JUSHA-C61 LCD Monitor is intended to be used in displaying and viewing digital images for review and analysis by trained medical practitioners. It does not support the display of mammography images for diagnosis.
Device Story
JUSHA-C61 is a 30-inch color TFT LCD monitor for medical image review. Input signals (gray/color) are processed via an internal Color/Gray Auto-identification (CGA) function; the device automatically applies a DICOM Look-Up Table (LUT) to gray areas and a GAMMA2.2 LUT to color areas. Features include ambient brightness adaptation and presence induction for automatic display adjustment. Used in clinical settings by medical practitioners to view high-resolution (3280x2048) images. Output facilitates clinical decision-making by providing calibrated, high-luminance (700cd/m²) visualization of diagnostic data.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including compliance with IEC 60601-1 and IEC 60601-1-2 standards, risk analysis, design reviews, and system verification/validation.
Technological Characteristics
30.0-inch a-Si TFT LCD; 3280x2048 resolution; 700cd/m² luminance; 12-bit color expansion; 7 DICOM LUTs and 1 GAMMA2.2 LUT; LED backlight. Includes ambient brightness sensor and presence induction. Connectivity via DVI-D and DisplayPort. Complies with IEC 60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for trained medical practitioners to display and view digital medical images for review and analysis. Not for mammography diagnostic use.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white and has a simple, clean design. The caduceus symbol is a common symbol for healthcare and medicine.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 1, 2014
Nanjing Jusha Display Technology Co., Ltd. % Ms. Li Le Certification Manager 8A, Block 1, Nanjing International Service Outsourcing Mansion, No. 301 Hanzhongmen Street, Nanjing 210036 CHINA
Re: K141679
Trade/Device Name: Jusha-C61 LCD Monitor Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: PGY Dated: June 24, 2014 Received: June 30, 2014
Dear Ms. Le:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
FDA
for Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
#### 510(k) Number (if known)
### K141679
Device Name
JUSHA-C61 LCD Monitor
Indications for Use (Describe)
JUSHA-C61 LCD Monitor is intended to be used in displaying digital images for review and analysis by trained medical practitioners. It does not support the display of mammography images for diagnosis.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Smh.7)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | May 6, 2014 |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Nanjing Jusha Display Technology Co., Ltd<br>Add: 301, 8F Block A, No.1, Nanjing International Service<br>Outsourcing Mansion, Hanzhongmen Street, Nanjing, 210036 China |
| Contact Person: | Li le<br>Certification Manager<br>Nanjing Jusha Display Technology Co., Ltd<br>Tel: +86-25- 83305050<br>Fax: +86-25- 58783271 |
| Device Trade Name: | JUSHA-C61 LCD Monitor |
| Common/Usual Name: | 6MP Color LCD Monitor |
| Classification Name: Product Code: | Display, Diagnostic Radiology 21CFR 892.2050<br>PGY |
| Predicate Device(s): | EIZO RX650;K134002 |
| Device Description: | JUSHA-C61 LCD Monitor is the display system with the high<br>resolution (3280*2048), high luminance ( $700cd/m^2$ ), 10 bit display<br>colors expanded to 12 bit color expansion technology, 7 DICOM look<br>up table and one GAMMA2.2 look up table inside. In particular,<br>JUSHA-C61 LCD Monitor contains CGA function, it is specially<br>made by JUSHA, it can automatic identify gray and color signals,<br>then gray area calls DICOM LUT and color area calls GAMMA2.2<br>LUT. JUSHA-C61 LCD Monitor has ambient brightness adapting and<br>presence induction system, with these this display can automatic<br>adjustment according to different requirements in order to achieve the<br>best results. |
| | The product is consisted of the following components: |
| | - 30.0 inch, a-Si TFT Liquid Crystal Display |
| | - JUSHA-C61 motherboard/FR-4/REV:2.0 |
| | - JUSHA-C61 LCD Monitor software |
| | - Power Adapter |
| | - Data Cable. |
| | The LCD Monitor is designed, tested, and will be manufactured in<br>accordance with both mandatory and voluntary standards: |
| | IEC 60601-1Medical equipment medical electrical equipment - Part 1:<br>General requirements for basic safety and essential performance<br>2005+CORR.1(2006)+CORR.2(2007) |
| | IEC 60601-1-2 Edition 3:2007, Medical electrical equipment - Part 1-<br>2: General requirements for basic safety and essential performance - |
| | Collateral standard: Electromagnetic compatibility - Requirements<br>and tests. |
| Intended Use: | JUSHA-C61 LCD Monitor is intended to be used in displaying and<br>viewing digital images for review and analysis by trained medical<br>practitioners. It does not support the display of mammography images<br>for diagnosis. |
