K141657 · Thd Spa · JAF · Feb 24, 2015 · Radiology
Device Facts
Record ID
K141657
Device Name
THD REVOLUTION
Applicant
Thd Spa
Product Code
JAF · Radiology
Decision Date
Feb 24, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1540
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K141657 · Feb 24, 2015
THD REVOLUTION
Thd Spa
Published clinical literature (Giordano et al., 2014; Morrett, 2012); Clinical patient cohorts
The sponsor utilized published clinical literature reporting on patient cohorts treated with the THD technique to demonstrate the safety and effectiveness of the THD Revolution device for the treatment of Grade IV hemorrhoids.
Literature review; Clinical cohort; Hemorrhoidal disease; Surgical treatment
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Transanal dearterialization with targeted mucopexy is effective for advanced haemorroids (Giordano et al., 2014); Clinical cohort study
31 patients with IV grade hemorrhoids; Sample Size: 31
Not applicable for this study
Intra-operative complications, recurrence
Transanal haemorrhoidal dearterialization as alternative surgical treatment to hemorrhoidal disease: initial report (Morrett, 2012); Clinical cohort study
250 patients with grade II-III and IV hemorrhoids; Sample Size: 250
The THD Revolution is intended to be used for the surgical treatment of hemorrhoidal disease. This treatment is based on the Doppler-guided Transanal Haemorrhoidal Dearterialization technique. The technique consists of locating and ligating the terminal branches of the superior haemorrhoidal artery using the THD Revolution device and the compatible single-use THD kits. THD Revolution and its accessories may exclusively be used by specialist physicians for ligation of the terminal branches of the superior haemorroidal artery in hospital facilities (or suitable aseptic outpatient facilities). Furthermore, by using the optic fiber accessory of the THD Revolution, it can be used in modality "Light Source", to support the different THD diagnostic devices.
Device Story
THD Revolution is an 8 MHz continuous wave (CW) Doppler detector used for surgical treatment of hemorrhoids. Input: ultrasonic signals from terminal branches of superior hemorrhoidal artery via Doppler probe. Processing: DSP microprocessor filters noise and enhances acoustic signal; device also functions as a 50W LED light source for surgical field illumination. Output: audible Doppler signals via loudspeaker; illumination via optic fiber. Used in hospital/outpatient settings by specialist physicians. Physician uses audible feedback to locate arteries for ligation and visual feedback for surgical guidance. Benefits: improved signal clarity via DSP, enhanced illumination via LED, and effective treatment of grade II-IV hemorrhoids.
Clinical Evidence
Clinical evidence includes two studies. 1) Prospective study (n=31) of grade IV hemorrhoids treated with THD Revolution; results showed no intra-operative complications and 1 recurrence. 2) Report (n=250) of grade II-IV hemorrhoids; results showed no intra-operative complications, low post-operative pain (73% reported 0-1 on AVS), and high patient satisfaction (8-10 on AVS).
Indicated for surgical treatment of hemorrhoidal disease (grades II, III, and IV) via Doppler-guided transanal hemorrhoidal dearterialization. For use by specialist physicians in hospital or aseptic outpatient facilities.
Regulatory Classification
Identification
A nonfetal ultrasonic monitor is a device that projects a continuous high-frequency sound wave into body tissue other than a fetus to determine frequency changes (doppler shift) in the reflected wave and is intended for use in the investigation of nonfetal blood flow and other nonfetal body tissues in motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing right, with a bird-like element above them, possibly representing an eagle or another bird of prey.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 24, 2015
THD SpA % Mr. Maurizio Pantaleoni Isemed srl Via Argentina Altobelli Bonetti, 3 40026, Imola BO Italy
Re: K141657
Trade/Device Name: THD Revolution Regulation Number: 21 CFR 892.1540 Regulation Name: Nonfetal ultrasonic monitor Regulatory Class: Class II Product Code: JAF Dated: December 12, 2014 Received: December 12, 2014
Dear Mr. Pantaleoni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141657
Device Name THD Revolution
### Indications for Use (Describe)
The THD Revolution is intended to be used for the surgical treatment of hemorrhoidal disease. This treatment is based on the Doppler-guided Transanal Haemorrhoidal Dearterialization technique.
