K141767 · Westridge Laboratories, Inc. · NUC · Mar 9, 2015 · Obstetrics/Gynecology
Device Facts
Record ID
K141767
Device Name
ID MILLENNIUM, ID MOMENTS SILICONE
Applicant
Westridge Laboratories, Inc.
Product Code
NUC · Obstetrics/Gynecology
Decision Date
Mar 9, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
ID® Millennium® is a personal lubricant, for penile and /or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyisoprene, and polyurethane condoms. ID® Moments® Silicone is a personal lubricant, for penile and /or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
Device Story
ID® Millennium® and ID® Moments® Silicone are non-sterile, silicone-based personal lubricants. Intended for over-the-counter use by consumers to moisturize and lubricate during intimate sexual activity. The device is applied topically to the penis or vagina. It is not a contraceptive or spermicide. The lubricant is supplied in plastic/PET bottles with various closures (screw-on, flip-top, pump) or foils. It functions by providing a physical barrier and reducing friction during sexual intercourse. The product is compatible with natural rubber latex, polyisoprene, and polyurethane condoms, ensuring the integrity of these barrier methods during use. The device benefits the patient by enhancing comfort and ease of sexual activity.
Clinical Evidence
No clinical data. Evidence consists of bench testing: biocompatibility (cytotoxicity per ISO 10993-5, sensitization per ISO 10993-10, vaginal irritation/systemic toxicity per ISO 10993-10/11) and condom compatibility per ASTM 7661-10. All results demonstrated safety and compatibility.
Technological Characteristics
Silicone-based liquid lubricant. Formulated as a blend of silicone fluids. Packaged in PET bottles or foils. Tested for viscosity, specific gravity, and microbial limits per USP <61>, <62>, <1111>, and <1112>. Condom compatibility tested per ASTM 7661-10. Non-sterile.
Indications for Use
Indicated for penile and/or vaginal application to moisturize and lubricate, enhance comfort during sexual activity, and supplement natural lubrication. Compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three abstract human profiles facing to the right, stacked one behind the other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 9, 2015
Westridge Laboratories, Inc. Albert Rego, Ph.D. Regulatory Consultant 27001 La Paz Road, Suite 312 Mission Vieio. CA 92691
Re: K141767
Trade/Device Name: ID® Millennium®, ID® Moments® Silicone Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: NUC Dated: February 21, 2015 Received: February 23, 2015
Dear Albert Rego,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
Page 2 - Albert Rego, Ph.D.
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K141767
Device Name ID® Millennium® . ID® Moments® Silicone
Indications for Use (Describe)
ID® Millennium® is a personal lubricant, for penile and /or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
ID® Moments® Silicone is a personal lubricant, for penile and /or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
| Submitted by: | Westridge Laboratories, Inc.<br>1671 E. Saint Andrew Place<br>Santa Ana, CA 92705-4932 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Robert Grant<br>Quality & Regulatory Affairs Manager<br>Tel: (714) 259-9400 Ext. 230<br>Fax: (714) 259-9550<br>Email: bGrant@westridgelabs.com |
| Date Prepared: | |
| Propriety Name: | ID® Millennium Personal Lubricant |
| Proposed Trade Name: | ID® Millennium Personal Lubricant |
| Common Name: | Personal Lubricant |
| Classification Name: | Condoms<br>21 CFR 884.5300 Class II<br>NUC |
| Predicate Device: | Wet Platinum Premium Lubricant®<br>510(K) No. (K130012) |
| Device Description: | ID® Millennium® is non-sterile, an over-the-counter silicone based<br>personal lubricant, formulated to be clear, non-irritating, non-greasy<br>natural and odorless. This device is silicone soluble liquid for use as a<br>personal lubricant. ID® Millennium® Personal Lubricant contains a<br>blend of silicone fluid ingredients similar to ingredients found in the predicate device. ID® Millennium® Personal Lubricant is neither a contraceptive nor a spermicide.<br><br>The device is available in the following variant personal lubricant formulas:<br>ID® Millennium® and ID Moments® Silicone.<br><br>ID® Millennium® Personal Lubricant is provided in plastic/PET bottles with screw-on cap, flip top closure, pump dispensers, and foils. |
