EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM

K141678 · Abbott Vascular · NTE · Jul 22, 2014 · Cardiovascular

Device Facts

Record IDK141678
Device NameEMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM
ApplicantAbbott Vascular
Product CodeNTE · Cardiovascular
Decision DateJul 22, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Emboshield NAV Embolic Protection System is indicated for use as a guide wire and embolic protection system to contain and remove embolic material (thrombus / debris) while performing angioplasty and stenting procedures in carotid arteries. The diameter of the artery at the site of the Filtration Element placement should be between 2.5 and 7.0 mm.

Device Story

The Emboshield NAV6 Embolic Protection System is a temporary percutaneous transluminal filtration system used during carotid artery angioplasty and stenting. The system comprises a BareWire Filter Delivery Wire (stainless steel with PTFE coating and radiopaque tip), an RX Delivery Catheter for deployment, a Filtration Element (nylon membrane with nitinol support), and an RX Retrieval Catheter. The device is operated by a physician in a clinical setting. The Filtration Element is deployed distal to the lesion to capture embolic debris released during the procedure. After the intervention, the Retrieval Catheter is used to collapse and remove the filter containing the captured material. This process prevents embolic material from traveling downstream, potentially reducing the risk of ischemic events during carotid interventions.

Clinical Evidence

Bench testing only. Testing included bioburden, wire movement, coating integrity, particulate testing, and comprehensive biocompatibility assessments (cytotoxicity, hemocompatibility, coagulation, systemic toxicity, irritation, sensitization, and material-mediated pyrogenicity).

Technological Characteristics

Materials: Stainless steel wire, PTFE coating, platinum/nickel radiopaque tips, nylon membrane, nitinol support structure. Dimensions: 0.014" wire, 135-139 cm catheter lengths, 5.0 mm or 7.2 mm filtration element sizes. Principle: Mechanical filtration via percutaneous catheter system. Sterilization: Not specified.

Indications for Use

Indicated for patients undergoing angioplasty and stenting procedures in carotid arteries with vessel diameters between 2.5 and 7.0 mm to contain and remove embolic material (thrombus/debris).

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # JUL 2 2 2014 K141678 # 510(k) Summary The 510(k) Summary is submitted in accordance with 21 CFR §807.92 and the requirements of the Safe Medical Device Act (SMDA) of 1990. Submitter's Name Submitter's Address Telephone Fax Contact Person Date of Submission Device Trade Name Device Common Name Device Classification Name Device Classification Product Code Predicate Device Names Summary of Substantial Equivalence Abbott Vascular 3200 Lakeside Drive, Santa Clara, CA 95054 (408) 845-0696 (408) 845-3743 Jessica Kao June 20, 2014 Emboshield NAV6 Embolic Protection System Embolic Protection System Catheter, Carotid, Temporary, for Embolization Capture Class II NTE - Emboshield NAV Embolic Protection System . (K081523) - Emboshield NAV6 Embolic Protection System . (K090665) - Emboshield NAV6 Embolic Protection System . (K110909) - . Emboshield NAV Embolic Protection System (K121015) - The proposed Emboshield NAV6 Embolic . Protection System is identical to the predicate except for the different PTFE coating on the Bare Wire Filter Delivery Wire. - The proposed Emboshield NAV6 Embolic . Protection System has the same intended use as the predicate devices. - Testing results have demonstrated that the . Emboshield NAV® Embolic Protection System is substantially equivalent to the predicate device. {1}------------------------------------------------ ## Device Description: The Emboshield NAV Embolic Protection System (EPS) is a temporary percutaneous transluminal filtration system designed to capture embolic material released during angioplasty and stent procedures within carotid arteries. The system consists of the following components: #### BareWire Filter Delivery Wire: . The BareWire Filter Delivery Wire is a 0.014" PTFE coated stainless steel guidewire with a 3 cm platinum/nickel radiopaque distal tip section. Three (3) Bare Wire designs are available as separately packaged items that offer different support levels and wire lengths, 315 cm and 190 cm. A 190cm BareWire Workhorse is supplied with the Emboshield NAV Embolic Protection System. #### . RX Delivery Catheter: The RX Delivery Catheter usable length is 135 cm. The crossing profile is either 0.0365" or 0.0415", depending on Filtration Element size. A pull handle is used to deploy the loaded Filtration Element from the pod. Two (2) pairs of indicator bands are provided along the catheter shaft; a proximal pair (90 cm and 100 cm from the catheter tip) to indicate the catheter tip position during advancement through the guide catheter; and a distal pair to indicate the proximity of the RX exit port during catheter retraction. A radiopaque marker band is positioned proximal to the pod. ### · Filtration Element: The Filtration Element consists of a nylon membrane with an internal radiopaque nitinol support structure . There are two proximal triangular entry ports and multiple 120 micron distal perfusion pores. There is also a proximal and a distal marker band. The Filtration Element is available in two sizes; small (5.0 mm) to treat vessel diameters of 2.5 - 4.8 mm and large (7.2 mm) to treat vessel diameters of 4.0 to 7.0 mm. #### RX Retrieval Catheter: . The RX Retrieval Catheter has a usable length of 139 cm and a molded expansile distal tip with a maximum outer diameter of 0.067''. A handle is situated at the proximal end. Two pairs of marker bands indicate the position of the Retrieval Catheter RX guidewire exit port and catheter tip. ### Indication for Use: The Emboshield NAV® Embolic Protection System is indicated for use as a guide wire and embolic protection system to contain and remove embolic material (thrombus / debris) while performing angioplasty and stenting procedures in carotid arteries. The diameter of the artery at the Filtration Element placement should be between 2.5 and 7.0 mm. {2}------------------------------------------------ ## Technological Characteristics: Comparisons to the predicate devices show that the technological characteristics such as product performance, design and intended use are substantially equivalent to the current marketed predicate devices. # Performance Data: Performance testing was successfully completed on the Emboshield NAV6 Embolic Protection System. The following tests were conducted: - Bioburden Testing . - . Wire Movement - . Coating Integrity - Particulate Testing ● - Vendor Operational Qualification and Performance Qualification . - Coating Adhesion Testing - o Friction Testing - Coating Thickness - o Coating Appearance - Biocompatibility ● - o Cytotoxicity - o Hemocompatibility - Hemolysis, Direct and Indirect . - . Complement Activation Test - Coagulation (PT and PTT) o - Acute Systemic Toxicity o - Intracutaneous Toxicity (Irritation) o - o Sensitization - Material-Mediated Pyrogenicity o {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird with three curved lines forming its body and wings. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 22, 2014 Abbott Vascular Ms. Jessica Kao Regulatory Affairs Specialist 3200 Lakeside Drive Santa Clara, CA 95054 Re: K141678 Trade/Device Name: Emboshield NAV6 Embolic Protection System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NTE Dated: June 20, 2014 Received: June 23, 2014 Dear Ms. Kao: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Ms. Jessica Kao If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Nicole Subrahim -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### Indications for Use K141678 510(k) Number (if known): Emboshield NAV6 Embolic Protection System Device Names: The Emboshield NAV Embolic Protection System is indicated for Indications for Use: use as a guide wire and embolic protection system to contain and remove embolic material (thrombus / debris) while performing angioplasty and stenting procedures in carotid arteries. The diameter of the artery at the site of the Filtration Element placement should be between 2.5 and 7.0 mm. Prescription Use X (Per 21 CFR 801 Subpart D) OR Over-The-Counter (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...