K141712 · Halyard Health · FRG · Dec 18, 2014 · General Hospital
Device Facts
Record ID
K141712
Device Name
KIMGUARD ONE STEP STERILIZATION WRAP
Applicant
Halyard Health
Product Code
FRG · General Hospital
Decision Date
Dec 18, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
KIMGUARD ONE-STEP® Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider using: - Sterilucent PSD-85 Hydrogen Peroxide Sterilizer that include: - Lumen Cycle and - Non-Lumen Cycle. KIMGUARD* ONE-STEP* Sterlization Wrap (KC100, KC200, KC300 and KC600) are intended to allow sterilization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used.
Device Story
KIMGUARD ONE-STEP Sterilization Wrap consists of two ultrasonically seamed sheets of nonwoven polypropylene fabric; used by healthcare providers to wrap medical devices prior to sterilization. The wrap acts as a barrier to maintain sterility of the enclosed items post-sterilization. It is designed for use with the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer (Lumen and Non-Lumen cycles). The wrap allows sterilant penetration and aeration during the sterilization process. Post-sterilization, the wrap maintains the sterility of the contents until the package is opened aseptically for clinical use. The device is single-use and provided non-sterile.
Clinical Evidence
Bench testing only. Testing included sterilant penetration, material compatibility/biocompatibility (cytotoxicity, intracutaneous reactivity, sensitization), performance testing post-sterilization, and maintenance of package integrity for 180 days. All tests met acceptance criteria.
Technological Characteristics
Two-sheet ultrasonically seamed nonwoven polypropylene fabric; three-layer SMS (spunbound-meltblown-spunbound) construction. Materials include polypropylene, <2% phthalocyanine blue pigment, <1% titanium dioxide, and <0.008% antistatic treatment. Thermal bonding with round pin, hexagonal, and triangle pattern. Single-use, non-sterile.
Indications for Use
Indicated for healthcare providers to enclose medical devices for sterilization in the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer (Lumen and Non-Lumen cycles) and to maintain sterility of the enclosed devices until use.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which is a stylized representation of a human figure. The symbol is composed of three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 18,2014
Halyard Health (Formerly known as Kimberly-Clark Health Care) C/O Mr. Peter Kalkbrenner Sterilucent, Inc. Director of Engineering 1400 Marshall Street, NE Minneapolis, MN 55413
Re: K141712
Trade/Device Name: KIMGUARD* ONE-STEP* Sterilization Wrap (Models KC100, KC200, KC300, KC400, KC500, and KC600) Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: November 19, 2014 Received: November 20, 2014
Dear Mr. Kalkbrenner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
## Page 2 - Mr. Kalkbrenner
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno DDS, MA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
## K141712
#### Device Name
KIMGUARD* ONE-STEP* Sterlization Wrap (Models KC100, KC300, KC400, KC500, and KC600)
#### Indications for Use (Describe)
KIMGUARD ONE-STEP® Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider using:
- . Sterilucent PSD-85 Hydrogen Peroxide Sterilizer that include:
- Lumen Cycle and o
- Non-Lumen Cycle. o
KIMGUARD* ONE-STEP* Sterlization Wrap (KC100, KC200, KC300 and KC600) are intended to allow sterilization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used.
Test results validated that KIMGUARD ONE-STEP® Sterilization Wraps (KC100, KC300, KC400, KC500, and KC600) allowed sterilization of the enclosed devices by the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer (i.e., both the Lumen and Non-Lumen Cycles). Additionally, the KIMGUARD ONE-STEP® Sterilization Wrap was validated to allow effective aeration under the pre-programmed PSD-85 Sterilization Cycles.
The PSD-85 Lumen Cycle has been validated to sterilize a load of up to ten (10) pounds (combined pouch and wrapped tray load) containing a maximum of ten (10) single channel stainless steel lumens per load with the following dimensions:
- · An inside diameter of 1 mm or larger and a length of 60 mm or shorter;
- An inside diameter of 2 mm or larger and a length of 250 mm or shorter;
· An inside diameter of 3 mm or larger and a length of 350 mm or shorter.
The PSD-85 Non-Lumen Cycle has been validated to sterilize a load of up to 25 pounds (combined pouch and wrapped tray load).
All models of the KIMGUARD ONE-STEP® Sterilization Wrap (KC100, KC300, KC400, KC500, and KC600) have been validated for use with the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer cycles in Table 1.
