K141879 · Sauter AG · FLE · Jun 17, 2015 · General Hospital
Device Facts
Record ID
K141879
Device Name
BELIMED STEAM STERILIZER MST-H
Applicant
Sauter AG
Product Code
FLE · General Hospital
Decision Date
Jun 17, 2015
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.6880
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Belimed Steam Sterilizers MST-H, models 9-6-12, 9-6-15, and 9-6-18 are designed for sterilization of non-porous and porous heat and moisture-stable materials used in healthcare facilities.
Device Story
Belimed Steam Sterilizer MST-H is a steam-based sterilization system for healthcare facilities. It processes non-porous and porous heat/moisture-stable materials using predefined cycles (PreVac, Gravity, Liquid, Bowie-Dick, Air Leak). The device uses demineralized water to generate steam; operates via PLC control with a 10" TFT color HMI on the loading side and LSC display on the unloading side. It features a cubic welded stainless steel chamber (AISI 316L) and motor-driven doors. The system monitors chamber temperature and pressure via independent sensors; errors trigger acoustic/visual alarms and are recorded on a thermal printer. The device ensures sterility assurance levels (SAL) of 10^-6. It is designed for use by healthcare personnel to sterilize instrument trays and fabric packs, improving workflow through automated cycle management and validated air removal.
Clinical Evidence
Bench testing only. Validation studies performed per ANSI/AAMI ST8:2013. Testing included empty chamber thermal distribution, biological indicator kill-time studies (SAL 10^-6), moisture retention testing (fabrics <3%, instruments <20%), and Bowie-Dick test verification. Liquid cycle validated for 121°C exposure for 12 minutes.
Technological Characteristics
Materials: AISI 316L stainless steel chamber/piping, AISI 321 door. Sensing: Absolute pressure transmitter, temperature sensors. Energy: Steam (250°F, 270°F, 275°F). Connectivity: Thermal printer, PLC-based control. Sterilization: Steam. Standards: ANSI/AAMI ST8:2013, ASME Section VIII Div 1, IEC 61010-1/2-040, UL 61010-1.
Indications for Use
Indicated for sterilization of non-porous and porous heat and moisture-stable materials in healthcare facilities. Not intended for liquids used for direct patient contact.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
Predicate Devices
Belimed Steam Sterilizer TOP 5000 Series 9-18 (K021223)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three human profiles facing right, connected by a flowing ribbon-like design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 17, 2015
Belimed Sauter AG Michael Hari Head of Medical Engineering Zelgstrasse 8 CH8583 Sulgen, Switzerland
Re: K141879
> Trade/Device Name: Belimed Steam Sterilizer MST-H Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: II Product Code: FLE Dated: May 18, 2015 Received: May 20, 2015
Dear Mr. Hari,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Hari
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
## 510(k) Number:
K141879
Device Name: Belimed Steam Sterilizer MST-H
Indications for Use
The Belimed Steam Sterilizers MST-H, models 9-6-12, 9-6-15, and 9-6-18 are designed for sterilization of non-porous and porous heat and moisture-stable materials used in healthcare facilities.
The Belimed Steam Sterilizers MST-H, models 9-6-15, and 9-6-18 are available in a single door (HS1) or double door (HS2) version.
