FastPack High Sensitivity C-Reactive Protein Immunoassay, High Sensitivity C-Reactive Protein Calibrator Kit, FastPack High Sensitivity C-Reactive Protein Controls, FastPack High Sensitivity C-Reactive Protein Method Verification Kit
K141689 · Qualigen, Inc. · DCK · Jan 7, 2015 · Immunology
Device Facts
Record ID
K141689
Device Name
FastPack High Sensitivity C-Reactive Protein Immunoassay, High Sensitivity C-Reactive Protein Calibrator Kit, FastPack High Sensitivity C-Reactive Protein Controls, FastPack High Sensitivity C-Reactive Protein Method Verification Kit
Applicant
Qualigen, Inc.
Product Code
DCK · Immunology
Decision Date
Jan 7, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5270
Device Class
Class 2
Indications for Use
FastPack® High Sensitivity C-Reactive Protein Immunoassay is to be used for evaluation of conditions thought to be associated with inflammation, in otherwise healthy individuals. The FastPack® High Sensitivity C-Reactive Protein Immunoassay is intended for use with the FastPack® Analyzer. Not intended for Point-of-Care use. FastPack® High Sensitivity C-Reactive Protein Calibrators are used for calibrating the quantitative FastPack® High Sensitivity C-Reactive Protein Immunoassay on the FastPack® Analyzer. FastPack® High Sensitivity C-Reactive Protein Controls are used for quality control of the FastPack® High Sensitivity C-Reactive Protein Immunoassay on the FastPack® Analyzer. FastPack® High Sensitivity C-Reactive Protein Verifiers are used in the quantitative verification of calibration and assay range of the quantitative FastPack® High Sensitivity C-Reactive Protein Immunoassay on the FastPack® Analyzer.
Device Story
FastPack® High Sensitivity C-Reactive Protein (hsCRP) Immunoassay is an automated, in-vitro diagnostic test for quantitative CRP measurement in serum or plasma (EDTA/lithium heparin). The device uses a sandwich immunoassay principle: patient sample binds to monoclonal anti-CRP antibody-alkaline phosphatase (ALP) conjugate and biotinylated anti-CRP antibody. The complex binds to streptavidin-coated paramagnetic particles. After washing, a chemiluminogenic substrate (indoxyl-3-phosphate and lucigenin) is added. ALP dephosphorylates the substrate, triggering oxidation and luminescence. A photomultiplier tube in the FastPack® Analyzer measures relative luminescence units (RLUs), which are proportional to CRP concentration. Used in clinical laboratories by professional staff; results aid clinicians in assessing inflammatory conditions. The system provides quantitative data to support clinical decision-making regarding patient health status.
Clinical Evidence
No clinical trials performed. Evidence consists of analytical bench testing: precision (total CV ≤ 9.0%), linearity (0.063–15.0 mg/L), limit of detection (0.032 mg/L), and interference studies. Method comparison study (n=143) against predicate showed high correlation (R=0.99, slope 0.98). Reference interval study (n=211) established 0.2–11.4 mg/L range.
Technological Characteristics
Sandwich immunoassay using monoclonal anti-CRP antibodies. Detection via chemiluminescence (ALP-labeled antibody, streptavidin-coated paramagnetic particles, indoxyl-3-phosphate/lucigenin substrate). Automated system (FastPack® Analyzer). Operates at 37 ± 0.5 °C. Reagents stored at 2-8 °C. Traceable to ERM-DA474/IFCC reference material.
Indications for Use
Indicated for evaluation of inflammation-associated conditions in otherwise healthy individuals. Not for Point-of-Care use.
Regulatory Classification
Identification
A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.
VITROS Chemistry Products hsCRP Performance Verifier I, II, and III (K041799)
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION MEMORANDUM
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k141689
B. Purpose for Submission:
New assay
C. Measurand:
C-reactive protein (CRP)
D. Type of Test:
Quantitative
E. Applicant:
Qualigen, Inc.
