K141893 · Bioinfinity (M) Sdn. Bhd. · HIR · Dec 18, 2014 · Obstetrics/Gynecology
Device Facts
Record ID
K141893
Device Name
VIBRANCE KEGEL DEVICE (VKD)
Applicant
Bioinfinity (M) Sdn. Bhd.
Product Code
HIR · Obstetrics/Gynecology
Decision Date
Dec 18, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1425
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Vibrance Kegel Device (VKD) is intended for the strengthening of the pelvic floor muscles, which has been found to help women with urinary incontinence.
Device Story
The Vibrance Kegel Device (VKD) is an intra-vaginal, battery-powered exercise device for pelvic floor muscle strengthening. It consists of a main body (ABS plastic) and three interchangeable silicone resistance sheaths of varying stiffness. The device provides active vibration biofeedback to help the user identify and engage the correct pelvic floor muscles during voluntary contractions. The user performs exercises according to an included manual; the device provides resistive training to strengthen muscles via physiologic principles. It is intended for over-the-counter use by women in a home setting. The device is non-sterile and reusable. It does not provide numerical data; rather, it uses vibration as a biofeedback mechanism to confirm muscle engagement. The device includes a winged base to prevent total ingestion into the vaginal canal.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including mechanical compression/pull-off tests, biocompatibility (ISO 10993-1, -5, -10, -12), and electrical safety (IEC 60601-1, IEC 60601-1-11).
Technological Characteristics
Materials: ABS plastic (chassis), medical-grade silicone (sheaths). Energy: Two 3V CR1220 lithium batteries. Principle: Resistive pelvic floor exerciser with electromechanical vibration biofeedback. Dimensions: 2.4 cm max diameter, 6 cm max inserted length. Connectivity: None. Sterilization: Non-sterile. Standards: ISO 10993, IEC 60601-1, IEC 60601-1-11, ISO 14971.
Indications for Use
Indicated for women with mild urinary incontinence for the strengthening of pelvic floor muscles.
Regulatory Classification
Identification
A perineometer is a device consisting of a fluid-filled sack for intravaginal use that is attached to an external manometer. The devices measure the strength of the perineal muscles by offering resistence to a patient's voluntary contractions of these muscles and is used to diagnose and to correct, through exercise, uninary incontinence or sexual dysfunction.
Predicate Devices
Colonial Medical Supply Pelvic Muscle Therapy (K002830)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a caduceus, rendered in black.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 18, 2014
Bioinflnity (M) Sdn, Bhd., LLC. % Carrie Hetrick Senior Consultant, RA Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701
Re: K141893
> Trade/Device Name: Vibrance Kegel Device (VKD) Regulation Number: 21 CFR 884.1425 Regulation Name: Perineometer Regulatory Class: II Product Code: HIR Dated: November 16, 2014 Received: November 21, 2014
Dear Carrie Hetrick,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K141893
# Device Name
Vibrance Kegel Device (VKD)
### Indications for Use (Describe)
The Vibrance Kegel Device (VKD) is intended for the strengthening of the pelvic floor muscles, which has been found to help women with urinary incontinence.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
for
# Vibrance Kegel Device (VKD)
### 1. Submission Submitter
Bioinfinity (M) Sdn. Bhd. No. 21, Jalan 4/62A, Bandar Menjalara, Kepong 52200 Kuala Lumpur Malaysia Phone: +603-6270236 Fax: +603-62720093 Contact: Mr. Ng Shea Kang, Market Development Director
### 2. Submission Correspondent
Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: (720) 838-4113 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Carrie Hetrick, Senior Consultant, RA Email: project.management@emergogroup.com
### 3. Date Prepared
December 18, 2014
# 4. Device Identification
| Trade/Proprietary Name: | Vibrance Kegel Device (VKD |
|----------------------------|----------------------------|
| Common/Usual Name: | Perineometer |
| Classification Name: | Perineometer |
| Classification Regulation: | 21 CFR § 884.1425 |
| Product Code: | HIR |
| Device Class: | Class II |
| Classification Panel: | Obstetrics/Gynecology |
### 5. Legally Marketed Predicate Device
Colonial Medical Supply Pelvic Muscle Therapy 510(k) Number K002830
### 6. Device Description
The Vibrance Kegel Device (VKD700) is an intra-vaginal exercise device intended to strengthen the pelvic floor musculature by offering resistance to an individual's voluntary contractions of these muscles. The Vibrance Kegel Device incorporates an active vibration
{5}------------------------------------------------
biofeedback mechanism for the identification of the pelvic floor muscles, and the sheaths provide graduated resistance levels for the progressive strengthening of the pelvic floor muscle.