The technique consists of locating and ligating the terminal branches of the superior haemorrhoidal artery using the THD Revolution device and the compatible single-use THD kits.
THD Revolution and its accessories may exclusively be used by specialist physicians for ligation of the terminal branches of the superior haemorroidal artery in hospital facilities (or suitable aseptic outpatient facilities). Furthermore, by using the optic fiber accessory of the THD Revolution, it can be used in modality "Light Source", to support the different THD diagnostic devices.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### THD Revolution
THD S.p.A. TRADITIONAL 510(K) Notification K141657
# 510(k) Summary for the THD Revolution
This 510(k) Summary is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.
# 2.1. General Information
Submitter:
THD S.p.A. Via dell'Industria, 1 42015 - Correggio (RE) Italy
Establishment Registration Number: 3006680097
Contact person:
Maurizio Pantaleoni ISEMED srl Via A. Bonetti 3/a 40026 Imola (BO) Italy Mob.phone: +39-348-4435155 Telephone: +39-0542-683803 Fax: +39-0542-698456 Email: regulatory@isemed.eu
Summary Preparation Date:
December 11, 2014
- 2.2. Names
Device Name: Classification Name: Product Code: Regulation number:
THD Revolution Nonfetal Ultrasonic monitor JAF 892.1540
# 2.3. Predicate Devices
The THD Revolution is equivalent to the following devices:
| Applicant | Device name | 510(k) Number |
|-------------|---------------|---------------|
| G.F. S.R.L. | THD | K070815 |
| THD SpA | THD Slide | K081429 |
| THD SpA | THD Slide One | K090009 |
{4}------------------------------------------------
### 2.4. Device Description
Identically to the predicate devices K090009, K081429 e K070815, the THD Revolution is a 8 MHz continuous wave (CW) Doppler detector with a loudspeaker and a light source.
Further, as the predicate device, the THD Revolution doppler is used with dedicated accessories (optical fiber and pneumatic foot pedal) and a dedicated kit containing a proctoscope, needle holder, knot tightener, sutures, doppler probe and Surgy proctoscope in order to allow the surgical operation through the Transanal Hemorroidal Dearterialization (THD) technique. In particular, the THD Revolution is compatible with any of the following kits:
- THD Slide,
- THD Slide One,
- THD Easy,
- THD Easy one,
- THD Basic,
- THD Basic plus.
The THD Slide and THD Slide one kits have been previously cleared (510(k) number: K081429 and K090009 respectively) and there have been no changes to the kits since their clearance. All the other kits consist of different combinations of one or more components characterizing the cleared THD Slide kit (K081429), THD Slide One kit (K090009). THD kit (K070815) or the Surgy proctoscope (K091490).
Furthermore, the THD Revolution can be used in "Light Source" mode to allow the doctor to perform inspections by using the optic fiber accessory.
### 2.5. Indications for Use
The THD Revolution is intended to be used for the surgical treatment of hemorrhoidal disease. This treatment is based on the Doppler-guided Transanal Haemorrhoidal Dearterialization technique.
The technique consists of locating and ligating the terminal branches of the superior haemorrhoidal artery using the THD Revolution device and the compatible single-use THD kits.
THD Revolution and its accessories may exclusively be used by specialist physicians for ligation of the terminal branches of the superior haemorroidal artery in hospital facilities (or suitable aseptic outpatient facilities). Furthermore, by using the optic fiber accessory of the THD Revolution, it can be used in modality "Light Source", to support the different THD diagnostic devices.
Concerning the surgical treatment of the hemorrhoids through the THD technique, the THD Revolution is equivalent to the predicate devices K090009, K081429 and K070815.
{5}------------------------------------------------
The "Light Source" mode represents an additional function of the THD Revolution allowing illumination and observation. In such modality the THD Revolution supports the THD family of diagnostic devices already cleared in K091490.