F. Proprietary and Established Names:
FastPack High Sensitivity C-Reactive Protein Immunoassay
FastPack High Sensitivity C-Reactive Protein Calibrator Kit
FastPack High Sensitivity C-Reactive Protein Controls
FastPack High Sensitivity C-Reactive Protein Method Verification Kit
G. Regulatory Information:
1. Regulation section:
21 CFR 866.5270 – C-reactive protein immunological test system
21 CFR 862.1150 – Calibrator
21 CFR 862.1660 – Quality Control Material (Assayed and Unassayed)
2. Classification:
Class II
Class II
Class I, reserved
{1}
3. Product code:
DCK – C-reactive, protein, Antigen, Antiserum
JIT – Calibrator, Secondary
JJX – Quality Control Material (Assayed and Unassayed)
4. Panel:
Clinical Chemistry
H. Intended Use:
1. Intended use(s):
See indication for use.
2. Indication(s) for use:
FastPack High Sensitivity C-Reactive Protein Immunoassay
FastPack High Sensitivity C-Reactive Protein Immunoassay is to be used for evaluation of conditions thought to be associated with inflammation, in otherwise healthy individuals. The FastPack High Sensitivity C-Reactive Protein Immunoassay is intended for use with the FastPack Analyzer. Not intended for Point-of-Care use.
FastPack High Sensitivity C-Reactive Protein Calibrator Kit
FastPack High Sensitivity C-Reactive Protein Calibrators are used for calibrating the quantitative FastPack High Sensitivity C-Reactive Protein Immunoassay on the FastPack Analyzer.
FastPack High Sensitivity C-Reactive Protein Controls
FastPack High Sensitivity C-Reactive Protein Controls are used for quality control of the FastPack High Sensitivity C-Reactive Protein Immunoassay on the FastPack Analyzer.
FastPack High Sensitivity C-Reactive Protein Method Verification Kit
FastPack High Sensitivity C-Reactive Protein Verifiers are used in the quantitative verification of calibration and assay range of the quantitative FastPack High Sensitivity C-Reactive Protein Immunoassay on the FastPack Analyzer.
3. Special conditions for use statement(s):
For prescription use only.
For in vitro diagnostic use only.
Not intended for Point-of-Care use.
{2}
4. Special instrument requirements:
FastPack Analyzer
I. Device Description:
Each FastPack High Sensitivity C-Reactive Protein Immunoassay Kit contains:
- 30 FastPack High Sensitivity C-Reactive Protein Reagent Packs
- Sample diluent A, 32 vials, 2.475 ml each
Each FastPack High Sensitivity C-Reactive Protein Reagent Pack contains:
- Paramagnetic Particles coated with streptavidin, 150 1-JL
- Murine monoclonal anti-CRP antibody covalently linked to alkaline phosphatase and Murine monoclonal anti-CRP antibody covalently linked to biotin, 100 1-JL
- Wash Buffer, 2.0 ml
Tris buffer containing surfactants
- Substrate, 145 1-JL lmmuGlow: lndoxyl-3-phosphate and lucigenin in buffer containing preservatives
The test reagents include:
- Conjugate solution –
- a murine monoclonal anti-C-reactive protein (CRP) monoclonal antibody conjugated to alkaline phosphatase
- a biotinylated murine monoclonal anti-CRP monoclonal antibody
- Solid support streptavidin coated paramagnetic particles
- Substrate solution – ImmuGlow
- Wash solution: Tris buffer containing detergents
- Sample diluent A – the defined protein (bovine serum albumin) matrix provided within the reagent kit for dilution of samples above the assay range
- FastPack High Sensitivity C-Reactive Protein Calibrator Kit
One level of calibrator material is provided ready to use in 1.0 mL/vial. Contains known quantities of human C-reactive protein. Contains 0.09% Sodium azide as preservative.
