← Product Code [ILY](/productcode/ILY) · K141892

# QK-SYSTEMIC THERAPEUTIC APPARATUS (K141892)

_Harbin Quanke Medi-Technical Development Co., Ltd. · ILY · Jan 28, 2015 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K141892

## Device Facts

- **Applicant:** Harbin Quanke Medi-Technical Development Co., Ltd.
- **Product Code:** [ILY](/productcode/ILY.md)
- **Decision Date:** Jan 28, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5500
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The subject device may be used for the temporary relief of minor muscle and joint pain and stiffness, the temporary relief of minor joint pain associated with arthritis, the temporary increase in local circulation where applied, and relaxation of muscles. In addition, the device may also help muscle spasms, minor sprains and strains, and minor muscular back pain.

## Device Story

QK Therapeutic Apparatus (models QK-C01A, QK-C01B, QK-C01CZ, QK-C01EZ, QK-S14, QK-C02FZ, QK-C02D) provides topical heating to the body. Device consists of energy generator, height adjustment knob, support, and control box. Energy generator uses semiconductor ceramic pieces to emit infrared waves (6K-12K nm range; peak 9300nm) to the treatment area. Used in clinical or home settings; operated by patient or clinician. Device is positioned 20-30cm from skin; provides 40-45°C surface temperature. Heat application promotes local circulation, muscle relaxation, and temporary pain relief. Desktop and floor-standing configurations available.

## Clinical Evidence

Bench testing only. Compliance with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC) standards demonstrated. Performance testing confirmed irradiation distance (20-30cm), irradiation time (30-60min/turn), and skin temperature (40-45°C). No clinical data provided.

## Technological Characteristics

Infrared therapeutic heating lamp. Energy source: AC120V, 50-60Hz. Emission spectrum: 6K-12K nm; peak 9300±465nm. Materials: Semiconductor ceramic energy generator. Form factor: Desktop and floor-standing models. Safety standards: IEC 60601-1, IEC 60601-1-2. Generator lifetime: 15,000 hours.

## Regulatory Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

## Predicate Devices

- TDP Heat Lamps ([K020851](/device/K020851.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are connected by a flowing line that forms the shape of a bird. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 28, 2015

Harbin Quanke Medi-Technical Development Co., Ltd. Xianglin Wang CEO No.42-1 Lane2 JingWei Str. Daoli Dist. Harbin, Heilongjiang, China 150010

Re: K141892

Trade/Device Name: QK-Therapeutic Apparatus; Models QK-C01A, QK-C01B, QK-C01CZ, QK-C01EZ, QK-S14, QK-C02FZ, QK-C02D Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: ILY Dated: December 10, 2014 Received: December 29, 2014

Dear Xianglin Wang,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Felipe Aquel -S

for Carlos Peña, Ph.D., M.S. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

#### 510(k) Number (if known) K141892

#### Device Name

QK Therapeutic Apparatus; Models QK-C01A, QK-C01CZ, QK-C01EZ, QK-S14, QK-C02FZ, and QK-C02D

#### Indications for Use (Describe)

The subject device may be used for the temporary relief of min and stiffness, the temporary relief of minor joint pain associated with arthritis, the temporary increase in local circulation where applied, and relaxation of muscles. In addition, the device may also help muscle spasms, minor sprains and minor muscular back pain.

| Type of Use (Select one or both, as applicable)                                       |
|---------------------------------------------------------------------------------------|
| <div> <span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
| <div> <span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows two logos. The first logo features a stylized letter "R" with a teal arc above it, accompanied by the registered trademark symbol. The second logo consists of red stylized letters "QK" enclosed in a red square frame, with Chinese characters below it. The logos appear to represent two different brands or organizations.

#### Section 5 510(K) Summary

#### 1. Prepared date: 2015-1-23

# 2. Submitter Information

| Name           | Harbin Quanke Medi-Technical Development Co.,Ltd.           |  |  |
|----------------|-------------------------------------------------------------|--|--|
| Address        | No.42-1 Lane2 JingWei Str .Daoli Dist. Harbin,Heilongjiang, |  |  |
|                | China, 150010                                               |  |  |
| Tel            | +86 451-84674063                                            |  |  |
| Fax            | +86 451-84671775                                            |  |  |
| Contact person | Wang xianglin                                               |  |  |
| E-mail         | qkgs@vip.163.com                                            |  |  |

