Retrospective clinical data was used to define the risk classification scoring algorithm (low, intermediate, high) by associating Prosigna scores with long-term clinical outcomes (10-year distant recurrence-free survival).
Proprietary algorithm using gene expression centroids and clinical variables
—
—
—
—
—
—
Indications for Use
The Prosigna™ Breast Cancer Prognostic Gene Signature Assay is an in vitro diagnostic assay which is performed on the NanoString nCounter Dx Analysis System using FFPE breast tumor tissue previously diagnosed as invasive breast carcinoma. This qualitative assay utilizes gene expression data, weighted together with clinical variables to generate a risk category and numerical score, to assess a patient's risk of distant recurrence of disease. The Prosigna Breast Cancer Prognostic Gene Signature Assay is indicated in female breast cancer patients who have undergone surgery in conjunction with locoregional treatment consistent with standard of care, either as: 1. A prognostic indicator for distant recurrence-free survival at 10 years in post-menopausal women with hormone receptor-positive (HR+), lymph node-negative, Stage I or II breast cancer to be treated with adjuvant endocrine therapy alone, when used in conjunction with other clinicopathological factors. 2. A prognostic indicator for distant recurrence-free survival at 10 years in post-menopausal women with hormone receptor-positive (HR+), lymph node-positive (1-3 positive nodes), Stage II breast cancer to be treated with adjuvant endocrine therapy alone, when used in conjunction with other clinicopathological factors. The device is not intended for patients with 4 or more positive nodes.
Device Story
Assay processes FFPE breast tumor tissue to quantify 58 gene RNA expression profiles. Workflow: microdissection; RNA isolation; multiplex hybridization with biotin-labeled magnetic capture probes and fluorescent barcode reporter probes; automated purification on Prep Station; digital counting on Digital Analyzer. System uses proprietary algorithm to integrate gene expression data with clinical variables (tumor size, proliferation score) to generate Prosigna score (0-100) and risk category (low/intermediate/high). Used in clinical labs by technicians; output provided as patient-specific report for physicians to assess 10-year distant recurrence risk and inform adjuvant endocrine therapy decisions.
Clinical Evidence
Bench testing only. Comparison study of 40 RNA samples extracted from FFPE tissue demonstrated linear correlation between 4-test and 10-test kit configurations (Deming regression: 10 sample = 1.094 + 0.982 * 4 sample). Bland-Altman analysis confirmed no bias and no changes in risk categorization between configurations. Clinical performance data relies on the predicate device (K130010).
Technological Characteristics
Nucleic acid hybridization and image analysis system. Components: RNA isolation kit, Prosigna reagents (CodeSet, Prep Pack, Cartridge), Prep Station, Digital Analyzer. Uses biotin-labeled magnetic probes and fluorescent barcode probes. Connectivity: standalone/networked. Software: Visual C++ web-based application. Sterilization: N/A (reagents).
Indications for Use
Indicated for female breast cancer patients post-surgery and locoregional treatment. Specifically for post-menopausal women with HR+, Stage I/II, node-negative breast cancer, or HR+, Stage II, node-positive (1-3 nodes) breast cancer, to be treated with adjuvant endocrine therapy. Not for patients with 4+ positive nodes.
Regulatory Classification
Identification
A gene expression profiling test system for breast cancer prognosis is a device that measures the ribonucleic acid (RNA) expression level of multiple genes and combines this information to yield a signature (pattern or classifier or index) to aid in prognosis of previously diagnosed breast cancer.
Special Controls
The device is classified as Class II under regulation 21 CFR 866.6040 with special controls. The special control guidance document "Class II Special Controls Guidance Document: Gene expression profiling test system for breast cancer prognosis" is available at http://www.fda.gov/cdrh/oivd/guidance/1627.html.
*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Gene Expression Profiling Test System for Breast Cancer Prognosis.” See § 866.1(e) for the availability of this guidance document.
