The True™ BAV Balloon Valvuloplasty Catheter is indicated for balloon aortic valvuloplasty.
Device Story
The True™ BAV Balloon Valvuloplasty Catheter is an over-the-wire, high-performance balloon catheter used for balloon aortic valvuloplasty. It features a proprietary, non-compliant, low-profile balloon fixed at the distal tip, designed for consistent diameters and lengths. Two radiopaque markers delineate the working length for placement. The coaxial catheter includes an atraumatic tip for valve advancement and is compatible with .035" guidewires. Proximal components include dual female luer-lock hubs for inflation and guidewire lumens. The device is supplied with a profile-reducing sheath and a re-wrapping tool. It is used by clinicians in a cardiovascular setting to dilate the aortic valve. The device is non-latex. Performance is verified through bench testing, including balloon distensibility, burst pressure, fatigue, and trackability, ensuring it functions equivalently to predicate devices.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing only. In vitro testing included visual inspection, balloon distensibility, tip length, balloon/shaft dimensions, marker band visibility, tip morphology, tensile testing (tip, balloon-to-shaft, hub-to-shaft), balloon nominal/burst pressure, fatigue (repeated inflation), shaft leaks/burst, media interaction, trackability, inflation/deflation time, marker band alignment, sheath compatibility, equipment interface, dye penetration, and pouch tensile strength.
Technological Characteristics
Over-the-wire balloon catheter; non-compliant, low-profile balloon; coaxial design; radiopaque markers; .035" guidewire compatibility; 100 cm shaft length; female luer-lock hubs; non-latex materials. Mechanical sensing/actuation via hydraulic inflation. No software or electronic components.
Indications for Use
Indicated for balloon aortic valvuloplasty in patients requiring this procedure.
Regulatory Classification
Identification
A balloon aortic valvuloplasty catheter is a catheter with a balloon at the distal end of the shaft, which is intended to treat stenosis in the aortic valve when the balloon is expanded.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible.
(2) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components.
(3) Non-clinical performance evaluation must demonstrate that the device performs as intended under anticipated conditions of use, including device delivery, inflation, deflation, and removal.
(4) In vivo evaluation of the device must demonstrate device performance, including the ability of the device to treat aortic stenosis.
(5) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to the use of the device.
In combination with the general controls of the FD&C Act, the NuMED NuCLEUS and NuCLEUS-X BAV Catheters are subject to the following special controls:
{0}------------------------------------------------
# True™ BAV Balloon Valvuloplasty Catheter 510(k) Summary 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
#### Submitter Information:
| Applicant: | Bard Peripheral Vascular, Inc<br>1625 West 3rd Street<br>Tempe, Arizona 85281 |
|------------|-------------------------------------------------------------------------------|
| Phone: | 480-350-6012 |
| Fax: | 480-449-2546 |
| Contact: | Aaron Conovaloff, Regulatory Affairs Associate |
| Date: | July 21, 2014 |
### Subject Device Name:
| Device Trade Name: | True™ BAV Balloon Valvuloplasty Catheter |
|-----------------------|---------------------------------------------------------------------|
| Common or Usual Name: | Balloon Aortic Valvuloplasty (21 CFR<br>870.1250, Product Code OZT) |
| Classification: | Class II |
| Classification Panel: | Cardiovascular |
### Predicate Devices:
- True Dilatation™ Balloon Valvuloplasty Catheter (K133569; cleared . December 20, 2013)
- Vida™ PTV Dilatation Catheter (K131002, cleared July 2, 2013) ●
Image /page/0/Picture/13 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, giving them a distinct and modern appearance. The overall design is simple yet impactful, making the word easily recognizable.
{1}------------------------------------------------
#### Device Description:
The True™ BAV Balloon Valvuloplasty Catheter is a high performance balloon catheter consisting of an over-the-wire catheter with a balloon fixed at the distal tip. The proprietary, non-compliant, low profile balloon is designed to provide consistent balloon diameters and lengths. Two radiopaque markers delineate the working length of the balloon and aid in balloon placement. The coaxial catheter includes an atraumatic tip to facilitate advancement of the catheter to and through the valve. The over-the-wire catheter is compatible with .035" quidewire and is available in 100 cm working length. The proximal portion of the catheter includes a female luer-lock hub connected to the inflation lumen, and a female luer-lock hub connected to the guidewire lumen. Packaged with every product is a profile reducing sheath that is positioned over the balloon for protection before use. A re-wrapping tool is also provided on the catheter shaft. This product is not manufactured with any latex.
| Attribute | True™ BAV Balloon Valvuloplasty Catheter<br>Product Offering |
|-------------------------------------------------------------------|--------------------------------------------------------------|
| Balloon Diameter<br>(mm) | 18, 20, 22, 24, 26 |
| Balloon Length<br>(cm) | 4 |
| Catheter Shaft<br>Length (cm) | 100 |
| Introducer Sheath<br>Compatibility by<br>Balloon Diameter<br>(mm) | 8F: 18 mm<br>9F: 20 mm<br>10F: 22 mm, 24 mm<br>12F: 26 mm |
#### Indications for Use of Device:
The True™ BAV Balloon Valvuloplasty Catheter is indicated for balloon aortic valvuloplasty.
