ZOLL X SERIES

K141774 · ZOLL Medical Corporation · MKJ · Nov 19, 2014 · Cardiovascular

Device Facts

Record IDK141774
Device NameZOLL X SERIES
ApplicantZOLL Medical Corporation
Product CodeMKJ · Cardiovascular
Decision DateNov 19, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5310
Device ClassClass 3
AttributesTherapeutic, Pediatric

Indications for Use

The X Series is intended for use by trained medical personnel who are familiar with basic monitoring, vital sign assessment, emergency cardiac care, and the use of the X Series, The X Series is also intended for use by (or on the order of) physicians at the scene of an emergency or in a hospital emergency room, intensive care unit, cardiac care unit, or other similar areas of a hospital. The usage may be in an ambulance or at the scene of an emergency. It is also intended to be used during the transport of patients. The X Series will be used primarily on patients experiencing symptoms of cardiac arrest or in post trauma situation. It may also be used whenever it is required to monitor any of those functions that are included (as options) in the device. The X Series unit can be used on pediatric patients (as described in the following table) and on adult patients (21 years of age or older) with and without heart dysfunction.

Device Story

Lightweight, portable device for emergency cardiac care and vital sign monitoring; used by trained medical personnel in ambulances, hospitals, or emergency scenes. Inputs include ECG leads, SpO2 sensors, NIBP cuffs, CO2 sensors, temperature probes, and invasive pressure transducers. Device processes physiological signals to provide real-time monitoring, defibrillation, and pacing therapy. New TBI Dashboard feature consolidates SBP, SpO2, EtCO2, and ventilation timer data into a graphical web-based format accessible via Bluetooth on external devices. Output includes visual/audio feedback for CPR, physiological waveforms, and trend data. Healthcare providers use output to guide resuscitation, ventilation, and cardiac therapy. Benefits include improved adherence to TBI guidelines and standardized resuscitation protocols.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated through bench testing, specifically software verification and system-level validation.

Technological Characteristics

Portable monitor/defibrillator/pacer. Materials: standard medical-grade plastics/electronics. Sensing: ECG, SpO2 (Masimo Rainbow SET), NIBP, IBP, CO2, temperature, respiration. Connectivity: Ethernet (Web Console), Bluetooth (TBI Dashboard). Software: embedded firmware with web-based data consolidation. Sterilization: N/A (external device).

Indications for Use

Indicated for patients ranging from neonate to adult. Used for manual/semiautomatic defibrillation in cardiac arrest (unconsciousness, apnea, pulselessness); synchronized cardioversion; ECG, SpO2, SpCO, SpMet, CO2, NIBP, IBP, temperature, and respiration monitoring; CPR feedback; and external transcutaneous pacing. TBI Dashboard indicated for monitoring SBP, SpO2, EtCO2, and ventilation timing in trauma patients. Contraindications: obstructive apnea (not an apnea monitor); specific transducer contraindications apply for invasive pressure.