The therapeutic effect is from the user using the device to exercise the pelvic floor muscles in accordance with the exercise program included in the instructions. The device provides biofeedback information that allows the user to verify the use of the correct muscles in the exercise routine and thus the effectiveness of the exercises undertaken. The device also includes a resistive component to the training program to strengthen the pelvic floor muscle via physiologic principles. The biofeedback mechanism works by directly sensing the force applied by the pelvic floor muscle.
The device consists of a Main Body and three Resistance Sheaths with varying degrees of stiffness. It has a PCB assembly, and is battery powered by two independent 3V CR 1220 Lithium Batteries operating on two independent, enclosed electronic circuits for the vibration biofeedback loop and the other for audio guided training. The maximum diameter of the Vibrance Kegel Device shaft is 2.4 cm. The maximum inserted length of the shaft is 6 cm.
### Main Body
The chassis of the device consists of an Acrylonitrile Butadiene Styrene (ABS) plastic frame with the biofeedback mechanism. It is incorporated within a tubular sheath made of medical grade silicone rubber which encloses the whole chassis. This main body and resistance sheath are the parts of the device that directly contact the user's intra-vaginal cavity. The base of the tubular sheath and chassis are winged to prevent total ingestion of the device into the vaginal canal.
### Resistance Sheath
Three Resistance Sheaths, each with varying levels of stiffness are included with the Vibrance Kegel Device. When a Resistance Sheath is placed over the Main Body, the challenge of the exercise is increased.
### 7. Indication for Use Statement
The Vibrance Kegel Device (VKD) is intended for the strengthening of the pelvic floor muscles, which has been found to help women with urinary incontinence.
### 8. Substantial Equivalence Discussion
The following table compares the Vibrance Kegel Device to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
{6}------------------------------------------------
| Manufacturer | Bioinfinity (M) Sdn. Bhd. | Colonial Medical Supply |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Vibrance Kegel Device | Pelvic Muscle Therapy (PMTx) |
| | Image: Vibrance Kegel Device | Image: Pelvic Muscle Therapy (PMTx) |
| 510(k) Number | K141893 | K002830 |
| Product Code | HIR | HIR |
| Regulation Number | 21 CFR § 884.1425 | 21 CFR § 884.1425 |
| Regulation Name | Perineometer | Perineometer |
| Indications for Use | The Vibrance Kegel Device (VKR) is intended for the strengthening of the pelvic floor muscles, which has been found to help women with urinary incontinence. | Pelvic muscle trainer assists the user to perform Kegel exercises, by offering resistance, which may help in the treatment of urinary incontinence. |
| Over the Counter (OTC) | Yes | Yes |
| Anatomical Sites | Vagina | Vagina |
| Feature | Resistive vaginal exerciser | Resistive vaginal exerciser |
| Target Population | Women with mild incontinence | Women with mild incontinence |
| Anatomical Site | Vagina | Vagina |
| Single Patient Device | Yes | Yes |
| Single Use or Reusable | Reusable | Reusable |
| Provided Sterile | Clean, but not sterile | Clean, but not sterile |
| Biofeedback display information | No | Numerical response to muscle contraction strength |
| Device Design | Handheld device consisting of a Main Body and three gradual Resistance Sheaths | Handheld pneumatically based device with vaginal silicone sensor |
| Material | Main Body: ABS Plastic, PCB Assembly, two low electrical energy coin cell batteries (CR1220)<br>Sheaths: Medical grade silicone (polydimethysiloxane) | Vaginal Probe: Medical grade silicone (polydimethysiloxane)<br>Tubing: Medical grade silicone (polydimethysiloxane)<br>Monitor: Not known |
| RoHs Compliant | Yes | Not applicable |
| Operating Principle | Resistive pelvic floor strengthener | Resistive pelvic floor strengthener |
| Resistance Component | The Vibrance Kegel Device<br>Electromechanical spring contact offers 3 levels of resistance sheaths with increasing stiffness | Balloon silicone sensor |
| Biocompatibility | Guidelines set forth in ISO 10993 testing results indicate the material is biocompatible, nontoxic, and well tolerated by mucosal membranes. | Guidelines set forth in ISO 10993 testing results indicate the material is biocompatible, nontoxic, and well tolerated by mucosal membranes. |
| Chemical Safety | Addressed by biocompatibility testing per ISO 10993 | Addressed by biocompatibility testing per ISO 10993 |
| Instructions for Use | Manual | Manual |
