PJUR SILICONE BASED LUBRICANT: PJUR BACKDOOR ANAL GLIDE AND ANALYSE ME!
Applicant
Pjur Group Luxembourg S.A.
Product Code
NUC · Obstetrics/Gynecology
Decision Date
Apr 6, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
pjur® Backdoor Anal Glide and pjur® Analyse Me! are personal lubricants for penile, vaginal, and/or anal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. These products are compatible with natural rubber latex, synthetic polyisoprene, and polyurethane condoms.
Device Story
Non-sterile, silicone-based personal lubricant; formulated as clear, non-irritating, non-greasy, odorless viscous liquid. Contains Cyclopentasiloxane, Dimethicone, Dimethiconol, and jojoba. Intended for over-the-counter use by consumers to moisturize and lubricate during intimate sexual activity. Applied topically to penile, vaginal, or anal areas. Enhances comfort by supplementing natural lubrication. Compatible with natural rubber latex, synthetic polyisoprene, and polyurethane condoms. No contraceptive or spermicidal properties.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, irritation) showed non-cytotoxic, non-sensitizing, and non-irritant results. Condom compatibility testing per ASTM D7661-10 confirmed compatibility with natural latex, polyurethane, and polyisoprene condoms. Shelf-life testing supports at least 1 year.
Technological Characteristics
Silicone-based lubricant; ingredients: Cyclopentasiloxane, Dimethicone, Dimethiconol, jojoba. Viscosity 600-1000 cps; density 0.900-0.950 g/ml; anhydrous. Condom compatibility tested per ASTM D7661-10. Biocompatibility tested per ISO 10993-1.
Indications for Use
Indicated for individuals requiring personal lubrication for penile, vaginal, and/or anal application to enhance comfort during intimate sexual activity and supplement natural lubrication. Compatible with natural rubber latex, synthetic polyisoprene, and polyurethane condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus or a series of human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 6, 2015
pjur group Luxembourg SA % Cheryl Wagoner Consultant Wagoner Consulting LLC PO Box 15729 Wilmington, NC 28408
Re: K141913
> Trade/Device Name: pjur® Backdoor Anal Glide and pjur® Analyse Me! Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: NUC Dated: July 14, 2014 Received: July 17, 2014
Dear Cheryl Wagoner,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Herbert P. Lerner -S
for Benjamin Fisher, PhD Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K141913
Device Name pjur® Backdoor Anal Glide or pjur® Analyse Me!
#### Indications for Use (Describe)
pjur® Backdoor Anal Glide and pjur® Analyse Me! are personal lubricants for penile, vaginal, and/or anal application . intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. These products are compatible with natural rubber latex, synthetic polyisoprene, and polyurethane condoms.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the "pjur group". The logo features the word "pjur" in a bold, sans-serif font, with a yellow circle above the letter "j". Below the word "pjur" is the word "group" in a smaller, sans-serif font. The logo also includes the registered trademark symbol.
# 510(k) Summary
(as required by 21 CFR 807.92)
| Submitter | pjur group Luxembourg SA |
|----------------|---------------------------------|
| | 87 esplanade de la Moselle |
| | L-6637 Wasserbillig, Luxembourg |
| | Telephone: +352 74-8989 |
| | Fax: +352 74-8990 |
| Contact Person | Patrick Giebel |
| | Quality Manager |
| | pgiebel@pjur.com |
| Date Prepared | 2/23/2015 |
| Trade Name | pjur® silicone based lubricant under the brand names of pjur®<br>Backdoor Anal Glide or pjur® Analyse Me! |
|---------------------|-----------------------------------------------------------------------------------------------------------|
| Common Name | Personal Lubricant |
| Classification Name | Condom<br>(21 CFR §884.5300, Product Code NUC) |
| Class | Class II |
Wet Platinum Premium Lubricant®, K130012 Predicate Devices
| Description | The pjur® Backdoor Anal Glide or pjur® Analyse Me! device is a non-<br>sterile, silicone-based personal lubricant. This over-the-counter<br>product is formulated to be clear, non-irritating, non-greasy, and<br>odorless. The pjur® Backdoor Anal Glide or pjur® Analyse Me! Device<br>is a silicone based lubricant contains neither a contraceptive nor a<br>spermicide. The product contains Cyclopentasiloxane, Dimethicone,<br>Dimethiconol, and jojoba. |
|-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | pjur® Backdoor Anal Glide and pjur® Analyse Me! are personal lubricants for penile, vaginal, and/or anal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. These products are compatible with natural rubber latex, synthetic polyisoprene, and polyurethane condoms. |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{4}------------------------------------------------
