Upcera Coloring Liquid (I and II) is a liquid used for the complete or partial coloration of milled Upcera zirconia substructure and anatomy before sintering.
Device Story
Upcera Coloring Liquid (I and II) are water-based, acidic metal salt solutions used by professional dental technicians in a laboratory setting. The liquids are applied to milled zirconia dental substructures or anatomy via immersion or brushing prior to the final sintering process. The coloring agents enable the technician to shade the zirconia to match a patient's natural tooth color. The liquid components (including HCl in Liquid I and Polydextrose in Liquid II) evaporate during the high-temperature sintering process, leaving no residue. The final output is a shaded zirconia restoration that provides aesthetic benefits to the patient. The device is non-sterile and intended for prescription use.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993 standards, including cytotoxicity (ISO 10993-5), oral mucosa irritation (ISO 10993-10), sensitization (ISO 10993-10), subacute/subchronic toxicity (ISO 10993-11), and genotoxicity (ISO 10993-3). Results showed no cytotoxic, irritant, sensitizing, toxic, or genotoxic effects. Heavy metal content (Pb, Cd, Hg, As, Cr) verified against USP limits.
Technological Characteristics
Water-based solution containing inorganic pigments (metal nitrates). Liquid I contains polyethylene glycol and HCl; Liquid II contains polydextrose. Non-sterile. Applied via brush or immersion. Sintering process removes liquid components. Biocompatibility verified per ISO 10993 series.
Indications for Use
Indicated for professional dental technicians to color milled zirconia substructures and anatomy prior to sintering to match natural tooth color in dental patients.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles, suggesting a sense of community and support. The logo is rendered in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 22, 2014
Liaoning Upcera Company Limited C/O Mr. Charles Shen Manton Business and Technology Services 853 Dorchester LN, Unit-B New Milford, NJ 07646
Re: K141723
Trade/Device Name: Upcera Coloring Liquid (I and II) Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: II Product Code: EIH Dated: July 24, 2014 Received: July 24, 2014
Dear Mr. Shen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Shen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runyon DDS, MA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
Section 4: Indication for Use
### Section 4: Indications for Use
510(k) Number (if known): N/A K141723
Device Name: Upcera Coloring Liquid (I and II)
Indications for Use:
Upcera Coloring Liquid (I and II) is a liquid used for the complete or partial coloration of milled Upcera zirconia substructure and anatomy before sintering.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
{3}------------------------------------------------
# Section 5: 510(k) Summary:
This summary of 510k safety and effectiveness information is being submitted In accordance with the requirements of 21CFR 807.92
#### 5.1 Submitter & Foreign Manufacture Identification
Liaoning Upcera Co., Ltd No.122 Xianghuai Road, Economic Development Zone, Benxi, Liaoning, China Tel: (086) 24-5565055 Submitter's FDA Registration Number: 3010582952 www.upcera-dental.com
#### 5.2 Contact Person
Charles Shen Manton Business and Technology Services 853 Dorchester LN, Unit-B New Milford, NJ 08534 Tel: 608-217-9358 Email: cyshen@aol.com
#### 5.3 Date of Summary: May 15, 2014
| 5.4<br>Device Name: | |
|------------------------|-----------------------------------|
| Proprietary Name: | Upcera Coloring Liquid (I and II) |
| Common Name: | Coloring Liquid |
| Classification Name: | Powder, Porcelain |
| Device Classification: | II |
| Regulation Number: | 21 CFR 872.6660 |
| Panel: General | Dental |
| Product Code: | EIH |
#### ર્ડ Predicate Device Information:
K123680, "CopranClor", manufactured by "White Peaks Dental Systems Gmbh (1) & Co. KG." located in Essen, Germany
#### ર્સ્વર્સ Device Description:
Upcera Coloring Liquid (I and II) are water based coloring liquids, which consist of watery, acidic metal salt solutions. They are used for the individual staining of dental zirconia frameworks and restorations prior to the final sintering of the restoration, enabling the user to adjust the restoration to match the natural color of the patient's teeth.
{4}------------------------------------------------
Upcera Coloring Liquid I is provided in 21 different shades, and Upcera Coloring Liquid II is provided in 16 different shades, which are corresponding to patient's closest tooth color. For staining, the zirconia materials have to be immerged into the liquids or to be brushed with the liquids, prior to sintering at high temperatures.
#### 5.7 Intended Use:
Upcera Coloring Liquid (I and II) is a liquid used for the complete or partial coloration of milled Upcera zirconia substructure and anatomy before sintering.
#### 5.8 Summary of Device Testing:
Bench testing was performed to ensure that the Upcera Coloring Liquid (I and II) met its specifications. All tests were verified to meet acceptance criteria. Biocompatibility testing was performed to verify the equivalent safety of the materials that are used.
#### રું. કે Substantial Equivalence Discussion
Upcera Coloring Liquid (I and II) is compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance.
