Digitek Titanium Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation. They are compatible with the ASTRA TECH Implant System OsseoSpeed™ 3.5mm, 4.0mm, 4.5mm, 5.0mm implants
Device Story
Digitek Titanium Abutments are premanufactured prosthetic components; designed for connection to endosseous dental implants to support prosthetic restorations. Device consists of titanium body and screw; features internal hexagonal prism at base for anti-rotation. Used by dental professionals in clinical settings to restore chewing function. Abutments are supplied non-sterile; require sterilization by user prior to installation. Device provides stable interface between implant and prosthesis; benefits patient by facilitating functional and aesthetic tooth replacement.
Clinical Evidence
No clinical data. Bench testing only: fatigue testing performed per ISO 14801:2007; cytotoxicity testing performed per ISO 10993-5; sterilization validation performed per ANSI/AAMI ST79.
Technological Characteristics
Material: Titanium alloy Ti-6Al-4V ELI (ASTM F136). Design: Straight abutment with internal hexagonal anti-rotation feature. Energy source: None (mechanical). Sterilization: Non-sterile, user-sterilized per ANSI/AAMI ST79. Connectivity: None.
Indications for Use
Indicated for use in prosthetic rehabilitation for patients requiring dental implant restoration; compatible with ASTRA TECH Implant System OsseoSpeed™ 3.5mm, 4.0mm, 4.5mm, and 5.0mm implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Inclusive® Titanium Abutment for - Astra OsseoSpeed™ Implants (K100993)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 31, 2015
Digitek Dental Solutions Limited Mr. Alwin Ngai Director Units B-C Flat D. 6/F, Dragon Ind. Bldg., 93 King Lam St., Cheung Sha Wan, Kowloon, Hong Kong
Re: K141970
Trade/Device Name: Digitek Titanium Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: June 19, 2015 Received: July 02, 2015
Dear Mr. Ngai,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Alwin Ngai
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K141970
Device Name Digitek Titanium Abutment
Indications for Use (Describe)
Digitek Titanium Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation. They are compatible with the ASTRA TECH Implant System OsseoSpeed™ 3.5mm, 4.0mm, 4.5mm, 5.0mm implants
Type of Use (Select one or both, as applicable)
ال Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5: 510(k) Summary
## 1. Submitter Information
| Company Name: | Digitek Dental Solutions Limited |
|------------------|-----------------------------------------------|
| Company Address: | Units B-C, Flat D, 6/F, Dragon Ind. Bldg., 93 |
| | King Lam St., Cheung Sha Wan, Kowloon, |
| | Hong Kong |
| Company Phone: | (852) 2742 3210 |
| Company Fax: | (852) 3590 8715 |
| Contact Person: | Alwin Ngai |
| Date Prepared: | July 29, 2015 |
### 2. Device Identification
| Device Model Name: | Digitek Titanium Abutment |
|----------------------|------------------------------------|
| Classification Name: | Endosseous Dental Implant Abutment |
| Regulation Number: | 872.3630 |
| Product Code: | NHA |
| Class | II |
| Panel | Dental |
### 3. Predicated Devices
| Primary | Predicate | Inclusive® Titanium Abutment for - Astra OsseoSpeed™ | | |
|---------|-----------|------------------------------------------------------|--|--|
| Device | | Implants, K100993 | | |
Reference predicate GC AADVA TI ABUTMENTS – BO AC SV BH, K103234 Device
### 4. Device Description
Digitek Titanium Abutments are endosseous implant abutments which are placed into a corresponding dental implant to provide support for a prosthetic restoration. These abutments are made of titanium grade Ti-6AI-4V ELI (meets ASTM Standard F136). The abutment is mounted into the implant with a screw. These abutments are compatible with the ASTRA TECH Implant System OsseoSpeed™ implants.
Digitek Titanium Abutments are provided straight only and are not intended to be modified to provide an angle correction.
{4}------------------------------------------------
Digitek Titanium Abutments are manufactured in two models "3.5" and "4.0". And a hexagonal prism at the base of the Digitek Titanium Abutments acts as an anti-rotation feature.
