The Stryker IVAS Bone Biopsy Kit can be used as a biopsy tool to remove a sample of bone tissue from a vertebral body for diagnostic purposes using a coring, cutting or aspiration technique.
Device Story
Disposable, single-use, sterile bone biopsy kit; includes coaxial biopsy needle, obturator, and syringe. Used with separately sold access cannula and stylet. Operates manually via percutaneous placement under fluoroscopic guidance. Biopsy needle removes bone tissue samples from vertebral bodies using coring, cutting, or aspiration techniques. Designed for use by clinicians in a clinical setting. Provides tissue samples for diagnostic evaluation; assists in clinical decision-making regarding vertebral pathology. Benefits patient by enabling minimally invasive tissue acquisition for diagnosis.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Stainless steel biopsy needle; manual operation; percutaneous access; fluoroscopic visualization. Sterilized via Gamma irradiation (VDmax25, 25 kGy, SAL 10^-6) per ISO 11137-1/2. 8-gauge diameter; non-swaged distal tip. Single-use, disposable. No software or electronic components.
Indications for Use
Indicated for patients requiring bone tissue biopsy from a vertebral body for diagnostic purposes.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
Stryker Bone and Vertebral Body Biopsy Kit (K032943)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three faces in profile, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 4, 2014
Stryker Instruments Ms. Brittney M. Larsen Senior Regulatory Affairs Representative 4100 E. Milham Avenue Kalamazoo, Michigan 49001
Re: K141673
Trade/Device Name: Stryker iVAS 8 Gauge Bone Biopsy Kit Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology-biopsy instrument Regulatory Class: Class II Product Code: KNW Dated: July 7, 2014 Received: August 8, 2014
Dear Ms. Larsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
for
Sincerely yours,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S.
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K141673
Device Name Stryker® iVAS® 8 Gauge Bone Biopsy Kit
### Indications for Use (Describe)
The Stryker IVAS Bone Biopsy Kit can be used as a biopsy tool to remove a sample of bone tissue from a vertebral body for diagnostic purposes using a coring, cutting or aspiration technique.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
4100 E. Milham Avenue Kalamazoo, MI 49001
t: 269 323 7700 f: 269 324 5412 www.stryker.com
Image /page/3/Picture/1 description: The image shows the Stryker Instruments logo. The word "stryker" is in bold, sans-serif font, with a registered trademark symbol to the right. Below the word "stryker" is a horizontal line, and below that is the word "Instruments" in a smaller, sans-serif font.
# 510(k) Summary
### I. Contact Details
| a. | 510(k) Owner: | Stryker Instruments<br>4100 E. Milham Avenue<br>Kalamazoo, Michigan 49001<br>USA<br>Ph: +1-269-323-7700<br>Fax: +1-269-324-5412 |
|-----|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| b. | FDA Establishment<br>Registration Number: | 1811755 |
| c. | Contact Person: | Brittney M. Larsen<br>Ph: +1-269-389-5293<br>Fax: +1-269-389-5412<br>Brittney.Larsen@Stryker.com |
| d. | Date Submitted: | June 20, 2014 |
| II. | Device Name | |
| a. | Trade Name: | Stryker® iVAS® 8 Gauge Bone Biopsy Kit |
| b. | Common Name: | Vertebral Bone Biopsy Needles |
| c. | Classification<br>Name: | Gastroenterology-urology biopsy instrument<br>(21 CFR 876.1075, Product code KNW) |
| d. | Classification: | II |
### III. Legally Marketed Predicate Device(s)
| 510(k) Number | Product Code | Trade Name | Manufacturer |
|---------------|--------------|-----------------------------------------------|------------------------|
| K032943 | KNW | Stryker Bone and Vertebral<br>Body Biopsy Kit | Stryker<br>Instruments |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Stryker Instruments. The word "stryker" is in bold, sans-serif font, with a registered trademark symbol to the right of the word. Below the word "stryker" is a horizontal line, and below that is the word "Instruments" in a smaller, sans-serif font.
4100 E. Milham Avenue Kalamazoo. MI 49001 t: 269 323 7700 f: 269 324 5412 www.stryker.com
### -Purpose of the Special 510(k) Submission
Stryker submits this Special 510(k): Device Modification to request clearance for a modification to our Stryker Bone and Vertebral Body Biopsy Kit(s). The predicate device(s) are currently cleared for 10-13gauge (G) sizes. The subject modification is to add an 8-G size product offering. which is outside the currently cleared size range. This device modification does not change the intended use, indications for use or the fundamental scientific technology of the predicate device(s).