- FastPack High Sensitivity C-Reactive Protein Control Kit
Two levels of hsCRP controls are provided ready to use in 1.0 mL/vial. hsCRP controls are prepared from human plasma and human proteins. Preservatives (0.09% sodium azide) and stabilizers have been added to maintain product integrity.
- FastPack High Sensitivity C-Reactive Protein Method Verification Kit
Three levels (Low, Mid and High) are provided ready to use in 0.5 mL/vial. hsCRP Verifiers contain components of human origin prepared with protein stabilizers and preservatives to yield predetermined concentrations. Contains 0.09% sodium azide as preservative.
The Fast Pack High Sensitivity C-Reactive Protein Calibrator Kit, Controls and Verification
3
{3}
Kit contain human source material. Each serum/plasma donor unit used in the manufacture of these products has been tested by FDA accepted methods and found non-reactive for the presence of HBsAg and antibody to HIV-1/2, HCV and HIV-1 Ag.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
- Olympus CRP Latex Reagent
- Bio-Rad Laboratories Liquichek™ Cardiac Markers Plus Control
- Ortho-Clinical Diagnostics, Inc. VITRO Chemistry Products hsCRP Performance Verifier, I, II, and III
2. Predicate 510(k) number(s):
- k051564
- k050537
- k041799
3. Comparison with predicate:
| Similarities and Differences between FastPack and Olympus hsCRP Assays | | |
| --- | --- | --- |
| Item | Qualigen FastPack High Sensitivity C-Reactive Protein Immunoassay | Olympus America, Inc. CRP Latex reagent k051564 |
| Intended Use/Indications for Use | High Sensitivity C-Reactive Protein Immunoassay is to be used for evaluation of conditions thought to be associated with inflammation, in otherwise healthy individuals. | Same |
| Sample Type | Serum or plasma (EDTA or lithium heparin) | Same |
| Reagent Storage Temperature | 2-8 °C | Same |
| Testing Environment | Professional use | Same |
| Precision (% CV) | Within-run: ≤ 1.0% Between-run: ≤ 5.2% Total: ≤9.0% | Within-run: ≤ 3.2% Total: ≤ 3.8% |
| Linearity | Assay linear from 0.2 mg/L to 15 mg/L in high sensitivity application | Assay linear from 0.2 - 160 mg/L |
{4}
5
| Interfering Substances/Specificity | No interference found at the below tested concentrations: Bilirubin (conjugated) up to 40 mg/dL Bilirubin (unconjugated) up to 40 mg/dL Hemoglobin up to 750 mg/dL Lipids up to 1000 mg/dL Human serum albumin up to 7.7 g/dL Transferrin up to 567 mg/dL Human IgG up to 2961 μg/mL | No interference found at the below tested concentrations: Bilirubin up to 40 mg/dL Hemoglobin up to 500 mg/dL Intralipid up to 1000 mg/dL |
| --- | --- | --- |
| Expected Values/Reference Intervals | 0.2 – 11.4 mg/L | Cardiac risk assessment categories: Low < 1 mg/L Average 1.0 to 3.0 mg/L High > 3.0 mg/L |
| Methodology | The FastPack High Sensitivity C-Reactive Protein Immunoassay is a paramagnetic particle, chemiluminescent immunoassay employing specific murine monoclonal antibodies. | The Olympus CRP Latex reagent is a turbidimetric assay employing rabbit antibodies coated on latex particles. |
| Assay principle | Chemiluminescence | Turbidimetry |
| Assay procedure | Automated | Automated |
| Assay range | 0.2 - 15 mg/L for High Sensitivity Application | 0.2 - 160 mg/L (provides measurements both for “Normal Application” and “Highly Sensitive Application”) |
| Traceability | Traceable to the ERM-DA474/IFCC reference which serves as the Primary Reference Material | Traceable to an external standard |
| Similarities and Differences between FastPack and Olympus CRP Calibrators | | |
| --- | --- | --- |
| Item | Qualigen FastPack High-Sensitivity C- Reactive Protein Immunoassay | Olympus America, Inc. CRP Latex reagent k051564 |
