K141788 · Prismatik Dentalcraft, Inc. · EIH · Aug 20, 2014 · Dental
Device Facts
Record ID
K141788
Device Name
OBSIDIAN MILLING BLOCKS
Applicant
Prismatik Dentalcraft, Inc.
Product Code
EIH · Dental
Decision Date
Aug 20, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.6660
Device Class
Class 2
Indications for Use
The Obsidian™ Milling Blocks is used to fabricate ceramic dental prostheses in the nature of crowns and bridges for posterior and anterior applications using CAD/CAM methods.
Device Story
Obsidian™ Milling Blocks are lithium silicate ceramic blocks, supplied with or without an attached mandrel, designed for fabrication of monolithic dental restorations (crowns, inlays, onlays, veneers, partial crowns, substructures). Used in dental clinics/laboratories; operated by dental professionals using CAD/CAM milling systems. Blocks are milled into final prosthetic shapes to provide esthetic, translucent, and high-strength restorations. Device benefits include restoration of dental function and esthetics. No software or electronic components integral to the block itself; milling process relies on external CAD/CAM hardware/software.
Clinical Evidence
Bench testing only. Performance verified against ISO 6872 standards for flexural strength, chemical solubility, radioactivity, and thermal expansion. Biocompatibility testing performed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (sensitization and irritation), with all tests resulting in a 'Pass'.
Technological Characteristics
Lithium silicate ceramic material. Form factor: milling blocks (with/without mandrel). Standards: ISO 6872 (flexural strength, solubility, radioactivity, thermal expansion), ISO 10993-5/10 (biocompatibility). Energy source: N/A (mechanical milling). Connectivity: N/A. Sterilization: Not specified.
Indications for Use
Indicated for fabrication of ceramic dental prostheses (crowns and bridges) for posterior and anterior applications in patients requiring dental restoration.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 20, 2014
Prismatik Dentalcraft, Incorporated Mr. Armin Zehtabchi Senior RA Specialist 2212 Dupont Drive, Suite IJK Irvine California 92612
Re: K141788
Trade/Device Name: Obsidian™ Milling Blocks Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain powder for clinical use Regulatory Class: II Product Code: EIH Dated: July 22, 2014 Received: July 23, 2014
Dear Mr. Zehtabchi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mary S. Runner -ഗ
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) TBD
41788
Device Name
Obsidian™ Milling Blocks
Indications for Use (Describe)
The Obsidian™ Milling Blocks is used to fabricate ceramic dental prostheses in the nature of crowns and bridges for posterior and anterior applications using CAD/CAM methods.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K 141788
Page 10 of 32
Image /page/3/Picture/1 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a triangle on the left and the company name on the right. The triangle is divided into smaller triangles, creating a prism-like effect. The text is in a simple, sans-serif font.
B.
## 007_510 (K) Summary-807.92(c)
This 510 (k) summary is being submitted in accordance with the requirements of SMDA of 1990 and 21 CFR 807.92.
#### SUBMITTER INFORMATION A.
| Company Name: | Prismatik Dentalcraft, Inc. |
|-------------------------------|------------------------------------------------------------|
| Company Address: | 2212 Dupont Dr., Suite IJK,<br>Irvine, CA 92612 |
| Company Phone: | 949-225-1269 |
| Company FAX: | 949-553-0924 |
| Facility Registration Number: | 3005477956 |
| Primary Contact Person: | Armin Zehtabchi, (949) 225-1234<br>Senior RA Specialist |
| Secondary Contact Person | Marilyn Pourazar, (949) 225-1269<br>Senior Director, RA/QA |
| Date Summary Prepared: | July 1, 2014 |
| Trade/Proprietary Name: | Obsidian™ Milling Blocks |
| 21 CFR Reference: | 21 CFR 872.6660 |
| 21 CFR Common Name: | Porcelain powder for clinical use |
| Classification: | Class II, EIH |
| Panel: | Dental |
### C. IDENTIFICATION OF PREDICATE DEVICE
Trade/Proprietary Name:
Obsidian™ Ceramic Blocks-K100781
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a triangle on the left and the text "PRISMATIK DENTALCRAFT, INC." on the right. The triangle is divided into three sections, each with a different pattern.
#### D. PROPOSED DEVICE DESCRIPTION
The Obsidian™ Milling Blocks are a lithium silicate ceramic to be supplied with or without an attached mandrel to be milled using CAD/CAM methods. The product can produce a variety of monolithic restorations with great esthetics, lifelike translucency, and high strength including full-contour crowns, inlays, onlays, veneers partial crowns and substructures due to its excellent machining properties. The milling blocks will be available in the commonly used VITA Classical and Chromascop Bleach shades.
#### E. INDICATIONS FOR USE
The Obsidian™ Milling Blocks is used to fabricate ceramic dental prostheses in the nature of crowns and bridges for posterior and anterior applications using CAD/CAM methods.
