K141826 · Symmetry Medical Inc- Manchester · FRG · Apr 2, 2015 · General Hospital
Device Facts
Record ID
K141826
Device Name
FLASHPAK STERILIZATION CONTAINER SYSTEM
Applicant
Symmetry Medical Inc- Manchester
Product Code
FRG · General Hospital
Decision Date
Apr 2, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
FlashPak® is a reusable rigid container system to be used during immediate use steam sterilization (IUSS or flash) by hospitals and healthcare facilities. It is intended to enable sterilization of the enclosed devices and prevent recontamination during immediate transport to the point of use. The container is compatible with gravity-displacement steam sterilization using a 10 minute cycle at 132° C for porous items with lumens or a 3 minute cycle at 132° C for nonporous items like routine metal instruments. The container is also compatible with pre-vacuum steam sterilization using a 4 minute cycle at 132° C for porous items with lumens or a 3 minute cycle for nonporous items like routine metal instruments. FlashPak® is recommended for sterilization of lumens with the following limits: gravity-displacement (5.5mm inner diameter or larger and up to 184mm in length), pre-vacuum (1mm inner diameter or larger and up to 203mm in length).
Device Story
FlashPak is a reusable rigid container system for immediate use steam sterilization (IUSS). It consists of a Radel lid and base, a removable stainless steel wire basket, and silicone gaskets/valves. The system encloses instruments during sterilization and protects them from recontamination during transport to the point of use. Steam enters and exits via pressure-actuated valves. Used in hospitals/healthcare facilities by clinical staff. The device is physically identical to the predicate; this submission adds validated 3-minute sterilization cycle parameters for nonporous metal instruments to the existing instructions for use. Benefits include standardized, validated sterilization and protection of instruments for immediate clinical use.
Clinical Evidence
Bench testing only. Sterilization efficacy was validated using the overkill method per ISO 17665-1. Testing utilized worst-case loading configurations (maximum weight of routine metal instruments). Biological indicators (Geobacillus stearothermophilus) were placed throughout the container and directly inoculated on devices. Half-cycle testing (three runs per mode) demonstrated no microbial growth, achieving a Sterility Assurance Level (SAL) of 10^-6. Positive and negative controls passed.
Technological Characteristics
Materials: Radel R5000 series (lid/base), 304 stainless steel (basket, latches, bellows), 316 stainless steel (bracket), medical-grade silicone (gaskets, valve seals, vent button). Principle: Rigid container with pressure-actuated valves for steam ingress/egress. Dimensions: Four sizes (9020, 9030, 9040, 9050). Sterilization: Gravity-displacement and pre-vacuum steam. No software or electronics.
Indications for Use
Indicated for use in hospitals and healthcare facilities for immediate use steam sterilization (IUSS) of medical instruments, including porous items with lumens and nonporous routine metal instruments, within a reusable rigid container system.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
FlashPak® Sterilization Container System (K113776)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three faces in profile, stacked on top of each other. The faces are black and are oriented to the right. The logo is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in black.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 2, 2015
Symmetry Surgical, Inc. Mr. Christopher Smith, RAC Senior Director of QA/RA 3034 Owen Drive Antioch, TN 37013
Re: K141826
Trade/Device Name: FlashPak® Sterilization Container System Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: February 18, 2015 Received: February 23, 2015
Dear Mr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Smith
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141826
Device Name FlashPak® Sterilization Container System
# Indications for Use (Describe)
FlashPak® is a reusable rigid container system to be used during immediate use steam sterilization (IUSS or flash) by hospitals and healthcare facilities. It is intended to enable sterilization of the enclosed devices and prevent recontamination during immediate transport to the point of use. The container is compatible with gravity-displacement steam sterilization using a 10 minute cycle at 132° C for porous items with lumens or a 3 minute cycle at 132° C for nonporous items like routine metal instruments. The containle with pre-vacum steam sterilization using a 4 minute cycle at 132° C for porous items with lumens or a 3 minute cycle for nonporous items like routine metal instruments. FlashPak® is recommended for sterilization of lumens with the following limits: gravity-displacement (5.5mm inner diameter or larger and up to 184mm in length), pre-vacuum (1mm inner diameter or larger and up to 203mm in length).
FlashPak® Model/Maximum Instrument Load Weight Recommendation:
9020 w/ 9020-08 Basket - Gravity 3 pounds (1.36 Kg) - Pre-Vacuum 3 pounds (1.36 Kg)
9030 w/ 9030-08 Basket - Gravity 10 pounds (4.55 Kg) - Pre-Vacuum 10 pounds (4.55 Kg)
9040 w/ 9040-08 Basket - Gravity 14 pounds (6.35 Kg) - Pre-Vacuum 14 pounds (6.35 Kg)
9050 w/ 9050-08 Basket - Gravity 16 pounds (7.27 Kg) - Pre-Vacuum 16 pounds (7.27 Kg)
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
区 Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
រៀត
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Symmetry Surgical. The logo consists of a stylized letter "S" made up of red and gray shapes, followed by the text "symmetry surgical" in a gray, sans-serif font. The logo is simple and modern, with a focus on the company name.
