Pasture F550S, Pasture F550CS, Pasture A520S, Pasture A520CS, Pasture E520S, and Pasture E520CS is a NIOSH certified N95 respirator, and is indicated to be used by the healthcare personnel during procedures to protect both the patient and the healthcare personnel from the transfer of microorganisms, body fluids, and particulate material.
Device Story
Disposable N95 respirators; duck-bill (F550 series) or cup-shaped (A520/E520 series) designs. Function as physical barriers to airborne particles, microorganisms, and body fluids. Used in clinical settings by healthcare personnel. Input is ambient air; output is filtered air for the wearer while containing wearer-generated particles. Features include multi-layer polypropylene construction, nose cushions, and elastic head straps; some models include adjustable buckles. Benefit is reduced risk of cross-contamination between patient and clinician during procedures.
Clinical Evidence
Bench testing only. Includes fluid resistance (ASTM F1862), flammability (16 CFR 1610), and biocompatibility (cytotoxicity, irritation, sensitization). Filtration efficiency and differential pressure verified via NIOSH certification (TC-84A-7504, TC-84A-7454, TC-84A-7453).
Indicated for use by healthcare personnel during medical/surgical procedures to protect both patient and wearer from transfer of microorganisms, body fluids, and particulate material.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
WILLSON ONE-FIT HEALTHCARE PARTICLE RESPIRATOR AND SURGICAL MASK, HC-NB095 (K070139)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, overlaid on three human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 28, 2015
Pasture Pharma Pte, Ltd C/O Ms. Sarah Hassan US MedPharm 38129 Spring Canyon Drive Murrieta, CA 92563
Re: K141876
> Trade/Device Name: Pasture F550S Surgical N95 Respirator, Pasture F550CS Surgical N95 Respirator, Pasture A520S Surgical N95 Respirator, Pasture A520CS Surgical N95 Respirator, Pasture E520S Surgical N95 Respirator, Pasture E520CS Surgical N95 Respirator Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: MSH Dated: October 15, 2015 Received: October 16, 2015
Dear Ms. Hassan,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K141876
#### Device Name
Pasture F550S Surgical N95 Respirator, Pasture F50CS Surgical N95 Respirator, Pasture A520S Surgical N95 Respirator, Pasture A520CS Surgical N95 Respirator, Pasture E520S Surgical N95 Respirator, Pasture E520CS Surgical N95 Respirator
#### Indications for Use (Describe)
Pasture F550S, Pasture F550CS, Pasture A520CS, Pasture E520S, and Pasture E520CS is a NIOSH certified N95 respirator, and is indicated to be used by the healthcare personnel during procedures to protect both the patient and the healthcare personnel from the transfer of microorganisms, body fluids, and particulate material.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
This 510(K) Summary is being submitted in accordance with requirements of 21 CFR 807.92(c)
The assigned 510(k) number is: K141876
#### Submitter Name and Address:
Pasture Pharma Pte, Ltd 8 Boon Lay Way #04-01 Trade Hub 21, Singapore 609964 Lloyd Soong, President & CEO Contact: +65-6515-6561 lloyd@pasturegroup.com
# US agent and correspondent:
Mrs. Sarah Hassan US Med Pharm Supplies, Inc 38129 Spring Canyon Drive Murrieta, CA 92563 Phone: 951-239-1933 Fax: 951-239-1933 E-mail: sarah@usmedpharm.com
Date Prepared: October 27, 2015
# Proprietary Name:
Pasture F550S Surgical N95 Respirator, Pasture F550CS Surgical N95 Respirator, Pasture A520S Surgical N95 Respirator, Pasture A520CS Surgical N95 Respirator, Pasture E520S Surgical N95 Respirator, Pasture E520CS Surgical N95 Respirator
Common name: Surgical N95 NIOSH certified Respirator
{4}------------------------------------------------
### Device Classification and Product Code
Classification Name: Surgical Apparel Regulation: 21 CFR §878.4040 Class: Class II Classification panel: General and Plastic Surgery Product Code: MSH
Predicate Device: K070139 WILLSON ONE-FIT HEALTHCARE PARTICLE RESPIRATOR AND SURGICAL MASK, HC-NB095 SURVIVAIR, INC
#### Intended Use:
Pasture F550S, Pasture F550CS, Pasture A520S, Pasture A520CS, Pasture E520S, and Pasture E520CS is a NIOSH certified N95 respirator, and is indicated to be used by the healthcare personnel during procedures to protect both the patient and the healthcare personnel from the transfer of microorganisms, body fluids, and particulate material.