Type of Use (Select one or both, as applicable) □ Prescription Use (Part 21 CFR 801 Subpart D) 区 Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
The Belimed Steam Sterilizer MST-H is offered with the following factory-programmed predefined sterilization cycles:
| Cycle | Pre-treatment | Sterilizing temperature | Sterilizing Time 1) | Drying time 2) | Recommended load 3) |
|----------------------------|-------------------------------------|-------------------------|---------------------|----------------|-------------------------------------------------------------------------------------------------------|
| PreVac 270<br>4S/30D | 5 pulses | 270 °F | 4 min | 30 min | Double wrapped instrument<br>trays, max. weight of 25 lbs<br>per tray, Fabric packs |
| PreVac 270<br>4S/5D | 5 pulses | 270 °F | 4 min | 5 min | Fabric packs<br>4) |
| PreVac 275<br>3S/30D | 5 pulses | 275 °F | 3 min | 30 min | Double wrapped instrument<br>trays, max. weight of 25 lbs<br>per tray, Fabric packs |
| Bowie-Dick<br>Test | 5 pulses | 273 °F | 3.5 min | 1 min | One Bowie-Dick-Test-Pack |
| Gravity 270<br>15S/30D | Purge | 270 °F | 15 min | 30 min | Double wrapped instrument<br>trays, with non-porous<br>instruments, max. weight of<br>25 lbs per tray |
| Liquid 250 45S | Purge | 250 °F | 45 min | 0 min | Liquids, up to 1000 ml in<br>vented or open containers 5) |
| Air leak Test | Vacuum: 65mbar<br>Test time: 15 min | - | - | - | Empty chamber |
| Warm up & Air<br>Leak Test | Vacuum: 65mbar<br>Test time: 15 min | 270 °F | 3 min | 3 min | Empty chamber |
| Warm up | 2 pulses | 270 °F | 3 min | 3 min | Empty chamber |
Table 1: Factory programmed Sterilization cycles
Notes for Table 1:
- 1. Only factory set sterilizing time should be utilized with medical devices intended for patient contact.
- 2. Factory set drying time. These are the minimum validated drying times, which may be extended depending on load conditions and wraps.
- 3. Recommended load: Refer to table 2 and 3. The load configurations listed in table 2 and 3 are those used during performance testing of the sterilizer models.
- 4. Fabric load should be preconditioned between 68°F and at a relative humidity of at least 35% to 60% for at least 2 hours.
- 5. Liquid cycle is not intended to sterilize liquids that are used for direct patient contact.
{4}------------------------------------------------
| Model | Instrument trays,<br>max. 25 Ibs each | Fabric Packs,<br>max. 3.3 lbs each |
|--------|---------------------------------------|------------------------------------|
| 9-6-12 | 16 | 32 |
| 9-6-15 | 20 | 40 |
| 9-6-18 | 24 | 48 |
The following tables show the recommended loads for the Belimed Steam Sterilizer MST-H:
Table 2: Recommended loads for non-porous and porous load cycles
| Model | Volume of liquid in one container | Number of containers |
|--------|-----------------------------------|----------------------|
| 9-6-12 | 1000 ml | 168 |
| 9-6-15 | 1000 ml | 210 |
| 9-6-18 | 1000 ml | 252 |
Table 3: Recommended loads for liquid cycle 250 °F
The Belimed Steam Sterilizer MST-H is offered in the following models:
| Model | Configuration | Net loading<br>capacity (L) | Chamber size<br>(H x W x D) (mm) | Overall Dimensions<br>(H x W x D) (mm) |
|------------|---------------|-----------------------------|----------------------------------|----------------------------------------|
| 9-6-12 HS1 | 1 door | 691 | 1080 x 660 x 1398 | 1970 x 1700 x 1714 |
| 9-6-12 HS2 | 2 doors | 691 | 1080 x 660 x 1398 | 1970 x 1700 x 1714 |
| 9-6-15 HS1 | 1 door | 864 | 1080 x 660 x 1706 | 1970 x 1700 x 2022 |
| 9-6-15 HS2 | 2 doors | 864 | 1080 x 660 x 1706 | 1970 x 1700 x 2022 |
| 9-6-18 HS1 | 1 door | 977 | 1080 x 660 x 2014 | 1970 x 1700 x 2330 |
| 9-6-18 HS2 | 2 doors | 977 | 1080 x 660 x 2014 | 1970 x 1700 x 2330 |
Table 4: Models and dimensions
{5}------------------------------------------------
# 510(k) Summary
| Date: | June 10, 2015 |
|-----------------------------|----------------------------------------------------------------------------------------------------|
| Submitter's Name / Address: | Belimed Sauter AG<br>Zelgstrasse 8<br>CH- 8583 Sulgen / Switzerland |
| Contact Person: | Michael Hari<br>Email: michael.hari@belimed.ch<br>Phone: +41 71 644 86 37<br>Fax: +41 71 644 85 11 |
| Trade Name: | Belimed Steam Sterilizer MST-H |
| Models: | 9-6-12, 9-6-15 and 9-6-18 |
| Classification: | Steam Sterilizer – Class II, as listed per 21 CFR 880.6880<br>Product Code: FLE |
| Predicate Device: | Belimed Steam Sterilizer TOP 5000<br>Series 9-18 (K021223) |
## INTENDED USE:
The Belimed Steam Sterilizers MST-H, models 9-6-12, 9-6-15, and 9-6-18 are designed for sterilization of non-porous and porous heat and moisture-stable materials used in healthcare facilities.
