K141837 · Carestream Health, Inc. · KPR · Mar 11, 2015 · Radiology
Device Facts
Record ID
K141837
Device Name
DRX-EVOLUTION
Applicant
Carestream Health, Inc.
Product Code
KPR · Radiology
Decision Date
Mar 11, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The device is a permanently installed diagnostic x-ray system for general radiographic x-ray imaging including tomography. The tomography feature is not to be used for imaging pediatric patients.
Device Story
Stationary digital radiography (DR) system; includes high-voltage generator, overhead tube crane, radiographic table, Bucky wall stand, and operator PC. Transforms X-ray radiation into digital images using flat-panel detectors (DRX-1, DRX-1C, DRX 2530C). Enables linear tomography via synchronized movement of tube head and table-mounted detector; utilizes software for image processing, including Bone Suppression and Pneumothorax Visualization. Used in clinical settings by radiology staff. Output displayed on operator PC for physician review to aid diagnosis. Benefits include high-quality diagnostic imaging with advanced processing features and expanded clinical utility through tomography and new detector integration.
Clinical Evidence
Bench testing evaluated linear tomography using tool and anthropomorphic phantoms (chest, hand, knee, pelvis), confirmed by board-certified radiologist. Clinical studies supported detector performance (DRX 2530C, DRX-1C) and software features (LLI, Bone Suppression). LLI software showed statistically equivalent/better diagnostic capability than predicate. Bone Suppression study showed companion images improved diagnostic rating compared to standard-of-care alone.
Technological Characteristics
Stationary digital X-ray system. Detectors: GOS scintillator (DRX-1) or CsI scintillator (DRX-1C, DRX 2530C). Features synchronized motion control for linear tomography. Includes optional high-voltage generator. Software-based image processing for bone suppression and pneumothorax visualization. Connectivity via operator PC interface. System is a modification of existing DRX-Evolution platform.
Indications for Use
Indicated for general radiographic x-ray imaging, including linear tomography, in adult patients. Tomography is contraindicated for pediatric patients.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, overlaid on three human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 11, 2015
Carestream Health, Inc. % Ms. Carolyn Wagner Regulatory Affairs Manager 150 Verona Street ROCHESTER NY 14608
Re: K141837
Trade/Device Name: DRX-Evolution Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: February 21, 2015 Received: February 24, 2015
Dear Ms. Wagner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known): K141837
Device Name: DRX-Evolution
Indications for Use:
The device is a permanently installed diagnostic x-ray system for general radiographic x-ray imaging including tomography. The tomography feature is not to be used for imaging pediatric patients.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
> K141837 510(k)
Page 1 of 1
{3}------------------------------------------------
# Carestream
# "510(k) Summary"
| 510(k) Owner Name: | Carestream Health, Inc. |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner Address: | 150 Verona Street<br>Rochester, New York 14608 |
| 510(k) Owner Phone: | 585-627-6588 |
| 510(k) Owner Fax: | 585-323-7643 |
| Contact Person & Info: | Carolyn Wagner<br>Sr. Regulatory Affairs Manager, X-Ray Systems<br>carolyn.wagner@carestreamhealth.com<br>585-627-6588 |
| Date Summary Prepared: | March 10, 2015 |
| Device Trade Name: | Carestream DRX-Evolution |
| Device Common Name: | diagnostic x-ray system |
| Classification Name: | Stationary x-ray system |
| Device Class: | Class II |
| Device (Product) Code: | KPR |
| Regulation Number: | 21 CFR 892.1680 |
| Predicate Device: | Carestream DRX-Evolution<br>Manufactured by Carestream Health, Inc.<br>510(k) No. K091889 (July 20, 2009)<br>Regulation No. 21 CFR 892.1680<br>Product Code: KPR |
| Reference Device: | Tomo-Link<br>Manufactured by GE Medical Systems<br>510(k) No. K944967 (12/09/1994)<br>Regulation No. 21 CFR 892.1740<br>Product Code: IZF |
## Device Description:
The DRX-Evolution is a diagnostic x-ray system utilizing digital radiography (DR) technology. The DRX-Evolution is designed for horizontal and upright projection exams. The system consists of a high voltage x-ray generator, overhead tube crane with
{4}------------------------------------------------
x-ray tube assembly, radiographic table with detector tray, Bucky image receptor on an upright wall stand, and x-ray controls containing a power distribution unit and operator PC (user interface).
The modifications to the device described in this submission include firmware and mechanical changes to facilitate use of the device for linear tomography exams and addition of a new generator option. Software updates to the DRX-Evolution include features such as Bone Suppression, Pneumothorax Visualization, and the ability to integrate the DRX-1C and DRX 2530C Detectors.
The new alternate generator has been implemented as an option for use with the DRX-Evolution system. Implementation of the new optional generator is transparent to the user and does not require any update to the user interface. The new generator is functionally equivalent to the generator cleared under the original 510(k) for this device.
The modifications to the DRX-Evolution device raise no new issues of safety or effectiveness.