Regulatory Classification

Identification

An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 19, 2014 Zoll Medical Corporation Tanmay Shukla Sr. Regulatory Affairs Specialist 269 Mill Road Chelmsford, Massachusetts 01824-4105 Re: K141774 Trade/Device Name: Zoll X series Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: Class III Product Code: MKJ, LIX, LDD, DRT, DRO, DXN, DSK, CCK, DQA Dated: October 20, 2014 Received: October 21, 2014 Dear Tanmay Shukla, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Melissa A. Torres -S For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Section 4 - Indications for Use 510(k) Number (if known): Device Name: X Series #### Intended Use: The X Series is intended for use by trained medical personnel who are familiar with basic monitoring, vital sign assessment, emergency cardiac care, and the use of the X Series, The X Series is also intended for use by (or on the order of) physicians at the scene of an emergency or in a hospital emergency room, intensive care unit, cardiac care unit, or other similar areas of a hospital. The usage may be in an ambulance or at the scene of an emergency. It is also intended to be used during the transport of patients. The X Series will be used primarily on patients experiencing symptoms of cardiac arrest or in post trauma situation. It may also be used whenever it is required to monitor any of those functions that are included (as options) in the device. The X Series unit can be used on pediatric patients (as described in the following table) and on adult patients (21 years of age or older) with and without heart dysfunction. | Pediatric Patient Subpopulation | Approximate Age Range | |---------------------------------|----------------------------| | Newborn (neonate) | Birth to 1 month of age. | | Infant | 1 month to 2 years of age. | | Child | 2 to 12 years of age. | | Adolescent | 12 to 21 years of age. | When the pediatric patient is less than 8 years of age or weighs less than 55 lbs. (25 kg.), use ZOLL pediatric defibrillation electrodes. Do not delay therapy to determine the patient's exact age or weight. #### Manual Defibrillation Use of the X Series in the manual mode for external and internal defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by: - · Unconsciousness. - · Absence of breathing. - · Absence of pulse. Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) AND/OR Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats. The unit can also be used for synchronized cardioversion of certain atrial or ventricular arrhythmias. Qualified medical personnel must decide when synchronized cardioversion is appropriate. The patient population will range from newborn (neonate) to adult. #### Semiautomatic Operation (AED) X Series products are designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient. They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically-approved patient care protocol. Use of the X Series in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by: - · Unconsciousness. - · Absence of breathing. - Absence of pulse. When the patient is less than 8 years of age or weighs less that 55 lbs. (25 Kg), you must use ZOLL pediatric defibrillation electrodes. Do not delay therapy to determine patient's exact age or weight. # ECG Monitoring The X Series is intended for use to monitor and/or record 3-. 5-. or 12-lead ECG waveform and heart rate, and to alarm when heart rate is above or below limits set by the operator. The patient population will range from newborn (neonate) to adult, with and without heart dysfunction. #### CPR Monitoring The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a minimum compression depth of at least 1.5 (3.8 cm) or 2.0 inches (5.0 cm), depending on the configuration, for adult patients. {4}------------------------------------------------ #### External Transcutaneous Pacing This product can be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation. The purposes of pacing include: - · Resuscitation from standstill or bradycardia of any etiology: - · As a standby when standstill or bradycardia might be expected: - · Suppression of tachycardia. - · Pediatric pacing. #### Non-Invasive Blood Pressure Monitoring The X Series is intended for use to make non-invasive measurements of arterial pressure and heart rate, and to alarm if either parameter is outside of the limits set by the user. Measurements are made using an inflatable cuff on the patient's arm or leg. The patient population will range from newborn (neonate) to adult. #### Temperature Monitoring The X Series is intended for use to make continuous temperature measurements of rectal, esophageal, or surface temperatures, and to alarm if the temperature is outside of the limits set by the user. The patient population will range from newborn (neonate) to adult. #### SpO2 Monitoring The X Series pulse CO-oximeter, with Masimo Rainbow SET technology and the Rainbow series of sensors, is intended for use for continuous nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), and/or methemoglobin saturation (SpMet). The pulse CO-oximeter and accessories are indicated for use on adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused, in hospitals, hospital-tvpe facilities, or in mobile environments. #### Respiration Monitoring The X Series is intended for use to continuously monitor respiration rate and to alarm if the rate falls outside of the range set by the operator. Because the measurement method actually measures respiratory effort, apnea episodes with continued respiratory effort (such as obstructive apnea) may not be detected. It is not intended to be used as an apnea monitor. The patient population will range from newborn (neonate) to adult. #### CO2 Monitoring The X Series is intended for use to make continuous noninvasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath rate. The patient population will range from newborn (neonate) to adult. {5}------------------------------------------------ #### Invasive Pressure Monitoring The X Series is intended for use to display and make continuous invasive pressure measurements from any compatible pressure transducer. The primary intended uses are arterial blood pressure, central venous pressure and intracranial pressure monitoring. Any contraindications of the particular transducer selected by the user shall apply. The patient population will range from newborn (neonate) to adult. #### 12-Lead Analysis The 12-lead ECG Analysis is intended for use in acquiring, analyzing and reporting ECG data, and to provide interpretation of the data for consideration by caregivers. The interpretations of ECG data offered by the device are only significant when used in conjunction with caregiver overread as well as consideration of all other relevant