#### Device Description:
Pasture F550S and Pasture F550CS respirators are disposable duck bill shaped N95 respirators with NIOSH certification number TC-84A-7504. They contain 5 layers composed of polypropylene, with a nose cushion, a synthetic elastic loop or strap, and a nosepiece which is the combination of zinc wires and embedded polyester inside of layers. F550CS is the same respirator as the F550S, but F550CS contains an adjustable buckle on the headband.
Pasture A520S and Pasture A520CS respirators are cup shaped N95 respirators with NIOSH certification number TC-84A-7454. They contain 4 layers composed of polypropylene with a synthetic elastic loop or a strap. A520CS is the same respirator as the A520S, but A520CS contains an adjustable buckle on the headband.
Pasture E520S and Pasture E520CS respirators are cup shaped N95 respirators with NIOSH certification number TC-84A-7453. They contain 4 layers composed of polypropylene with a
{5}------------------------------------------------
nose cushion, synthetic elastic loop or strap, and an aluminum nosepiece. E520CS is the same respirator as E520S, but E520CS contains an adjustable buckle on the headband.
| | Predicate | Pasture | Pasture | Pasture | Pasture | Pasture | Pasture | Test | K070139 | F550S | A520S | E520S |
|-------------------------|-------------------------------------------------------------|--------------------------------------------------------------------------------|--------------------------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------|----------------------------------|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|---------------------------------------------|
| Description | K070139 | F550S | F550CS | A520S | A520CS | E520S | E520CS | | | F550CS | A520CS | E520CS |
| NIOSH | N95 | N95 | N95 | N95 | N95 | N95 | N95 | NIOSH | N95 | N95 | N95 | N95 |
| certification<br>number | TC-84A-4516 | TC-84A-7504 | TC-84A-7504 | TC-84A-7454 | TC-84A-7454 | TC-84A-7453 | TC-84A-7453 | Certification | TC-84A-4516 | TC-84A-7504 | TC-84A-7454 | TC-84A-7453 |
| Outer layer | Polyester | Polypropylene<br>Spunbond | Polypropylene<br>Spunbond | Polypropylene<br>Spunbond | Polypropylene<br>Spunbond | Polypropylene<br>Spunbond | Polypropylene<br>Spunbond | Fluid Resistance<br>(ASTM F1862) | Not available | Pass 120mmHg | Pass 120mmHg | Pass 160mmHg |
| Filter<br>Media | Polypropylene | Polypropylene<br>Meltblown | Polypropylene<br>Meltblown | Polypropylene<br>Meltblown | Polypropylene<br>Meltblown | Polypropylene<br>Meltblown | Polypropylene<br>Meltblown | Flammability<br>(16 CFR 1610) | Not available | Class 1 | Class 1 | Class 1 |
| Inner layer | Polypropylene | Polypropylene<br>Spunbond | Polypropylene<br>Spunbond | Polypropylene<br>Spunbond | Polypropylene<br>Spunbond | Polypropylene<br>Spunbond | Polypropylene<br>Spunbond | In vitro cytotoxicity | Not available | Non-cytotoxic under<br>conditions of study | Non-cytotoxic under<br>conditions of study | Non-cytotoxic under<br>conditions of study |
| Nose Piece | Not<br>Applicable | combination<br>of zinc wires<br>and embedded<br>polyester | combination<br>of zinc wires<br>and embedded<br>polyester | Not<br>Applicable | Not<br>Applicable | aluminum | aluminum | Sensitization | Not available | Non-sensitizer under<br>conditions of study | Non-sensitizer under<br>conditions of study | Non-sensitizer under<br>conditions of study |
| Ear<br>Attachment | synthetic<br>elastic | synthetic<br>elastic | synthetic<br>elastic | synthetic<br>elastic | synthetic<br>elastic | synthetic<br>elastic | synthetic<br>elastic | Irritation | Not available | Non-irritant under<br>conditions of study | Non-irritant under<br>conditions of study | Non-irritant under<br>conditions of study |
| Respirator<br>style | cup shaped | duck<br>shaped | bill<br>shaped | cup shaped | cup shaped | cup shaped | cup shaped | Size | | 90±3*200±3*95±3mm | 150±3*130±3*60±3mm | 140±3*125±3*50±3mm |