Predicate Devices
Prosigna™ Breast Cancer Prognostic Gene Signature Assay (K130010)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is positioned to the right of a circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### May 28, 2015
Nanostring Technologies, Inc. Sylva Krizan, Ph.D. Regulatory Affairs Manager 530 Fairview Avenue North, Suite 2000 Seattle, WA 98109
Re: K141771
Trade/Device Name: Prosignatin Breast Cancer Prognostic Gene Signature Assay Regulation Number: 21 CFR §866.6040 Regulation Name: Gene expression profiling test system for breast cancer prognosis Regulatory Class: Class II Product Code: NYI Dated: June 30, 2014 Received: July 1, 2014
Dear Dr. Krizan:
This letter corrects our substantially equivalent letter of November 7, 2014.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical
{1}------------------------------------------------
Page 2 - Sylva Krizan, Ph.D.
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Reena Philip -S
Reena Philip, Ph.D. Director Division of Molecular Genetics and Pathology Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure: Corrected Indications for Use Statement
{2}------------------------------------------------
#### 510(k) Number (if known): K141771
Device Name: Prosigna™ Breast Cancer Prognostic Gene Signature Assay
#### Indications for Use:
The Prosigna® Breast Cancer Prognostic Gene Signature Assay is an in vitro diagnostic assay which is performed on the NanoString nCounter® Dx Analysis System using FFPE breast tumor tissue previously diagnosed as invasive breast carcinoma. This qualitative assay utilizes gene expression data, weighted together with clinical variables to generate a risk category and numerical score, to assess a patient's risk of distant recurrence of disease.
The Prosigna Breast Cancer Prognostic Gene Signature Assay is indicated in female breast cancer patients who have undergone surgery in conjunction with locoregional treatment consistent with standard of care, either as:
1. A prognostic indicator for distant recurrence-free survival at 10 years in postmenopausal women with Hormone Receptor-Positive (HR+), lymph node-negative, Stage I or II breast cancer to be treated with adjuvant endocrine therapy alone, when used in conjunction with other clinicopathological factors.
2. A prognostic indicator for distant recurrence-free survival at 10 years in postmenopausal women with Hormone Receptor-Positive (HR+), lymph node-positive (1-3 positive nodes), Stage II breast cancer to be treated with adjuvant endocrine therapy alone, when used in conjunction with other clinicopathological factors. The device is not intended for patients with 4 or more positive nodes
Prescription Use ___x (Part 21 CFR 801 Subpart
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Division Sign-Off Office of In Vitro Diagnostics and Radiological Health
510(k) K141771
Page 1 of 1_
{3}------------------------------------------------
### 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
### A. 510(k) Number:
K141771
### B. Purpose for Submission:
Modification of device configuration and software
### C. Measurand:
58 gene RNA expression profile
#### D. Type of Test:
Gene expression profile system based upon non-amplified RNA hybridization, visualization, and image analysis
#### E. Applicant:
NanoString Technologies
#### F. Proprietary and Established Names:
Prosigna™ Breast Cancer Prognostic Gene Signature Assay
### G. Regulatory Information:
- 1. Regulation section:
21 CFR §866.6040 Gene expression profiling test system for breast cancer prognosis
- 2. Classification:
Class II
- 3. Product code:
NYI, Classifier, prognostic, recurrence risk assessment, RNA gene expression, breast cancer
- 4. Panel:
Immunology (82)
#### H. Intended Use:
- 1. Intended use(s):
The Prosigna™ Breast Cancer Prognostic Gene Signature Assay is an in vitro diagnostic assay which is performed on the NanoString nCounter Dx Analysis System using FFPE
{4}------------------------------------------------
breast tumor tissue previously diagnosed as invasive breast carcinoma. This qualitative assay utilizes gene expression data, weighted together with clinical variables to generate a risk category and numerical score, to assess a patient's risk of distant recurrence of disease.
The Prosigna Breast Cancer Prognostic Gene Signature Assay is indicated in female breast cancer patients who have undergone surgery in conjunction with locoregional treatment consistent with standard of care, either as:
- 1. A prognostic indicator for distant recurrence-free survival at 10 years in post-menopausal women with hormone receptor-positive (HR+), lymph node-negative, Stage I or II breast cancer to be treated with adjuvant endocrine therapy alone, when used in conjunction with other clinicopathological factors.
