← Product Code [GBM](/productcode/GBM) · K141642

# VAPRO INTERMITTENT CATHETER (K141642)

_Hollister Incorporated · GBM · Aug 22, 2014 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K141642

## Device Facts

- **Applicant:** Hollister Incorporated
- **Product Code:** [GBM](/productcode/GBM.md)
- **Decision Date:** Aug 22, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5130
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The VaPro™ intermittent catheter is a flexible tubular device that is inserted through the urethra by male, female and pediatric patients who need to drain urine from the bladder. The VaPro Plus™ intermittent catheter is a flexible tubular device that is inserted through the urethra by male, female and pediatric patients who need to drain urine from the bladder.

## Device Story

Hydrophilic-coated, single-use intermittent catheter; used for urinary incontinence management. Features protective sleeve and introducer tip to shield catheter from distal urethral bacteria; vapor strip in packaging hydrates coating for lubrication. VaPro Plus variant includes integrated collection bag for use when external receptacles are unavailable. Designed for ease of use by patients with limited dexterity. Operated by patient or caregiver; used in home or clinical settings. Output is urine drainage; facilitates bladder emptying; reduces risk of urinary retention.

## Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993 and FDA G95-1 standards. Product evaluation confirms device functionality.

## Technological Characteristics

Hydrophilic PVP (polyvinylpyrrolidone) coated catheter; water vapor hydration system; protective sleeve and introducer tip; funnel end design (with optional collection bag for Plus model). Sterile, single-use. Biocompatibility per ISO 10993.

## Regulatory Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

## Predicate Devices

- VaPro intermittent catheter ([K090960](/device/K090960.md))
- VaPro Plus intermittent catheter ([K110862](/device/K110862.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine. The caduceus is depicted with a staff entwined by two snakes and topped with wings.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 22, 2014

Hollister Incorporated Jeanne Lee Senior Manager, Regulatory Affairs 2000 Hollister Drive Libertyville, IL 60048

Re: K141642

> Trade/Device Name: VaPro™ intermittent catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: GBM Dated: June 16, 2014 Received: June 19, 2014

Dear Jeanne Lee,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number

#### Device Name

VaPro™ intermittent catheter VaPro Plus™ intermittent catheter

#### Indications for Use

The VaPro™ intermittent catheter is a flexible tubular device that is inserted through the urethra by male, female and pediatric patients who need to drain urine from the bladder.

The VaPro Plus™ intermittent catheter is a flexible tubular device that is inserted through the urethra by male, female and pediatric patients who need to drain urine from the bladder.

| Type of Use |  |
|-------------|--|
|-------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY -------------------------------------------------------------------------------------------------------------------------------------------------------------

Concurrence of Center for Devices and Radiological Health (CDRH)

This section applies only to requirements of the Paperwork Reduction Act of 1995,

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@)fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Hollister logo. The logo consists of a red geometric symbol on the left and the word "Hollister" in black font on the right. The geometric symbol is a stylized cross or star shape.

## 510(k) Summary

| Submitted By:                     | Jeanne Lee<br>Hollister Incorporated<br>2000 Hollister Drive<br>Libertyville, IL 60018<br>(t) 847-996-6350<br>(f) 847-918-3981                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Summary Prepared:            | June 16, 2014                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Device Name:                      | Classification Name- Urological catheter and accessories<br>Common/Usual Name- Catheter, Urethral<br>Proprietary Name- VaProTM intermittent catheter<br>VaPro PlusTM intermittent catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Predicate Device:                 | The VaPro intermittent catheter (with modification) is substantially<br>equivalent to its original design, VaPro intermittent catheter,<br>K090960.<br><br>The VaPro Plus intermittent catheter (with modification) is<br>substantially equivalent to its original design, VaPro Plus<br>intermittent catheter, K110862.                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Description:               | The VaPro intermittent catheter is a hydrophilic coated, single use<br>catheter to be used as a means of managing urinary incontinence<br>by draining urine from the bladder. The catheter comes in a<br>protective sleeve and is offered with a protective introducer tip as a<br>way to shield the sterile catheter from bacteria in the distal urethra<br>during insertion. The packaging contains a vapor strip that hydrates<br>the catheter coating which then lubricates the catheter. The outer<br>packaging was designed to facilitate access for those with limited<br>dexterity.<br>The VaPro Plus has the addition of a collection bag that is<br>connected to the catheter for use when drainage into a suitable<br>receptacle is not feasible or practical. |
| Intended Use:                     | The VaPro intermittent catheter is a flexible tubular device that is<br>inserted through the urethra by male, female, and pediatric patients<br>who need to drain urine from the bladder.<br><br>The VaPro Plus intermittent catheter is a flexible tubular device<br>that is inserted through the urethra by male, female, and pediatric<br>patients who need to drain urine from the bladder.                                                                                                                                                                                                                                                                                                                                                                           |
| Technological<br>Characteristics: | The table below summarizes the technological characteristics of<br>the device as compared to the predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |

{4}------------------------------------------------

|                     | Modified Device:<br>VaPro intermittent<br>catheter                                                                                                                                             | Predicate Device:<br>VaPro intermittent<br>catheter (K090960) | Modified Device:<br>VaPro Plus<br>intermittent catheter | Predicate Device:<br>VaPro Plus<br>intermittent catheter<br>(K110862) |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|---------------------------------------------------------|-----------------------------------------------------------------------|
| Intended Use        | The VaPro/VaPro Plus intermittent catheter is a flexible tubular device that is inserted through the urethra by male, female, and pediatric patients who need to drain urine from the bladder. |                                                               |                                                         |                                                                       |
| Condition of<br>Use | Single Use                                                                                                                                                                                     | Single Use                                                    | Single Use                                              | Single Use                                                            |
| Prelubricated       | Yes-by water vapor<br>hydration                                                                                                                                                                | Yes-by water vapor<br>hydration                               | Yes-by water vapor<br>hydration                         | Yes-by water vapor<br>hydration                                       |
| Ready to use        | Yes                                                                                                                                                                                            | Yes                                                           | Yes                                                     | Yes                                                                   |
| End Design          | Funnel                                                                                                                                                                                         | Funnel                                                        | Catheter funnel<br>attached to collection<br>bag        | Catheter funnel<br>attached to collection<br>bag                      |
| Sterile             | Yes                                                                                                                                                                                            | Yes                                                           | Yes                                                     | Yes                                                                   |
| No touch<br>design  | Yes-contains sleeve                                                                                                                                                                            | Yes-contains sleeve                                           | Yes-contains sleeve                                     | Yes-contains sleeve                                                   |
| Lubricant           | PVP Based<br>(polyvinylpyrollidone)<br>Coating                                                                                                                                                 | PVP Based<br>(polyvinylpyrollidone)<br>Coating                | PVP Based<br>(polyvinylpyrollidone)<br>Coating          | PVP Based<br>(polyvinylpyrollidone)<br>Coating                        |
| Protective Tip      | Yes                                                                                                                                                                                            | Yes                                                           | Yes                                                     | Yes                                                                   |
| Collection Bag      | No                                                                                                                                                                                             | No                                                            | Yes                                                     | Yes                                                                   |

Performance Testing Conclusions:

Biocompatibility testing was performed based on the United States Food and Drug Administration Office of Device Evaluation General program Memorandum #G95-1 and ISO 10993 biocompatibility testing standards. Results indicate compliance to the standard.

Product evaluation also supports device functionality.

---

**Source:** [https://fda-staging.innolitics.com/device/K141642](https://fda-staging.innolitics.com/device/K141642)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).