## DESCRIPTION OF DEVICE:
The Belimed Steam Sterilizers MST-H, models 9-6-12, 9-6-15, and 9-6-18 are available in a single door (HS1) or double door (HS2) version.
The sterilizers are equipped with Pre-vacuum, Gravity, Liquid, Air Leak Test and Bowie-Dick Test cycles.
{6}------------------------------------------------
The Belimed Steam Sterilizer MST-H is offered with the following factory-programmed predefined sterilization cycles:
| Cycle | Pre-treatment | Sterilizing<br>temperature | Sterilizing<br>Time 1) | Drying<br>time 2) | Recommended load<br>3) |
|----------------------------|-------------------------------------|----------------------------|------------------------|-------------------|-------------------------------------------------------------------------------------------------------|
| PreVac 270<br>4S/30D | 5 pulses | 270 °F | 4 min | 30 min | Double wrapped instrument<br>trays, max. weight of 25 lbs<br>per tray, Fabric packs |
| PreVac 270<br>4S/5D | 5 pulses | 270 °F | 4 min | 5 min | Fabric packs<br>4) |
| PreVac 275<br>3S/30D | 5 pulses | 275 °F | 3 min | 30 min | Double wrapped instrument<br>trays, max. weight of 25 lbs<br>per tray, Fabric packs |
| Bowie-Dick<br>Test | 5 pulses | 273 °F | 3.5 min | 1 min | One Bowie-Dick-Test-Pack |
| Gravity 270<br>15S/30D | Purge | 270 °F | 15 min | 30 min | Double wrapped instrument<br>trays, with non-porous<br>instruments, max. weight of<br>25 lbs per tray |
| Liquid 250 45S | Purge | 250 °F | 45 min | 0 min | Liquids, up to 1000 ml in<br>vented or open containers 5) |
| Air leak Test | Vacuum: 65mbar<br>Test time: 15 min | - | - | - | Empty chamber |
| Warm up & Air<br>Leak Test | Vacuum: 65mbar<br>Test time: 15 min | 270 °F | 3 min | 3 min | Empty chamber |
| Warm up | 2 pulses | 270°F | 3 min | 3 min | Empty chamber |
Table 1: Factory programmed Sterilization cycles
Notes for Table 1:
- 1. Only factory set sterilizing time should be utilized with medical devices intended for patient contact.
- 2. Factory set drying time. These are the minimum validated drying times, which may be extended depending on load conditions and wraps.
- 3. Recommended load: Refer to table 2 and 3. The load configurations listed in table 2 and 3 are those used during performance testing of the sterilizer models.
- 4. Fabric load should be preconditioned between 68°F and at a relative humidity of at least 35% to 60% for at least 2 hours.
- 5. Liquid cycle is not intended to sterilize liquids that are used for direct patient contact.