- 2. A prognostic indicator for distant recurrence-free survival at 10 years in post-menopausal women with hormone receptor-positive (HR+), lymph node-positive (1-3 positive nodes), Stage II breast cancer to be treated with adjuvant endocrine therapy alone, when used in conjunction with other clinicopathological factors. The device is not intended for patients with 4 or more positive nodes.
- 2. Indication(s) for use:
Same as intended use
- 3. Special conditions for use statement(s): For Prescription Use Only
- 4. Special instrument requirements: nCounter Dx Analysis System
### I. Device Description:
The required components for the Prosigna™ Breast Cancer Prognostic Gene Signature Assay include the RNA Isolation kit (manufactured by Roche), Prosigna reagents (Reference Sample, CodeSet, Prep Pack, Cartridge(s) and Prep Plate) and the instruments that comprise the nCounter Dx Analysis System; the Prep Station and Digital Analyzer.
The assay requires microdissection of tumor from FFPE biopsies, isolation of RNA using a Roche RNA isolation kit, transfer of RNA to PCR tubes for hybridization before placing onto the prep station. Two sets of probes specific to each of 58 RNAs are added to the hybridization reaction. These consist of biotin-labeled magnetic probes to purify the RNAs and capture them on the assay cartridge and fluorescent "barcode" probes to detect and quantify individual RNAs. The patient sample and probes are pipetted automatically into the Prosigna test cartridge by the Prep Station. The prep station uses magnetic bead capture and washing to remove excess RNA and un-hybridized probes. The isolated and hybridized RNA species are then bound via biotin on the capture probe randomly to streptavidin on the cartridge. The fluorescent molecules are then aligned on the cartridge by addition of an electric current. The cartridge is then transferred to the Digital Analyzer where the cartridge
{5}------------------------------------------------
is scanned and digital analysis software is used to count the number of each RNA species present. The amount of each RNA is then put into a proprietary algorithm to produce a Prosigna score.
The test output is a patient specific report which includes a Prosigna score (0-100) and risk category (low/intermediate/high).
#### J. Substantial Equivalence Information:
- 1. Predicate device name(s):
ProsignaTM Breast Cancer Prognostic Gene Signature Assay
- 2. Predicate 510(k) number(s): k130010
- 3. Comparison with predicate:
Table 1: Comparison with Predicate
| Item | Device | Predicate |
|--------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Intended Use | The Prosigna™ Breast Cancer<br>Prognostic Gene Signature<br>Assay is an <i>in vitro</i> diagnostic<br>assay which is performed on the<br>NanoString nCounter Dx<br>Analysis System using FFPE<br>breast tumor tissue previously<br>diagnosed as invasive breast<br>carcinoma. This qualitative<br>assay utilizes gene expression<br>data, weighted together with<br>clinical variables to generate a<br>risk category and numerical<br>score, to assess a patient's risk<br>of distant recurrence of disease.<br><br>The Prosigna Breast Cancer<br>Prognostic Gene Signature<br>Assay is indicated in female<br>breast cancer patients who have<br>undergone surgery in<br>conjunction with locoregional<br>treatment consistent with<br>standard of care, either as: | Same |
{6}------------------------------------------------
| | A prognostic indicator for<br>1. distant recurrence-free<br>survival at 10 years in<br>post-menopausal women<br>with hormone<br>receptor-positive (HR+),<br>lymph node-negative, Stage<br>I or II breast cancer to be<br>treated with adjuvant<br>endocrine therapy alone,<br>when used in conjunction<br>with other<br>clinicopathological factors.<br>2. A prognostic indicator for<br>distant recurrence-free<br>survival at 10 years in<br>post-menopausal women<br>with hormone<br>receptor-positive (HR+),<br>lymph node-positive (1-3<br>positive nodes), Stage II<br>breast cancer to be treated<br>with adjuvant endocrine<br>therapy alone, when used in<br>conjunction with other<br>clinicopathological factors.<br>The device is not intended<br>for patients with 4 or more<br>positive nodes. | |
|------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Prescription Use | Yes | Yes |
| Device Description | Prosigna™ Breast Cancer<br>Prognostic Gene Signature<br>Assay and nCounterDx<br>Analysis Platform; all elements<br>cleared by FDA as a distributed<br>test and platform | Same |
| Test Sample | FFPE breast tumor tissue | Same |