| Intended Use: | ID® Millennium® Personal Lubricant is non-sterile, over-the-counter<br>personal lubricant. This device is compatible with natural rubber latex,<br>polyisoprene, and polyurethane condoms. |
{4}------------------------------------------------
# 510(K) Summary
| Indications for Use: | ID® Millennium®:<br>ID® Millennium® is a personal lubricant, for penile and /or vaginal<br>application, intended to moisturize and lubricate, to enhance the ease<br>and comfort of intimate sexual activity and supplement the body's<br>natural lubrication. This product is compatible with natural rubber<br>latex, polyisoprene, and polyurethane condoms. |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | ID® Moments® Silicone:<br>ID® Moments® Silicone is a personal lubricant, for penile and /or vaginal<br>application, intended to moisturize and lubricate, to enhance the ease<br>and comfort of intimate sexual activity and supplement the body's<br>natural lubrication. This product is compatible with natural rubber<br>latex, polyisoprene, and polyurethane condoms. |
| | ID® Millennium® Personal Lubricant and predicate device have the same<br>Indications for Use Statement. |
| Technological<br>Characteristics: | There are no differences in the fundamental technological<br>characteristics of ID® Millennium ® Personal Lubricant and predicate<br>Wet Platinum Premium Lubricant®. ID® Millennium® Personal<br>Lubricant consists mainly of silicone; it contains a blend of silicone fluid<br>ingredients similar to other lubricants currently on the U.S. markets and<br>is substantially equivalent to the predicate device. ID® Millennium®<br>Personal Lubricant formula is neither a contraceptive nor a spermicide. |
## BIOCOMPATIBILITY:
As with the predicate, testing for cytotoxicity, vaginal irritation, sensitization, and systemic toxicity demonstrate that the device is biocompatible.
| Testing Performed | Standard Reference # | Results |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|--------------------------------------------------------------|
| Cytotoxicity (Agar Overlay) | ISO 10993-5:2009 | The device is not cytotoxic |
| ISO Guinea Pig Maximization Sensitization | ISO 10993-10:2010 | The device is non-sensitizing |
| Vaginal Irritation and Systemic Toxicity<br>Study Following Repeated Exposure in<br>Rabbits (The study utilizes FDA<br>recommended hybrid ISO Irritation / Acute<br>Systemic Toxicity test) | ISO 10993-10:2010<br>ISO 10993-11:2006 | The device is not an irritant and not<br>systemically toxic. |
{5}------------------------------------------------
# 510(K) Summary
| Specifications: | ID® Millennium line of products has the following specifications:<br>viscosity, specific gravity, appearance, color and odor, absence of<br>pathogenic organisms (Pseudomonas aeruginosa, Staphylococcus<br>aureus, and Candida albicans) per USP <62>, total aerobic microbial<br>count, and total yeast and mold count per USP <61> and <1111>, and<br>water activity test per USP <1112>. |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Condom<br>Compatibility: | Condom Compatibility Testing was performed in accordance with<br>ASTM 7661-10 on major brand name latex, polyurethane, and<br>polyisoprene condoms.<br><br>The condom compatibility testing demonstrate that ID® Millennium®<br>lubricants are compatible with commercially available male condoms<br>made from natural rubber latex, polyurethane, and polyisoprene<br>materials. |
| Shelf Life Testing: | ID® Millennium Personal Lubricant has three year shelf-life. Retains<br>were obtained and tested per Westridge Laboratories, Inc.<br>specifications. |
| Conclusion: | ID® Millennium® Personal Lubricant has the same intended use and<br>basic technological characteristics as the predicate device and is as safe<br>and effective as its predicate device. Therefore, ID® Millennium<br>Personal Lubricant is substantially equivalent to the predicate device. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.