K141644 · Innovasis, Inc. · GWO · Nov 7, 2014 · Neurology
Device Facts
Record ID
K141644
Device Name
INNOVASIS CRANIAL SYSTEM
Applicant
Innovasis, Inc.
Product Code
GWO · Neurology
Decision Date
Nov 7, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Innovasis Cranial System is intended for use in selective trauma of cranial skeleton, cranial surgery and reconstructive procedures.
Device Story
Innovasis Cranial System consists of low-profile titanium bone plates, mesh, and anchoring screws; used for reconstruction, stabilization, and rigid fixation of non-loadbearing cranial bone segments post-craniotomy, craniectomy, or fracture. Devices are single-use; provided non-sterile for steam sterilization by healthcare providers. Instrumentation included for customizing implant size, shape, and placement. Surgeons utilize these implants to fixate bone pieces, facilitating healing and structural restoration in cranial surgery. Benefits include rigid fixation and biocompatible material construction.
Clinical Evidence
Bench testing only. No clinical data. Performance testing conducted per ASTM F543-13 (torsion, axial pullout), ASTM F382-99 (bending stiffness), and ASTM F2129-08 (corrosion). Biocompatibility testing performed per ISO 10993-1, including pyrogen (LAL) and cytotoxicity (MEM) testing; all results passed.
Indicated for reconstruction, stabilization, and rigid fixation of non-loadbearing areas following craniotomy, craniectomy, and cranial fractures in adults and adolescents (age 12 and older).
Regulatory Classification
Identification
A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.
Predicate Devices
Jeil Medical Products LeForté Neuro System Bone Plate (K112812)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure with three faces in profile, one behind the other. The figures are connected by a flowing ribbon-like shape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 7, 2014
Innovasis, Inc. Marshall McCarty Director QA/RA 614 E. 3900 South Salt Lake City, Utah 84107
> K141644 Trade/Device Name: Innovasis Cranial System Regulation Number: 21 CFR 882.5320 Regulation Name: Preformed Alterable Cranioplasty Plate Regulatory Class: Class II Product Code: GWO, GXR, HBW Dated: October 8, 2014 Received: October 9, 2014
Dear Mr. McCarty,
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -5)/△
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141644
Device Name Innovasis Cranial System
Indications for Use (Describe)
The Innovasis Cranial System is intended for use in selective trauma of cranial surgery and reconstructive procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Innovasis. The logo is in black and features the word "INNOVASIS" in a bold, sans-serif font. Below the main logo, there is a tagline that reads "INNOVATE / INVOLVE / INVENT" in a smaller font. The logo appears to be for a company that focuses on innovation and invention.
# 510(k) Summary
- l. Submitter: Innovasis, Inc. 614 E. 3900 South Salt Lake Citv. UT 84107
- Contact: Marshall C. McCarty Phone: (801) 261-2236 Fax: (801) 261-3758 mmccarty@innovasis.com
Date Prepared: November 7, 2014
### II. Device
| Name of Device: | Innovasis® Cranial System |
|-----------------------|---------------------------------------------|
| Common or Usual Name: | Bone Plate, Bone Screw for Cranial Fixation |
- Classification: Regulation No.: 21 CFR 882.5320 Class 2, Product Code: GWO Regulation No.: 21 CFR 882.5360 Class 2, Product Code: HBW Regulation No.: 21 CFR 882.5250 Class 2, Product Code: GXR
#### III. Predicate Device
Jeil Medical Products LeForté Neuro System Bone Plate K112812 This predicate has not been subject to a design-related recall.