{5}
| Similarities and Differences between FastPack and Predicate CRP Controls | | |
| --- | --- | --- |
| Item | Qualigen FastPack High-Sensitivity C-Reactive Protein Immunoassay | Bio-Rad Laboratories Liquichek™ Cardiac Markers Plus Control k050537 |
| Intended Use/Indication for Use | For in-vitro diagnostic use to monitor the precision and accuracy of the High-Sensitivity C-Reactive Protein | Same |
| Antigen used in controls | Human CRP | B-type Natriuretic Peptide, Creatine Kinase (Total), C-Reactive Protein, Homocysteine, Digitoxin, N-terminal pro-B-type Natriuretic Peptide, CK-MB, Myoglobin, Troponin I, Troponin T |
| Matrix | Liquid human serum matrix containing a predetermined level of human CRP | Prepared from human serum with added constituents of human and animal origin, preservatives, and stabilizers. The controls are in liquid form. |
| Storage temperature | 2-8 °C | 2-8 °C (Opened), or -20 to -70 °C (Unopened) |
| Number of levels | 2 | 3 |
{6}
| Similarities And Differences between FastPack CRP Verifiers and Predicate Verifiers | | |
| --- | --- | --- |
| Item | Qualigen FastPack High-Sensitivity C-Reactive Protein Immunoassay | Ortho-Clinical Diagnostics, Inc. VITRO Chemistry Products hsCRP Performance Verifier, I, II, and III k041799 |
| Intended Use/Indication for Use | For use in the quantitative verification of calibration and assay range of the High-Sensitivity C-Reactive Protein. | Same |
| Antigen used | Human CRP | Same |
| Storage temperature | 2-8 °C | Same |
| Matrix | Low Verifier: HEPES buffer with Bovine Serum Albumin (BSA) and Detergent Mid and High Verifiers: Liquid human serum matrix containing a predetermined level of human CRP | A base matrix of human plasma proteins to which stabilizers and preservative have been added. |
| Number of levels | 3 | Same |
K. Standard/Guidance Document Referenced (if applicable):
- CLSI EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline, Second Edition, 2005
- CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline 2014
- CLSI EP7-A2: Interference Testing in Clinical Chemistry, Approved Guideline, Second Edition, 2014
- CLSI EP9-A3: Method Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline, Third Edition 2014
- CLSI EP14-A2: Evaluation of Matrix Effects; Approved Guideline, Second Edition 2014
- CSLI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline, Second Edition 2013
- CLSI EP28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline, Third Edition 2014
- CLSI-I/LA21-A2: Clinical Evaluation of Immunoassays; Approves Guideline, Second Edition 2008
{7}
- CLSI-I/LA30-A: Immunoassay Interference by Endogenous Antibodies; Approved Guideline 2009
# L. Test Principle:
The FastPack High Sensitivity C-Reactive Protein Immunoassay employs a Sandwich immunoassay principle. Endogenous CRP in a patient sample, calibrator, control, or verifier is dispensed into a FastPack reagent pack. In the reagent pack, the sample binds with a murine monoclonal anti-CRP antibody covalently linked to alkaline phosphatase (ALP) and a different murine monoclonal anti-CRP antibody linked to biotin. After incubation, immunoreacted complex (Monoclonal anti-CRP antibody-ALP conjugate and anti-CRP antibody linked to biotin reacted with CRP in the sample) is mixed with streptavidin coated paramagnetic particles. After washing steps (using a Tris buffer containing detergents) to separate bound from unbound anti-CRP monoclonal antibody-ALP, a chemiluminogenic substrate mixture is added to the system. This mixture contains indoxyl-3-phosphate, a substrate for ALP, and lucigenin (N,N'- dimethyl-9,9'-biacridinium dinitrate). ALP dephosphorylates indoxyl-3-phosphate to indol-3-ol, which subsequently undergoes oxidation. As a result, lucigenin is reduced to form a dioxetane structure that is cleaved to yield N-methylacridone. This compound produces a sustained luminescent glow following excitation. The raw relative luminescence units (RLUs) generated are measured by a photomultiplier tube in the FastPack Analyzer and are directly proportional to the concentration of CRP in the sample.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