| Technological<br>Characteristics | The pjur® Backdoor Anal Glide or pjur® Analyse Me! device is a<br>silicone based lubricant that contains a blend of silicone fluid<br>ingredients similar to ingredients found in other lubricants currently on |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | the U.S. markets and substantially equivalent to the predicate device.<br>Both the Subject and predicate devices contain Cyclopentasiloxane,<br>Dimethicone, Dimethiconol. Both devices also contained a plant-<br>derived oil additive that is present in very small amounts and has no |
| | impact upon safety or effectiveness. In the case of The pjur® Backdoor<br>Anal Glide or pjur® Analyse Me! device the additive is jojoba which is<br>commonly used in food and cosmetic devices. The additive for Wet<br>Platinum Premium Lubricant®, K130012 is Vitamin E. This plant based<br>additive, as with the Subject device, does not negatively impact risk or<br>biocompatibility. The presence of jojoba oil in the Subject device and<br>Vitamin E in the Predicate device do not raise any new questions of<br>safety or effectiveness nor does it represent a new technology. |
| | These products are compatible with natural rubber latex, synthetic<br>polyisoprene, and polyurethane condoms. |
| | The intended use of the Subject device is the same as its predicate.<br>Both are intended as non-sterile, over-the-counter personal lubricants.<br>The indications for use for the Subject and Predicate devices are not<br>identical, but are substantially equivalent. Both the Subject device and<br>the predicate are indicated as a personal lubricant for intimate sexual<br>activity and have been fully tested to ensure that there are no new<br>questions of safety or effectiveness related to this usage nor does it<br>represent new technology. |
{5}------------------------------------------------
| Performance Data | Biocompatibility testing was performed in accordance with ISO 10993–<br>1, "Biological evaluation of medical devices – Part 1: Evaluation and<br>testing" including: | |
|------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| | Test Performed | Results |
| | ISO 10993-5 Cytotoxicity (Direct<br>contact) | Non-cytotoxic |
| | ISO 10993-10 Guinea Pig<br>Maximization Sensitization test | No sensitization response. |
| | ISO 10993-10 | Non-irritant |
| | • Irritation (Direct contact) | |
| | • Irritation( Intracutanteous injection) | |
| | • Vaginal Irritation (repeat insult) | |
| | Shelf-Life Testing | At least 1 year |
| | Viscosity | 600-1000 cps |
| | Density | 0.900-0.950 g/ml |
| | Appearance | Viscous liquid |
| | Turbidity | Clear, no turbidity |
| | Color | Slightly yellow |
| | Odor | None |
| | Microbial Limits | <100 cfu/g TAMC, <10 cfu/g<br>TYMC, absence of<br>pathogenic organisms |
| | Water Activity | <0.3 Aw |
| | pH and Osmolaltity | N/A since product is<br>anhydrous |
| | Condom compatibility testing was performed using the methods<br>outlined in ASTM D7661-10 including burst volume, burst pressure,<br>force of break and elongation until break. Testing results demonstrate<br>that the pjur® pjur® Backdoor Anal Glide or pjur® Analyse Me! device<br>is compatible with natural latex, , polyurethane, and polyisoprene<br>condoms. | |
| Conclusion | The pjur® Backdoor Anal Glide or pjur® Analyse Me! device has the<br>same intended use as that of its predicate. The ingredients in the pjur®<br>formulation are the similar to the Wet Platinum Premium Lubricant,<br>K130012. The labeling claims and indications for use of the pjur®<br>Backdoor Anal Glide or pjur® Analyse Me! device are the similar to<br>those of the predicates. |
|------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The labeling of the pjur® Backdoor Anal Glide or pjur® Analyse Me!<br>device formulation contains the same warnings and precautions as<br>those in the labeling of the predicate. |
| | Any differences that exist between t The pjur® Backdoor Anal Glide<br>or pjur® Analyse Me! device and the predicates have no significant<br>effect on the safety or effectiveness. |
{6}------------------------------------------------
| T The pjur® Backdoor Anal Glide or pjur® Analyse Me! device is |
|-----------------------------------------------------------------------|
| substantially equivalent to other personal lubricant products cleared |
| lin the US in terms of biocompatibility, technology, intended use, |
| lindications, and suitability characteristics. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.