# 510K Summary of Safety and Effectiveness (Per 21 CFR 807.92)
# General Company Information
| Company Name: | Symmetry Surgical |
|---------------------------------------------------|---------------------------------------------------|
| Company Address: | 3034 Owen Drive |
| | Antioch, TN 37013 |
| Company Telephone: | 615-964-5392 |
| | |
| Contact: | Christopher M. Smith |
| | Sr. Director of QA/RA |
| | |
| Contact Address & Telephone: | 3034 Owen Drive |
| | Antioch, TN 37013 |
| | 615-964-5392 |
| | |
| Date: | April 2, 2015 |
| | |
| Device Trade Name: | FlashPak® Sterilization Container System |
| | |
| Common Name: | Wrap, Sterilization |
| | |
| Classification, Regulation # and<br>Product Code: | Class II - 880.6850 - FRG |
| | |
| Predicate Device: | FlashPak® Sterilization Container System, K113776 |
Image /page/3/Picture/5 description: The image is a solid red color. The shape is a rectangle that is slightly distorted. The right side of the rectangle is curved inward.
3034 Owen Drive, Antioch, TN 37013 Office 800.251.3000 615,964,5567 Fax = symmetrysurgical.com
Image /page/3/Picture/7 description: The image is a close-up shot of a dark, textured surface. The surface appears to be made of a rough material, possibly stone or concrete. The lighting is dim, which accentuates the texture and creates shadows. The left side of the image has a curved edge, while the rest of the image is more rectangular.
{4}------------------------------------------------
# ymmetry surgical
### Device Description:
The FlashPak Container System consists of a Radel® lid and base that together fully enclose a removable metal wire basket that holds items to be sterilized. A silicone gasket creates an air tight seal between the lid and base by means of metal latches. Two pressure actuated valves; one in the center of the lid and one in the center of the base, allow ingress and egress of steam sterilant into the container during the sterilization cycle and when closed form an airtight enclosure against separate silicone seals. The container comes in four sizes and is indicated for use with gravity-displacement and dynamic-air-removal sterilization modes recommended in ANSI/AAMI ST79 for immediate use steam sterilization (IUSS). Approved chemical indicators are recommended for use during each sterilization cycle.
## Intended Use:
The intended use of the predicate device was changed to add additional steam sterilizations. The 3 minute gravity displacement cycle at 132 ºC and 3 minute dynamic-air removal cycle at 132 ºC were validated and added to the following intended use statement.
"FlashPak® is a reusable rigid container system to be used during immediate use steam sterilization (IUSS or flash) by hospitals and healthcare facilities. It is intended to enable sterilization of the enclosed devices and prevent recontamination during immediate transport to the point of use. The container is compatible with gravitydisplacement steam sterilization using a 10 minute cycle at 132° C for porous items with lumens or a 3 minute cycle at 132° C for nonporous items like routine metal instruments. The container is also compatible with pre-vacuum steam sterilization using a 4 minute cycle at 132° C for porous items with lumens or a 3 minute cycle for nonporous items like routine metal instruments. FlashPak is recommended for sterilization of lumens with the following limits: gravity-displacement (5.5mm inner diameter or larger and up to 184mm in length), pre-vacuum (1mm inner diameter or larger and up to 203mm in length).
FlashPak® Model/Maximum Instrument Load Weight Recommendation:
| 9020 w/9020-08 Basket - Gravity 3 pounds (1.36 Kg) - Pre-Vacuum 3 pounds (1.36 Kg) |
|--------------------------------------------------------------------------------------|
| 9030 w/9030-08 Basket - Gravity 10 pounds (4.55 Kg) - Pre-Vacuum 10 pounds (4.55Kg) |
| 9040 w/9040-08 Basket - Gravity 14 pounds (6.35 Kg) - Pre-Vacuum 14 pounds (6.35 Kg) |
| 9050 w/9050-08 Basket - Gravity 16 pounds (7.27 Kg) - Pre-Vacuum 16 pounds (7.27 Kg) |
# Substantial Equivalence Discussion
The predicate device is the FlashPak Sterilization Container System cleared under 510(k) #K113776. The predicate device and the subject device are physically identical. They are the same device. The differences between the predicate and the subject device involve the steam sterilization parameters recommended in the instructions for use (IFU) and intended use statement. The subject device has additional immediate use steam sterilization parameters recommended for its use that were taken from ANSI/AAMI ST79 and were validated for efficacy.
Image /page/4/Picture/10 description: The image is a solid red color. The shape is rectangular, but the right side is slightly curved. The red color is uniform throughout the image.
34 Owen Drive, Antioch, TN 37013 615 964 5567 symmetry surgical.com
Image /page/4/Picture/12 description: The image is a close-up shot of a dark, textured surface. The surface appears to be slightly curved on the left side, creating a rounded edge. The texture is grainy and uneven, with subtle variations in shading that suggest depth and dimension. The overall impression is one of a solid, unyielding material with a rough, natural finish.