| Extraction/amplification<br>reagents/amplification<br>procedures | No amplification required;<br>procedure for processing FFPE<br>tumor samples provided;<br>includes RNA isolation,<br>multiplex<br>hybridization in solution,<br>automated purification on a | Same |
·
{7}------------------------------------------------
| liquid handling robot and<br>analysis on an automated<br>epifluorescence microscope | | |
|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Differences | | |
| Item | Device | Predicate |
| Kit Stability/Shelf Life | 8 months, based upon real time<br>stability data available from<br>original testing protocol | 7 months, based on testing<br>completed at time of<br>clearance |
| Device Configuration | Reagents configured and<br>software programmed to<br>prepare 2 reference samples and<br>10 test sample -10 test<br>configuration<br>And<br>Reagents configured and<br>software programmed to<br>prepare 2 reference samples and<br>4 test sample -4 test<br>configuration | Reagents configured and<br>software programmed to<br>prepare 2 reference<br>samples and 10 test sample<br>-10 test configuration |
| Instrument Software | Version 1.3 | Version 1.0 |
| Instrument Functionality | FLEX configuration allows for<br>IVD or research use of the<br>device with different modes<br>separated through user<br>permissions and required log-<br>out and log-in when changing<br>modes | IVD use only |
## K. Standard/Guidance Document Referenced (if applicable):
Class II Special Controls Guidance Document: "Gene Expression Profiling Test System for Breast Cancer Prognosis, issued on May 9, 2007"
### L. Test Principle:
Used together, the Prosigna™ Breast Cancer Prognostic Gene Signature Assay and nCounter Dx Analysis System are a nucleic acid hybridization and image analysis system based upon coded probes designed to detect the messenger RNA transcribed from 58 genes. The test input is purified RNA from FFPE breast tumor specimens which are acquired from surgical resection. The Prosigna assay uses gene-specific probe pairs that hybridize directly to the mRNA transcripts in solution. The nCounter Dx Analysis System delivers direct, multiplexed measurements of gene expression through digital readouts of the relative abundance of the mRNA transcripts.
{8}------------------------------------------------
Specifications are included as part of the Prosigna Assay to control for sample quality, RNA quality, and process quality. The Prosigna assay utilizes prototypical expression profiles (centroids) for breast cancer. Patients RNA signatures are categorized into one of four centroids (not reported) based upon how close their gene expression pattern is to each of the centroids. The software algorithm produces a Prosigna score based on the similarity of the expression profile to each centroid, as well as the pathological tumor size and a proliferation score computed from a subset of genes. Three risk categories (low, intermediate and high) were defined based on a study with over 1007 patient samples associating Prosigna score with long-term outcome, defined by distance recurrence free survival at 10 years (DRFS) (Table 2).
| Nodal Status | Prosigna Score Range | Risk Classification |
|------------------------------|----------------------|---------------------|
| Node-Negative | 0-40 | Low |
| | 41-60 | Intermediate |
| | 61-100 | High |
| Node-Positive<br>(1-3 nodes) | 0-40 | Low |
| | 41-100 | High |
Table 2: Risk Classification Scoring Algorithm Using Prosigna Score
#### Performance Characteristics (if/when applicable): M.
- 1. Analytical performance:
See Predicate Device K130010 for Analytical Performance Data.
### 2. Comparison studies:
- a. Method comparison with predicate device:
A comparison of Prosigna Scores was conducted using the two available configurations of the kit (4 sample and 10 sample kits). Forty samples of RNA previously extracted from FFPE breast tissue, covering a range of Prosigna scores were tested with both configurations and the data plotted using Deming regression (Figure 1 below).
{9}------------------------------------------------
Image /page/9/Figure/0 description: This image is a scatter plot comparing two samples, labeled '10 sample' on the y-axis and '4 sample' on the x-axis. The plot shows a strong positive correlation between the two samples, with data points clustered tightly around a straight line. A Deming regression equation is provided, stating '10 sample = 1.094 + 0.982 * 4 sample', indicating the linear relationship between the two samples.