#### 3. Device Information

| Trade name            | QK Therapeutic Apparatus                                        |
|-----------------------|-----------------------------------------------------------------|
| Model                 | QK-C01A, QK-C01B, QK-C01CZ, QK-C01EZ, QK-S14, QK-C02FZ, QK-C02D |
| Common name           | Infrared therapeutic heating lamp                               |
| Classification name   | lamp, infrared, therapeutic heating                             |
| Regulatory class      | Class 2                                                         |
| Production regulation | 21CFR Part 890.5500                                             |
| Product code          | ILY                                                             |
| Panel                 | Physical Medicine                                               |

#### 4. Predicate Device

| <b>Submitter</b> | C&H International Inc                 |
|------------------|---------------------------------------|
| Manufacturer     | Chongqing Silicate Research Institute |
| Trade name       | TDP Heat Lamps                        |
| Model            | CQG-111A/B,CQG-222A/B/D,CQG-270A/B    |
| Product code     | ILY                                   |
| 510(K)           | K020851                               |

# 5. Device Description

QK Therapeutic Apparatus, including 7 different models, consists of Energy generator,Height adjustment knob,Support ,and Control box. It is used to provide topical heating to the body by the energy generator.The energy generator is composed of many semiconductor ceramic piece, and it can emit multiple peak wide-band nanometer waves which can be absorbed by human.

The device C01A,C01B,C01CZ,C01EZ and QK-S14 are desktop style,while the other models are Floor Standing. The Emission spectrum ranges from 6K to 12K nm, the Energy Peak Value Wave Range is 9300nm±465nm and the All-direction

{4}------------------------------------------------

# Harbin Quanke Medi-Technical Development Co.,Ltd.

Eradiation Rate is at least 83 %,The energy generator shall be replaced after 15000 hours of usage.

# 6. Indications for Use

The subject device may be used for the temporary relief of minor muscle and joint pain and stiffness, the temporary relief of minor joint pain associated with arthritis, the temporary increase in local circulation where applied, and relaxation of muscles. In addition, the device may also help muscle spasms, minor sprains and strains, and minor muscular back pain.

# 7. Summary of Substantially Equivalent

The subject device and the predicate device have the same intended use, operation principle and Conformance standard, and have the similar technical data, see the Table 1