{3}------------------------------------------------
| TABLE 1: Validated Sterilucent PSD-85 Hydrogen Peroxide Sterilizer Cycle<br>(Note: The instructions provided below are not intended to replace the detailed Instructions for Use provided<br>with the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer.) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PSD-85 Cycle | Intended Loads |
| Lumen | Reusable metal and nonmetal devices including devices with diffusion-restricted spaces such<br>as the hinged portion of forceps and scissors and up to 10 single channel stainless steel<br>lumened devices of the following dimensions per chamber load:<br>An inside diameter of 1 mm or larger and a length of 60 mm or shorter An inside diameter of 2 mm or larger and a length of 250 mm or shorter An inside diameter of 3 mm or larger and a length of 350 mm or shorter (Refer to the PSD-85 User Manual for complete instructions on load(s) and cycle(s), including<br>chamber loading instructions (i.e. 10 lbs per load) |
| Non-Lumen | Non-lumened reusable metal and nonmetal devices including devices with stainless steel<br>diffusion-restricted spaces such as the hinged portion of forceps and scissors.<br>(Refer to the PSD-85 User Manual for complete instructions on load(s) and cycle(s), including<br>chamber loading instructions (i.e. 25 lbs per load)) |
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X | Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### Applicant's Name, Address, Telephone, FAX, Contact Person Halyard Health (formerly known as Kimberly-Clark Health Care) 1400 Holcomb Bridge Road Roswell, GA 30076-2190, USA Establishment Registration Number: 1033422 Contact Name: Thomas Kozma, Director of Regulatory Affairs E-mail: thomas.kozma@hyh.com Phone: 770.587.8393 FAX: 920.225.3408 Date Prepared: November 19, 2014 TRADE NAME: KIMGUARD ONE-STEP* Sterilization Wrap (Models KC100, KC200, KC300, KC400, KC500, and KC600) CLASSIFICATION NAME: Sterilization Wrap Common/Usual Name: Sterilization Wrap FRG Product Code: DEVICE CLASSIFICATION: Class II per 21 CFR §880.6850 Predicate Devices: K112805 - KIMGUARD® ONE-STEP* Sterilization Wrap (Models KC100, KC200, KC300, KC400, KC500, and KC600).
### INDICATIONS FOR USE
KIMGUARD ONE-STEP® Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider using:
- Sterilucent PSD-85 Hydrogen Peroxide Sterilizer that include: ●
- o Lumen Cycle and
- o Non-Lumen Cycle.
KIMGUARD* ONE-STEP* Sterilization Wrap (KC100, KC200, KC300, KC400, KC500 and KC600) are intended to allow sterilization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used.
Test results validated that KIMGUARD ONE-STEP® Sterilization Wraps (KC100, KC200, KC300, KC400, KC500, and KC600) allowed sterilization of the enclosed devices by the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer (i.e., both the Lumen and Non-Lumen Cycles). Additionally, the KIMGUARD ONE-STEP® Sterilization Wrap was validated to allow effective aeration under the pre-programmed PSD-85 Sterilization Cycles.
{5}------------------------------------------------
The PSD-85 Lumen Cycle has been validated to sterilize a load of up to ten (10) pounds (combined pouch and wrapped tray load) containing a maximum of ten (10) single channel stainless steel lumens per load with the following dimensions:
- · An inside diameter of 1 mm or larger and a length of 60 mm or shorter;
- An inside diameter of 2 mm or larger and a length of 250 mm or shorter:
- An inside diameter of 3 mm or larger and a length of 350 mm or shorter.
The PSD-85 Non-Lumen Cycle has been validated to sterilize a load of up to 25 pounds (combined pouch and wrapped tray load).
All models of the KIMGUARD ONE-STEP® Sterilization Wrap (KC100, KC200, KC300, KC400, KC500, and KC600) have been validated for use with the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer cycles in Table 1.