The data was linear over the range of the assay with no outliers between the two methods indicating that the 4-kit and 10-kit assays produce substantially equivalent results.
No value deviated by more than 2 units from the average score when run using either configuration.
Image /page/9/Figure/3 description: The image is a Bland Altman plot showing the average ROR vs. ROR difference. The x-axis represents the average ROR, ranging from 0 to 100. The y-axis represents the ROR difference, ranging from -4 to 4. The plot includes horizontal lines indicating the upper acceptance limit, upper limit of agreement, average bias, lower limit of agreement, and lower acceptance limit.
Bland-Allman Plot; Average ROR vs. ROR Difference
Figure 3: Bland-Altman comparison plo: of ROR scores obtained from each kit configuration for the analysis data set (40 samples).
Bland Altman Analysis (above) showed that variations in data using the 4-kit versus the 10-kit configuration did not bias results and no changes in risk categorization occurred.
7
{10}------------------------------------------------
- b. Matrix comparison:
Not Applicable. FFPE tissue is the only matrix indicated for this device.
- 3. Clinical studies:
See Predicate Device K130010 for Clinical Performance Data
- 4. Clinical cut-off:
Same as assay cut-off
- 5. Expected values/Reference range:
Risk assessment is reported as Low Risk, Intermediate Risk, or High Risk for node negative patients or as Low Risk or High Risk for Node positive patients (see Table 3).
Table 3: Risk Classification Scoring Algorithm Using Prosigna Score
| Nodal Status | Prosigna Score Range | Risk Classification |
|---------------------------|----------------------|---------------------|
| Node-Negative | 0-40 | Low |
| | 41-60 | Intermediate |
| | 61-100 | High |
| Node-Positive (1-3 nodes) | 0-40 | Low |
| | 41-100 | High |
#### N. Instrument Name:
The nCounter Dx Analysis System consists of a liquid handling robot Prep Station 5s and an epifluorescent scanner Digital Analyzer 5s.
### O. System Descriptions:
- 1. Modes of Operation:
Automated
- 2. Software:
The Digital Analyzer measures and sorts multiple signals (reporter probes bound to mRNA transcript) from the clinical sample to establish an indicator (Prosigna score and risk category) to aid in determining patient prognosis. The Prep Station automates posthybridization sample processing while the Digital Analyzer includes signal reading, raw data storage, data acquisition software and software to process the detected targets (algorithm).
The Software is a Visual C++ web-based application developed by Nanostring.
{11}------------------------------------------------
The current version of the Software is v1.3 and includes validation for the 4-test kit configuration in a dual IVD/RUO mode.
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes ___x_ ____________________________________________________________________________________________________________________________________________________________________
### 3. Specimen Identification:
Specimen identifying information is entered into a computer application manually.
- 4. Specimen Sampling and Handling:
Samples are handled individually until RNA is extracted from FFPE tissue. RNA samples are then handled in batches of 6 or 12 on the instrument. These include 2 control samples and either 4 or 10 test samples, depending upon the configuration of reagent plates used.
- 5. Calibration:
Installation, calibration and preventative maintenance of instrumentation are performed by the instrument manufacturer. No user calibration required.
- 6. Quality Control:
Quality control includes testing of the mixed Reporter CodeSet and Capture ProbeSet for the following performance characteristics:
- Signal level of the geometric mean of housekeeping gene probes .
- . Signal levels of each of the 50 classifier genes
- Background level of the negative controls .
- Linearity of positive controls .
- . Probe cross-contamination
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
None
#### Q. Proposed Labeling:
- 1. Labeling was modified to reflect
- . the new software version,
- to indicate the presence of the 4-test kit configuration .
- . to include acceptable specifications for low volume spectrophotometers to be used with the assay
- . to alert physicians to a trend in the data noted by post-hoc analysis whereby most distant recurrence appears to occur after 5 years.
{12}------------------------------------------------
- 2. The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
### R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
### S. Other Supportive Device and Instrument Information:
- 1. See predicate device K130010 for additional information concerning device performance.
