summary).
## Performance Data
The Emprint™ Procedure Planning Application was tested in accordance with a test plan to evaluate all functions performed by the software as configured on the computer workstation. The system was tested and passed all criteria established by the design specifications and verification/validation test plans.
### Substantial Equivalence Discussions
Based on the indications for use, technological characteristics, performance testing, and comparison to the predicates, the Emprint™ Procedure Planning Application has been
{5}------------------------------------------------
shown to be substantially equivalent to the predicate devices identified in this submission, and does not present any new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.