### 2.6. Technological characteristics
Identically to the predicate devices K090009, K081429 and K070815, the THD Revolution consists of a doppler device and the following accessories:
- THD Optical fiber
- Pneumatic foot pedal ●
- Back-up doppler probe ●
- Kit
#### ● Doppler device
The THD Revolution doppler represents an evolution of the original Doppler called THD Evolution which was cleared together with its specific kit under the 510K numbers: K07815 (THD kit), K081429 (THD Slide kit) and K090009 (THD Slide one kit).
From a technical and technological point of view, the THD Revolution Doppler has some new features compared to the predicate devices:
- Design: The THD Revolution consists of a double thermoplastic shell containing . an 8 MHz continuous wave (CW) Doppler detector with loudspeaker and a light source.
These characteristics are identical to the predicate devices K07815, K081429 and K090009.
- . General Wiring: Schematically the THD Revolution consists of: 1) a power input module; 2) two power management suppliers (one supplying the LED and the other one supplying all the other parts of the device) working within the range 115-230V and compliant with IEC 60601-1:2005; 3) a Doppler module controlling the Doppler functions (ultrasound signal); 4) a main board consisting of a DSP microprocessor which manages all the utilities (Doppler module, loudspeaker, Led on/off and Led light intensity), 5) a Push buttons board with a display representing the user interface; 6) a LED module consisting of a 50W LED emitter, a thermal switch, a fan system and a sensor detecting the optical fiber; 7) a loudspeaker with the electronic circuitry installed on the Doppler module.
The general wiring of the predicate devices K07815, K081429 and K090009 is identically to the THD Revolution with regards to: 1) the power input module; 2) the Doppler module; 3) the Push buttons board and 4) the loudspeaker.
The predicate devices K07815, K081429 and K090009 differ from the THD Revolution with regards to the following aspects: 1) the predicates devices have a voltage selector for the manual regulation of the electric power (115V or 230V) and consequently they are provided with a power supply transformer instead of the two power managements boards characterizing the THD Revolution. Identically to the THD Revolution the transformer guarantees the compliance with IEC 60601-1:2005; 2) the predicate devices do not have a main board with DSP microprocessor since all the signals are processed through analogic circuitry. Therefore, the
{6}------------------------------------------------
predicate devices have a power supply board, which supplies all the utilities (Doppler module, loudspeaker and light); 3) the light source of the predicate devices is an analogic lamp and not a LED.
- Power supply: The THD Revolution is characterized by a power supply of 115-● 230Vac +/-10% (Voltage); 50-60Hz (Frequency) and 100VA (Absorption). These characteristics are identical to the predicate devices K07815, K081429 and K090009.
- Light source: Both the THD Revolution and the predicate device have an internal light source to illuminate the surgical field during the procedure.
However, the THD Revolution's light source is a LED (50W) while the predicate devices light source is a tungsten halogen lamp (50W). The LED technology improves the illumination of the surgical area and allows to minimize the maintenance operations that by contrary are required to change the lamp of the predicate devices if damaged. Furthermore, the THD Revolution allows to control the light of the LED either by manually regulating its intensity or by automatically turning the light off if the optic fiber is not connected correctly to the device. Such features satisfy higher standards compared to the predicate devices.
Therefore, the THD revolution's new features offer better control and higher safety compared to the predicate devices without raising new questions.
- Acoustic output level: The THD Revolution's acoustic output is characterized by the . following values: 347 mW/cm2 (Max Ispta3); 0,347 W/cm2 (Max Isppa3) and 0,032 (Max MI).
These parameters are the same as the predicate device K090009.
- . Digital Signal Processor (DSP): The THD Revolution is equipped with DSP technology that improves the acoustic signal quality reducing the noise and thus guaranteeing a clearer signal compared to the predicate devices K07815, K081429 and K090009 that by contrary do not have DSP.
Specific performance test demonstrates that the acoustic signal of the THD Revolution is at least as good as the predicate devices, confirming the substantial equivalence.