| Model | Instrument trays,<br>max. 25 lbs each | Fabric Packs,<br>max. 3.3 lbs each |
|--------|---------------------------------------|------------------------------------|
| 9-6-12 | 16 | 32 |
| 9-6-15 | 20 | 40 |
| 9-6-18 | 24 | 48 |
The following tables show the recommended loads for the Belimed Steam Sterilizer MST-H:
Table 2: Recommended loads for non-porous and porous load cycles
{7}------------------------------------------------
| Model | Volume of liquid in one<br>container | Number of containers |
|--------|--------------------------------------|----------------------|
| 9-6-12 | 1000 ml | 168 |
| 9-6-15 | 1000 ml | 210 |
| 9-6-18 | 1000 ml | 252 |
Table 3: Recommended loads for liquid cycle 250 °F
The Belimed Steam Sterilizer MST-H is offered in the following models:
| Model | Configuration | Net loading<br>capacity (L) | Chamber size<br>(H x W x D) (mm) | Overall Dimensions<br>(H x W x D) (mm) |
|------------|---------------|-----------------------------|----------------------------------|----------------------------------------|
| 9-6-12 HS1 | 1 door | 691 | 1080 x 660 x 1398 | 1970 x 1700 x 1714 |
| 9-6-12 HS2 | 2 doors | 691 | 1080 x 660 x 1398 | 1970 x 1700 x 1714 |
| 9-6-15 HS1 | 1 door | 864 | 1080 x 660 x 1706 | 1970 x 1700 x 2022 |
| 9-6-15 HS2 | 2 doors | 864 | 1080 x 660 x 1706 | 1970 x 1700 x 2022 |
| 9-6-18 HS1 | 1 door | 977 | 1080 x 660 x 2014 | 1970 x 1700 x 2330 |
| 9-6-18 HS2 | 2 doors | 977 | 1080 x 660 x 2014 | 1970 x 1700 x 2330 |
Table 4: Models and dimensions
{8}------------------------------------------------
## NONCLINICAL COMPARISON TO PREDICATE DEVICE:
The Belimed Steam Sterilizers MST-H, models 9-6-12, 9-6-15, and 9-6-18 are very similar to the predicate device.
A summary of the differences to the predicate device is included in Table 5 :
| General Sterilizer<br>Features | Belimed Steam Sterilizer<br>TOP5000<br>Series 9-18 (K021223) | Belimed Steam Sterilizer<br>MST-H | Substantially<br>Equivalent or<br>Different |
|-----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Terminal Sterilization of non-<br>porous and porous heat and<br>moisture-stabile materials used in<br>healthcare facilities | Terminal Sterilization of non-<br>porous and porous heat and<br>moisture-stabile materials used<br>in healthcare facilities | Same |
| Operating Principle | The sterilizing agent of the<br>Belimed Sterilizers MST-H is<br>steam. Steam is made from<br>demineralized water and is<br>therefore non-toxic. Steam is the<br>preferred method to sterilize<br>porous an non-porous heat<br>resistant materials and textiles in<br>hospitals for more than 100 years.<br>Preferred sterilization<br>temperatures are 250 °F, 270 °F<br>and 275 °F. | The sterilizing agent of the<br>Belimed Sterilizers MST-H is<br>steam. Steam is made from<br>demineralized water and is<br>therefore non-toxic. Steam is the<br>preferred method to sterilize<br>porous an non-porous heat<br>resistant materials and textiles in<br>hospitals for more than 100<br>years. Preferred sterilization<br>temperatures are 250 °F, 270 °F<br>and 275 °F. | Same |
| Product code | FLE | FLE | |
| C.F.R: code | 880.6880 Steam Sterilizer | 880.6880 Steam Sterilizer | |
| Class | II | II | |
| Built according to<br>Standard | ANSI/AAMI ST8:1994<br>Hospital Steam Sterilizers | ANSI/AAMI ST8:2013<br>Hospital Steam Sterilizers | Substantially<br>Equivalent,<br>(ST8:2013 uses<br>heavier instrument<br>load and refers to<br>new standard<br>versions) |
| Electrical safety<br>standard | UL 3101-1<br>IEC 61010-1 ed1, am1<br>IEC 61010-2-041 de1 | UL IEC 61010-1: 2004,<br>IEC 61010-2-040:2005 | Substantially<br>Equivalent |