Precision was evaluated following the CLSI EP5-A2 guidance. Six serum patient samples with concentrations of 0.5, 1.0, 2.5, 5.0, 7.5 and $12.5\mathrm{mg / L}$ CRP were tested in duplicate determinations in each of two runs per day on each of two FastPack Reagent lots, two FastPack Analyzers per reagent lot, over a period of 20 days to yield 320 replicate determinations of each sample. A single FastPack calibrator lot was utilized for all runs. Within-run, between-run, and between-day components of variation were calculated and total imprecision using general linear model (GLM). The table below presents the results:
| Sample | Average | Within-run | | Between-run | | Between-day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | CV% | SD | CV% | SD | CV% | SD | CV% |
| 1 | 0.38 | 0.002 | 0.53 | 0.014 | 3.68 | 0.005 | 1.32 | 0.033 | 8.68 |
| 2 | 1.00 | 0.004 | 0.43 | 0.038 | 3.80 | 0.013 | 1.30 | 0.087 | 8.70 |
| 3 | 2.06 | 0.008 | 0.39 | 0.071 | 3.45 | 0.030 | 1.46 | 0.180 | 8.74 |
| 4 | 5.01 | 0.023 | 0.46 | 0.206 | 4.11 | 0.073 | 1.46 | 0.362 | 7.23 |
| 5 | 7.67 | 0.035 | 0.46 | 0.314 | 4.09 | 0.111 | 1.45 | 0.684 | 8.92 |
| 6 | 12.54 | 0.073 | 0.58 | 0.649 | 5.18 | 0.230 | 1.83 | 1.091 | 8.70 |
{8}
b. Linearity/assay reportable range:
Linearity:
This linearity study followed CLSI EP6-A. A high patient sample was intermixed with a low sample to generate 8 concentration levels each tested in triplicate determinations using one FastPack reagent lot on one FastPack analyzer using one FastPack calibrator lot. Linear results were compared to 2nd and 3rd order polynomial fits against a pre-specified allowable error. The linearity range was found to extend from the LOQ (0.063 mg/L) to 15.0 mg/L.
The linear fit yields an equation of $y = 0.00 + 0.97x$ with percent deviation from linearity of $\leq \pm 10\%$.
Samples recovering above the range may be diluted using Sample Diluent A. Dilutional linearity has been validated by the sponsor in their Hook effect study.
Hook effect:
Four patient serum samples with endogenous values of approximately 450, 300, 300, and 400 mg/L, respectively were run neat and diluted 1:100 in Sample Diluent A in triplicate determinations using one FastPack Reagent lot, one calibrator lot on two FastPack Analyzers. The objective of the study was to demonstrate that the Neat values exceeded the upper limit of the assay (15 mg/L).
The results demonstrate that all high level samples recover > 15 mg/L when run neat (range: 21.51 - 35.03 mg/L) and that back calculated actual values range from 309.5 - 442.6 mg/L CRP.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The sponsor bases its traceability process for hsCRP on the International Standard prEN ISO 17511:2003, Metrological Traceability of Values Assigned to Calibrator and Control Material. It is traceable to the ERMDA474/IFCC reference which serves as the Primary Reference Material. The value assignment process for calibrators, controls and verifiers was reviewed and found to be adequate.
Shelf Life and Open Vial Stability for Calibrators:
The FastPack calibrator materials have a shelf life stability of 12 months stored at 2-8°C, and have open-vial stability of 30 days when stored at 2-8°C.