patient data. The 12-lead ECG Analysis is intended for use on adults (> 18 years of age). #### Web Console X Series Web Console functionality allows medical personnel to view physiological data available on a connected X Series device. Physiological data may include ECG, noninvasive blood pressure, SpO2/SpCO/SpMet, end-tidal carbon dioxide, respiration rate, invasive blood pressure, temperature, and Heart Rate. For each patient connected to an X Series, the X Series collects patient data, consolidates the data on an embedded web page that can be requested, through an Ethernet connection, by a web browser on a PC for display. #### TBI Dashboard X Series TBI Dashboard functionality creates an embedded web page that consolidates concurrently displayed data on the X Series for SBP (systolic blood pressure), SpO2, EtCO2 data and timer into graphical trend formats and a countdown timer for ventilations. The web page can be requested, through an Bluetooth connection, using a standard web browser on a viewing device for display. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the word "ZOLL" in a sans-serif font. The letters are a light blue color. A small registered trademark symbol is located to the right of the second "L" in the word. ZOLL Medical Corporation Worldwide Headquarters 269 Mill Road Chelmsford, MA 01824 U.S.A. 978-421-9655 978-421-0025 Main Fax # 510(k) Summarv: | Applicant's Name and Address: | ZOLL Medical Corporation<br>269 Mill Road<br>Chelmsford, MA 01824 | |-------------------------------|-------------------------------------------------------------------| | Application Correspondent: | Tanmay Shukla<br>978-421-9171 | June 30, 2014 Date Summary Prepared: Classification: Class III Device Name Product Code Automated External Defibrillators (MKJ) Cardiopulmonary Resuscitation Aid (LIX) Low-Energy - Defibrillators (LDD) Cardiac Monitors - including Cardiotachometer and Rate Alarms (DRT) External Transcutaneous Cardiac Non-Invasive Pacemaker (DRO) Noninvasive Blood Pressure Measurement System (DXN) Blood Pressure Computer (DSK) Carbon Dioxide Gas Analyzer (CCK) Oximeter (DQA) | Predicate Devices | ZOLL X Series (K133269) | |-------------------|----------------------------------------| | | ZOLL RescueNet Link (K111296) | | | L770-CPR Resuscitation Timer (K062080) | ZOLL X Series {7}------------------------------------------------ # Description: The predicate X Series is a lightweight, portable device designed to be used by trained medical personnel who are familiar with vital signs monitoring and emergency cardiac care. As in its previously cleared configuration, the modified X Series combines the functions of a Semi-automatic (AED)/ manual defibrillator, external transcutaneous pacer and patient monitors (including ECG, SpO2, SpCO, SpMet, CO2, NIBP, IBP, temperature and respiration monitoring). With the current application, we are proposing to revise the X Series communication processor software to create a web page with a graphical trend format and countdown timer for ventilations that consolidates the existing SBP (systolic blood pressure from the NIBP monitoring), SpO2, EtCO2 data and the timer on the display of the X Series. The access to this web page will be marketed as the Traumatic Brain Injury (TBI) Dashboard. This TBI Dashboard assists the operator who is ventilating a patient to follow recommendations in the TBI guidelines. Through an encrypted and password protected communications protocol, a suitable separate device with touchscreen, web browser capability and Bluetooth communications can display the TBI Dashboard. As described above, the proposed software change impacts only the communication aspects of the X Series device. Therapeutic and monitoring capabilities of the X Series device as cleared under K133269 have remained unchanged and are not impacted by the proposed change. The X Series therapeutic and monitoring functionality is entirely controlled by access directly through the X Series device, itself, and not through the TBI Dashboard. TBI Dashboard is intended to supplement and not replace any part of the device monitoring. #### Indications for Use: The X Series is intended for use by trained medical personnel who are familiar with basic monitoring, vital sign assessment, emergency cardiac care, and the use of the X Series. The X Series is also intended for use by (or on the order of) physicians at the scene of an emergency or in a hospital emergency room, intensive care unit, cardiac care unit, or other similar areas of a hospital. The usage may be in an ambulance or at the scene of an emergency. It is also intended to be used during the transport of patients. The X Series will be used primarily on patients experiencing symptoms of cardiac arrest or in post trauma situation. It may also be used whenever it is required to monitor any of those functions that are included (as options) in the device. The X Series unit can be used on pediatric patients (as described in the following table) and on adult patients (21 years of age or older) with and without heart dysfunction. | Pediatric Patient Subpopulation | Approximate Age Range | |---------------------------------|----------------------------| | Newborn (neonate) | Birth to 1 month of age. | | Infant | 1 month to 2 years of age. | | Child | 2 to 12 years of age. | | Adolescent | 12 to 21 years of age. | {8}------------------------------------------------ Page 3 of 6 When the pediatric patient is less than 8 years of age or weighs less than 55 lbs. (25 kg.), use ZOLL pediatric defibrillation electrodes. Do not delay therapy to determine the patient's exact age or weight. # Manual Defibrillation Use of the X Series in the manual mode for external and internal defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by: - · Unconsciousness. - · Absence of breathing. - Absence of pulse. This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats. The unit can also be used for synchronized cardioversion of certain atrial or ventricular arrhythmias. Qualified medical personnel must decide when synchronized cardioversion is appropriate. The patient population will range from newborn (neonate) to adult. # Semiautomatic Operation (AED) X Series products are designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient. They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically-approved patient care protocol. Use of the X Series in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by: - · Unconsciousness. - · Absence of breathing. - · Absence of pulse. When the patient is less than 8 years of age or weighs less that 55 lbs. (25 Kg), you must use ZOLL pediatric defibrillation electrodes. Do not delay therapy to determine patient's exact age or weight. K141774 {9}------------------------------------------------ # ECG