← Product Code [MSH](/productcode/MSH) · K141876

# Pasture F550S Surgical N95 Respirator, Pasture F550CS Surgical N95 Respirator, Pasture A520S Surgical N95 Respirator, Pasture A520CS Surgical N95 Respirator, Pasture E520S Surgical N95 Respirator, Pasture E520CS Surgical N95 Respirator (K141876)

_Pasture Pharma Pte, Ltd. · MSH · Oct 28, 2015 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K141876

## Device Facts

- **Applicant:** Pasture Pharma Pte, Ltd.
- **Product Code:** [MSH](/productcode/MSH.md)
- **Decision Date:** Oct 28, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

Pasture F550S, Pasture F550CS, Pasture A520S, Pasture A520CS, Pasture E520S, and Pasture E520CS is a NIOSH certified N95 respirator, and is indicated to be used by the healthcare personnel during procedures to protect both the patient and the healthcare personnel from the transfer of microorganisms, body fluids, and particulate material.

## Device Story

Disposable N95 respirators; duck-bill (F550 series) or cup-shaped (A520/E520 series) designs. Function as physical barriers to airborne particles, microorganisms, and body fluids. Used in clinical settings by healthcare personnel. Input is ambient air; output is filtered air for the wearer while containing wearer-generated particles. Features include multi-layer polypropylene construction, nose cushions, and elastic head straps; some models include adjustable buckles. Benefit is reduced risk of cross-contamination between patient and clinician during procedures.

## Clinical Evidence

Bench testing only. Includes fluid resistance (ASTM F1862), flammability (16 CFR 1610), and biocompatibility (cytotoxicity, irritation, sensitization). Filtration efficiency and differential pressure verified via NIOSH certification (TC-84A-7504, TC-84A-7454, TC-84A-7453).

## Technological Characteristics

Multi-layer polypropylene non-woven fabric; meltblown filter media. F550 series: 5 layers, duck-bill shape, zinc/polyester nosepiece. A520/E520 series: 4 layers, cup shape, aluminum nosepiece (E520) or none (A520). Synthetic elastic straps; optional ABS adjustable buckles. Single-use, non-sterile. NIOSH N95 certified.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- WILLSON ONE-FIT HEALTHCARE PARTICLE RESPIRATOR AND SURGICAL MASK, HC-NB095 ([K070139](/device/K070139.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, overlaid on three human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 28, 2015

Pasture Pharma Pte, Ltd C/O Ms. Sarah Hassan US MedPharm 38129 Spring Canyon Drive Murrieta, CA 92563

Re: K141876

> Trade/Device Name: Pasture F550S Surgical N95 Respirator, Pasture F550CS Surgical N95 Respirator, Pasture A520S Surgical N95 Respirator, Pasture A520CS Surgical N95 Respirator, Pasture E520S Surgical N95 Respirator, Pasture E520CS Surgical N95 Respirator Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: MSH Dated: October 15, 2015 Received: October 16, 2015

Dear Ms. Hassan,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

*Tejashri Purohit-Sheth, M.D.*

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K141876

#### Device Name

Pasture F550S Surgical N95 Respirator, Pasture F50CS Surgical N95 Respirator, Pasture A520S Surgical N95 Respirator, Pasture A520CS Surgical N95 Respirator, Pasture E520S Surgical N95 Respirator, Pasture E520CS Surgical N95 Respirator

#### Indications for Use (Describe)

Pasture F550S, Pasture F550CS, Pasture A520CS, Pasture E520S, and Pasture E520CS is a NIOSH certified N95 respirator, and is indicated to be used by the healthcare personnel during procedures to protect both the patient and the healthcare personnel from the transfer of microorganisms, body fluids, and particulate material.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) SUMMARY

This 510(K) Summary is being submitted in accordance with requirements of 21 CFR 807.92(c)

