K141854 · Osung Mnd Co., Ltd. · KCT · Oct 29, 2015 · General Hospital
Device Facts
Record ID
K141854
Device Name
INSTRUMENT CASSETTE (EFCCL1-F)
Applicant
Osung Mnd Co., Ltd.
Product Code
KCT · General Hospital
Decision Date
Oct 29, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
OSUNG INSTRUMENT CASSETTE is intended to hold instruments and accessories during storage and steam sterilization. This case cassette is intended to be used with a sterilized wrap in order to maintain sterility. This case cassette is intended to be used with an FDA cleared sterilization wrap in order to maintain sterility.
Device Story
Instrument cassette designed to hold reusable dental/surgical instruments (e.g., scissors, forceps, elevators) during steam sterilization and storage. Comprised of base, lid, silicone mat/rail, holders, and spring lock system. Used in clinical settings; operated by healthcare staff. Device is placed inside FDA-cleared sterilization wrap; subjected to gravity steam sterilization (132°C for 15 minutes). Protects instruments during processing; maintains sterility post-sterilization. Benefits include organized instrument handling and effective sterilization maintenance.
Clinical Evidence
Bench testing only. No clinical data. Validation included half-cycle steam sterilization (ISO 17665-1/2), dry time validation (ANSI/AAMI ST79), cleaning validation (AAMI TIR 30), and limits of reuse testing (FDA 2011 guidance, AAMI TIR12/30). Results demonstrated no growth at half-cycle and successful dry time performance.
Technological Characteristics
Perforated lid and bottom cassette; materials include silicone mat/rail. Dimensions: 205 x 305 x 31 mm. Sterilization method: Gravity steam (132°C, 15 min exposure, 30 min dry time). Reusable. Standards: ISO 17665-1/2, ANSI/AAMI ST79, AAMI TIR12, AAMI TIR30.
Indications for Use
Indicated for holding dental/surgical instruments and accessories during storage and steam sterilization cycles. Intended for use with FDA-cleared sterilization wraps to maintain sterility. No specific patient population or contraindications listed.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, stacked on top of each other. The logo is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 29, 2015
OSUNG MND CO., LTD c/o Mr. Kiran Sheikh Brite Sources USA 180 S. Western Avenue, #205 Carpentersville, IL 60110
Re: K141854
Trade/Device Name: Instrument Cassette (EFCCL1-F) Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: September 11, 2015 Received: September 15, 2015
Dear Mr. Sheikh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Sheikh
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K141854
Device Name
Instrument Cassette (EFCCL1-F)
Indications for Use (Describe)
OSUNG INSTRUMENT CASSETTE (EFCCL1-F) is intended to hold instruments and accessories during storage and steam sterilization. This case cassette is intended to be used with a sterilized wrap in order to maintain sterility.
This case cassette is intended to be used with an FDA cleared sterilization wrap in order to maintain sterility.
Validated sterilization parameters:
| Cycle Type | Temperature | Exposure Time | Dry Time |
|---------------|--------------|---------------|------------|
| Gravity Steam | 132℃ (270°F) | 15 minutes | 30 minutes |
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(K) - safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: 29 October, 2015
#### 1. Applicant/submitter:
Manufacturer Name: OSUNG MND Co.,Ltd. 57, Hwangguem-ro, 109beon-gil, Yangchon-eup, Gimpo-si, Gyeonggi-do, Republic of Korea Tel: +82-31-987-5395 Fax: +82-31-987-5397 E-mail:qa@osung.co.kr Establishment registration number: 9616119 Contact name: Taerim Lee
#### 2. Submission Correspondent
Correspondent Name: Kiran Sheikh
Official correspondent: Brite Sources USA 180 S. Western Avenue, #205 Carpentersville, IL 60118 Tel: 312-895-1773 Fax: 847-589-1298 E-mail:support(@britesourcs.com
#### 3. Device Name and Classification
| Trade/Proprietary Name: | Instrument Cassette (EFCCL1-F) |
|-----------------------------|--------------------------------------------------------|
| Common Name: | Instrument Cassette |
| Classification Information: | General Hospital<br>21 CFR 880.6850<br>Class II<br>KCT |
#### 4. Predicated Device
Miltex Thompson Cassettes(K101653) Miltex Inc.
#### 5. Device Description
The OSUNG INSTRUMENT CASSETTE is designed to handle reusable dental instruments for sterilization and storage. This device is intended to hold instruments and accessories during storage and steam sterilization, and to be used with a sterilized wrap in order to maintain sterility.
The OSUNG INSTRUMENT CASSETTE is comprised of a base, lid, silicone mat/rail, holders, and spring lock system.
