K141634 · Romedex International Srl · LJS · Jan 27, 2015 · General Hospital
Device Facts
Record ID
K141634
Device Name
NAUTILUS DELTA
Applicant
Romedex International Srl
Product Code
LJS · General Hospital
Decision Date
Jan 27, 2015
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Software as a Medical Device, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K141634 · Jan 27, 2015
NAUTILUS DELTA
Romedex International Srl
European hospital clinical usage data; Post-market clinical experience records; Side-by-side clinical comparison data
The sponsor utilized post-market clinical experience from hundreds of patients in European hospitals and side-by-side clinical comparisons with predicate devices to support the substantial equivalence of the Nautilus Delta device.
European Post-market Clinical Experience; Observational / Post-market surveillance; Study Period: Post-CE mark
Patients in European hospitals; Sample Size: Hundreds of patients; Number of Sites: Several hospitals
Legally marketed similar devices (including predicate devices)
Safety and performance (no adverse events reported)
Indications for Use
The intended use of Nautilus DeltaTM is to support navigation and tip positioning of central venous access devices. Nautilus DeltaTM can be used as an alternative method to fluoroscopy and chest X-ray for central venous catheter tip placement confirmation.
Device Story
Nautilus Delta system facilitates real-time central venous access device (CVAD) tip positioning. System inputs: patient cardiac electrical activity (ECG) via sterile extension cable connected to catheter stylet/guidewire. Patient module acquires/processes ECG data; transmits wirelessly (Bluetooth) to mobile medical application. Application displays surface ECG, R-peak markers, heart rate, and navigation signal to guide catheter from access point to cavo-atrial junction (CAJ). Used in clinical settings by healthcare providers; replaces/augments traditional fluoroscopy or chest X-ray. Provides visual feedback on graphical user interface; assists clinicians in accurate tip placement; reduces radiation exposure by minimizing need for imaging.
Clinical Evidence
Clinical data from European post-market use in hundreds of patients across multiple hospitals. No adverse events reported. Side-by-side clinical comparisons with predicate devices demonstrated substantial equivalence. Bench testing confirmed compliance with standards including ES60601-1, IEC 60601-1-2, IEC 60601-2-27, IEC 62304, IEC 62366, ISO 11737-1, ISO 11135-1, ISO 11135-2, EN 55022, and EN 300328.
Technological Characteristics
System includes sterile ECG extension cable, non-sterile ECG cable, patient module, and mobile application. Uses Bluetooth for wireless data transmission. Features include real-time ECG acquisition, R-peak detection, heart rate calculation, and navigation signal computation. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-27 (ECG monitoring), IEC 62304 (software), and ISO 11135 (sterilization).
Indications for Use
Indicated for navigation/positioning of CVADs (≥3 Fr) using cardiac electrical activity as alternative to X-ray/fluoroscopy. Population: Adults/adolescents (>12-21y) with PICCs, CVCs, ports, hemodialysis catheters; children (>2-12y) with PICCs/CICCs; infants (>1mo-2y) and neonates (birth-1mo) with CICCs. Contraindicated/limited in patients with altered cardiac rhythm (atrial fibrillation, flutter, severe tachycardia, pacemaker-driven rhythm); additional confirmation method required for these patients.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is a stylized image of three faces in profile, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 27, 2015
Romedex International SRL Sorin Grunwald, PhD. Managing Director 625 Clayton Street San Francisco, CA 94117
Re: K141634
Trade/Device Name: Nautilus Delta™ Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, implanted, long- term intravascular catheter Regulatory Class: II Product Code: LJS Dated: December 15, 2015 Received: December 17, 2015
Dear Dr. Grunwald:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Sorin Grunwald, Ph.D.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K141634
Device Name Nautilus Delta
#### Indications for Use (Describe)
Nautilus Delta is indicated for navigation and positioning of central venous access devices (CVADs) of at least 3 Fr in size. Nautilus Delta provides real-time catheter tip location by using the patient's cardiac electrical activity and is indicated for use as an alternative method to chest X-ray and fluoroscopy for CVAD tip placement confirmation. In adult patients and in adolescents (greater than 12 through 21 years of age), Nautilus Delta can be used with CVADs such as peripherally inserted central catheters (PICCs), central venous catheters (CVCs), implantable ports, and hemodialysis catheters; in children (greater than 2 to 12 years of age), Nautilus Delta can be used with centrally inserted central catheters (CICCs); in infants (greater than 1 month to 2 years of age) and in neonates (from birth to 1 month of age), Nautilus Delta can be used with CICCs. In each specific age group, the CVAD type and size must be chosen and the CVAD must be used according to the CVAD's indications for use. Limiting but not contraindicated situations for this method are in patients where alterations of cardiac rhythm change the presentation of the P wave as in atrial fibrillation, atrial flutter, severe tachycardia, and pacemaker driven rhythm. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Proprietary Name: | Nautilus Delta™ also known as Handy Nautilus™ |
|-------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name: | Nautilus Delta™ |
| Applicant Name: | Romedex International SRL |
| Applicant Address: | 74 Fundeni Str., 022325, Bucharest, Romania<br>+40.21.255.0385, info@romedex.com, www.romedex.com |
| Contact person: | Sorin Grunwald PhD, 625 Clayton Str., San Francisco, CA 94117<br>sorin@romedex.com, Tel: +1.650.209.4838, Fax: +1.650.636.9666 |
| Date of Preparation: | June 16, 2014 |
