K141994 · Novocol, Inc. · EBF · Nov 14, 2014 · Dental
Device Facts
Record ID
K141994
Device Name
BULK FILL FLOWABLE COMPOSITE
Applicant
Novocol, Inc.
Product Code
EBF · Dental
Decision Date
Nov 14, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Base under Class I and II direct restorations Liner under direct restorative materials Class III and V restorations Undercut Blockout Minimally invasive cavities (including small non stress-bearing occlusal restorations) Tunnel preparations Repair of small enamel defects Repair of small defects in esthetic indirect restorations Repair of resin and acrylic temporary materials
Device Story
Low-viscosity, radiopaque flowable composite; used by dentists for direct dental restorations. Input: material applied via syringe or single-dose capsule to tooth cavity. Operation: light-cured using high-intensity blue light; methacrylate resins and fillers undergo polymerization to form hard, bonded composite. Output: cured dental restoration. Benefit: low polymerization stress on tooth bond; high conversion rate; improved physical/mechanical properties; hydrophobic; moisture-resistant.
Clinical Evidence
Bench testing only. Evaluated per ISO 4049:2009. Tested properties: depth of cure, conversion, compressive strength, diametral tensile strength, microhardness, volume shrinkage stress, water absorption, and solubility. Biocompatibility testing performed per ISO 10993-1:2009 and ISO 7405:2008.
Technological Characteristics
Resin system: DDCDMA (Dimer Dicarbamate Dimethacrylate), EBPADMA, and UDMA. Filler system: Ytterbium Fluoride nano-particles, Silica, and acid-resistant silanated Barium glass. Light-cured polymerization. Radiopaque. Supplied in syringes or single-dose capsules. Shades: A1, A2, A3, A3.5.
Indications for Use
Indicated for dental patients requiring restorative procedures including Class I, II, III, and V restorations, cavity lining/base, undercut blockout, minimally invasive cavity repair, tunnel preparations, and repair of enamel defects or temporary resin/acrylic materials.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 14,2014
Septodent-Louisville Cora Bracho- Troconis Director of Secintific and Regulatory Affairs 416 S. Taylor Avenue Louisville, Co 80027
Re: K141994
Trade/Device Name: Bulk Fill Flowable Composite Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth shade resin materal Regulatory Class: II Product Code: EBF Dated: August 15, 2014 Received: August 19, 2014
Dear Ms. Bracho-Troconis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/0/Picture/10 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is a symbol of three stylized human profiles facing to the right, overlapping each other.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno DDS, mA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Section 004: Indications for Use Statement
(As Required by 21 CFR § 807.87)
510(k) Number: K141994
# Device Name: Bulk Fill Flowable Composite
Indications for Use:
- Base under Class I and II direct restorations
- Liner under direct restorative materials
- Class III and V restorations
- Undercut Blockout
- Minimally invasive cavities (including small non stress-bearing occlusal restorations)
- Tunnel preparations
- Repair of small enamel defects
- Repair of small defects in esthetic indirect restorations
- Repair of resin and acrylic temporary materials
Prescription Use X (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) 10
{3}------------------------------------------------
# Section 005: 510(k) Summary
(As Required By 21 CFR 807.92)
| 510(k) Applicant: | Septodont, Inc.<br>416 S. Taylor Ave<br>Louisville, Co. 80027<br>(303) 655-7535 |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Cora Bracho-Troconis<br>Director of Scientific and Regulatory Affairs<br>Phone: (303) – 665-7535<br>Fax: (303) – 666-4320<br>cbracho@septodont.com |
| Date Summary was Prepared: | July 14th, 2014 |
| Trade Name: | Bulk Fill Flowable Composite |
| Common Name: | Bulk Fill Flowable Composite |
| Classification Name: | Material, Tooth Shade, Resin |
Predicate Devices:
| 510(k)# | Product | Manufacturer |
|---------|------------------------------------------|----------------|
| K091635 | Venus Diamond Flow | Heraeus Kulzer |
| K120453 | Filtek Bulk Fill Flowable<br>Restorative | 3M ESPE |
| K123773 | X-Tra Base | VOCO (GmbH) |
#### Device Description:
The new Bulk Fill Flowable Composite manufactured by Septodont Louisville is a low viscosity, radiopaque flowable composite which sets upon exposure to a high-intensity blue light source. After polymerization, it exhibits low stress on the bond with the tooth, and it has a high polymerization conversion. This high degree of polymerization is achieved with short curing times and creates improved physical and mechanical properties.
