Propper Insight™ Binocular Indirect Ophthalmoscope is an AC-adapter powered or rechargeable battery powered device for medical professionals containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, vitreous) and the retina of the eye
Device Story
Propper Insight™ Binocular Indirect Ophthalmoscope (BIO) is a head-worn optical instrument for eye examinations. Device consists of illumination and viewing optics housed in a metal module attached to an adjustable headband via a pivot bracket. Illumination system utilizes an LED source, lenses, filters (red-free, amber, cobalt blue), light apertures, and a diffuser. Illumination intensity is adjusted via two polarizer filters. Viewing system uses adjustable lenses and mirrors to visualize the patient's fundus. Magnetic securement mechanism allows pivoting between in-use and out-of-use positions; magnetic contact automatically toggles power to the LED. Powered by AC adapter or rechargeable battery. Operated by physicians or eye care professionals in clinical settings. Provides direct visualization of ocular structures to assist in diagnosis and clinical decision-making regarding eye health.
Clinical Evidence
Bench testing only. Device complies with ISO 10943:2012 (Ophthalmic Instruments - Indirect Ophthalmoscopes). No clinical data provided.
Technological Characteristics
Head-worn binocular indirect ophthalmoscope. LED light source. Optomechanical illumination control via dual polarizers. Metal housing (BIO module) with magnetic pivot/power-switching mechanism. Powered by AC adapter or rechargeable battery. Complies with ISO 10943:2012. No software or digital connectivity.
Indications for Use
Indicated for medical professionals to examine the ocular media (cornea, aqueous, lens, vitreous) and the retina of the eye in patients.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
Predicate Devices
Heine OMEGA 500 binocular Indirect ophthalmoscope, LED version (K123316)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, connected by a flowing line that resembles a bird in flight. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 17, 2014
Propper Manufacturing Co., Inc. Andrew Sharavara, Ph.D. Chief Technical Officer 36-04 Skillman Avenue Long Island City, NY 11101
Re: K141638
Trade/Device Name: Propper Insight" Binocular Indirect Ophthalmoscope (BIO) Kits (Models 199185 and 199285) Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: II Product Code: HLI Dated: June 12, 2104 Received: June 19, 2014
Dear Dr. Sharavara:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Y. Alexander - S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known): ____________________
Device Name: Insight™ Binocular Indirect Ophthalmoscope.
Indications for Use:
Propper Insight™ Binocular Indirect Ophthalmoscope is an AC-adapter powered or rechargeable battery powered device for medical professionals containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, vitreous) and the retina of the eye
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
## 510(k) Summary
Submitted by Propper Manufacturing Company, Inc. 36-04 Skillman Avenue, Address: Long Island City, New York 11101 Telephone: (800) 832-4300 or (718) 392-6650 Facsimile: (718) 482-8909 Contact Name: Andrew Sharavara
## Date Submitted: June 12, 2014
| Trade Name: | Insight™ Binocular Indirect Ophthalmoscope |
|--------------|--------------------------------------------|
| Common Name: | Binocular Indirect ophthalmoscope |
Product Code / Regulation: HLI (AC), HLJ (DC) / 21 C.F.R. 866.1570
Description: The Propper Insight™ Binocular Indirect Ophthalmoscope is an AC-powered or rechargeable battery-powered indirect ophthalmoscope that complies with standard ISO 10943:2012, Ophthalmic Instruments - Indirect Ophthalmoscopes.
The Propper Insight 100 is a Binocular Indirect Ophthalmoscope (abbreviated - BIO), worn on the medical professional's head containing illumination and viewing optics intended to examine the media and the retina of the eye when used in conjunction with an ophthalmic lens.
The illumination part of the Propper Insight™ Binocular Indirect Ophthalmoscope consists of a LED (Light Emitting Diode) source, lenses, a selection of red-free, amber and cobalt blue filters, three sizes of light apertures, diffuser and illumination mirror. The device has optomechanical system for adjustment of illumination level, which is based on the relative positions of two polarizer filters.
The viewing part consists of viewing lenses, and mirrors that are adjustable to obtain views of the patient eye fundus.
The illumination part and the viewing part are combined in the metal housing (BIO Module) which is attached to the headband with the pivot bracket mechanism. The attachment mechanism allows the BIO Module to be pivoted between in-use (down) and out-of-use positions (up). The attachment mechanism also allows adjustment of the BIO Module relative to the user's eyes for the most optimal viewing path. The attachment mechanism includes a magnetic securement of the BIO Module in both in-use and out-of-use positions. Part of the magnetic securement operates an electric contact to automatically provide power to the illumination source in the in-use position.
The BIO Module and the power cable connector (AC-powered version), or the BIO Module and the rechargeable battery are attached to the adjustable headband.
{4}------------------------------------------------
- Indication: Propper Insight™ Binocular Indirect Ophthalmoscope is an AC-adapter powered or rechargeable battery powered device for medical professionals containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, vitreous) and the retina of the eye.
The Insight™ Binocular Indirect Ophthalmoscope is similar in Substantial Equivalence: intended use, design and operating principles to the following device:
| Predicate device | 510k number |
|-------------------------------------------------------------------|-------------|
| Heine OMEGA 500 binocular Indirect<br>ophthalmoscope, LED version | K123316 |
Substantial equivalence to the predicate device was evaluated according to the FDA guidance document "guidance for Industry. Ophthalmoscope guidance (Direct and Indirect)" issued on July 8, 1998.
Propper Insight™ Binocular Indirect Ophthalmoscope is equivalent to the predicate device Heine Omega-500 Binocular Indirect Ophthalmoscope because both devices are designed for the same purpose, use similar light sources (LED), have the same indications for use, very similar operational principles and design, and the new device does not introduce new potential hazard or safety risks.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.