The internal insert cassette has the ability to hold individualized pieces and accessories which include scissors, needle, elevators, forceps, pincers, retractors, mallets etc.
{4}------------------------------------------------
Description and Dimensions
| Model No | Description | Max no of instruments | Dimensions (mm) [ L X W X H ] | Max Weight (with instruments) | Total Surface area (mm²) | Perforation (%) | Silicone Surface area (mm²) |
|----------|-----------------------------------------------------------------------------------|-----------------------|-------------------------------|-------------------------------|--------------------------|-----------------|-----------------------------|
| EFCCL1-F | Perforated lid & bottom,<br>Mid length instruments &<br>tools(forceps, elevators) | 10 | 205 X 305 X 31 | 1,302.0 | 150,118 | 29.0 | 71,622 |
{5}------------------------------------------------
# 6. Intended Uses
OSUNG INSTRUMENT CASSETTE is intended to hold instruments and accessories during storage and steam sterilization. This case cassette is intended to be used with a sterilized wrap in order to maintain sterility.
This case cassette is intended to be used with an FDA cleared sterilization wrap in order to maintain sterility.
Validated sterilization parameters:
| Cycle Type | Temperature | Exposure Time | Dry Time |
|---------------|---------------|---------------|------------|
| Gravity Steam | 132°C (270°F) | 15 minutes | 30 minutes |
# 7. Substantial Equivalence
The Table below shows that the substantial equivalence through side-by-side comparisons between the OSUNG Instrument Cassette and predicate device of Miltex Thompson Cassettes (K101653). The comparison analysis consists of the products' intended use, technological characteristics and performance testing to support the substantial equivalency to its corresponding predicate devices.
| Subject device | INSTRUMENT CASSETTE(EFCCL1-F) | Miltex Thompson Cassettes |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | new | K101653 |
| Manufacturer | OSUNG MND Co.,Ltd. | Miltex Inc. |
| Indication for use | OSUNG INSTRUMENT CASSETTE is intended<br>to hold instruments and accessories during<br>storage and steam sterilization. This case<br>cassette is intended to be used with a<br>sterilized wrap in order to maintain sterility.<br><br>This case cassette is intended to be used with<br>an FDA cleared sterilization wrap in order to<br>maintain sterility. | Miltex Thompson Cassettes are general<br>dented/surgical instrument cassettes indicated<br>to hold instruments and accessories in place<br>during storage and the sterilization cycle.<br>Suitable for Gravity Steam and Pre-Vacuum<br>Steam, the cassettes are intended to be used in<br>conjunction with sterilization wrap in order to<br>maintain sterility. |
| Shape | Image: INSTRUMENT CASSETTE | Image: Miltex Thompson Cassettes |
{6}------------------------------------------------
| Dimensions(LXWXH) | Model No | Dimensions(mm) | Model No | Dimensions(mm, inch) | Maximum number of instruments | Model No | Max no. of Instrument | Model No | Max no. of Instrument |
|-------------------|----------|---------------------------------------------|-----------------------|------------------------------|-------------------------------|----------|-----------------------|------------|-----------------------|
| | EFCCL1-F | 205 X 305 X 31 | 3-083005 | 204X81X30 | | EFCCL1-F | 10 | 3-083005 | 5 |
| | | | 3-083105 | 204X81X30 | | | | 3-083105 | 5 |
| | | | 3-083210 | 204X81X30 | | | | 3-083210 | 5 |
| | | | 3-084007 | 204X124X28.5 | | | | 3-084007 | 7 |
| | | | 3-084107 | 204X124X28.5 | | | | 3-084107 | 7 |
| | | | 3-084214 | 204X124X28.5 | | | | 3-084214 | 7 |
| | | | 3-085009 | 204X154X30 | | | | 3-085009 | 9 |
| | | | 3-085109 | 204X154X30 | | | | 3-085109 | 9 |
| | | | 3-085218 | 204X154X30 | | | | 3-085218 | 9 |
| | | | 3-089110 | 204X229X28.5 | | | | 3-089110 | 10 |
| | | | 3-118114 | 204X276.5X30 | | | | 3-118114 | 14 |
| | | | 3-118116 | 204X276.5X30 | | | | 3-118116 | 16 |
| | | | 3-118122 | 204X276.5X30 | | | | 3-118122 | 22 |
| | | | 3-139126 | 232X330X30 | | | | 3-139126 | 26 |