| Compared<br>item             | Subject Device<br>Present application                                                                                                                                                                                                                                                                                                                                                                                | Predicate Device<br>K020851                                                                                                                                                                                                                                                                                                                                                                                          | Comm<br>ent |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Indications<br>for use       | The subject device may be<br>used for the temporary relief<br>of minor muscle and joint<br>pain and stiffness, the<br>temporary relief of minor<br>joint pain associated with<br>arthritis, the temporary<br>increase in local circulation<br>where applied, and<br>relaxation of muscles. In<br>addition, the device may also<br>help muscle spasms, minor<br>sprains and strains, and<br>minor muscular back pain. | The TDP Heat Lamp may<br>be used for the temporary<br>relief of minor muscle and<br>joint pain and stiffness,<br>the temporary relief of<br>minor joint pain<br>associated with arthritis,<br>the temporary increase in<br>local circulation where<br>applied, and relaxation of<br>muscles. In addition, the<br>lamp may also help<br>muscle spasms, minor<br>sprains and strains, and<br>minor muscular back pain. | Same        |
| Operation<br>principle       | The device's energy<br>generator can emit infrared<br>wave to provide topical<br>heating for the purpose of<br>temporary pain relief of<br>muscular pain, arthritis,<br>shoulder pain, back pain and<br>joint stiffness.                                                                                                                                                                                             | The mineral curing plate<br>of the TDP heat lamp can<br>emit a infrared wave to<br>provide topical heating for<br>the purpose of<br>temporary pain relief of<br>muscular pain, arthritis,<br>shoulder pain, back pain<br>and joint stiffness                                                                                                                                                                         | Same        |
| Spectrum                     | 6K~12Knm                                                                                                                                                                                                                                                                                                                                                                                                             | 2K~25Knm                                                                                                                                                                                                                                                                                                                                                                                                             | Similar     |
| Ranges                       |                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                      |             |
| Peak wave<br>length          | 9300±465nm                                                                                                                                                                                                                                                                                                                                                                                                           | 10000±1000 nm                                                                                                                                                                                                                                                                                                                                                                                                        | Similar     |
| Length of<br>generator       | 18.5~35.5cm                                                                                                                                                                                                                                                                                                                                                                                                          | 16.6-12.4cm                                                                                                                                                                                                                                                                                                                                                                                                          | Similar     |
| Lifetime of<br>generator     | 15000h                                                                                                                                                                                                                                                                                                                                                                                                               | 1000~1500h                                                                                                                                                                                                                                                                                                                                                                                                           | Similar     |
| Control<br>Timer             | No or Yes                                                                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                  | Similar     |
| Power<br>Supply              | AC120V                                                                                                                                                                                                                                                                                                                                                                                                               | 220~240V                                                                                                                                                                                                                                                                                                                                                                                                             | Similar     |
| Power                        | 160~230W                                                                                                                                                                                                                                                                                                                                                                                                             | 120~300W                                                                                                                                                                                                                                                                                                                                                                                                             | Similar     |
| Power<br>Frequency           | 50~60Hz                                                                                                                                                                                                                                                                                                                                                                                                              | 50~60Hz                                                                                                                                                                                                                                                                                                                                                                                                              | Same        |
| Warm-Up<br>Time              | 3 min                                                                                                                                                                                                                                                                                                                                                                                                                | 10min                                                                                                                                                                                                                                                                                                                                                                                                                | Similar     |
| Irradiation<br>distance      | 20-30cm                                                                                                                                                                                                                                                                                                                                                                                                              | 30-40cm                                                                                                                                                                                                                                                                                                                                                                                                              | Similar     |
| Irradiation<br>time          | High: 20-40min/turn<br>Low:30-60min/turn                                                                                                                                                                                                                                                                                                                                                                             | 15-60min/turn                                                                                                                                                                                                                                                                                                                                                                                                        | Similar     |
| Maintain<br>time             | 15-20min                                                                                                                                                                                                                                                                                                                                                                                                             | 15-20min                                                                                                                                                                                                                                                                                                                                                                                                             | Similar     |
| Skin<br>Temperatu<br>re      | 40-45°C                                                                                                                                                                                                                                                                                                                                                                                                              | 40-45°C                                                                                                                                                                                                                                                                                                                                                                                                              | Same        |
| Style                        | Desktop&Floor standing                                                                                                                                                                                                                                                                                                                                                                                               | Desktop&Floor standing                                                                                                                                                                                                                                                                                                                                                                                               | Same        |
| Operating<br>Environme<br>nt | Temperature: 5℃~40℃<br>Humidity: ≤85%RH                                                                                                                                                                                                                                                                                                                                                                              | Temperature: 10°C ~ 40°C<br>Humidity: 10%~ 95% RH                                                                                                                                                                                                                                                                                                                                                                    | Similar     |
| Storage<br>Environme<br>nt   | Temperature: -40°C ~55°C<br>Humidity : ≤85%RH                                                                                                                                                                                                                                                                                                                                                                        | Temperature: 10°C ~40°C<br>Humidity: 10% ~ 95% RH                                                                                                                                                                                                                                                                                                                                                                    | Similar     |
| Conforman<br>ce<br>standard  | IEC60601-1(Safety),<br>IEC60601-1-2(EMC)                                                                                                                                                                                                                                                                                                                                                                             | IEC60601-1(Safety),<br>IEC60601-1-2(EMC)                                                                                                                                                                                                                                                                                                                                                                             | Same        |

Table 1

{5}------------------------------------------------

# Harbin Quanke Medi-Technical Development Co.,Ltd.

{6}------------------------------------------------

# Harbin Quanke Medi-Technical Development Co.,Ltd.

From the comparison form above, the subject device and predicate device have the same Indications for use, Operation principle, Power Frequency, skin temperature , maintain time, Style and Conformance standard.

In the Spectrum Range, length of generator lifetime, Warm-up time, Irradiation distance, Irradiation time items, the subject device is slightly different from the predicate device, but the subject meets the requirements of IEC60601-1, IEC60601-1-2 and performance, so these difference does not bring any problem of effectiveness and safety.

# 8. Performance Data

- 8.1 The subject device has tested and complied with the following voluntary recognized standards:
	- IEC 60601-1 Medical Electrical Equipment-Part 1: General Requirements for . Safety, 1988+A1:1991+A2:1995
	- IEC 60601-1-2 Medical Electrical Equipment-Part 1-2: General Requirements . for Safety -Collateral Standard: Electromagnetic Compatibility-Requirements and Tests, 2007.
- 8.2 According to the performance test result in section 11of this submission ,the irradiation distance is 20-30cm, the irradiation time is 30-60min/turn, and the skin Temperature is 40-45°C.

### 9. Conclusion

The QK Therapeutic Apparatus is substantially equivalent to the predicate devices in K020851.

---

**Source:** [https://fda-staging.innolitics.com/device/K141892](https://fda-staging.innolitics.com/device/K141892)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).