All models of the KIMGUARD ONE-STEP® Sterilization Wrap (KC100, KC200, KC300, KC400, KC500, and KC600) have been validated for use with the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer cycles in Table 1.
| TABLE 1: Validated Sterilucent PSD-85 Hydrogen Peroxide Sterilizer Cycle<br>(Note: The instructions provided below are not intended to replace the detailed<br>Instructions for Use provided with the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer.) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PSD-85 Cycle | Intended Loads |
| Lumen | Reusable metal and nonmetal devices including devices with diffusion-restricted<br>spaces such as the hinged portion of forceps and scissors and up to 10 single<br>channel stainless steel lumened devices of the following dimensions per chamber<br>load:<br>An inside diameter of 1 mm or larger and a length of 60 mm or shorter An inside diameter of 2 mm or larger and a length of 250 mm or shorter An inside diameter of 3 mm or larger and a length of 350 mm or shorter (Refer to the PSD-85 User Manual for complete instructions on load(s) and<br>cycle(s), including chamber loading instructions (i.e. 10 lbs per load) |
| Non-Lumen | Non-lumened reusable metal and nonmetal devices including devices with<br>stainless steel diffusion-restricted spaces such as the hinged portion of forceps<br>and scissors.<br>(Refer to the PSD-85 User Manual for complete instructions on load(s) and<br>cycle(s), including chamber loading instructions (i.e. 25 lbs per load)) |
### DESCRIPTION OF DEVICE
KIMGUARD* ONE-STEP* Sterilization Wrap is comprised of two sheets of KIMGUARD* Sterilization Wrap that is ultrasonically seamed on two edges. This seamed configuration allows for convenient wrapping of an article using two sheets simultaneously.
The sheets of sterilization wrap are square or rectangular fabric produced using a threelayer SMS (spunbound-meltblown-spunbound) process. The wrap fabric is composed of polypropylene with the addition of less than 2% by weight of phthalocyanine blue pigment, less than 1% by weight titanium dioxide pigment, and less than 0.008% by weight of antistatic treatment. The wrap allows a sterilized package to be opened aseptically.
{6}------------------------------------------------
### Substantial Equivalence to Predicate Device
KIMGUARD ONE-STEP* Sterilization Wrap (i.e., subject of this Premarket Notification) is substantially equivalent to the predicate Kimberly-Clark KIMGUARD* ONE-STEP* Sterilization Wraps (K112805) in technology, design, and materials.
The following table compares the subject KIMGUARD* ONE-STEP* Sterilization Wrap to the predicate KIMGUARD* ONE-STEP* Sterilization Wrap.
| Characteristics | Predicate Devices:<br>KIMGUARD* ONE-STEP* Sterilization<br>Wrap (KC100, KC200, KC300, KC400,<br>KC500, & KC600) (K112805) | Proposed Device:<br>KIMGUARD* ONE-STEP* Sterilization Wrap<br>(KC100, KC200, KC300, KC400, KC500, &<br>KC600) |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Kimberly-Clark Corporation | Kimberly Clark Corporation [SAME] |
| Regulation/Product<br>Code | Sterilization Wrap: 880.6850 / FRG | Sterilization Wrap: 880.6850 / FRG<br>[SAME] |
| Indications for Use | The device is intended to be used to enclose<br>another medical device that is to be sterilized by a<br>healthcare provider by the Amsco® V-PRO™ 1<br>Plus Low Temperature Sterilization System's<br>Lumen (identical to the V-PRO™ 1 Cycle) and<br>Non-Lumen Cycles, and the V-PRO™ Low<br>Temperature Sterilization System's Flexible Cycle.<br>The wrap is intended to allow sterilization of the<br>enclosed medical device(s) and also to maintain<br>sterility of the enclosed device(s) until opened<br>within the period of time for which performance<br>data demonstrating maintenance of sterility has<br>been provided. The KIMIGUARD ONE-STEP*<br>Sterilization Wrap was validated to be effectively<br>aerated during the pre-programmed V-PRO™, V-<br>PRO™ 1 Plus, and V-PRO™ Flexible Cycles. | KIMGUARD* ONE-STEP* Sterilization Wrap<br>(KC100, KC200, KC300, KC400, KC500 and<br>KC600) are intended to enclose another medical<br>device that is to be sterilized by a healthcare<br>provider using:<br>• Sterilucent PSD-85 Hydrogen Peroxide<br>Sterilizer that includes:<br>• Lumen Cycle<br>• Non-Lumen Cycle<br>KIMGUARD* ONE-STEP* Sterilization Wrap<br>(KC100, KC200, KC300, KC400, KC500 and<br>KC600) are intended to allow sterilization of the<br>enclosed medical device(s) and also maintain<br>sterility of the enclosed device(s) until used. |