- 2. Sponsor also included an update to the specifications for spectrophotometers that may be used to prepare RNA for the Prosigna device. The data revealed that at least two low volume spectrophotometers would provide acceptable quantitation of RNA for later use in the Prosigna assay.
- 3. Additional Clinical Claims
- a. Sponsor originally sought additional claims, through changes to the patient Case Report Form in order to claim the Prosigna device could distinguish risk groups by risk of early recurrence versus late recurrence. That is to say that breast cancer that is expected to recur within a 10 year period, but that had a higher risk of arising within 5 years after surgery could be distinguished from those at higher risk of arising between 5 and 10 years after surgery. The Sponsor's calculations and pre-defined acceptance criteria only included those patients in the 5-10 year (late) recurrence group and did not take into account the early recurrence group.
- b. Sponsor indicated that this information was provided to avoid confusion and were only included for "descriptive" purposes. FDA noted that the Sponsor provided precise calculations of data that had overlapping confidence intervals and was not statistically significant.
- c. FDA noted to Sponsor that the information that they intended to convey was already present in the Kaplan-Meier survival curves presented on the case report forms and that a textual instructions to take care in the interpretation of survival was truthful and accurate without being false or misleading.
- d. Sponsor removed charts of early v. late recurrence from the patient case report form.
- e. Sponsor has altered the final version of labeling to remove a specific claim of early v. late recurrence.
- Sponsor did include a text description of the 0-5 year early versus 5-10 year late ﺕ recurrence and presented the Kaplan-Meier curve for this population
- 4. Additional Stability Claims
- a. Sponsor took a retrospective look at their stability data that was based upon reduction of signal of the control material geomean value once the stability study was completed.
- b. Sponsor noted that if the lot release criteria for the control material were increased from a geomean value of 2289 to 3308, the stability data previously collected would support a stability claim of 11 months.
- c. FDA noted that while this hypothesis was scientifically sound, it was not properly
{13}------------------------------------------------
validated as the new lot release criteria was neither specified, nor in place when the stability testing began. As such, the stability remains established at 8 months, as was established via an add to file for the original 510(k): K130010..
- 5. Additional Labeling Changes
- Sponsor validated a new software package that allowed the nCounter elements a. instrument run in two modes. The first is the cleared "IVD" mode. The other was variously described as "non-IVD" or "life Sciences" mode.
- b. FDA indicated to Sponsor that the change in software was acceptable but that all labeling and software must indicate that the FLEX device configuration is acceptable for IVD use only when used in "IVD" mode.
#### T. Administrative Information:
- 1. Applicant Contact Information:
- a. Name of applicant: Nanostring Technologies
- b. Mailing address: 530 Fairview Avenue North, Suite 2000
Seattle, Washington 98109
- c. Phone #: (206)432-8854
- d. Fax #: (206)378-6288
- e. E-mail address (optional): skrizan(@nanostring.com
- Contact: Sylva Krizan, Ph.D. f.
#### 2. Review Documentation:
| July 2, 2014 | Special 510(k) received. Lead Reviewer assigned-Kevin Lorick. |
|--------------------|--------------------------------------------------------------------------|
| July 14, 2014 | RTAA designation. 510(k) accepted for review. |
| July 14, 2104 | Special 510(k) converted to a traditional 510(k) due to new claims |
| July 31, 2014 | Request for Additional Information |
| September 8, 2014 | Supplement S001, Response to Request for Additional Information received |
| September 26, 2014 | FDA sent Email Request for clarification. |
| September 29, 2014 | FDA sent 2nd Email Request for clarification. |
{14}------------------------------------------------
- October 1, 2014 Sponsor sent response to September 26, 2014 FDA email
- October 2, 2014 Email and phone calls to Sponsor to discuss possible disallowance of claims.
- October 2, 2014 Phone call from Sponsor. Sponsor agreed in principle to attempt to modify submission by COB on October 3, 2014.
- October 9, 2014 Teleconference with Sponsor to discuss timing and content of their response to FDA inquiries.