{2}------------------------------------------------
## Comparison of Indications for Use to Predicate Devices:
The indications for use statement for the True™ BAV Balloon Valvuloplasty Catheter does not raise any new issues of safety and effectiveness as demonstrated through the risk analysis process based on the proposed indications for use statement as compared to the predicate devices. Therefore, the subject device, the True™ BAV Balloon Valvuloplasty Catheter, is substantially equivalent to the predicate devices.
# Technological Comparison to Predicate Devices:
The True™ BAV Balloon Valvuloplasty Catheter has the following similarities to the predicate device, the True Dilatation™ Balloon Valvuloplasty Catheter (clearance to market via K133569 on December 20, 2013):
- . Same intended use
- . Same indications for use
- Same target population .
- Same operating principle .
- Same fundamental scientific technology ●
- Same sterility assurance level and method of sterilization .
The True™ BAV Balloon Valvuloplasty Catheter has the following similarities to the predicate device, the Vida™ PTV Dilatation Catheter (clearance to market via K131002, on July 2, 2013):
- Same operating principle .
- . Same fundamental scientific technology
- Same packaging materials and configurations .
- . Same sterility assurance level and method of sterilization
### Performance Data:
To demonstrate substantial equivalence of the subject device to the predicate devices, its technological characteristics and performance criteria were evaluated. Using FDA Guidance Documents on non-clinical testing of medical devices and internal Risk Assessment procedures, the following in vitro tests were performed on the subject device:
- . Visual Inspection
Image /page/2/Picture/21 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, with a white fill. The letters are evenly spaced and of uniform size. The overall impression is clean and modern.
{3}------------------------------------------------
- o Balloon Distensibility
The following in vitro tests were leveraged from the predicate Vida™ PTV Dilatation Catheter:
- 0 Tip Length
- Balloon Outer Diameter (i.e. Diameter and Profile Test) o
- Balloon Working Length .
- Catheter Shaft Outer Diameter o
- 0 Catheter Shaft Inner Diameter
- 0 Tip Visibility
- 0 Catheter Shaft Visibility
- Marker Band Visibility 0
- Tip Morphology o
- Tip Tensile (i.e. Tip Pull and Torque Test, and Bond Strength Test) 0
- Balloon to Shaft Tensile 0
- 0 Hub to Shaft Tensile
- Catheter Shaft Elongation 0
- o Balloon Nominal (Operating) Pressure
- Rated Burst Pressure (i.e. Balloon Minimum Burst Strength) 0
- o Balloon Burst Mode
- Fatigue (i.e. Repeated Balloon Inflation) 0
- Catheter Shaft Leaks (i.e. Catheter Body Maximum Pressure Test) 0
- Catheter Shaft Burst (i.e. Catheter Body Maximum Pressure Test) 0
- 0 Media Interaction (i.e. Balloon Preparation Test)
- Catheter Shaft Length .
- Trackability
- Inflation/Deflation Time (i.e. Balloon Inflation/Deflation Test, and Balloon . Inflatability Test)
- Marker Band Alignment .
- Sheath Compatibility 0
- Equipment Interface o
- 0 Visual Inspection of Packaging
Image /page/3/Picture/32 description: The image shows the word "BARD" in a stylized, outlined font. The letters are large and bold, with a slight 3D effect created by subtle color variations within the outlines. The overall impression is clean and modern.
{4}------------------------------------------------
- 0 Dye Penetration
- Pouch Tensile Strength .
The results from these tests demonstrate that the technological characteristics and performance criteria of the True™ BAV Balloon Valvuloplasty Catheter are substantially equivalent to the predicate devices, and that it can perform in a manner equivalent to devices currently on the market for the same intended use.
#### Conclusions:
The subject device, the True™ BAV Balloon Valvuloplasty Catheter, met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The True™ BAV Balloon Valvuloplasty Catheter is substantially equivalent to the legally marketed predicate devices, the True Dilatation™ Balloon Valvuloplasty Catheter and the Vida™ PTV Dilatation Catheter.
Image /page/4/Picture/9 description: The image shows the word "BARD" in a stylized, blocky font. Each letter is outlined in black, giving it a bold and distinct appearance. The "A" in "BARD" is designed with a unique triangular shape, setting it apart from the other letters.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with healthcare, featuring a staff with a serpent entwined around it.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 20, 2015
Bard Peripheral Vascular, Inc. Dr. Aaron Conovaloff Regulatory Affairs Associate 1625 West 3rd Street Tempe, Arizona 85281
Re: K141985
Trade/Device Name: True BAV Balloon Valvuloplasty Catheter Regulation Number: 21 CFR 870.1255 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: OZT Dated: July 21, 2014 Received: July 22, 2014
Dear Dr. Conovaloff:
This letter corrects our substantially equivalent letter of September 18, 2014.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
Page 2 - Dr. Aaron Conovaloff
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Nicole G. Ibrahim -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K141985
Device Name: True™ BAV Balloon Valvuloplasty Catheter
Indications for Use: The True™ BAV Balloon Valvuloplasty Catheter is indicated for balloon aortic valvuloplasty.
Prescription Use X (Part21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/7/Picture/12 description: The image shows the word "BARD" in a stylized, blocky font. The letters are bold and outlined in black. The "A" is designed with a triangular shape in the middle, and the "D" has a rounded shape.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.