{5}------------------------------------------------
etry surgical"
A comparison of the similarities and differences between the predicate device and the subject device is provided in the table below. The table follows the elements of interest defined in the guidance for Industry and FDA—Premarket Notification [510(k)] Submissions for Medical Sterilization Packaging Systems in Health Care Facilities of March 7, 2002.
| ELEMENT | PREDICATE | SUBJECT DEVICE |
|---------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | FlashPak® is a reusable rigid<br>container system to be used during<br>flash sterilization by hospitals and<br>healthcare facilities. It is intended<br>to enable flash sterilization of the<br>enclosed devices and prevent<br>recontamination during immediate<br>transport to the point of use. The<br>container is compatible with<br>gravity-displacement flash<br>sterilization using a 10 minute cycle<br>at 132° C and with pre-vacuum<br>flash sterilization using a 4 minute<br>cycle at 132° C. FlashPak is<br>recommended for surface<br>sterilization of stainless steel<br>instruments and for lumens with<br>the following limits: gravity-<br>displacement (5.5mm inner<br>diameter or larger and up to<br>184mm in length), pre-vacuum<br>(1mm inner diameter or larger and<br>up to 203mm in length). | FlashPak® is a reusable rigid<br>container system to be used during<br>immediate use steam sterilization<br>(IUSS or flash) by hospitals and<br>healthcare facilities. It is intended<br>to enable sterilization of the<br>enclosed devices and prevent<br>recontamination during immediate<br>transport to the point of use. The<br>container is compatible with<br>gravity-displacement steam<br>sterilization using a 10 minute cycle<br>at 132° C for porous items and<br>items with lumens or a 3 minute<br>cycle at 132° C for nonporous items<br>like routine metal instruments. The<br>container is also compatible with<br>pre-vacuum steam sterilization<br>using a 4 minute cycle at 132° C for<br>porous items and items with<br>lumens or a 3 minute cycle for<br>nonporous items like routine metal<br>instruments. FlashPak is<br>recommended for sterilization of<br>lumens with the following limits:<br>gravity-displacement (5.5mm inner<br>diameter or larger and up to<br>184mm in length), pre-vacuum<br>(1mm inner diameter or larger and<br>up to 203mm in length). |
| | Container:<br>Radel R5000 series base and lid;<br>SS (304) Instrument Basket; | |
| ELEMENT | PREDICATE | SUBJECT DEVICE |
| Material Composition | SS (304) Latches; | Same as predicate |
| | Clear Medical Grade Silicone Gasket<br>and Valve Seal; | |
| | SS (300) Hardware; | |
| | Sintered SS Filter Disk; | |
| | Silicone Lid Vent Button | |
| | Valve: | |
| | SS (304) Bellows; | |
| | SS (316) Bracket | |
| | | |
| Physical Properties | Materials used in the FlashPak are<br>commonly used in many other<br>legally marketed sterilization trays<br>and cases. Physical properties of<br>the system materials and<br>construction have a demonstrated<br>long history of safe and effective<br>use for this application. | Same as predicate |
| | | |
| | | |
| Chemical Properties | Materials used in the FlashPak are<br>commonly used in many other<br>legally marketed sterilization trays<br>and cases. Chemical properties of<br>the system materials and<br>construction have a demonstrated<br>long history of safe and effective<br>use for this application. | Same as predicate |
| | | |
| | | |
| Configurations/Dimensions | Container Sizes:<br>(11" x 11" x 7") #9020<br>(20" x 11" x 8"), #9030<br>(24" x 11" x 9") #9040<br>(24" x 13" x 9") #9050 | Same as predicate |
| | | |
| | | |
| ELEMENT | PREDICATE | SUBJECT DEVICE |
| Air Permeance | N/A<br>Container operates with valves, no<br>filters, wraps or perforations | N/A |
| Percent of Surface<br>Perforations | N/A<br>Container is not perforated | N/A |
| PERFORMANCE | PREDICATE | NEW DEVICE |
| Sterilant Penetration | Sterilization Efficacy<br>Gravity-displacement Cycle:<br>10 minutes at 132°C(270°F) for porous and cannulated devicesDynamic-air-removal Cycle4 minutes at 132°C(270°F) for porous and cannulated devices | Same as predicate with the added indications for:<br>Gravity-displacement Cycle:3 minutes at 132°C(270°F) for nonporous and routine metal devicesDynamic-air-removal Cycle3 minutes at 132°C(270°F) for nonporous and routine metal devices |
| | Sterilant Penetration<br>Three thermal mapping studies for each of the two sterilization modes demonstrating a precise match (within seconds) between the chamber and container temperature profiles. | |
| Microbial Barrier Properties<br>(Packaging Integrity) | The container completely encapsulates the items sterilized inside with the use of a silicone lid gasket, a silicone lid vent, sintered SS microfilter and silicone valve seals to provide an effective microbial barrier during transport to the point of use immediately | Same as predicate |
| ELEMENT | PREDICATE | SUBJECT DEVICE |
| | after sterilization. | |
| | Container passed a microbial aerosol challenge test by preventing microbial ingress during a one hour exposure to aerosolized microbes in a closed chamber. | |
| Material Compatibility | Materials used in the FlashPak are commonly used in many other legally marketed sterilization trays and cases. The system materials and construction have a demonstrated long history of safe and effective use for this application. | Same as predicate…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.