| EMC conformity | IEC / EN 61326:2001 / 2002 | IEC / EN 61326:2001 / 2002 | Same |
| Electrical supply | 3 Phase 208V, 60Hz | 3 Phase 208V, 60Hz | |
| 9-6-9 &9-6-12 | Max. current: 13A | Max. current: 13A | Same |
| 9-6-15 & 9-6-18 | Max. current: 20A | Max. current: 20A | |
| Utility supply pressure<br>requirement<br>Steam<br>Water<br>Compressed air | 2.5 - 3 bar g<br>2 – 5 bar g<br>5 - 7 bar g | 2.5 - 3.5 bar g<br>2 – 5 bar g<br>5 – 7 bar g | Substantially<br>Equivalent |
| Steam capacity<br>requirement:<br>Model 9-6-9<br>Model 9-6-12<br>Model 9-6-15<br>Model 9-6-18 | 120 kg/h<br>160 kg/h<br>200 kg/h<br>240 kg/h | n/a<br>100 kg/h<br>120 kg/h<br>170 kg/h | Lower,<br>Substantially<br>Equivalent |
| Compressed air<br>capacity requirement: | 10 Nm³/h | 10 Nm³/h | Same |
| Cooling water<br>requirement:<br>Model 9-6-9<br>Model 9-6-12<br>Model 9-6-15<br>Model 9-6-18 | 2 m³/h<br>2.5 m³/h<br>3 m³/h<br>3.5 m³/h | n/a<br>2.4 m³/h<br>2.8 m³/h<br>3.2 m³/h | Lower,<br>Substantially<br>Equivalent |
| Control devices | | | |
| Vacuum device<br>Model 9-6-9, 9-6-12<br>Model 9-6-15, 9-6-18 | Water-ring vacuum pump<br>90 m³/h<br>125 m³/h | Water-ring vacuum pump<br>90 m³/h<br>125 m³/h | Same |
| Additional gas ejector /<br>sterilizer runs with<br>max. cooling water<br>temperatures | No/<br>80 °F | Yes, if optional cooling water<br>loop is not installed<br>90 °F | Substantially<br>Equivalent |
| Steam condenser | Shell-and-tube heat exchanger,<br>stainless steel | Shell-and-tube heat exchanger,<br>stainless steel | Substantially<br>Equivalent |
| Air Filtration | 0,2µm, efficiency >99.5%<br>>99.97% for 0.3-micron particles | 0,2µm, efficiency >99.5%<br>>99.97% for 0.3-micron particles | Same |
| Control valves<br>for steam, vacuum and<br>condensate | Stainless steel grade 316 L<br>pneumatically driven piston-valves | Stainless steel grade 316 L<br>pneumatically driven piston-<br>valves | Substantially<br>Equivalent |
| Control valves for<br>cooling water | Brass<br>Electric solenoid valves | Fiber-reinforced plastic material<br>Electric solenoid valves | Substantially<br>Equivalent |
| Piping material grade | AISI 316L | AISI 316L | Same |
| Pressure vessel | | | |
| Chamber and jacket<br>material grade | AISI 316L | AISI 316L | Same |
| Chamber construction | Cubic welded structure with all-<br>round heating ducts | Cubic welded structure with all-<br>round heating ducts | Same |
| Chamber design<br>pressure | 39psi | 40psi | Substantially<br>Equivalent |
| Safety valves | Bronce, ASME approved | Bronce, ASME approved | Same |
| Chamber design<br>standard | ASME Section VIII, Division 1, for<br>unfired pressure vessels | ASME Section VIII, Division 1, for<br>unfired pressure vessels | Same |
| Door material grade | AISI 321 | AISI 321 | Same |
| Door opening | Single or double door<br>Horizontal door operation<br>Motor driven | Single or double door<br>Horizontal door operation<br>Motor driven | Substantially<br>Equivalent |
| Door safety:<br>Door safety bar<br>Additional door locking<br>fixture<br>Number of door limit<br>switches | Yes<br>No<br>Two 'door closed' switches per<br>door | Yes<br>Yes<br>one 'door closed' switch per door<br>and one 'door closed &<br>mechanically locked' switch per<br>door | Substantially<br>Equivalent |
| Door seal | Compressed air activated | Compressed air activated | Same |
| Chamber sizes<br>Model:<br>9-6-9<br>9-6-12<br>9-6-15<br>9-6-18<br>Control | (H x W x D) in mm:<br>1080x660x1040<br>1080x660x1400<br>1080x660x1700<br>1080x660x2000 | (H x W x D) in mm:<br>Not available<br>1080x660x1398<br>1080x660x1706<br>1080x660x2014 | Substantially<br>Equivalent |