### IV. Device Description:
The Innovasis« Cranial System Bone Plates are comprised of a variety of shapes and sizes intended for reconstruction, stabilization and/or rigid fixation of non loadbearing areas subsequent to craniotomy, craniectomy and cranial fractures in adults and adolescents (age 12 and higher). The single use devices are a variety of low profile plates and anchoring screws that fixate bone pieces together in the cranial area of the patient. The design features include:
- · Commercially Pure Titanium or 6 Al 4V Titanium Alloy
- · Straight and Square plates
- · Double Y plates
- Burr hole plates
- · Plates with Shunts
- · Self-Drilling/Self-Tapping Screws
- Mesh
- · Instrumentation to customize size, shape and placement of implants
- · Compact sterilization tray to optimize organization of implants
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Innovasis. The logo is black and features the word "INNOVASIS" in bold, sans-serif font. The "A" in "INNOVASIS" is stylized to look like a triangle with a line extending from the top right corner. Below the main logo, in a smaller font, are the words "INNOVATE / INVOLVE / INVENT".
## V. Indications for Use
The Innovasis® Cranial System is intended for use in selective trauma of cranial skeleton, cranial surgery and reconstructive procedures.
### VI. Comparison of Technological Characteristics with the Predicate Device
| | Innovasis® ICS | Jeil LeForté Neuro System |
|---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications | The Innovasis® Cranial System is intended for use<br>in selective trauma of cranial skeleton, cranial<br>surgery and reconstructive procedures. | The Leforté Neuro System is intended for use in<br>selective trauma of cranial skeleton, cranial surgery<br>and reconstructive procedures. |
| Material | ASTM F67 Commercially Pure Grade 2 Titanium<br>(plates), Grade 3 Titanium (mesh) and ASTM F136<br>6AI4V Titanium for Surgical Implants (Screws) | ASTM F67 Commercially Pure Grade 2 Titanium<br>(plates), Grade 3 Titanium (mesh) and ASTM F136<br>6AI4V Titanium for Surgical Implants (Screws)<br>Additional plates offered in ASTM F67 Grade 1 and 3 |
| Features | Straight and Square plates<br>Double Y plates<br>Burr hole plates<br>Plates with Shunts<br>Self-Drilling/Self-Tapping Screws<br>Mesh<br>Instrumentation to customize size, shape and<br>placement of implants | Straight and Square plates<br>Double Y plates<br>Burr hole plates<br>Plates with Shunts<br>Self-Drilling/Self-Tapping Screws<br>Mesh<br>Gap, Curved, X-shape, Y, Quad and Calvarium plates<br>Instrumentation to customize size, shape and<br>placement of implants |
| Dimensions | Plates Straight, Double Y and Square, 2, 4, 6 and<br>12 holes; Burr hole plates 6 holes 15mm diameter,<br>5 holes 15mm, 6 holes 22mm and 5 holes 22mm.<br>All plates and mesh are 0.648 +/-0.025mm<br>Mesh 98mm x 98mm<br>Screws: 1.6 x 3mm, 1.6 x 4mm, 1.6 x 5mm, 1.9 x<br>4mm | Plates Straight, Double Y and Square, 2, 4, 6 and 12<br>holes; Burr hole plates 6 holes 15mm diameter, 5<br>holes 15mm, 6 holes 22mm and 5 holes 22mm.<br>Mesh 98mm x 98mm<br>Above plates and mesh are 0.62mm<br>Additional plates and mesh offered in thickness 0.3 to<br>0.5mm, including Gap Plates (3 sizes)<br>Curved, X-shape, Quad and Calvarium plates are<br>offered in other Jeil LeForté configurations<br>Screws: 1.6 x 3mm, 1.6 x 4mm, 1.6 x 5mm, 1.9 x<br>4mm<br>2.2 x 5mm and 1.4 x 1.9mm screws are offered in<br>other Jeil LeForté configurations. |
| Surface | Plate: Electropolish | Plate: Anodized |
| Treatment | Screw: N/A | Screw: N/A |
| Sterilization | Sold Non-Sterile to be<br>Steam Sterilized | Sold Non-Sterile to be<br>Steam Sterilized |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Innovasis. The logo is in black and features the company name in bold, sans-serif font. Below the company name is the tagline "Innovate / Involve / Invent" in a smaller font size.