| Technology: | JUSHA-C61 LCD Monitor is the display system with the high<br>resolution (3280*2048) with electronic capabilities for evaluation of<br>high resolution medical images, high luminance (700cd/m²), 10 bit<br>display colors expanded to 12 bit color expansion technology, 7<br>DICOM look up table and one GAMMA2.2 look up table inside. In<br>particular, JUSHA-C61 LCD Monitor contains CGA function, it is<br>specially made by JUSHA, it can automatic identify gray and color<br>signals, then gray area calls DICOM LUT and color area calls<br>GAMMA2.2 LUT. JUSHA-C61 LCD Monitor has ambient brightness<br>adapting and presence induction system, with these this display can<br>automatic adjustment according to different requirements in order to<br>achieve the best results. |
| Determination of Substantial<br>Equivalence: | Summary of Non-Clinical Tests:<br>The LCD Monitor complies with voluntary standards as following:<br>IEC 60601-1Medical equipment medical electrical equipment - Part 1:<br>General requirements for basic safety and essential performance<br>2005+CORR.1(2006)+CORR.2(2007)<br>IEC 60601-1-2 Edition 3:2007, Medical electrical equipment - Part 1-<br>2: General requirements for basic safety and essential performance -<br>Collateral standard: Electromagnetic compatibility - Requirements<br>and tests.<br>JUSHA-C61 is substantially equivalent to EIZO RX650. JUSHA-<br>C61 employs the maximum resolution values same as that of EIZO<br>RX650. Comparison table of the principal characteristics of 2 devices<br>is shown in the Attachment 1.<br>The following quality assurance measures were applied to the<br>development of the system:<br>Risk Analysis<br>Requirements Reviews<br>Design Reviews<br>Raw materials verification<br>Testing on unit level (Module verification)<br>Integration testing (System verification)<br>Final acceptance testing (Validation)<br>Performance testing (Verification)<br>Safety testing (Verification) |
| | Summary of Clinical Tests:<br>The subject of this premarket submission, LCD Monitor, did not<br>require clinical studies to support substantial equivalence. |
| | The proposed device is Substantially Equivalent (SE) to the predicate<br>device which is US legally market device. Therefore, the subject<br>device is determined as safe and effectiveness. |
| Conclusion: | Nanjing Jusha Display Technology Co., Ltd Considers the JUSHA-<br>C61 LCD Monitor to be as safe, as effective, and performance is<br>substantially equivalent to the predicate device(s). |
{4}------------------------------------------------
{5}------------------------------------------------
# 510(k) Premarket Notification Submission
# Attachment 1:
| Attributes | Predicate Device | Proposed Device |
|--------------------------|---------------------------------------|---------------------------------------|
| Product | EIZO RX650 | JUSHA-C61 LCD Monitor |
| 510(k) Number | K134002 | K141679 |
| Screen technology | 30.0" Color TFT LCD Panel | 30.0" Color TFT LCD Panel |
| Viewing angle (H, V) | Horizontal 176°, Vertical 176° | Horizontal 170°, Vertical 170° |
| Resolution | 2048 x 3280/3280x 2048 | 2048 x 3280/3280x 2048 |
| Display area | 645.5(H) x 403.0 (V) mm | 645.5 (H) x 403.0 (V) mm |
| Contrast Ratio | 1000:1 | 1000:1 |
| Recommended<br>Luminance | 400cd/m² | 400cd/m² |
| Pixel Pitch | 0.197x0.197 mm | 0.197x0.197 mm |
| Backlight | LED | LED |
| Display Colors | 10-bit ,1.07 billion colors | 12-bit, 68.7billion colors |
| Luminance<br>calibration | Built in calibration sensor provided | Built in calibration sensor provided |
| Input signals | DVI standard 1.0,<br>DisplayPort 1.1a | DVI standard 1.0,<br>DisplayPort 1.1a |
| Input terminational | DVI-D Dual Link x2<br>Display Port x2 | DVI-D Dual Link x2<br>Display Port x2 |
| Display controller | Off the shelf | Off the shelf |
| Power Requirement | AC 100~240V 50~60Hz | AC 100~240V 50~60Hz |
{6}------------------------------------------------
| | 510(k) Premarket Notification Submission | | | | |
|--|------------------------------------------|--|--|--|--|
| Power<br>Consumption/Save<br>Mode | 108W/less than 6W | 100W/less than 1.5W |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Power Management | DVI DMPM | DVI DMPM<br>DisplayPort 1.1a |
| USB Ports/standard | 1 upstream (endpoint), 2<br>downstream/ Rev. 2.0 | 1 upstream (endpoint), 2<br>downstream/ Rev. 2.0 |
| Dimensions w/o stand<br>(W x H x D) | Without stand:<br>692mmx466mmx109 mm With<br>stand:<br>692mmx478~643mmx302 mm | Without stand:<br>704mm x 478mm x 81mm<br>With stand:<br>704mm x 611mm x292mm |
| Applicable standard | 1 IEC 60601-1Medical equipment<br>medical electrical equipment - Part 1:<br>General requirements for basic safety<br>and essential performance<br>2005+CORR.1(2006)+CORR.2(2007<br>)<br>2 IEC 60601-1-2 Edition 3:2007,<br>Medical electrical equipment - Part<br>1-2: General requirements for basic<br>safety and essential performance -<br>Collateral standard: Electromagnetic<br>compatibility - Requirements and<br>tests. | 1 IEC 60601-1Medical<br>equipment medical electrical<br>equipment - Part 1: General<br>requirements for basic safety<br>and essential performance<br>2005+CORR.1(2006)+CORR.2(<br>2007)<br>2 IEC 60601-1-2 Edition<br>3:2007, Medical electrical<br>equipment - Part 1-2: General<br>requirements for basic safety<br>and essential performance -<br>Collateral standard:<br>Electromagnetic compatibility -<br>Requirements and tests. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.