| Design<br>features | 3 layers<br>molded-cone<br>with filter web<br>in the middle | 5 layers of<br>non-woven<br>fiber<br>containing<br>filter web in<br>the middle | 5 layers of<br>non-woven<br>fiber<br>containing<br>filter web in<br>the middle | 4 layers<br>molded-cone<br>with filter web<br>in the middle | 4 layers<br>molded-cone<br>with filter web<br>in the middle | 4 layers<br>molded-cone<br>with filter web<br>in the middle | 4 layers<br>molded-cone<br>with filter web<br>in the middle | Sterility | Non-sterile<br>Single use | Non-sterile<br>Single use | Non-sterile<br>Single use | Non-sterile<br>Single use |
| Adjustable<br>buckle | Not<br>Applicable | | Acrylonitrile<br>Butadiene<br>Styrene | | Acrylonitrile<br>Butadiene<br>Styrene | | Acrylonitrile<br>Butadiene<br>Styrene | Indication for Use | Please see<br>below *1 | Pasture F550S, Pasture F550CS, Pasture A520S, Pasture A520CS, Pasture E520S, and Pasture E520CS is a NIOSH certified N95 respirator, and is indicated to be used by the healthcare personnel during procedures to protect both the patient and the healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. | | |
{6}------------------------------------------------
* 1 The Willson® ONE-Fit™ HC-NB095 Healthcare Particulate Respirator and Surgical Mask is a NIOSH-approved N95 single use respirator intended for use by healthcare personnel during medical/ surgical procedures to protect both the wearer against the spatter of blood and other potentially infectious materials and reducing the transfer of microorganisms and other airborne particulate matter.
{7}------------------------------------------------
## Comparison to Predicated Device:
Pasture F550S, Pasture F550CS, Pasture A520S, Pasture A520CS, Pasture E520S and Pasture E520CS respirators are substantially equivalent to WILLSON ONE-FIT HEALTHCARE PARTICLE RESPIRATOR AND SURGICAL MASK, HC-NB095.
## Performance Testing
| Test Performed | Result |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Fluid Resistance | The Pasture F550S, Pasture F550CS, Pasture A520S, Pasture<br>A520CS, Pasture E520S, and Pasture E520CS met the<br>requirements of ASTM F1862. |
| Flammability | The Pasture F550S, Pasture F550CS, Pasture A520S, Pasture<br>A520CS, Pasture E520S, and Pasture E520CS met the<br>requirements of 16 CFR 1610. |
| Particulate Filtration Efficiency | NIOSH Certification: TC-84A-7504 (F550S and F550CS)<br>NIOSH Certification: TC-84A-7454 (A520S and A520CS)<br>NIOSH Certification: TC-84A-7453 (E520S and E520CS) |
| Bacterial Filtration Efficiency | NIOSH Certification: TC-84A-7504 (F550S and F550CS)<br>NIOSH Certification: TC-84A-7454 (A520S and A520CS)<br>NIOSH Certification: TC-84A-7453 (E520S and E520CS) |
| Differential Pressure | NIOSH Certification: TC-84A-7504 (F550S and F550CS)<br>NIOSH Certification: TC-84A-7454 (A520S and A520CS)<br>NIOSH Certification: TC-84A-7453 (E520S and E520CS) |
{8}------------------------------------------------
| Cytotoxicity | The Pasture F550S, Pasture F550CS, Pasture A520S, Pasture A520CS, Pasture E520S, and Pasture E520CS were non-cytotoxic under the conditions of the study. |
|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Irritation | The Pasture F550S, Pasture F550CS, Pasture A520S, Pasture A520CS, Pasture E520S, and Pasture E520CS were non-irritating under the conditions of the study. |
| Sensitization | The Pasture F550S, Pasture F550CS, Pasture A520S, Pasture A520CS, Pasture E520S, and Pasture E520CS were non-sensitizing under the conditions of the study. |
## Conclusions:
The test data submitted in this submission demonstrate that the subject device is as safe and as effective as the predicate and technological characteristics do not raise any new questions of safety and effectiveness. Pasture F550S, Pasture F550CS, Pasture A520S, Pasture A520CS, Pasture E520S and Pasture E520CS are substantially equivalent to the predicate device cleared in K070139.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.