Digitek titanium Abutments are supplied in a non-sterile state and should be sterilized prior to installation.
# 5. Indication for Use
Digitek Titanium Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation. They are compatible with the ASTRA TECH Implant System OsseoSpeed" 3.5mm, 4.0mm, 4.5mm, 5.0mm implants.
# 6. Substantial Equivalence
The proposed Digitek Titanium Abutments are substantially equivalent to the Inclusive® Titanium Abutments for - Astra OsseoSpeed™ Implants, and the GC AADVA TI ABUTMENTS - BO AC SV BH. These abutments are substantially equivalent in indications for use, material of body and screw, design and sterility status.
| Elements of<br>Comparison | Subject Device | Primary Predicate<br>Device | Reference Predicate<br>Device |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Digitek Titanium<br>Abutments | Inclusive® Titanium<br>Abutments for - Astra<br>OsseoSpeed™<br>Implants | GC AADVA TI<br>ABUTMENTS - BO<br>AC SV BH |
| 510(k)<br>Number | Applying | K100993 | K103234 |
| Indications<br>for Use | Digitek Titanium<br>Abutments are<br>premanufactured<br>prosthetic<br>components directly<br>connected to<br>endosseous dental<br>implants and are<br>intended for use as an<br>aid in prosthetic<br>rehabilitation. They<br>are compatible with<br>the ASTRA TECH<br>Implant System | Inclusive® Titanium<br>Abutments for - Astra<br>OsseoSpeed™<br>Implants are<br>premanufactured<br>prosthetic components<br>directly connected to<br>endosseous dental<br>implants and are<br>intended for use as an<br>aid in prosthetic<br>rehabilitation. They are<br>compatible with the<br>Astra Tech | "GC AADVA TI<br>ABUTMENTS - BO<br>AC SV BH" are dental<br>implant abutments for<br>use with partially or<br>fully edentulous<br>patients to restore<br>chewing function by<br>attachment to a dental<br>implant fixture placed<br>in the maxilla or<br>mandible. Each<br>abutment is<br>accompanied by a |
{5}------------------------------------------------
| | OsseoSpeed™<br>3.5mm, 4.0mm,<br>4.5mm, 5.0mm<br>implants. | OsseoSpeed™ 3.0, 3.5,<br>4.0, 4.5, 5.0 implants. | screw in order to<br>engage corresponding<br>dental implant fixture. |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material of<br>Body and<br>Screw | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V |
| Design | Abutment (main body)<br>assembly with<br>abutment screw.<br>Abutment connection<br>to implant is an<br>internal hexagon | Implant/Abutment<br>assembly with<br>abutment screw.<br>Abutment connection to<br>implant is an internal<br>hexagon | Allows the prosthesis<br>to be retained to the<br>abutment; abutment<br>screw is intended to<br>secure the abutment<br>to the endosseous<br>dental implant |
| Abutment<br>Platform<br>Diameters<br>(mm) | 3.5, 4.0, 4.5, 5.0mm | 3.0, 3.5, 4.0, 4.5,<br>5.0mm | Max. 6.0mm<br>Min. 2.5mm |
| Sterility<br>Status | Supplied in non-sterile<br>state | Supplied in non-sterile<br>state | Supplied in non-sterile<br>state |
# 7. Performance Testing
Fatigue testing was performed according to ISO 14801:2007, Dentistry --Implants -- Dynamic fatigue test for endosseous dental implants. Testing was performed on Digitek Titanium Abutment with the implants that they are intended to fit.
Final finished sterilized Digitek Titanium Abutment has been tested for cytotoxicity according to ISO 10993-5.
Sterilization method has been validated according to ANSI/AAMI ST79.
Reverse engineering analysis was performed on OEM abutments to obtain data used in the design the Digitek Titanium Abutment.
The Digitek Titanium Abutment conforms to the FDA Guidance Document for Endosseous Dental Implants and Abutments.
# 8. Conclusion
There are no known technological differences between Digitek Titanium
{6}------------------------------------------------
Abutments and those of the predicate devices. Thus, Digitek Titanium Abutments is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.