K142025 · Neurolumen, LLC · ILY · Jan 21, 2015 · Physical Medicine
Device Facts
Record ID
K142025
Device Name
NEUROLUMEN PN-1000
Applicant
Neurolumen, LLC
Product Code
ILY · Physical Medicine
Decision Date
Jan 21, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Neurolumen is indicated for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm, relieving stiffness, promoting relaxation of muscle tissue, and to temporarily increase local blood circulation where heat is indicated. The Neurolumen is also indicated for the symptomatic relief and management of chronic, intractable pain and adjunctive treatment for post-surgical and post-traumatic acute pain.
Device Story
Neurolumen PN-1000 is a medical therapy device delivering Transcutaneous Electrical Nerve Stimulation (TENS) and Low Level Laser Therapy (LLLT) to patient limbs. Device consists of a control unit and six wraps; each wrap contains two laser diodes, four light-emitting diodes (LEDs), and TENS gel pads. Operated by patient or clinician; wraps applied to feet, ankles, and legs. Control unit powered by internal rechargeable lithium-ion battery. Device provides therapy for up to 30 minutes per charge. Healthcare providers use the device to manage pain and stiffness; output affects clinical decision-making by providing non-pharmacological pain management. Benefits include temporary pain relief, increased local circulation, and muscle relaxation.
Clinical Evidence
Bench testing only. Laser diodes tested on three subjects to demonstrate capability to raise skin temperature between 40-45 degrees C for at least 10 minutes. Results confirmed unit maintains these temperatures.
Technological Characteristics
Materials: TENS gel pads, laser/LED diodes. Energy: 3.6V Li-Ion battery. Output: 808nm laser (1mW), 660nm red LED (90μW), 940nm IR LED (22μW). TENS: 2Hz, 100mA max, 75VDC max, 100us biphasic pulse. Connectivity: Standalone. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-10, IEC 60825-1. Software: 8-bit microcontroller.
Indications for Use
Indicated for temporary relief of minor muscle/joint pain, arthritis, muscle spasm, stiffness; promotion of muscle tissue relaxation; temporary increase in local blood circulation; symptomatic relief/management of chronic, intractable pain; adjunctive treatment for post-surgical and post-traumatic acute pain.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 21, 2015
Neurolumen, LLC % Mr. Andy Thompson Systec Engineering, LLC P.O. Box 1123 Euless. Texas 76039
Re: K142025
Trade/Device Name: Neurolumen PN-1000 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: ILY, GZJ Dated: December 17, 2014 Received: December 18, 2014
Dear Mr. Thompson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
for
Sincerely yours,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K142025
Device Name
Neurolumen PN-1000
Indications for Use (Describe)
The Neurolumen is indicated for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm, relieving stiffness, promoting relaxation of muscle tissue, and to temporarily increase local blood circulation where heat is indicated.
The Neurolumen is also indicated for the symptomatic relief and management of chronic, intractable pain and adjunctive treatment for post-surgical and post-traumatic acute pain.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
| <span style="font-family: sans-serif;"> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for SYSTEC Innovative Technologies LLC. The logo features the word "SYSTEC" in large, bold letters, with the words "INNOVATIVE TECHNOLOGIES LLC" underneath in a smaller font. To the right of the text is a circular graphic with a star-like design inside.
### Section C
## 510(k) Summary
12/12/2014
#### Contact Information:
| 510(k) Owner | 510(k) Submitter |
|----------------------------|-----------------------------------------|
| Neurolumen, LLC | Systec Engineering, LLC |
| 9636 N. May Ave, Suite 230 | dba Systec Innovative Technologies, LLC |
| Oklahoma City, OK 73120 | PO Box 1123 |
| 405-463-6525 phone | Euless, TX 76040 |
| 405-463-6528 fax | 817-864-9197 phone |
| Doug Durning | 817-684-1019 fax |
| doug@neurolumen.com | Andy Thompson |
| | andy@systeceng.com |
#### Device Identification:
Trade Name: Neurolumen PN-1000 Requiatory Class: II Requlation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Product Code: ILY, GZJ
### Claiming Equivalence to:
Trade Name: NeurolumenPN 510(k) Number: K082223 Regulatory Class: II Regulation Number: 21 CFR 890.5500 Requlation Name: Infrared Lamp Product Code: ILY, GZJ
Trade Name: TerraQuant MQ2000 and HandyRx Laser MQ2007 510(k) Number: K061614 Regulatory Class: II Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Product Code: ILY, GEX
Trade Name: TerraQuant MQ2000 with TQ-1TENS
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for SYSTEC Innovative Technologies LLC. The logo is composed of the company name in bold, stylized font, with the words "Innovative Technologies LLC" in a smaller font size below. To the right of the company name is a circular graphic containing a five-pointed star with small circles at the end of each point.
510(k) Number: K071445 Regulatory Class: II Regulation Number: 21 CFR 890.5500 Requlation Name: Infrared Lamp Product Code: ILY, GEX, GZJ
Trade Name: ComboCare 2000 510(k) Number: K081141 Regulatory Class: II Regulation Number: 21 CFR 890.5500 Regulation Name: Transcutaneous electrical nerve stimulation for pain relief. Product Code: GZJ, NHN, ISA
#### Device Description:
The Neurolumen is a medical therapy device that delivers Transcutaneous Electrical Nerve Stimulation (TENS) and Low Level Laser Therapy (LLLT) to a patient's limbs.