Monitoring The X Series is intended for use to monitor and/or record 3-, 5-, or 12-lead ECG waveform and heart rate, and to alarm when heart rate is above or below limits set by the operator. The patient population will range from newborn (neonate) to adult, with and without heart dysfunction. # CPR Monitoring The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a minimum compression depth of at least 1.5 (3.8 cm) or 2.0 inches (5.0 cm), depending on the configuration, for adult patients. # External Transcutaneous Pacing This product can be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation. The purposes of pacing include: - . Resuscitation from standstill or bradycardia of any etiology: - As a standby when standstill or bradycardia might be expected: ● - Suppression of tachycardia. - Pediatric pacing. ● # Non-Invasive Blood Pressure Monitoring The X Series is intended for use to make non-invasive measurements of arterial pressure and heart rate, and to alarm if either parameter is outside of the limits set by the user. Measurements are made using an inflatable cuff on the patient's arm or leg. The patient population will range from newborn (neonate) to adult. # Temperature Monitoring The X Series is intended for use to make continuous temperature measurements of rectal, esophageal, or surface temperatures, and to alarm if the temperature is outside of the limits set by the user. The patient population will range from newborn (neonate) to adult. # SpO2 Monitoring The X Series pulse CO-oximeter, with Masimo Rainbow SET technology and the Rainbow series of sensors, is intended for use for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), and/or methemoglobin saturation (SpMet). The pulse CO-oximeter and accessories are indicated for use on adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused, in hospitals, hospital-type facilities, or in mobile environments. {10}------------------------------------------------ ### Respiration Monitoring The X Series is intended for use to continuously monitor respiration rate and to alarm if the rate falls outside of the range set by the operator. Because the measurement method actually measures respiratory effort, apnea episodes with continued respiratory effort (such as obstructive apnea) may not be detected. It is not intended to be used as an apnea monitor. The patient population will range from newborn (neonate) to adult. ### CO2 Monitoring The X Series is intended for use to make continuous noninvasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and breath rate. The patient population will range from newborn (neonate) to adult. ### Invasive Pressure Monitoring The X Series is intended for use to display and make continuous invasive pressure measurements from any compatible pressure transducer. The primary intended uses are arterial blood pressure, central venous pressure and intracranial pressure monitoring. Any contraindications of the particular transducer selected by the user shall apply. The patient population will range from newborn (neonate) to adult. ### 12-Lead Analysis The 12-lead ECG Analysis is intended for use in acquiring, analyzing and reporting ECG data, and to provide interpretation of the data for consideration by caregivers. The interpretations of ECG data offered by the device are only significant when used in conjunction with caregiver overread as well as consideration of all other relevant patient data. The 12-lead ECG Analysis is intended for use on adults (> 18 years of age). # Web Console X Series Web Console functionality allows medical personnel to view physiological data available on a connected X Series device. Physiological data may include ECG, noninvasive blood pressure, SpO2/SpCO/SpMet, end-tidal carbon dioxide, respiration rate, invasive blood pressure, temperature, and Heart Rate. For each patient connected to an X Series, the X Series collects patient data, consolidates the data on an embedded web page that can be requested, through an Ethernet connection, by a web browser on a PC for display. # TBI Dashboard X Series TBI Dashboard functionality creates an embedded web page that consolidates concurrently displayed data on the X Series for SBP (systolic blood pressure). SpO2, EtCO2 data and timer into graphical trend formats and a countdown timer for ventilations. The web page can be requested, through an Bluetooth connection, using a standard web browser on a viewing device for display. {11}------------------------------------------------ #### Substantial Equivalence - Non-Clinical Evidence: Features and functions cleared with the ZOLL X Series (K133269) device has remained unchanged in the proposed version of the device. TBI Dashboard functionality to the X Series device, added through a software revision, is substantially equivalent to the ZOLL RescueNet Link (K111296) and L770-CPR Resuscitation Timer (K062080). Safety, efficacy and substantial equivalence was shown through software verification and system level validation. ### Substantial Equivalence - Clinical Evidence: N/A - Clinical evidence was not necessary to show substantial equivalence ### Comparison of Technological Characteristics Features and functions cleared with the ZOLL X Series (K133269) have remained unchanged in the proposed version of the device. The proposed TBI Dashboard functionality and RescueNet Link (K111296) both collect and display data, that is entered by the user (caregiver) or captured from the device(s) to which they are connected, in the form of a web page that can be accessed by the authorized users through standard web browser technology. As with the predicate RescueNet link, TBI Dashboard is intended to supplement and not replace any part of the device monitoring. Ventilation rate timer functionality in the TBI Dashboard is substantially equivalent to the the L770-CPR Resuscitation Timer (K062080). Both the devices are visual aids for a rescuer trained in the protocol they support; which can automatically set the target ventilation rate to the guideline-recommended rate based upon the selections made at initiation. # Performance Testing: Extensive performance testing in the form of the software verification and system level validation ensures that the ZOLL X Series performs as well as the indicated predicate devices and meets all of its functional requirements and performance specifications. Safety testing assures that the device complies with applicable sections of recognized industry and safety standards. #### Conclusion The information provided in this 510(k) demonstrates that the ZOLL X Series' features and functions are substantially equivalent to those of the indicated commercially distributed devices with regard to performance, safety and effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...