{4}------------------------------------------------
The new Bulk Fill Flowable Composite is packaged in single dose capsules and syringes. The shades offered with Bulk Fill Flowable Composite are:
· A1, A2, A3, A3.5
#### Indications for Use:
- Base under Class I and II direct restorations
- Liner under direct restorative materials
- Class III and V restorations
- Undercut Blockout
- Minimally invasive cavities (including small non stress-bearing occlusal restorations)
- Tunnel preparations
- Repair of small enamel defects
- Repair of small defects in esthetic indirect restorations
- Repair of resin and acrylic temporary materials
# Technological Characteristics
The new Bulk Fill Flowable Composite resin system consists of 3 major components. The resin system was designed using our crosslinkable proprietary monomer DDCDMA (Dimer Dicarbamate Dimethacrylate). A system has been designed using DDCDMA where EBPADMA (Ethoxylated Bisphenol "A" Dimethacrylate) is the base resin and UDMA (Urethacrylate) is added as a minor component. This is similar to the resin system used in N'Durance, a Septodont Universal composite cleared by the FDA K072150.
This 3-component resin polymerizes using a modified initiating system The chemistry of this resin system results in low stress on the bond with the tooth and high conversion of monomers into polymers. Additionally, these resins impart a greater hydrophobicity and are less sensitive to changes in atmospheric moisture.
The filler system in the new Bulk Fill Flowable Composite contains Ytterbium Fluoride mono dispersed nano sized particles for dental application. Silica, as well as an acid resistant silanated Barium glass.
When irradiated by light, the methacrylate functionalities of the resins and fillers undergo, in conjunction with the photoinitiator system, a light-induced polymerization to form a hard composite that is bonded to the tooth structure with a permanent dental adhesive.
#### Substantial Equivalence:
#### Biocompatibility:
Bulk-Fill Flowable composite was tested following ISO 10993-1:2009 and ISO 7405:2008. The conclusion of the assessment is that the product is biocompatible and safe for its intended use.
{5}------------------------------------------------
#### Summary of Physical Tests
This 510(k) submission includes data from bench testing to evaluate the performance of Bulk Fill Flowable Composite in comparison with its predicates. The standard utilized was ISO 4049:2009. The properties that were tested include Depth of Cure, Conversion, Compressive Strength, Diametral Tensile Strength, Microhardness, Volume Shrinkage Stress, Water Absorption and Solubility.
# Indications
Bulk Fill Flowable Composite has very similar indications as the claimed predicate devices Venus Diamond Flowable Composite (K091635), Filtek Bulk Fill (K120453) and X-Tra Base VOCO (K123773). Specifically, Bulk Fill Flowable Composite and Filtek Bulk Fill (K120453) are very similar in their indications, with the main difference being that Filtek Bulk Fill also includes Pit and Fissure and Core build-up indications.
# Technology
The delivery and curing methods for the predicate devices are similar to that of Bulk Fill Flowable composite as well. They are all flowable composites which are applied directly to the site from a syringe or single unit dose tips. Additionally, all of these devices polymerize when irradiated by light to form a hard composite that is bonded to the tooth structure with a permanent dental adhesive.
# Conclusion:
Information provided in this 510(k) submission shows that Bulk Fill Flowable Composite is substantially equivalent to the identified predicate devices in terms of intended use, indications for use, physical properties and technological characteristics. In addition to being substantially equivalent, the Bulk Fill Flowable Composite has also been determined to be biocompatible and safe for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.