The assigned 510(k) number is: K141876

#### Submitter Name and Address:

Pasture Pharma Pte, Ltd 8 Boon Lay Way #04-01 Trade Hub 21, Singapore 609964 Lloyd Soong, President & CEO Contact: +65-6515-6561 lloyd@pasturegroup.com

# US agent and correspondent:

Mrs. Sarah Hassan US Med Pharm Supplies, Inc 38129 Spring Canyon Drive Murrieta, CA 92563 Phone: 951-239-1933 Fax: 951-239-1933 E-mail: sarah@usmedpharm.com

Date Prepared: October 27, 2015

# Proprietary Name:

Pasture F550S Surgical N95 Respirator, Pasture F550CS Surgical N95 Respirator, Pasture A520S Surgical N95 Respirator, Pasture A520CS Surgical N95 Respirator, Pasture E520S Surgical N95 Respirator, Pasture E520CS Surgical N95 Respirator

Common name: Surgical N95 NIOSH certified Respirator

{4}------------------------------------------------

### Device Classification and Product Code

Classification Name: Surgical Apparel Regulation: 21 CFR §878.4040 Class: Class II Classification panel: General and Plastic Surgery Product Code: MSH

Predicate Device: K070139 WILLSON ONE-FIT HEALTHCARE PARTICLE RESPIRATOR AND SURGICAL MASK, HC-NB095 SURVIVAIR, INC

#### Intended Use:

Pasture F550S, Pasture F550CS, Pasture A520S, Pasture A520CS, Pasture E520S, and Pasture E520CS is a NIOSH certified N95 respirator, and is indicated to be used by the healthcare personnel during procedures to protect both the patient and the healthcare personnel from the transfer of microorganisms, body fluids, and particulate material.

#### Device Description:

Pasture F550S and Pasture F550CS respirators are disposable duck bill shaped N95 respirators with NIOSH certification number TC-84A-7504. They contain 5 layers composed of polypropylene, with a nose cushion, a synthetic elastic loop or strap, and a nosepiece which is the combination of zinc wires and embedded polyester inside of layers. F550CS is the same respirator as the F550S, but F550CS contains an adjustable buckle on the headband.

Pasture A520S and Pasture A520CS respirators are cup shaped N95 respirators with NIOSH certification number TC-84A-7454. They contain 4 layers composed of polypropylene with a synthetic elastic loop or a strap. A520CS is the same respirator as the A520S, but A520CS contains an adjustable buckle on the headband.

Pasture E520S and Pasture E520CS respirators are cup shaped N95 respirators with NIOSH certification number TC-84A-7453. They contain 4 layers composed of polypropylene with a 

{5}------------------------------------------------

nose cushion, synthetic elastic loop or strap, and an aluminum nosepiece. E520CS is the same respirator as E520S, but E520CS contains an adjustable buckle on the headband.

