The Orthofix Galaxy Wrist external fixator is an orthopedic device intended to be used for bone stabilization in trauma and orthopedic procedures, both on adults and pediatric patients as required. The indications for use include: - intra-articular or extra-articular fractures and dislocations of the wrist with or without soft tissue damage; - polytrauma; - carpal dislocations; - unreduced fractures following conservative treatment; - bone-loss or other reconstructive procedures; - infection.
Device Story
Orthofix Galaxy Wrist is an external fixation system for wrist bone stabilization. Device comprises frames, bars, clamps, accessories, and instruments; used with commercially available Orthofix pins. Surgeon secures pins/wires to bone via clamps to create rigid frame. System allows fracture reduction, alignment restoration, and joint distraction/compression or mobilization. Used in clinical/surgical settings for trauma and orthopedic procedures. Provides mechanical stability to facilitate healing in fractures, dislocations, and reconstructive cases. Benefits patient by maintaining bone alignment during recovery.
Clinical Evidence
Bench testing only. No clinical data required. Mechanical testing performed per ASTM F1541. MRI compatibility testing performed per ASTM F2182, ASTM F2213, and ASTM F2052.
Technological Characteristics
External skeletal fixation device. Metallic components. Mechanical fixation via clamps and pins/wires. Tested per ASTM F1541 (mechanical), ASTM F2182 (RF heating), ASTM F2213 (torque), and ASTM F2052 (displacement force).
Indications for Use
Indicated for adult and pediatric patients requiring bone stabilization for intra-articular or extra-articular wrist fractures/dislocations (with/without soft tissue damage), polytrauma, carpal dislocations, unreduced fractures post-conservative treatment, bone-loss/reconstructive procedures, or infection.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 8, 2014
Orthofix Srl % Ms. Cheryl Wagoner Principal Consultant Cheryl Wagoner Consulting Po Box 15729 Wilmington, North Carolina 28408
Re: K141760
Trade/Device Name: Orthofix Galaxy Wrist Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT Dated: November 3, 2014 Received: November 4, 2014
Dear Ms. Wagoner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141760
Device Name Orthofix Galaxy Wrist
Indications for Use (Describe)
The Orthofix Galaxy Wrist external fixator is an orthopedic device intended to be used for bone stabilization in trauma and orthopedic procedures, both on adults and pediatric patients as required.
The indications for use include:
- intra-articular or extra-articular fractures and dislocations of the wrist with or without soft tissue damage:
- polytrauma;
- carpal dislocations;
- unreduced fractures following conservative treatment;
- bone-loss or other reconstructive procedures:
- infection.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Orthofix. The logo consists of a blue abstract shape above the word "ORTHOFIX" in a bold, sans-serif font. The "R" in "ORTHOFIX" has a registered trademark symbol next to it.
| Submitter | Orthofix Srl<br>Via delle Nazioni, 9<br>37012 Bussolengo (VR) - Italy | |
|-------------------------------------------|--------------------------------------------------------------------------------|--------------|
| Telephone | + 39 045 6719.000 | |
| Fax | + 39 045 6719.380 | |
| Contact Person | Gianluca Ricadona<br>Quality & Regulatory Affairs Manager | |
| Address | Via delle Nazioni, 9<br>37012 Bussolengo (VR) - Italy | |
| Telephone | + 39 045 6719.000 | |
| Fax | + 39 045 6719.380 | |
| email | gianlucaricadona@orthofix.com | |
| Date Prepared | November 25, 2014 | |
| Trade Name | Orthofix Galaxy Wrist | |
| Common Name | External Fixation Device and Accessories | |
| Panel Code | Orthopaedics/87 | |
| Classification Name | Single/multiple component metallic bone fixation appliances and<br>accessories | |
| Class | Class II | |
| Regulation Number | 21 CFR 888.3030 | |
| Product Code | KTT | |
| Device Name | 510(k) Number | Manufacturer |
| Synthes Small External<br>Fixation System | K963618 | Synthes |
| Device Name | 510(k) Number | Manufacturer | Description | The Orthofix Galaxy Wrist is a complimentary system for the current<br>Galaxy Fixation system, engineered specifically for the wrist. The<br>primary objective of the Galaxy Wrist is to offer the possibility of<br>distraction/compression and the mobilization of the joint. The device<br>includes various frames, bars, clamps, accessories and instruments. |
|----------------------------------------------------------------|---------------|--------------|-------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Synthes Small External<br>Fixation System | K963618 | Synthes | | The system is designed to be used with commercially available Orthofix<br>pins. The clamps enable the frame to be coupled to bone by securing<br>the pins/wires for the intended use. |
| Stryker Hoffmann II MRI<br>Compact External Fixation<br>System | K053038 | Stryker | | |
| Orthofix Penning Dynamic<br>Wrist Fixator | K955848 | Orthofix | | |
| Orthofix Galaxy Fixation<br>system | K113770 | Orthofix | | |
{4}------------------------------------------------
| | The system allows the surgeon to:<br>- Position screws where the condition of the bone and soft tissues<br>permits<br>- Reduce the fracture in order to restore alignment |
|--------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications and<br>Intended Use | The Orthofix Galaxy Wrist external fixator is an orthopedic device,<br>intended to be used for bone stabilization in trauma and orthopedic<br>procedures, both on adults and pediatric patients as required.<br>The indications for use include:<br>- intra-articular or extra-articular fractures and dislocations of the wrist<br>with or without soft tissue damage;<br>- polytrauma;<br>- carpal dislocations;<br>- unreduced fractures following conservative treatment;<br>- bone-loss or other reconstructive procedures;<br>- infection. |
| Technological<br>Characteristics<br>and Substantial<br>Equivalence | Documentation was provided to demonstrate that the Orthofix Galaxy<br>Wrist is substantially equivalent to the legally marketed Predicates. The<br>Galaxy Wrist is substantially equivalent to the predicate devices in<br>intended use, indications for use, technological characteristics, and<br>labeling. The Galaxy Wrist is comparable to its predicate in size, shape<br>and materials. |
| Performance Data | The potential hazards have been evaluated and controlled through a<br>Risk Management Plan.<br>All testing met or exceeded the requirements as established by the test<br>protocols and applicable standards. A review of the mechanical data<br>indicates that the components of the Subject device are capable of<br>withstanding expected loads without failure. The Subject device was<br>therefore found to be substantially equivalent to the Predicates. Clinical<br>data was not needed to support the safety and effectiveness of the<br>Subject Device.<br>The following mechanical testing was performed:<br>- ASTM F 1541 "Standard Specification and Test Methods for<br>External Skeletal Fixation Devices"<br>MRI compatibility testing was also conducted per:<br>- ASTM F2182 "Standard test method for measurement of radio<br>frequency induced heating near passive implants during magnetic<br>resonance imaging"<br>- ASTM F2213 "Standard Test Method for Measurement of<br>Magnetically Induced Torque on Medical Devices in the Magnetic<br>Resonance Environment"<br>- ASTM F2052 "Standard Test Method for Measurement of<br>Magnetically Induced Displacement Force on Medical Devices in<br>the Magnetic Resonance Environment" |
| Conclusion | Based on the indications for use, technological characteristics,<br>materials, and comparison to predicate devices, the Subject Orthofix<br>Galaxy Wrist has been shown to be substantially equivalent to legally<br>marketed predicate devices. |
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.