| | | | 3-072007 | 178X67X17 | | | | 3-072007 | 7 |
| | | | 3-082008 | 203X127X17 | | | | 3-082008 | 7 |
| | | | STDSL27 | 178X67X25 | | | | STDSL27 | 7 |
| | | | 3-072014 | 178X115X17 | | | | 3-072014 | 14 |
| | | | STD222 | 204X276.5X30 | | | | STD222 | 22 |
| | | | STD209 | - | | | | STD209 | |
| | | | STDSTAT | 178X270X38 | | | | STDSTAT | 14 |
| | | | STDSTAT8 | 171X206X29 | | | | STDSTAT8 | 8 |
| | | | STDSTAT814 | 171X206X29 | | | | STDSTAT814 | 14 |
| | | | STDORTHO | 127X178X51 | | | | STDORTHO | - |
| | | | STDBOS | 178X76X35 | | | | STDBOS | - |
| | | | STDBOM | 254X140X102 | | | | STDBOM | - |
| | | | 3-080205 | 204X71X37 | | | | 3-080205 | - |
| | | | 4-6835 | 6-1/2 x 6-1/4 x 1-3/16(inch) | | | | 4-6835 | |
| | | | 4-083000 | 8 x 3-13/16 x 1-3/16(inch) | | | | 4-083000 | |
| | | | 4-084000 | 8 x 4-7/8 x 1-3/16(inch) | | | | 4-084000 | |
| | | | 4-085000 | 8 x6 x 1-3/16(inch) | | | | 4-085000 | |
| | | | 4-089100 | 8 x9 x 1-3/16(inch) | | | | 4-089100 | |
| | | | 4-008122 | 8 x11 x 1-3/16(inch) | | | | 4-008122 | |
| | | | 4-008115 | 8 x11x 2(inch) | | | | 4-008115 | |
| | | | 4-009126 | 9-1/8 x13 x 1-3/16(inch) | | | | 4-009126 | |
| | | | 4-6815 | 15 x8-1/4x 1-3/16(inch) | | | | 4-6815 | |
| | | | STDBH | - | | | | STDBH | - |
| | | | STDBH6 | - | | | | STDBH6 | - |
| | | | STDBHS | - | | | | STDBHS | - |
| | | | STDBB2 | - | | | | STDBB2 | - |
| | | | STDPB | 42X33X26 | | | | STDPB | - |
| | | | STDES48 | - | | | | STDES48 | - |
| | | | STDES60 | - | | | | STDES60 | - |
| | | | STDES72 | - | | | | STDES72 | - |
| Maximum weight | Model No | Weight of each cassette with instruments(g) | Not stated in summary | | | | | | |
| | EFCCL1-F | 1,302.0 | | | | | | | |
{7}------------------------------------------------
{8}------------------------------------------------
| Sterilization<br>method | Gravity steam:<br>ExposureTime:15 minutes<br>Temperature: 132 °C<br>Drying: 30 minutes. | Gravity steam:<br>ExposureTime:30 minutes<br>Temperature: 121°C<br>Drying: 20 minutes.<br>Pre-Vacuum steam:<br>ExposureTime:4 minutes<br>Temperature: 132 °C<br>Drying: 20 minutes. |
|-------------------------|-----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Reusable | Yes | Yes |
# 8. Performance data
Validation test for steam sterilization of ultrasonic cleaning and limit of reuse test for the OSUNG INSTRUMENT CASSETTE(EFCCL1-F) has been carried out as the results were shown by test reports.
All of tests demonstrated compliance to both safety and effectiveness requirements that apply to the device, in terms of sterilization performance, maintenance of sterility and reusability.
Hence, the OSUNG INSTRUMENT CASSETTE(EFCCL1-F) demonstrated appropriate performance that is substantially equivalent to the predicate device and that the design output meets the design input requirements.
| Performance Test | Standard(s) used | Results |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| Half cycle sterilization<br>validation at 132°C | ISO 17665-1(2006)<br>ISO 17665-2(2009) | No growth at half cycle. |
| Dry Time Validation | ANSI/AAMI ST79(2010) &<br>A1(2010) & A2(2011) &<br>A3(2012) & A4(2013) | The minimum weight difference was<br>-0.05% (Pass) |
| Cleaning Validation | AAMI TIR 30 (2011) | |
| Limits of Reuse | FDA Guideline:<br>Processing/Reprocessing<br>Medical Devices in Health<br>Care Settings: Validation<br>Methods and Labeling<br>(2011)<br>AAMI TIR12(2010)<br>AAMI TIR30(2011) | |
# 9. Conclusions
Based on the intended use, indications for use, technological characteristics, and performance data, the Instrument Cassette is substantially equivalent to the Miltex Thompson Cassettes (K101653).
Following consideration of all information presented to support substantial equivalence of the Instrument Cassette to the predicate device, similarities in intended use, design, principles of operation, and performance specifications were identified. The performance data following validation tests carried out for the Instrument Cassette showed that it is substantially equivalent to the known performance of the predicate device, Miltex Thompson Cassettes
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.