The Neurolumen consists of a control unit and six wraps. Each wrap contains two laser diodes, 4 light emitting diodes, and one or two TENS gel pads. After applying the wraps to the feet, ankles, and legs, the control unit is able to provide up to 30 minutes of therapy on a single charge of the internal lithium-ion battery. A charger is provided for recharging the internal battery.
Image /page/4/Picture/7 description: The image shows a collection of electronic components and accessories. There are two black heating pads, each with a power cord. A digital display unit with multiple buttons is also present, along with a power adapter. The components appear to be part of a heating system or electronic device setup.
PN-1000 Components
Image /page/4/Picture/9 description: In the image, a person is lying down with a device attached to their leg. The device has a digital display and several buttons with arrows pointing in different directions. The person's hands are holding the device, and there are tubes connecting the device to the leg. The device appears to be a medical device used for therapy or treatment.
PN-1000 Typical Usage
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for SYSTEC Innovative Technologies LLC. The logo is in black and white and features the word "SYSTEC" in large, bold letters. Below that is "INNOVATIVE TECHNOLOGIES LLC" in smaller letters. To the right of the text is a circular design with a star inside.
Image /page/5/Picture/2 description: The image shows a black device with two textured rectangular areas on top. A cable is connected to the device, and a separate textured rectangular piece is placed next to it. The device appears to be some sort of electronic equipment, possibly a charger or adapter, with the textured areas potentially serving as contact points or sensors.
PN-1000 Leg Wrap with Gel Pad
Image /page/5/Picture/4 description: The image shows two electronic devices connected by a cable. Both devices appear to be rectangular with a screen or display area in the center. The device on the left has a slightly larger base or bottom section compared to the one on the right. The cable connecting the two devices is black and has a connector on one end.
PN-1000 Foot/Ankle Wrap
| Unit | |
|------------------------------------|------------------------------------------------|
| Model | Neurolumen PN-1000 |
| Part Number | 0904000-00 |
| Battery | Lithium-Ion 3.6V |
| Charger | 220/110 VAC, 50/60Hz input<br>9VDC 1.5A output |
| Laser Diodes | |
| Quantity per wrap | 2 |
| Wavelength | 808nm |
| Output Power | 1mW |
| LEDs, Red | |
| Quantity per wrap | 2 |
| Wavelength | 660nm |
| Output Power | 90μW |
| LEDs, IR | |
| Quantity per wrap | 2 |
| Wavelength | 940nm |
| Output Power | 22μW |
| TENS (Load Impedance 500-550 ohms) | |
| Frequency | 2 Hz |
| Output Current | 100mA max |
| Pulse Voltage | 75VDC max |
| Pulse width | 100us biphasic |
#### Intended Use:
The Neurolumen is indicated for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm, relieving stiffness, promoting relaxation of muscle tissue, and to temporarily increase local blood circulation where heat is indicated.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for SYSTEC Innovative Technologies LLC. The logo features the word "SYSTEC" in large, bold, black letters. Below that, the words "INNOVATIVE TECHNOLOGIES LLC" are written in a smaller font. To the right of the text is a black circle with a white star inside.
The Neurolumen is also indicated for the symptomatic relief and management of chronic, intractable pain and adjunctive treatment for post-surgical and post-traumatic acute pain.
There are no changes to the intended use from the predicate device.
#### Comparison of technological characteristics:
| Characteristic | Predicate Device | Neurolumen PN-1000 |
|------------------------|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Control circuit | K082223: Digital circuit | Microcontroller.<br>A 8-bit microcontroller was<br>added to drive a newly<br>added user interface,<br>monitor the device<br>functions, and generate the<br>TENS pulse timing. |
| Liquid Crystal Display | K082223: Not present<br>K071445: Present | Added to display TENS<br>intensities, battery status,<br>and therapy countdown<br>timer. |
| Firmware | K082223: Not present<br>K071445: Present | Added for microcontroller.<br>Firmware developed per the<br>FDA guidance document on<br>medical devices containing<br>software. |
| Battery | K082223: 4 AA NiMH<br>batteries | 1 Li-Ion battery pack added<br>for improved charge density<br>and cycle life. |
| TENS Pulse | K082223:<br>Pulse width: 100us biphasic.<br>Frequency: 2 Hz repetition<br>rate.<br>Current: 100mA max. | Identical |
| Laser Classification | K082223: Not present<br>K061614: IIIa | Laser classification testing<br>to IEC 60825-1 rated laser<br>and LED output at class 3R<br>(IIIa). Equivalent to other<br>predicate devices cleared<br>as non-heating lamps. |
| Laser Diodes | K082223: Not present<br>K061614: 905nm<br>K071445: 905nm | 808nm |
| Red LEDs | K082223: 660nm | 660nm |
| | K061614: 680nm | |
| | K081141: 640nm | |
| | K071445: 660nm | |
| IR LEDs | K082223: 905nm | 940nm |
| | K061614: 840nm | |
| | K081141: 870nm | |
| | K071445: 875nm | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for SYSTEC Innovative Technologies LLC. The logo is in black and white. To the right of the text is a circle with a star inside.
The determination of substantial equivalence is derived from the fact that the new device is based on the predicate device with improvements to the user interface and laser diode. It contains the same basic construction in the control unit and wrap assemblies, contains the identical TENS output waveform, and contains the same basic types of LED/laser diode outputs as predicate devices.
The New Device has been fully tested to national and international safety standards for medical devices, including IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-10, and IEC 60825-1.
#### Device Testing Results
The laser diodes were tested on three subjects to demonstrate that the device is able to raise the skin temperature between 40-45 degrees C for at least 10 minutes. The testing results confirmed that the unit is capable of raising and maintaining these temperatures.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.