Shelf Life and Open Vial Stability for Controls and Verifiers:
The FastPack controls and verifiers have a shelf life stability of 24 months stored at 2-8°C, and have open-vial stability of 30 days when stored at 2-8°C.
The stability study protocols and acceptance criteria were reviewed and found to be adequate.
9
{9}
10
d. Detection limit:
The limit of blank (LOB), limit of detection (LOD), and limit of quantitation (LOQ) were determined according to CLSI EP17.
LoB:
In this study, the limit of blank was determined from 80 replicate determinations of a blank sample tested on four different FastPack instruments using two reagent lots. The LOB was determined as the upper 95th percentile of the distribution. This value was 0.005 mg/L CRP.
LoD:
The LOD was estimated from 80 replicate determinations of three low samples. Per the CLSI EP17-A guideline, LOD was determined by the following equation:
$$
\mathrm {L O D} = \mathrm {L O B} + (\mathrm {c} \beta * \mathrm {S D S}),
$$
where $\mathrm{c}\beta = 1.645 / (1 - (1 / (4 * \mathrm{f}))$, where $\mathrm{f}$ is the degrees of freedom, and SDS is the pooled standard deviation of the observations. In this study, the LOD was found to be $0.032\mathrm{mg / L}$ CRP.
LoQ:
The LoQ was determined as the lowest sample which provided $< 20\%$ CV. The LOQ was set to $0.063\mathrm{mg / L}$ CRP.
e. Analytical specificity:
Endogenous substance interference was evaluated by spiking two serum samples (1.0 and $6.0\mathrm{mg / L}$ CRP, respectively) with hemoglobin, lipid, bilirubin, albumin, rheumatoid factor (RF), transferrin, human anti-mouse IgG HAMA and Heterophile/HAMA. The samples testing interference with hemoglobin, lipid, bilirubin, albumin, rheumatoid factor (RF), transferrin were analyzed in 5 replicate determinations using one FastPack analyzers and one lot of Fast Pack reagents. The samples testing interference with human anti-mouse IgG HAMA were tested in duplicate and samples testing interference with Heterophile/HAMA in triplicate determinations, respectively. There was no significant interference for the following endogenous substances in the ranges tested below:
Bilirubin up to $40\mathrm{mg / dL}$
Hemoglobin up to $750\mathrm{mg / dL}$
Lipids up to $1000\mathrm{mg / dL}$
Albumin up to $7.7\mathrm{g / dL}$
Rheumatoid factor (RF) up to 1000 IU/mL
Transferrin up to $567\mathrm{mg / dL}$
Human IgG up to $2961~\mu \mathrm{g / mL}$
Human anti-mouse IgG HAMA up to $4\mathrm{mg / dL}$
Heterophile/HAMA up to $3641\mathrm{ng / mlL}$.
{10}
The sponsor also tested potential interference to the exogenous substances in 5 replicate determinations using one FastPack analyzer and one lot of Fast Pack reagents. No significant interference was found when these substances were tested in the concentration ranges indicated below:
L-Ascorbic Acid up to 200 mg/L
Oxaloacetic Acid up to 300 µM
Glutathione up to 300 µM
Isoniazid up to 300 µM
L-DOPA up to 300 µM
The sponsor’s definition of non-significant interference is <10% difference between the spiked and unspiked samples.
The labeling includes a limitation that heterophilic antibodies may interfere with this method.