|                         | Predicate                                                   | Pasture                                                                        | Pasture                                                                        | Pasture                                                     | Pasture                                                     | Pasture                                                     | Pasture                                                     | Test                             | K070139                   | F550S                                                                                                                                                                                                                                                                                                                                               | A520S                                       | E520S                                       |
|-------------------------|-------------------------------------------------------------|--------------------------------------------------------------------------------|--------------------------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------|----------------------------------|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|---------------------------------------------|
| Description             | K070139                                                     | F550S                                                                          | F550CS                                                                         | A520S                                                       | A520CS                                                      | E520S                                                       | E520CS                                                      |                                  |                           | F550CS                                                                                                                                                                                                                                                                                                                                              | A520CS                                      | E520CS                                      |
| NIOSH                   | N95                                                         | N95                                                                            | N95                                                                            | N95                                                         | N95                                                         | N95                                                         | N95                                                         | NIOSH                            | N95                       | N95                                                                                                                                                                                                                                                                                                                                                 | N95                                         | N95                                         |
| certification<br>number | TC-84A-4516                                                 | TC-84A-7504                                                                    | TC-84A-7504                                                                    | TC-84A-7454                                                 | TC-84A-7454                                                 | TC-84A-7453                                                 | TC-84A-7453                                                 | Certification                    | TC-84A-4516               | TC-84A-7504                                                                                                                                                                                                                                                                                                                                         | TC-84A-7454                                 | TC-84A-7453                                 |
| Outer layer             | Polyester                                                   | Polypropylene<br>Spunbond                                                      | Polypropylene<br>Spunbond                                                      | Polypropylene<br>Spunbond                                   | Polypropylene<br>Spunbond                                   | Polypropylene<br>Spunbond                                   | Polypropylene<br>Spunbond                                   | Fluid Resistance<br>(ASTM F1862) | Not available             | Pass 120mmHg                                                                                                                                                                                                                                                                                                                                        | Pass 120mmHg                                | Pass 160mmHg                                |
| Filter<br>Media         | Polypropylene                                               | Polypropylene<br>Meltblown                                                     | Polypropylene<br>Meltblown                                                     | Polypropylene<br>Meltblown                                  | Polypropylene<br>Meltblown                                  | Polypropylene<br>Meltblown                                  | Polypropylene<br>Meltblown                                  | Flammability<br>(16 CFR 1610)    | Not available             | Class 1                                                                                                                                                                                                                                                                                                                                             | Class 1                                     | Class 1                                     |
| Inner layer             | Polypropylene                                               | Polypropylene<br>Spunbond                                                      | Polypropylene<br>Spunbond                                                      | Polypropylene<br>Spunbond                                   | Polypropylene<br>Spunbond                                   | Polypropylene<br>Spunbond                                   | Polypropylene<br>Spunbond                                   | In vitro cytotoxicity            | Not available             | Non-cytotoxic under<br>conditions of study                                                                                                                                                                                                                                                                                                          | Non-cytotoxic under<br>conditions of study  | Non-cytotoxic under<br>conditions of study  |
| Nose Piece              | Not<br>Applicable                                           | combination<br>of zinc wires<br>and embedded<br>polyester                      | combination<br>of zinc wires<br>and embedded<br>polyester                      | Not<br>Applicable                                           | Not<br>Applicable                                           | aluminum                                                    | aluminum                                                    | Sensitization                    | Not available             | Non-sensitizer under<br>conditions of study                                                                                                                                                                                                                                                                                                         | Non-sensitizer under<br>conditions of study | Non-sensitizer under<br>conditions of study |
| Ear<br>Attachment       | synthetic<br>elastic                                        | synthetic<br>elastic                                                           | synthetic<br>elastic                                                           | synthetic<br>elastic                                        | synthetic<br>elastic                                        | synthetic<br>elastic                                        | synthetic<br>elastic                                        | Irritation                       | Not available             | Non-irritant under<br>conditions of study                                                                                                                                                                                                                                                                                                           | Non-irritant under<br>conditions of study   | Non-irritant under<br>conditions of study   |
| Respirator<br>style     | cup shaped                                                  | duck<br>shaped                                                                 | bill<br>shaped                                                                 | cup shaped                                                  | cup shaped                                                  | cup shaped                                                  | cup shaped                                                  | Size                             |                           | 90±3*200±3*95±3mm                                                                                                                                                                                                                                                                                                                                   | 150±3*130±3*60±3mm                          | 140±3*125±3*50±3mm                          |
| Design<br>features      | 3 layers<br>molded-cone<br>with filter web<br>in the middle | 5 layers of<br>non-woven<br>fiber<br>containing<br>filter web in<br>the middle | 5 layers of<br>non-woven<br>fiber<br>containing<br>filter web in<br>the middle | 4 layers<br>molded-cone<br>with filter web<br>in the middle | 4 layers<br>molded-cone<br>with filter web<br>in the middle | 4 layers<br>molded-cone<br>with filter web<br>in the middle | 4 layers<br>molded-cone<br>with filter web<br>in the middle | Sterility                        | Non-sterile<br>Single use | Non-sterile<br>Single use                                                                                                                                                                                                                                                                                                                           | Non-sterile<br>Single use                   | Non-sterile<br>Single use                   |
| Adjustable<br>buckle    | Not<br>Applicable                                           |                                                                                | Acrylonitrile<br>Butadiene<br>Styrene                                          |                                                             | Acrylonitrile<br>Butadiene<br>Styrene                       |                                                             | Acrylonitrile<br>Butadiene<br>Styrene                       | Indication for Use               | Please see<br>below *1    | Pasture F550S, Pasture F550CS, Pasture A520S, Pasture A520CS, Pasture E520S, and Pasture E520CS is a NIOSH certified N95 respirator, and is indicated to be used by the healthcare personnel during procedures to protect both the patient and the healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. |                                             |                                             |