| Product Classification: | Class II, 21 CFR §880.5970<br>LJS - Accessory to Percutaneous, Implanted, Long-Term<br>Intravascular Catheters |
| Classification Panel: | General Hospital |
| Predicate devices: | Epidural Catheter Connector (K973371)<br>Sherlock 3CG Tip Confirmation System (K113808) |
| Device Description: | The Nautilus Delta™ system consists of the following elements:<br>single use sterile ECG extension cable (also known as Nautilus<br>Electrical Adaptor also known as NautilusE), non-sterile ECG<br>cable, patient module (ECG data acquisition and processing and<br>integrated remote control), and mobile medical application<br>running on any mobile platform which complies with the minimum<br>requirements. Nautilus Delta™ provides real-time catheter tip<br>location information by using the patient's cardiac electrical<br>activity (ECG). This information can be used to position any<br>central venous access devices (CVADs) at or around the cavo-<br>atrial junction (CAJ). Nautilus Delta supports navigation of central<br>venous catheters from the vascular access point towards the CAJ<br>by computing and displaying a navigation signal. Nautilus Delta<br>also displays on its graphical user interface the surface (skin) ECG<br>signal, a marker identifying the R-peak, and the patient's heart<br>rate. In order to obtain intravascular ECG information at the tip<br>(distal end) of a catheter, a stylet or a guidewire inserted in the<br>catheter can be connected to Nautilus Delta patient module via the<br>sterile Nautilus Delta ECG extension cable. Nautilus Delta and<br>NautilusE do not have any direct or indirect contact with the<br>patient. |
| Intended Use: | The intended use of Nautilus DeltaTM is to support navigation and<br>tip positioning of central venous access devices. Nautilus DeltaTM<br>can be used as an alternative method to fluoroscopy and chest X-<br>ray for central venous catheter tip placement confirmation. |
| Indications for Use: | Nautilus Delta is indicated for navigation and positioning of central<br>venous access devices (CVADs) of at least 3 Fr in size. Nautilus<br>Delta provides real-time catheter tip location information by using<br>the patient's cardiac electrical activity and is indicated for use as an<br>alternative method to chest X-ray and fluoroscopy for CVAD tip<br>placement confirmation.<br>In adult patients and in adolescents (greater than 12 through 21<br>years of age), Nautilus Delta can be used with CVADs such as<br>peripherally inserted central catheters (PICCs), central venous<br>catheters (CVCs), implantable ports, and hemodialysis catheters;<br>in children (greater than 2 to 12 years of age), Nautilus Delta can<br>be used with PICCs and with centrally inserted central catheters<br>(CICCs); in infants (greater than 1 month to 2 years of age) and<br>in neonates (from birth to 1 month of age), Nautilus Delta can be<br>used with CICCs. In each specific age group, the CVAD type and<br>size must be chosen and the CVAD must be used according to the<br>CVAD's indications and instructions for use.<br>Limiting but not contraindicated situations for this method are in<br>patients where alterations of cardiac rhythm change the<br>presentation of the P wave as in atrial fibrillation, atrial flutter,<br>severe tachycardia, and pacemaker driven rhythm. In such patients,<br>who are easily identifiable prior to central venous catheter<br>insertion, the use of an additional method is required to confirm<br>catheter tip location. |
| Summary of Technological<br>Characteristics compared<br>to Predicate Devices: | Nautilus Delta has features, materials and technological<br>characteristics which are similar to the ones of the predicate<br>devices. The most notable differences are:<br>a) Nautilus Delta uses wireless technology (Bluetooth) instead of<br>an USB cable for the ECG data transmission between the<br>Patient and the Operator modules.<br>b) Nautilus Delta uses a mobile platform instead of a netbook PC.<br>c) Nautilus Delta provides a Bluetooth remote control integrated<br>in the Patient Module instead of a standalone Bluetooth remote<br>control.<br>d) Nautilus Delta provides an additional optional navigation<br>signal to support catheter tip navigation from the vascular<br>access point towards the cavo-atrial junction.<br>e) Nautilus Delta displays the patient's heart rate and markers to<br>identify the R-peak of the QRS complex of the ECG<br>waveform. |
{4}------------------------------------------------
{5}------------------------------------------------
Summary of Romedex has performed extensive testing in order to demonstrate Performance Testing: that the new characteristics do not affect safety and effectiveness when compared to the predicate devices and to demonstrate compliance with the following standards: ES60601-1, IEC 60601-1-2, IEC 60601-2-27, IEC 62304, IEC 62366, ISO 11737-1, ISO 11135-1, ISO 11135-2, EN 55022, and EN 300328. Summary of Since obtaining the CE mark, Nautilus Delta also known as Handy Nautilus was successfully used in Europe in hundreds of patients in Clinical Data: several hospitals in several countries by many different users of different backgrounds and experience levels. No adverse events were reported. Side-by-side comparisons with legally marketed similar devices, including the predicate devices, were performed in clinical settings to demonstrated substantial equivalence of Nautilus Delta with the predicate devices. Summary of Based on the analysis of the indications for use, intended use, Substantial Equivalence: technological characteristics, performance tests, and post-market clinical experience, Romedex International has concluded that the new characteristics of the subject device do not affect the safety or effectiveness of the subject device as compared to the predicate devices. Romedex International considers that the new device Nautilus Delta is substantially equivalent to the predicate devices K973371 and K113808.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.