### II. Device Description
The Stryker® Inflatable Vertebral Augmentation System (iVAS®) 8 Gauge Bone Biopsy Kit mav be used as a biopsy tool to remove a sample of bone tissue from a vertebral body for diagnostic purposes using a coring, cutting or aspiration technique.
The Stryker® iVAS® 8 qauge bone biopsy kit is a disposable device, intended for single-use, individually packaged and provided sterile. The biopsy device will be sold in a kit consisting of a coaxial biopsy needle, obturator and syringe. The biopsy kit is used with access cannula and stylet accessories of compatible size. The access cannula and stylet are sold separately and are Class I exempt under 21 CFR 888.4200, product code OAR (Injector, vertebroplasty (does not contain cement).
### III. Indications for use
The Stryker iVAS Bone Biopsy Kit can be used as a biopsy tool to remove a sample of bone tissue from a vertebral body for diagnostic purposes using a coring, cutting or aspiration technique.
### IV. Substantial Equivalence Comparison
The Stryker® iVAS® 8 Gauge Bone Biopsy Kit is substantially equivalent to the Stryker Bone and Vertebral Body Biopsy Kit predicate device(s) (K032943). Stryker claims this equivalence because the Stryker® iVAS® 8 Gauge Bone Biopsy Kit has an equivalent intended use, mechanism for use, and mode of action as compared to the predicate device(s). The differences between the Stryker® iVAS® 8 Gauge Bone Biopsy Kit and the predicate device(s) have been identified and explained in Table 2, the substantial equivalence comparison matrix.
{5}------------------------------------------------
4100 E. Milham Avenue
Kalamazoo, MI 49001
t: 269 323 7700 f: 269 324 5412
www.stryker.com
Image /page/5/Picture/1 description: The image shows the Stryker Instruments logo. The word "stryker" is written vertically in a bold, sans-serif font. To the right of the word "stryker" is the word "Instruments" written vertically in a smaller, sans-serif font.
Table 2. Substantial equivalence comparison matrix
| Table 2. Substantial equivalence comparison matrix | | | |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description | Stryker Bone and Vertebral Body Biopsy Kit(s) (Predicate) | Stryker® IVAS® 8 Gauge Bone Biopsy Kit (Subject) | Explanation of Differences |
| General Information | | | |
| Classification | Class II | Class II | Identical |
| Regulation | 21 CFR 876.1075, Gastroenterology-<br>urology biopsy instrument | 21 CFR 876.1075, Gastroenterology-<br>urology biopsy instrument | Identical |
| Product code | KNW, Instrument, Biopsy | KNW, Instrument, Biopsy | Identical |
| Review panel | 78 Gastroenterology/Urology | 78 Gastroenterology/Urology | Identical |
| Indications for use | Stryker's Bone and Vertebral Body Biopsy Kits can be used as a biopsy tool to remove a sample of bone tissue from a vertebral body for diagnostic purposes using a coring, cutting or aspiration technique. | The Stryker iVAS Bone Biopsy Kit can be used as a biopsy tool to remove a sample of bone tissue from a vertebral body for diagnostic purposes using a coring, cutting or aspiration technique. | Identical |
| Use | Single-use | Single-use | Identical |
| Anatomical sites | Bone and vertebral body | Bone and vertebral body | Identical |
| Visualization technique | Fluoroscopy | Fluoroscopy | Identical |
| Method of placement | Percutaneous | Percutaneous | Identical |
| Mechanics of action | Manual | Manual | Identical |
| Mode of action | Suction | Suction | Identical |
| Shelf life | 2 years | 1 year | Similar – the intent is to extend the shelf life for the subject device upon successful completion of aging studies. |
| Description | Stryker Bone and Vertebral Body<br>Biopsy Kit(s)<br>(Predicate) | Stryker® IVAS® 8 Gauge Bone<br>Biopsy Kit<br>(Subject) | Difference |
| Size Information | | | |
| Penetration<br>depth | 5 - 7.5-in. | 5 - 7.5-in. | Identical |
| Biopsy needle<br>and Obturator | Compatible with 10, 11 and 13-gauge<br>access cannulae/stylets | Compatible with 8-gauge access<br>cannula/stylet | The subject device biopsy needle<br>diameter increased to accommodate<br>user preference for an 8-gauge biopsy<br>system. Verification and validation<br>testing have been conducted to ensure<br>there are no new concerns of safety<br>and effectiveness. |
| Patient Contacting Material Information | | | |