- K123680, "CopranClor", manufactured by "White Peaks Dental Systems Gmbh (1) & Co. KG." located in Essen, Germany
The following table shows similarities and differences of use, design, and material between our device and the predicate devices.
| Description | Our Device (Upcera Coloring Liquid I and<br>II) | Predicate Device (K123680) |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for<br>Use | Upcera Coloring Liquid (I and II) is a<br>liquid used for the complete or partial<br>coloration of milled Upcera zirconia<br>substructure and anatomy before<br>sintering. | CopranColor is intended to be used as<br>the coloring agent for the shading of<br>white zirconia ceramic materials that<br>are to be used by professional dental<br>technicians to fabricate all-ceramic<br>dental restorations for dental patients. |
| Technology | Water based with inorganic pigments | Water based with inorganic pigments |
| Operating<br>Principle | Brush or immerse zirconia ceramic<br>materials with coloring liquid before<br>sintering | Brush or immerse with zirconia<br>ceramic material before sintering |
| Ingredient | Upcera Coloring Liquid I:<br>Water, Polyethylene glycol, HCl | Water, Polyethylene glycol, inorganic<br>salts |
## Table 5.1: Comparison of Intended Use, Design, Material, and Processing
{5}------------------------------------------------
## Section 5: 510k Summary
| | inorganic salts<br>Upcera Coloring Liquid II:<br>Water, Polydextrose, inorganic salts | |
|--------------|---------------------------------------------------------------------------------------|-------------|
| Bottle Size | Various | Various |
| Shade | Various | Various |
| Prescription | Yes | Yes |
| Sterile | Non-sterile | Non-sterile |
Our device is similar to the predicate device in terms of indications for use, technology, and principle of operation. It is similar with predicate device that water and inorganic pigments are the main ingredients.
The indication for use of the subject device is phrased differently than the predicate device for business reasons. However, both share exactly the same expected use: as coloring agent for shading of dental zirconia blanks. Same as predicate device, the subject device is also restricted to professional dental technicians, as illustrated on its product label.
There are some minor differences between Upcera Coloring Liquid I and the predicate device. First, the predicate device does not contain acid (HCl). HCl will be evaporated with no residue left at the sintering step after the coloring liquid is applied to the zirconia ceramics, so it will not cause any effect to the patients. Second, Upcera Coloring Liquid I uses different inorganic pigments (metal nitrates). This will not raise any safety issues as the metal nitrates we use all have long established safety profiles, and we have conducted biocompatibility testing with the products.
There are minor differences between Upcera Coloring Liquid II and the predicate device. First, the predicate device does not contain Polydextrose. Polydextrose is an indigestible synthetic polymer of glucose. It is a food ingredient classified as soluble fiber by the U.S. Food and Drug Administration, so it will not cause any effect to the patients. Second, Upcera Coloring Liquid II uses different inorganic pigments (metal nitrates). This will not raise any safety issues as the metal nitrates we use all have long established safety profiles, and we have conducted biocompatibility testing with the products.
The following table shows similarities and differences of the key performance between our device and the predicate devices. Tests were conducted following the recommended procedures outlined in USP or ISO 10993-1, and results met all relevant requirements in the test standards, and are comparable to the predicate device.
{6}------------------------------------------------
## Section 5: 510k Summary
| Description | Our Device (Upcera Coloring Liquid I and II) | Predicate Device (K123680) |
|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------|------------------------------------|
| Heavy Metals (USP) | $Pb \le 10mg/kg$ , $Cd \le$<br>$3mg/kg$ , $Hg \le 2mg/kg$ , As<br>$\le 2mg/kg$ , $Cr \le 2mg/kg$ | Information not available |
| Cytotoxicity<br>(ISO 10993-5:2009) | No cyteotoxicity effect | |
| Irritation Oral Mucosa<br>Irritation<br>(ISO 10993-10: 2010) | Not a primary oral mucosa<br>irritant under the conditions of<br>the study | |
| Sensitization (ISO<br>10993-10: 2010) | Not a sensitizer under the<br>conditions of the study | Claims biocompatible in<br>general |
| Subacute and<br>Subchronic Toxicity<br>(ISO 10993-11: 2006) | No subacute and subchronic<br>toxic effects observed | |
| Genotoxicity (ISO<br>10993-3: 2003) | No genotoxic effects observed | |
| Table 5.2: Comparison of Performance Testing | | | |
|----------------------------------------------|--|--|--|
|----------------------------------------------|--|--|--|
Therefore, the Upcera Coloring Liquid (I and II) manufactured by "Liaoning Upcera Co., Ltd." meet requirements per internal standard, USP, ISO 10993-1. It is safe and effective, and its performance meets the requirements of its pre-defined acceptance criteria and intended uses. The test results are also comparable to the predicate device.
### 5.10 Substantial Equivalence Conclusion
It has been shown in this 510(k) submission that Upcera Coloring Liquid (I and II) and its predicate devices have the similar indications for use, technology, principle of operation, similar composition, performance, and biocompatibility.
The difference between Upcera Coloring Liquid (I and II) and its predicate device do not raise any question regarding its safety and effectiveness.
Upcera Coloring Liquid (I and II), as designed and manufactured, are as safe and effective as its predicate device, and therefore are substantially equivalent as its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.