| Control Technology | PLC | PLC | same |
| HMI Technology<br>loading side | LCD color / 5,7" | TFT color 10" | Different, larger<br>display |
| HMI Technology<br>unloading side | BT2-2004 LCD display /<br>h x w = 35mm x 68mm | LSC display /<br>h x w=27mmx100mm | Substantially<br>Equivalent |
| Printer Technology | Matrix Dot technology<br>42 char /line, paper width 57mm | Thermal technology<br>42 char /line, paper width 57mm | Substantially<br>Equivalent<br>(same Information<br>on printout) |
| Process control | Automatic through all phases of a<br>cycle | Automatic through all phases of a<br>cycle | Same |
| Sterilization process<br>control | Pressure controlled by an<br>absolute pressure transmitter | Pressure controlled by an<br>absolute pressure transmitter | Same |
| Sterilization process<br>recording | Independent temperature and<br>pressure sensors for recording | Independent temperature and<br>pressure sensors for recording | Same |
| Sterilization process<br>monitoring | Temperature sensor for exposure<br>temperature monitoring | Temperature sensor for exposure<br>temperature monitoring | Same |
| Process alarms | Detected errors cause an acoustic<br>and visual alarm. Alarms are<br>recorded on the printer.<br>At cycle end the door on the<br>unloading side (double door<br>models) can't be opened. Alarms<br>must be manually acknowledged<br>on the load side operating panel. | Detected errors cause an<br>acoustic and visual alarm. Alarms are<br>recorded on the printer.<br>At cycle end the door on the<br>unloading side (double door<br>models) can't be opened. Alarms<br>must be manually acknowledged<br>on the load side operating panel. | Same |
| Cycle phases | All cycle phases and sterilizer status are monitored on display. Display of:<br>-Chamber temperature & pressure<br>-Process status<br>-Door status<br>-Alarms, if pending | All cycle phases and sterilizer status are monitored on display. Display of:<br>-Chamber temperature & pressure<br>-Process status<br>-Door status<br>-Alarms, if pending<br>Additional process status display for cycle state above the chamber (LED bar display) | Substantially<br>Equivalent<br>(Additional<br>information) |
| Cycle end | Acoustic signal at cycle end<br>Cycle end Indication on operating panel display. | Acoustic signal at cycle end<br>Cycle end Indication on operating panel display. | Same |
| Performance: | | | |
| Biological performance | SAL of 10-6 reduction | SAL of 10-6 reduction | |
| (half cycle) | Through achievement of no growth at half cycle with validation loads.<br>AAMI ST8:1994 | Through achievement of no growth at half cycle with<br>validation loads.<br>AAMI ST8:2013 | Same |
| Comply with | | | |
| Moisture retention | Fabrics:<br><3% increase in pre-sterilization textiles test pack weight | Fabrics:<br><3% increase in pre-sterilization textiles test pack weight | |
| Comply with | Instruments<br><20% increase in pre-sterilization towel weight of the test load<br>AAMI ST8: 1994 | Instruments<br><20% increase in pre-sterilization towel weight of the test load<br>AAMI ST8:2013 | Same |
| Factory predefined<br>cycles | | | |
| Prevac 270°F | Yes | Yes | |
| Sterilize temperature | 132°C / 270°F | 132°C / 270°F | |
| Sterilize time | 4 minutes | 4 minutes | |
| Dry time | 20 minutes | 30 minutes | |
| No. of pre-vacuums | 4 | 5 (improved air removal) | |
| Qualified according to | AAMI ST8: 1994 | AAMI ST8:2013 | Substantially<br>Equivalent |
| Test Load | Instrument load in double wrapped trays, full load | Instrument load in double wrapped trays, full load | |
| | 17lbs gross weight per tray…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.