| Biopsy needle | Stainless steel | Stainless steel | Identical |
| Performance Information | | | |
| Tensile force<br>(Pull test) | Withstand < 15-lbs. of force | Withstand < 15-lbs. of force | Identical |
| Biopsy needle<br>tip | Swaged | Not swaged | Subject biopsy needle does not exhibit<br>a swaged, tapered, distal tip in order to<br>increase sample size. Verification and<br>validation testing have been conducted<br>to ensure there are no new concerns of<br>safety and effectiveness. |
| Sterilization Information | | | |
| Sterilization<br>method | Gamma irradiation, VDmax25 | Gamma irradiation, VDmax25 | Identical |
| Sterilization<br>assurance level<br>(SAL) | 10-6 | 10-6 | Identical |
| Minimum<br>sterilization<br>dose | 25-kilograys (kGy) | 25-kilograys (kGy) | Identical |
{6}------------------------------------------------
4100 E. Milham Avenue
Kalamazoo, Ml 49001
t: 269 323 7700 f: 269 324 5412
www.stryker.com
Instrument
# Table 2 continued. Substantial equivalence comparison matr
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Stryker Instruments. The word "Stryker" is in a bold, sans-serif font, with a registered trademark symbol to the right of the word. Below the word "Stryker" is a horizontal line, and below that is the word "Instruments" in a smaller, sans-serif font.
4100 E. Milham Avenue Kalamazoo. MI 49001 t: 269 323 7700 f: 269 324 5412 www.stryker.com
### V. Non-clinical Testing
A risk Management File (RMF), in compliance with EN ISO 14971:2012, Medical devices - Application of risk management to medical devices, was completed to assess the impact of the subject device modification. In accordance with the risk assessment, verification and validation testing was performed in an effort to mitigate risk where possible. Testing performed includes: functional, packaging integrity and sterilization testing. See Table 3, for a summary of the non-clinical testing and acceptance criteria used to evaluate the subject device modification. Stryker has determined that the device modification to the predicate device(s) raises no new questions of safety or effectiveness.
| Non-Clinical Testing | Acceptance Criteria |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Impact testing | Biopsy needle to withstand impact loads of at least 25 cycles at 2-in-lb. |
| Tensile force testing | Biopsy needle and obturator to withstand a force measuring less than 15-lbs. |
| Non-axial torque testing | Biopsy needle handle to withstand a force measuring less than 25-in-lb. |
| Axial torque testing | Biopsy needle handle to withstand a force measuring less than 3.25-in-lb. |
| Functional testing | Non-swaged biopsy needle must retain an equivalent or greater average subjective size measurement than the swaged biopsy needle. |
| Sterilization testing | Sterilization method is VDmax25, in compliance with the ISO 11137-1 and ISO11137-2 standards, acceptance criteria includes:<br>Bioburden less than 1000 CFU Less than 2 positive tests of sterility in product verification dose experiment SAL 10-6 |
### Table 3. Non-clinical testing and acceptance criteria summary
### VI. Clinical Testing
No clinical testing was deemed necessary for this 510(k).
### VII. Conclusion / Substantial Equivalence (SE) Rationale
The Stryker® iVAS® 8 Gauge Bone Biopsy Kit is substantially equivalent in intended use, indications for use, technological characteristics, safety and effectiveness to the previously cleared Stryker Bone and Vertebral Body Biopsy
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Stryker Instruments. The word "stryker" is in bold, black letters, with a registered trademark symbol to the right of the word. Below the word "stryker" is a horizontal line, and below that is the word "Instruments" in a smaller font.
4100 E. Milham Avenue Kalamazoo, MI 49001 t: 269 323 7700 f: 269 324 5412 www.stryker.com
### Conclusion/Substantial Equivalence (SE) Rationale continued.
Kit. The predicate device(s) have the same fundamental scientific technology, basic design, functional characteristics and applications. The device modification does not introduce any new concerns of safety or effectiveness. Therefore, the Stryker® iVAS® 8 Gauge Bone Biopsy Kit is substantially equivalent to the existing predicate device(s).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.