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
In this method comparison study, one hundred and thirty three human serum samples with CRP values across the measuring range were analyzed on the candidate device and the results were compared to those determined with the predicate device. Deming regression results from the method comparison study are summarized in the table below:
| No. Samples | Range Tested (mgl/L) | Slope (95% CI) | y-intercept (95% CI) | R (95% CI) |
| --- | --- | --- | --- | --- |
| 143 | 0.21 – 15.00 | 0.98 (0.95-1.01) | -0.12 (-0.21 to -0.02) | 0.99 (0.99 - 0.99) |
b. Matrix comparison:
Serum and lithium-heparin plasma
The sponsor performed a matrix comparison study to assess the performance of the assay when different sample types/tubes (serum vs. EDTA plasma and. Lithium Heparin plasma) were tested. Forty one human blood samples were processed to serum, lithium-heparin plasma, or EDTA plasma in parallel. The samples were then
{11}
tested in duplicate determinations each in the FastPack High Sensitivity C-Reactive Protein Immunoassay using two lots of FastPack Reagents and one lot of calibrators. The data analysis was performed using singlicate results. Deming regression results for comparisons of EDTA and lithium-heparin plasma to serum are summarized in tables below:
Comparison of serum and EDTA plasma:
| Parameter | Result |
| --- | --- |
| N compared | 41 |
| Range of observations, mg/L | Serum: 0.33 - 14.72
EDTA Plasma: 0.29 - 14.76 |
| Absolute bias, mg/L | -0.225 |
| % Bias | -6.1 |
| Deming regression results | |
| Slope | 0.94 |
| y-intercept | 0.0 |
| R | 0.984 |
| R² | 0.967 |
Comparison of serum and Lithium-Heparin plasma:
| Parameter | Result |
| --- | --- |
| N compared | 41 |
| Range of observations, mg/L | Serum: 0.33 - 14.72
Lithium-Heparin Plasma: 0.31 - 14.86 |
| Absolute bias, mg/L | 0.002 |
| % Bias | 0.5 |
| Deming regression results | |
| Slope | 1.00 |
| y-intercept | 0.00 |
| R | 0.993 |
| R² | 0.986 |
The studies support the sponsor's claims that EDTA and Lithium-Heparin are acceptable anticoagulants to be used with The FastPack hsCRP Assay.
{12}
13
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
A reference interval study was carried out with serum samples from 211 subjects representing 4 different geographic regions of the United States yielded the results in the table below. The non-parametric 2.5th - 97.5th percentile of 0.2 - 11.4 mg/L provides the reference interval determined from this study, which is in accord with literature¹⁻⁴.
| Observed values | |
| --- | --- |
| Mean (SD) | 3.2 (3.1) mg/L |
| Median (Min - Max) | 1.9 (0.2 - 13.1) mg/L |
| 2.5th - 97.5th percentile | 0.2 - 11.4 mg/L |
Newborns with no evidence of infection have CRP concentrations of < 2 mg/L.⁵
References:
¹ Aziz N, Fahey JL, Detels R, Butch AW. Analytical performance of a highly sensitive C-reactive protein-based immunoassay and the effects of laboratory variables on levels of protein in blood. Clinical and Diagnostic Laboratory Immunology 2003;10:652-7.
² Imhof A, Frohlich M, Loewel H, et al. Distributions of C-reactive protein measured by high-sensitivity assays in apparently healthy men and women from different populations in Europe. Clin Chem 2003;49:669-72.
³ Sennels HP, Jacobsen S, Jensen T, et al. Biological variation and reference intervals for circulating osteopontin, osteoprotegerin, total soluble receptor activator of nuclear factor kappa B ligand and high-sensitivity C-reactive protein. Scand J Clin Lab Invest 2007;67:821-35.
⁴ Charuruks N, Laohajinda B, Rujiwanitgun S, Chaiworaporn M. Reference interval for C-reactive protein and its distribution pattern in thai adults. Circ J 2005;69:339-44.
⁵ Soldin OP, Bierbower LH, Choi JJ, et al. Serum iron, ferritin, transferrin, total iron
{13}
binding capacity, hs-CRP, LDL cholesterol and magnesium in children; new reference intervals using the Dade Dimension Clinical Chemistry System; Clin Chim Acta 2004;342:211-7.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.