### VII. Performance Data
(Non-clinical)—Performance testing per ASTM F543-13, F582-99 for 3 point bend (or 4 point bend), axial compression, foam axial pullout, and torque to failure and F2129-08 for corrosion support the safety of the subject devices and demonstrate that the Innovasis Cranial System devices should perform as intended in the specified use conditions.
| Test | Test Method Summary | Results |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Torsion<br>Properties | ASTM F543-13 The torsional testing of the bone screws<br>was performed by hand as these were very small screw<br>samples. Both subject and predicate sample groups<br>were tested identically so comparisons could be made<br>accurately. | The mean torsional strength for the subject screws<br>exceeded the strength of the predicate screws<br>thereby demonstrating substantial equivalence. |
| Static Axial<br>Pullout | ASTM F543-13 | The mean pullout force for the subject screws<br>exceeded the pullout force of the predicate screws<br>thereby demonstrating substantial equivalence. |
| Static Axial<br>Compression | A domed superior test load applied to center of<br>specimen to measure Stiffness, Peak Force, and<br>Displacement at Peak Force. | The mean bending stiffness, Peak Force and<br>Displacement at deformation of the Rev. 3 subject<br>plate exceeded the bending stiffness, Peak Force<br>and Displacement at deformation of the predicate<br>plate demonstrating substantial equivalence. |
| Static 3 Point<br>Bend | ASTM F382-99 (08) Three-point bend instead of four-<br>point bend testing was performed on the predicate and<br>subject double "Y" samples due to their shortened<br>length. Comparisons between the two groups could still<br>be made to determine equivalence. | The mean bending stiffness of the subject plates met<br>or exceeded the bending stiffness of the predicate<br>plate demonstrating substantial equivalence. |
| Static 4 Point<br>Bend | ASTM F382-99 (08) | The mean bending stiffness of the subject plates met<br>or exceeded the bending stiffness of the predicate<br>plate demonstrating substantial equivalence. |
| Corrosion | ASTM F2129-08 | The data show the corrosion resistance of the<br>subject plates met or exceeded the corrosion<br>resistance of the predicate plate thereby passing the<br>requirement in the protocol and demonstrating<br>substantial equivalence. |
| Biocompatibility | A biocompatibility evaluation for the ICS implants was<br>conducted in accordance with the FDA Blue Book<br>Memorandum #G95-1 "Use of International Standard<br>ISO-10993, 'Biological Evaluation of Medical Devices<br>Part 1: Evaluation and Testing," May 1, 1995, and<br>International Standard ISO 10993-1 "Biological<br>Evaluation of Medical Devices -Part 1: Evaluation and<br>Testing Within a Risk Management Process," as<br>recognized by FDA. | The implants are composed of known biocompatible<br>material [ASTM F67 Commercially Pure Titanium<br>(plates and mesh) and ASTM F136 6Al4V Titanium<br>for Surgical Implants (Screws)]. Testing was<br>conducted to ensure manufacturing process did not<br>add contamination. Test battery included Pyrogen<br>Testing (LAL) and Cytotoxicity testing (MEM) All test<br>results were below cutoff (pass). |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "INNOVASIS" in bold, black letters. The "A" in the word is stylized with a line going through it. Below the word "INNOVASIS" are the words "INNOVATE / INVOLVE / INVENT" in a smaller font size.
#### VIII. Conclusions
The Innovasis Cranial System has been subjected to risk analysis, engineering analysis and testing to recognized standards. The predicate device, K112812 LeForte Neuro System was cleared based on the results of non-clinical data. In consideration of the following evaluated parameters subject and predicate device performance data were compared to support the safety of the subject devices:
- · Design configurations
- · Applied mechanical loads
- · Product sizes and shapes
- · Materials used
- · Biocompatibility requirements
- · Manufacturing and processing methods
- Shelf life
The evaluations demonstrate that the Innovasis Cranial System devices should perform as intended in the specified use conditions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.