- October 24, 2014 Email from Sponsor with preliminary responses to FDA inquiries.
- October 30, 2014 Sponsor request to discuss proposed "Special Indications for Use"
- November 3, 2014 FDA call to Sponsor to seek modification of redlines version of PI.
- Sponsor email providing updates to labeling removing changes to November 4, 2014 Special Conditions for Use
- November 4, 2014 FDA Email to Sponsor indicating where the new version was deficient
- November 5, 2014 Email to Sponsor requesting updated data comparison figures and validation for support instrument specifications.
- November 5, 2014 Sponsor email request for additional feedback based upon the November 3. PI.
- Email to Sponsor with suggested PI changes. November 6, 2014
- November 7, 2014 Sponsor email with Final Labeling
- November 8, 2014 SE determination made
- April 17, 2015 Amendment received by DCC
- April 17, 2015 Lead Reviewer Kevin Lorick Assigned
- Internal Discussion about necessary documents and procedures May 1, 2015
- May 6, 2015 510(k) Amendment determined to be appropriate.
- Updated labeling, 510(k) Summary and Indications for Use May 22, 2015 uploaded to DocMan.
{15}------------------------------------------------
### 3. Substantial Equivalence Discussion:
| | | Yes | No | |
|----|------------------------------------------------------------------------------------|-----|----|-------------------------------------|
| 1. | Same Indication Statement? | x | | If YES = Go To 3 |
| 2. | Do Differences Alter The Effect Or Raise<br>New Issues of Safety Or Effectiveness? | | | If YES = Stop NSE |
| 3. | Same Technological Characteristics? | x | | If YES = Go To 5 |
| 4. | Could The New Characteristics Affect<br>Safety Or Effectiveness? | | | If YES = Go To 6 |
| 5. | Descriptive Characteristics Precise<br>Enough? | x | | If NO = Go To 8<br>If YES = Stop SE |
| 6. | New Types Of Safety Or Effectiveness<br>Questions? | | | If YES = Stop NSE |
| 7. | Accepted Scientific Methods Exist? | | | If NO = Stop NSE |
| 8. | Performance Data Available? | | | If NO = Request Data |
| 9. | Data Demonstrate Equivalence? | | | Final Decision: SE |
Note: See
http://eroom.fda.gov/eRoomReq/Files/CDRH3/CDRHPremarketNotification510kProgra m/0 4148/FLOWCHART%20DECISION%20TREE%20.DOC for Flowchart to assist in decision-making process. Please complete the following table and answer the corresponding questions. "Yes" responses to questions 2, 4, 6, and 9, and every "no" response requires an explanation.
- a. Explain how the new indication differs from the predicate device's indication: NA
- b. Explain why there is or is not a new effect or safety or effectiveness issue: NA
- c. Describe the new technological characteristics:
{16}------------------------------------------------
NA
- d. Explain how new characteristics could or could not affect safety or effectiveness: NA
- e. Explain how descriptive characteristics are not precise enough: NA
- f. Explain new types of safety or effectiveness question(s) raised or why the question(s) are not new: N/A
- g. Explain why existing scientific methods cannot be used: NA
- h. Explain what performance data is needed: Demonstration that 4 and 10-test kit configurations produce identical results on the device with the v1.3 software are sufficient. No additional performance data is required.
- ...: Explain how the performance data demonstrates that the device is or is not substantially equivalent:
The changes to the devices are minor software and configuration changes. These effect only the sample preparation steps in terms of reagent positioning. No changes to reagents, hardware, analysis, scoring algorithm or significant changes to patient disease state result from the device modification. One additional piece of information is clarified in package insert but this was done for physician information purposes in a manner that may reduce confusion and aid patient safety.
- U. Reviewer Name and Signature:
Kevin
Lorick -S
Digitally signed by Kevin Lorick -S
DN: c=US, o=U.S. Government,
ou=HHS, ou=FDA, ou=People,
cn=Kevin Lorick -S,
0.9.2342.19200300.100.1.1=200035
0803
Date: 2015.05.29 13:59:27-04'00'
Kevin Lorick, Ph.D. CDRH/OIR/PACB
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.