| Sterilization<br>Parameters | Hydrogen Peroxide-based Amsco® V-PRO™<br>1 Plus Low Temperature Sterilization<br>System's Lumen (identical to the V-PRO™ 1<br>Cycle) and Non-Lumen Cycles, and the V-<br>PRO™ Low Temperature Sterilization<br>System's Flexible Cycle | Sterilucent PSD-85 Hydrogen Peroxide<br>Sterilizer that includes:<br>• Lumen Cycle<br>• Non-Lumen Cycle |
| Maintenance of<br>Package Sterility | For models KC100, KC200, KC300, K400,<br>KC500, and KC600 for at least 30 days. | Real-time testing following sterilization using<br>Sterilucent PSD-85 Hydrogen Peroxide<br>Sterilizer supports maintenance of package<br>sterility for 180 days for all models of<br>KIMGUARD* ONE-STEP* Sterilization Wrap. |
| Technology | Tortuous sheet material used to enclose<br>medical devices that are to be sterilized by a<br>healthcare provider to allow sterilization of<br>the enclosed medical device(s) and maintain<br>sterility of the enclosed device(s) until used. | Tortuous sheet material used to enclose<br>medical devices that are to be sterilized by a<br>healthcare provider to allow sterilization of<br>the enclosed medical device(s) and maintain<br>sterility of the enclosed device(s) until used.<br>[SAME] |
| Device Design | Two sheets of nonwoven polypropylene<br>fabric. Each sheet is composed of three<br>thermally- bonded layers consisting of a<br>Meltblown polypropylene layer surrounded by<br>Spunbound polypropylene layers (SMS) | Two sheets of nonwoven polypropylene<br>fabric. Each sheet is composed of three<br>thermally- bonded layers consisting of a<br>Meltblown polypropylene layer surrounded<br>by Spunbound polypropylene layers (SMS)<br>[SAME] |
| Characteristics | Predicate Devices:<br>KIMGUARD* ONE-STEP* Sterilization<br>Wrap (KC100, KC200, KC300, KC400,<br>KC500, & KC600) (K112805) | Proposed Device:<br>KIMGUARD* ONE-STEP* Sterilization Wrap<br>(KC100, KC200, KC300, KC400, KC500, &<br>KC600) |
| Method for bonding<br>SMS layers | Thermal bonding with round pin, hexagonal,<br>triangle bond pattern ("daisy" pattern) | Thermal bonding with round pin, hexagonal,<br>triangle bond pattern ("daisy" pattern)<br>[SAME] |
| Materials | Polypropylene with blue and white<br>pigments | Polypropylene with blue and white<br>pigments<br>[SAME] |
| Distribution | Non-Sterile and<br>Over-the-Counter | Non-Sterile and<br>Over-the-Counter<br>[SAME] |
| Single Use Device | Yes | Yes<br>[SAME] |
Device Comparison Table (Technological. Design, & Materials)
{7}------------------------------------------------
## SUMMARY OF NONCLINICAL TESTS
Performance of KIMGUARD* ONE-STEP* Sterilization Wrap (KC100. KC200. KC300, KC400. KC500, KC600) has been tested in accordance with the applicable requirements recommended in Pre-Market Notification [510(k)] Submissions for Medical Sterilization Packaging Systems in Health Care Facilities; Draft Guidance for Industry and FDA (March 7, 2002). All results of testing met acceptance criteria demonstrating that the KIMGUARD* ONE-STEP* Sterilization Wrap allows sterilization of contents by Sterilucent PSD-85 Hydrogen Peroxide Sterilizer and maintains sterility of contents until used.
| Summary of Testing Performed | Results |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Sterilucent System Sterilant Penetration | Passed |
| Material Compatibility/Biocompatibility - post-sterilization<br>(Cytotoxicity- ISO Elution, ISO Intracutaneous Reactivity,<br>ISO guinea Pig Maximization Sensitization) | Passed |
| Performance Testing - Post-Sterilization | Passed |
| Maintenance of Package Integrity (180 Days) | Passed |
## OVERALL PERFORMANCE CONCLUSIONS
The nonclinical studies demonstrate that the KIMGUARD* ONE-STEP* Sterilization Wrap performs as intended as a sterilization packaging system of medical devices when terminally sterilized in the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer (Lumen and Non-Lumen Cycles). These studies demonstrate that the KIMGUARD* ONE-STEP* Sterilization Wrap met the same criteria as the predicate devices and are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.