The THD Revolution' acoustic signal is equivalent to the predicate devices in term of audibility and its even clearer thanks to DSP technology.
- . Adjustable parameters: The THD Revolution allows the regulation of the light intensity within the range 40-100% (step 10%) and of the volume of the loudspeaker The predicate devices K07815, K081429 and K090009 allow the regulation of the volume of the loudspeaker but not of the light intensity.
Identically to the predicate devices K07815, K081429 and K090009, the THD Revolution allows the regulation of the volume of the loudspeaker. Also, the THD Revolution allows the regulation of the light intensity guaranteeing higher performance standards compared to the predicate devices.
### Doppler probe
THD Revolution is supplied with two identical Doppler probes called "THD Slide Doppler probe".
{7}------------------------------------------------
One THD Slide Doppler probe is included into each compatible kit and represents the probe to be used for the surgical procedure. Such probe can be sterile and disposable as well as non-sterile and reusable (5 treatments).
The other THD Slide Doppler probe represents the back-up Doppler probe to be used as substitute in the event the probe provided with any of the THD Revolution's kits malfunctions. The back-up Doppler probe is supplied sterile and disposable.
The THD Slide Doppler probe has been already cleared in K081429 and K090009 and no changes have been made since its clearance.
#### ● Optical fiber and pneumatic pedal
The optical fiber and the pneumatic foot pedal provided with the THD Revolution are exactly the same as those of the predicate devices (K07815, K081429 and K090009).
#### ● Kit
The THD Revolution Doppler is compatible with any of the following kits:
- 1. THD Slide,
- 2. THD Slide One,
- 3. THD Easy,
- 4. THD Easy one,
- 5. THD Basic,
- 6. THD Basic plus.
- 7.
The THD Slide and THD Slide one kits have been previously cleared (510(k) number: K081429 and K090009 respectively) and there have been no changes to the kits since their clearance. All the other kits consist of different combinations of one or more components characterizing the cleared THD Slide kit (K081429), THD Slide One kit (K090009). THD kit (K070815) or the Surgy proctoscope (K091490).
### 2.7. Material in contact with patients
Identically to the predicate devices K07815, K081429 and K090009, the THD Revolution's components that can come in contact with the patient are the THD slide Doppler probe (made of Makrolon 2458), the proctoscope (made of Makrolon 2458), the needle holder (made of AISI 410 steel) and sutures (K041515) characterizing the THD Slide kit, THD Slide one Kit, THD Easy kit, THD Easy one Kit, THD Basic kit and THD Basic Plus Kit. All the three components are exactly the same as the predicate devices (K07815, K081429 and K090009) with regards to technology and materials they are manufactured with.
Further, the THD Basic and Basic Plus kits have also the Surgy proctoscope that comes in contact with the patient. Such device is made of biocompatible Makrolon 2458 as the proctoscope previously mentioned. The Surgy proctoscope is alreadycleared in K091490 and no changes have been made since its clearance.
{8}------------------------------------------------
### 2.8 Performance data
- 1. Identically to the predicate devices K07815. K081429 and K090009. the THD Revolution has been manufactured according to the following standards:
- o IEC 60601-1
- o IEC 60601-2-37
- o EC 60601-1-2
- 2. Acoustic signal:
The THD Revolution has been provided with DSP technology in order to improve the quality and control of the acoustic signal reducing the noise. It means that the acoustic signal is processed through the DSP microprocessor and then emitted through the loudspeaker. The emission of the acoustic signal through the loudspeaker of the THD Revolution is exactly the same as the predicate devices' Doppler (THD Evolution) cleared in K07815, K081429 and K090009.
- · A performance test was conducted to verify the acoustic output level considering the THD Evolution Doppler and the THD Slide Doppler probe cleared in K081429 and K090009 which is exactly the same probe provided with the THD Revolution. Therefore, given that the THD Revolution's Doppler probe is exactly the same as the predicate devices K090009 and the acoustic signal broadcasting system of the THD Revolution is exactly the same as the THD Evolution Doppler cleared in K07815, K081429 and K090009, the test conducted on the THD Evolution is valid for the THD Revolution as well.