| Energy Use and/or | Energy is supplied by two independent | None - pneumatic device provides an |
| Manufacturer | Bioinfinity (M) Sdn. Bhd. | Colonial Medical Supply |
| Trade Name | Vibrance Kegel Device | Pelvic Muscle Therapy (PMTx) |
| Delivered | replaceable 3.0 Volt CR-1220 lithium<br>/manganese dioxide (Li/MnO2) Button<br>Cell batteries | air pressure reading that measures the<br>strength of the pelvic floor muscle<br>contractions |
| Packaging | Device in sealed plastic bag and<br>manual in cardboard box | Device in sealed plastic bag and<br>manual in cardboard box |
| Dimensions | 6 cm x 2 cm | 7.5 cm x 3 cm |
#### Table 5A – Comparison of Characteristics
{7}------------------------------------------------
### 9. Non-Clinical Performance Data
The following testing has been performed to support substantial equivalence:
Mechanical Testing – Compression testing was conducted on the Vibrance Kegel Device to measure the forces required to vibrate the sensor for each of the three levels of resistance sheath. A pull off test was also conducted to evaluate how much force is needed for the sheath to be removed from the main body when the device is in use, but is adequate for removal of a sheath after use. A correlation measurement test was then conducted to evaluate the test results of the compression test. The mechanical testing concluded that the different components of the device would function properly and safely when exposed to loads and configurations representative of normal use.
Biocompatibility - The biological safety of the Vibrance Kegel Device was evaluated in accordance with ISO 10993-1:2009 and guidance document entitled Blue Book Memo, G95-1, Use of International Standard ISO-10993, and Biological Evaluation of Medical Devices: Part 1: Evaluation and Testing. Under these, for the stated indications for use, each component of the device's biological safety was evaluated for in vitro cytotoxicity, skin sensitization, and vaginal irritation.
- In Vitro Cytotoxicity test was performed according to ISO 10993-12 and ISO 10993-5, with the results suggesting that the test articles did not induce a cytotoxic effect.
- Skin Sensitization (Maximization Test) was performed according to ISO 10993-10, with the results suggesting that the device extracts did not produce skin sensitization in guinea pigs.
- Irritation (Vaginal Irritation Study) was performed on New Zealand White Rabbits according to ISO 10993-10, with the results showing that there were no significant clinical signs and gross findings in either the control or test group, and there were no mortalities.
Electromagnetic Compatibility - The Vibrance Kegel Device complies with the applicable voluntary standards which include IEC 60601-1 and IEC 60601-1-11. The Vibrance Kegel Device complies with these voluntary standards including IEC 60601-1-11 for Electrical Safety and for systems used in the home healthcare environment.
Risk Analysis - Formal Risk Assessment of the Vibrance Kegel Device was performed in accordance with ISO 14971. With respect to perceivable conditions in which the device would be subjected to a worst-case environmental or human error scenario, Bioinfinity believes the outcomes of these risks are considered acceptable within the context of ISO 14971, and that all potential risks have been mitigated to the lowest form.
{8}------------------------------------------------
### Performance Testing Summary
As part of demonstrating safety and effectiveness of Vibrance Kegel Device and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, Bioinfinity (M) Sdn. Bhd. completed a number of tests. The Vibrance Kegel Device meets all the requirements for overall design, biocompatibility, and electrical safety testing performance confirms that the output meets the design inputs and specifications. The Vibrance Kegel Device passed all testing stated above as shown by the acceptable results obtained.
The Vibrance Kegel Device complies with the applicable voluntary standards for biocompatibility. The device passed all the testing in accordance with national and international standards.
### 10. Clinical Performance Data
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.
### 11. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device; or the device has the same intended use and different technological characteristics that can be demonstrated as substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) submission that the differences between the Vibrance Kegel Device and the predicate device do not raise any questions regarding its safety and effectiveness. The Bioinfinity (M) Sdn. Bhd. Vibrance Kegel Device, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.