#### Indications for Use / Intended Use:
The Indications for Use for the modified device, as described in its labeling, are: "The device is a permanently installed diagnostic x-ray system for general radiographic x-ray imaging including tomography. The tomography feature is not to be used for imaging pediatric patients." The Indications for Use for the modified device are similar to the originally cleared DRX-Evolution (predicate device) except for the addition of the tomography functionality that is not to be used for pediatric patients.
The intended use for the modified device, as determined by descriptions and the proposed labeling contained in this submission, is similar to the Indications for Use statement provided above. The linear tomography exam is considered a general radiography exam and is normally performed using a stationary x-ray system such as the DRX-Evolution. This exam requires accurate movement of the x-ray tube head with respect to the image capture device (and patient). This functionality is similar to the existing long length imaging function of the DRX-Evolution that controls accurate movement of the tube head and detector with respect to the patient.
#### Comparison of Technological Characteristics:
The modified DRX-Evolution device has the same technological characteristics as the originally cleared DRX-Evolution predicate device (K091889). In addition, firmware and hardware have been modified to enable accurate synchronized movement of the flat panel imager in the table with respect to movement of the tube head to add linear tomography functionality. Other added features such as the optional generator and new software options do not introduce new technological characteristics and are briefly described below.
{5}------------------------------------------------
Implementation of the alternative (optional) generator is transparent to the user and does not require changes to the existing selections pertaining to the generator available on the system console user interface.
The capability to perform linear tomography exams employs similar technology to the long length imaging feature that ensures proper synchronization of the tube head and detector for long length imaging exams. A longer exposure (approximately 3 seconds) is required for a linear tomography exam but the amount of radiation exposure is still within the window of some non linear tomography exams (i.e. exams for larger patients that require longer exposure to obtain a usable image).
Additional software features cleared by FDA for use on Carestream DR devices are extensions of the original image processing software incorporated in the DRX-Evolution system. These additional software features provide companion views to the original image than can be used to aid diagnosis. Incorporation of these optional features in the DRX-Evolution is consistent with its existing technological characteristics.
The DRX-Evolution can be used with the legally marketed Carestream DRX-1, DRX-1C and DRX 2530C detectors that are commercially available for use with other Carestream products or as standalone devices. The Carestream DRX-1 Detector utilizes a gadolinium oxysulfide (GOS) scintillator while the Carestream DRX-1C and DRX 2530C Detectors utilize a cesium iodide (CsI) scintillator, with all other aspects of the design and function remaining the same except for the size.
Ability to integrate Carestream's DRX-1C and DRX 2530C detectors with the DRX-Evolution required only minor updates to the user interface to select the detector being used so that exposure parameters are calculated appropriately. These detectors are functionally identical to the DRX-1 Detector except that they provide equal or superior image quality with respect to noise and spatial resolution at equivalent doses.
#### Discussion of Performance Testing - Bench:
The performance characteristics and operation / usability of the Carestream DRX-Evolution system were evaluated in non-clinical (bench) testing. These studies demonstrated the intended workflow, related performance, overall function, verification and validation of requirements for intended use, shipping performance, and reliability of the DRX-Evolution system including both software and hardware requirements. Nonclinical test results have demonstrated that the device conforms to its specifications. Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate device.
The linear tomography functionality was evaluated in bench testing using a tool phantom designed to evaluate this functionality, and also using anthropomorphic phantoms. Test results using the tool phantom were as expected, demonstrating accuracy of the tube head movement with respect to the capture device. In addition, linear tomography images were generated using four different anthropomorphic phantoms (chest, hand, knee and
{6}------------------------------------------------
pelvis). The images were evaluated by a board-certified radiologist for diagnostic quality. Results of this evaluation demonstrated that the linear tomography images acquired using the DRX-Evolution system are of acceptable diagnostic quality. The tests described above and the performance data collected are consistent with the testing performed to substantiate the safety and effectiveness of linear tomography in the previously cleared reference device, Tomo-Link (K944967). The results of bench testing of new software / hardware components and phantom image evaluation of the linear tomography functionality support a substantial equivalence determination.
### Discussion of Performance Testing - Clinical
Clinical studies were performed in accordance with FDA guidance document "Guidance for the Submission of 510(k)'s for Solid State Imaging Devices" to demonstrate the diagnostic capability of the DRX 2530C Detector (K130464) and DRX-1C Detector (K120062). Results of these studies demonstrated equivalent or superior image quality to the DRX-1 Detector (predicate device).
Clinical studies were also performed to evaluate the DR Long Length Imaging Software (K130567), and Bone Suppression software (K133442). Results of the DR Long Length Imaging study demonstrated that the investigational software produced LLI images with statistically equivalent or better diagnostic capability to the predicate software. Results of the Bone Suppression clinical study demonstrated that the software generates a companion image that, when presented to the physician along with the standard-of-care image, is rated substantially equivalent or improved as compared to that of the predicate product (standardof-care image without the bone-suppressed companion image).
The results of the clinical testing described above support a substantial equivalence determination.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.