#### F. DETERMINATION OF SUBSTANTIAL EQUIVALENCE
The comparison table below outlines and provides the similarities and the substantial-equivalency of the predicate-Obsidian™ Ceramic-Blocks-K100781-(cleared by Prismatik Dentalcraft, Inc. on 6/21/2010) and the proposed Obsidian™ Milling Blocks, and Prismatik believes that the comparative data presented in the preceding paragraphs, demonstrate that proposed Obsidian™ Milling Blocks is essentially the same as currently marketed devices for the same indications for use, and supports our claim of substantial equivalence to predicate Class II devices under the classification of porcelain powder for clinical use (21 CFR 872.6660) that have previously been found to be substantially equivalent. and that any differences between the proposed device Obsidian™ Milling Blocks and the predicate device do not introduce any new issues of safety or effectiveness. Both the proposed device and the predicate device consist of general glass ceramic material and have the same intended use.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of the company name in a simple, sans-serif font. To the left of the text is a geometric shape that resembles a prism, which is likely a visual representation of the company's name. The overall design is clean and professional, suggesting a focus on precision and quality.
| Attributes | Predicate Device<br>Obsidian Ceramic Blocks<br>510(k)-K100781 | Proposed Device<br>Obsidian Milling Blocks | Similarities and<br>Difference<br>Between the<br>Predicate<br>and the Proposed<br>Device |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Indications for Use | This device is used to fabricate<br>ceramic dental prostheses in the<br>nature of crowns and bridges for<br>posterior and anterior applications<br>using CAD/CAM or hot-press<br>methods. | This device is used to fabricate<br>ceramic dental prostheses in the<br>nature of crowns and bridges<br>for posterior and anterior<br>applications using CAD/CAM<br>methods. | Same |
| Composition | The average composition of the<br>proposed device is provided. Refer<br>to section 011. | The average composition of the<br>proposed device is provided<br>Refer to section 011. | Additional materials were<br>added. Refer to section 011. |
| Shades | A1, A2, A3, A35, B1, B2, B3, C1,<br>C2, C3, D2, D3, BL1 and BL4 | A1, A2, A3, A35, B1, B2, B3,<br>C1, C2, D2, D3, BL1 and BL4 | Same |
| Flexural Strength | >300 MPa (meeting ISO 6872<br>requirements) | >300 MPa (meeting ISO<br>6872 requirements) | Same |
| Chemical Solubility | < 100µg/cm² (meeting ISO 6872<br>requirements) | < 100µg/cm² (meeting ISO<br>6872 requirements) | Same |
| Freedom from<br>Extraneous Material | Shall be free from extraneous<br>materials when assessed by visual<br>inspection (meeting ISO 6872<br>requirements) | Shall be free from extraneous<br>materials when assessed by<br>visual inspection (meeting ISO<br>6872 requirements) | Same |
| Radioactivity | Activity concentration of uranium238<br>less than 1.0Bq g-1 (meeting ISO<br>6872 requirements) | Activity concentration of<br>uranium238 less than 1.0Bq g-1<br>(meeting ISO 6872<br>requirements) | Same |
| Coefficient of Thermal<br>Expansion (25-500°C) | 12.2+/-0.5 x10-6 °C (meeting ISO<br>6872 requirements) | 12.2+/-0.5 x 10-6 °C (meeting<br>ISO 6872 requirements) | Same |
| Packaging | Boxes of 5 blocks for Milling<br>Blocks | Boxes of 5 blocks for Milling<br>Blocks | Same |
| Biocompatibility | Non-toxic and biocompatible<br>(Meeting the ISO 10993-5 and<br>10993-10 Requirements) | Non-toxic and biocompatible<br>(Meeting the ISO 10993-5 and<br>10993-10 Requirements) | Same |
| Accessories | Veneering/Add-On Powders, Paste<br>Stains and Glaze, Low Fusing | Veneering/Add-On Powders,<br>Paste Stains and Glaze, Low | Same |
# Table 1 - Comparison between Predicate and Proposed Device
#### G. SUMMARY OF NON-CLINICAL TESTING
To meet the ISO 6872 requirements, various non-clinical and applicable tests were performed in accordance with the Design Verification Plan including a Risk Analysis addressing the impact of changes to the device. The test results for the Flexural Strength, Chemical Solubility, Freedom from Extraneous Material, Radioactivity, and Coefficient Thermal Expansion indicate the Obsidian™ Milling Blocks is comparable to the predicate device.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a triangle on the left and the company name on the right. The triangle is filled with a pattern of lines and shapes. The company name is written in a sans-serif font.
In addition, the proposed device, Obsidian™ Milling Blocks, have been tested for Cytotoxicity (ISO 10993-5), Sensitization (ISO 10993-10), and Irritation (ISO 10993-10) to meet the biocompatibility requirement and the reports are as follow:
- 0 The Cytotoxicity Report shows that there was no reaction on any of the cells.
- The Sensitization Report shows that there was no reaction on the tested o subject.
- The Irritation Report shows that there was no erythema or edema on the test o subject.
| Test Description | Results |
|------------------------------------------------------------------------------------------------------|---------|
| Cytotoxicity Study using the IX MEM extraction method at 37°C | Pass |
| ISO Intracutaneous Study, Extract 0.9% sodium chloride USP solution<br>(SC) and sesame oil, NF (SO) | Pass |
| Irritation and Skin Sensitization Study, Extract 0.9% sodium chloride<br>USP and sesame oil, NF (SO) | Pass |
Copies of the tests results are attached. Refer to Attachment A.
#### H. CONCLUSION
The proposed device, Obsidian™ Milling Blocks has the same performance specifications, fundamental scientific technology and intended use as that of the predicate device, Obsidian™ Ceramic Blocks. The changes to the device do not raise any new questions regarding safety or efficacy. The performance data and a declaration of conformity with design controls support a determination of continuing substantial equivalence of the proposed device to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.