{6}------------------------------------------------

* 1 The Willson® ONE-Fit™ HC-NB095 Healthcare Particulate Respirator and Surgical Mask is a NIOSH-approved N95 single use respirator intended for use by healthcare personnel during medical/ surgical procedures to protect both the wearer against the spatter of blood and other potentially infectious materials and reducing the transfer of microorganisms and other airborne particulate matter.

{7}------------------------------------------------

## Comparison to Predicated Device:

Pasture F550S, Pasture F550CS, Pasture A520S, Pasture A520CS, Pasture E520S and Pasture E520CS respirators are substantially equivalent to WILLSON ONE-FIT HEALTHCARE PARTICLE RESPIRATOR AND SURGICAL MASK, HC-NB095.

## Performance Testing

| Test Performed                    | Result                                                                                                                                                            |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Fluid Resistance                  | The Pasture F550S, Pasture F550CS, Pasture A520S, Pasture<br>A520CS, Pasture E520S, and Pasture E520CS met the<br>requirements of ASTM F1862.                     |
| Flammability                      | The Pasture F550S, Pasture F550CS, Pasture A520S, Pasture<br>A520CS, Pasture E520S, and Pasture E520CS met the<br>requirements of 16 CFR 1610.                    |
| Particulate Filtration Efficiency | NIOSH Certification: TC-84A-7504 (F550S and F550CS)<br>NIOSH Certification: TC-84A-7454 (A520S and A520CS)<br>NIOSH Certification: TC-84A-7453 (E520S and E520CS) |
| Bacterial Filtration Efficiency   | NIOSH Certification: TC-84A-7504 (F550S and F550CS)<br>NIOSH Certification: TC-84A-7454 (A520S and A520CS)<br>NIOSH Certification: TC-84A-7453 (E520S and E520CS) |
| Differential Pressure             | NIOSH Certification: TC-84A-7504 (F550S and F550CS)<br>NIOSH Certification: TC-84A-7454 (A520S and A520CS)<br>NIOSH Certification: TC-84A-7453 (E520S and E520CS) |

{8}------------------------------------------------

| Cytotoxicity  | The Pasture F550S, Pasture F550CS, Pasture A520S, Pasture A520CS, Pasture E520S, and Pasture E520CS were non-cytotoxic under the conditions of the study.   |
|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Irritation    | The Pasture F550S, Pasture F550CS, Pasture A520S, Pasture A520CS, Pasture E520S, and Pasture E520CS were non-irritating under the conditions of the study.  |
| Sensitization | The Pasture F550S, Pasture F550CS, Pasture A520S, Pasture A520CS, Pasture E520S, and Pasture E520CS were non-sensitizing under the conditions of the study. |

## Conclusions:

The test data submitted in this submission demonstrate that the subject device is as safe and as effective as the predicate and technological characteristics do not raise any new questions of safety and effectiveness. Pasture F550S, Pasture F550CS, Pasture A520S, Pasture A520CS, Pasture E520S and Pasture E520CS are substantially equivalent to the predicate device cleared in K070139.

---

**Source:** [https://fda-staging.innolitics.com/device/K141876](https://fda-staging.innolitics.com/device/K141876)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).