The acoustic output levels resulted from the test are: 347 mW/cm- (Max Ispta3); 0,347 W/cm- (Max Isppa3) and 0,032 (Max MI).
- Given that the DSP technology characterizes the THD Revolution but not the . predicate devices's THD Evolution Doppler (K081429, K070815 and K090009), a specific test was conducted comparing the acoustic signal emitted by the THD Revolution Doppler with the output signal characterizing the Doppler of the predicate devices (THD Evolution Doppler cleared in K081429, K070815 and K090009). The obtained results demonstrate the equivalence of the signal underling that the THD revolution's acoustic signal is at least as good as the predicate devices, and better yet, it's clearer thanks to the DSP that sharply reduces the background noise.
- 3. Light source: luminous efficacy and photobiological safety
The THD Revolution light source is represented by a LED while the light source of the predicate devices K081429, K070815 and K090009 is an halogen lamp. Both the light source have the same watt (50 W).
- The LED of the THD Revolution has been tested according to IEC 62471 (1st ● Edition) to evaluate the photobiological safety. The device under test was a THD
CONFIDENTIAL
{9}------------------------------------------------
Revolution prototype representative of the standard THD SpA production. The device was connected to the dedicated optical fiber. The results demonstrate the compliance with the standard IEC 62471 and the LED source has been classified as "Risk group: Exempt".
- . In order to test the luminous efficacy of the THD Revolution's LED compared to the hanalogic lamp of the predicate devices a specific test was conducted. The test was performed using the THD Revolution and the predicate device THD Evolution (cleared in K081429. K070815 and K090009). Both the devices were connected to the optical fiber and the luminous efficacy was measured by means of a dedicated digital illuminance meter. The obtained results underline greater luminous efficacy for the THD Revoution's LED (10300 lux) than for the predicate device's halogen lamp (1580 Lux), demonstrating higher illumination in case of the THD Revolution.
- 4. Performance data-Clinical:
Specific clinical studies have been conducted in order to demonstrate that the THD Revolution is effectiveness not only for the surgery of haemorrhoids of second and third degree like the predicate devices K081429, K070815 and K090009 but also for grade IV haemorrhoids characterized by prolapse that cannot be manually reduced.
- Transanal dearterialization with targeted mucopexy is effective for advanced . haemorroids. Giorndano P, Tomasi I, Pascariello A, Mills E and Elahi S. Colorectal Dis. 2014 May;16(5):373-6.
Methods: 31 patients with IV grade haemorroids characterized by constant prolapse were operated using the THD Revolution and consequently the THD technique (transanal haemorroidal dearterialization).
Results: the study demonstrates that the THD device is effective in the treatment of grade IV hemorrhoids revealing no intra-operative complications and only one case of recurrence that was treated by surgical removal.
Conclusions: the THD device allows accurate transfixaction and ligation of the artery and further it guarantees the area of prominent prolapsed to be targeted by dividing the procedure into two phases, dearterialization and plication.
- · Transanal haemorrhoidal dearterialization as alternative surgical treatment to hemorrhoidal disease:initial report. Morrett R.S. Crujano General. Vol 34 Num 2; 2012.
Methods: 250 patients with grade II-III and IV (21 patients) haemorroids were operated using the THD technique (transanal haemorroidal dearterialization) consisting in ligating the branches of the superior rectal artery and performing the mucosapexis to correct the prolapse
Results: there were no intra-operative complications and the post operative pain was very low (73% of patients reported 0-1 according to AVS). The satisfactoriness grade (8-10 according to AVS) was reported to be high (8-10 according to AVS).
{10}------------------------------------------------
### THD Revolution
THD S.p.A. TRADITIONAL 510(K) Notification K141657
# Conclusion:
The THD Technique resulted effective in the treatment of haemorroids of second, third and forth degree (characterized by constant prolapsed). The THD technique resulted safe and effective and with low incidence of complications.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.