← Product Code [QEW](/productcode/QEW) · K141617

# CLEANER ROTATIONAL THROMBECTOMY SYSTEM (K141617)

_Rex Medical, L.P. · QEW · Nov 14, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K141617

## Device Facts

- **Applicant:** Rex Medical, L.P.
- **Product Code:** [QEW](/productcode/QEW.md)
- **Decision Date:** Nov 14, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5150
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Cleaner™ rotational thrombectomy system is indicated for mechanical declotting of native vessel dialysis fistulae and synthetic dialysis access grafts. The Cleaner™ Rotational Thrombectomy System is indicated for mechanical declotting and controlled and selective infusion of physician-specified fluids, including thrombolytics, in the peripheral vasculature.

## Device Story

Percutaneous infusion system; utilizes rotating sinusoidal wire at 4000 RPM to create fluid vortex; facilitates mechanical declotting and dispersion of physician-specified fluids (e.g., thrombolytics) in peripheral vasculature; contrast media/solutions infused through catheter lumen to distal side hole; wire rotation enhances solution penetration into clot; residual clot aspirated via introducer sheath; sterile, single-use disposable; used by physicians in clinical settings; provides mechanical declotting and targeted drug delivery to restore vessel patency.

## Clinical Evidence

No human clinical data. Evidence based on bench testing (infusion performance, electrical safety, EMC) and a GLP animal study assessing device performance in restoring vessel patency in peripheral vasculature. Animal study confirmed successful restoration of patency in occluded vessels.

## Technological Characteristics

Percutaneous catheter with rotating sinusoidal wire; 4000 RPM rotation; radiopaque a-traumatic distal tip; sterile, single-use disposable; infusion lumen with distal side hole; mechanical energy source for rotation.

## Regulatory Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

## Predicate Devices

- Trellis Peripheral Infusion System ([K013635](/device/K013635.md))
- Pulse*Spray Infusion System ([K950907](/device/K950907.md))
- Cleaner™ Rotational Thrombectomy System ([K091029](/device/K091029.md))
- Cleaner™ Rotational Thrombectomy System ([K120346](/device/K120346.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a circular seal with an eagle emblem and text around it, representing the Department of Health & Human Services - USA. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" in a bold, sans-serif font, followed by "ADMINISTRATION" in a smaller font.

July 29, 2020

Rex Medical, L.P. % MDI Consultants, Inc. Susan Goldstein-Falk 55 Northern Boulbevard Great Neck, New York 11021

Re: K141617

Trade/Device Name: Cleaner Rotational Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEW, KRA

Dear Susan Goldstein-Falk:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated November 14, 2014. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075, Gregory.Oconnell(@FDA.HHS.gov.

Sincerely,

Digitally signed by Gregory Gregory W. W. O'connell -S O'connell -S Date: 2020.07.29 19:20:18 -04'00

Gregory O'Connell Assistant Director Plaque Modification Devices Team DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features an emblem with three stylized human profiles facing right, overlaid and slightly offset from each other, representing the department's focus on people. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 14, 2014

Rex Medical, L.P. % Susan Goldstein-Falk MDI Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, NY 11021

Re: K141617

> Trade/Device Name: Cleaner Rotational Thrombectomy System Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: October 6, 2014 Received: October 7, 2014

Dear Ms. Goldstein-Falk,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{2}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K141617

Device Name

Cleaner™ Rotational Thrombectomy System

| Indications for Use (Describe) |  |
|--------------------------------|--|
|--------------------------------|--|

The Cleaner™ Rotational Thrombectomy System is indicated for mechanical declotting of native vessel dialysis fistulae and synthetic dialysis access grafts.

The Cleaner™ Rotational Thrombectomy System is indicated for mechanical declotting and controlled and selective infusion of physician-specified fluids, including thrombolytics, in the peripheral vascular.

### Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary

The assigned 510(k) number is:_____K141617

| Submitter:      | Rex Medical, L.P.<br>555 East North Lane, Suite 5035<br>Conshohocken, PA 19428                                                 |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Colin Valentis<br>Development Engineer<br>Phone: (610) 940-0665 x106<br>Fax: (610) 940-1590<br>Email: cvalentis@rexmedical.com |
| Date Prepared:  | September 19th, 2014                                                                                                           |
| Trade Name:     | Cleaner™ Rotational Thrombectomy System                                                                                        |
| Common Name:    | Thrombectomy catheter                                                                                                          |

Classification Name: Continuous Flush Catheter (21 CFR 870.1210, Product Code KRA) Peripheral Atherectomy Catheter (21 CFR 870.4875, Product Code MCW)

# Predicate Device(s):

| K013635 | Trellis Peripheral Infusion System      |
|---------|-----------------------------------------|
| K950907 | Pulse*Spray Infusion System             |
| K091029 | Cleaner™ Rotational Thrombectomy System |
| K120346 | Cleaner™ Rotational Thrombectomy System |

# Device Description:

The Cleaner™ Rotational Thrombectomy System is a percutaneous infusion system that utilizes a rotating sinusoidal wire to increase the dispersion of delivered solution. The devices employ mechanical rotation of a flexible "S" shaped dispersion wire at 4000 RPM which creates a fluid vortex within the treatment site. Contrast media and physician specified solutions, including thrombolytics, may be infused through the catheter lumen to a side hole at the distal end. The dispersion wire uses mechanical rotation to allow the infused solution to penetrate the clot increasing the effectiveness of the treatment. Any residual clot can be aspirated through an introducer sheath prior to restoration of flow. The distal soft tip is a radiopaque a-traumatic tip on the distal end of the dispersion wire. The devices are sterile and single-use disposable.

{5}------------------------------------------------

# Intended Use:

The Cleaner™ rotational thrombectomy system is indicated for mechanical declotting of native vessel dialysis fistulae and synthetic dialysis access grafts.

The Cleaner™ Rotational Thrombectomy System is indicated for mechanical declotting and controlled and selective infusion of physician-specified fluids, including thrombolytics, in the peripheral vasculature.

# Technological Characteristics:

The Cleaner™ Rotational Thrombectomy System is similar with regard to materials, intended use, principles of operation and technological characteristics to the predicate device. Any differences that may exist do not significantly affect the safety and efficacy of the device. Results of bench testing and animal studies demonstrate Cleaner™ Rotational Thrombectomy System is as safe and effective as the legally marketed predicate device.

# Non-Clinical Performance Testing:

The expanded indications for use to include controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature. are based on the bench testing and animal data presented in the performance section of the submission. A GLP animal study was conducted utilizing the Cleaner Rotational Thrombectomy System to assess device performance at restoring vessel patency in the peripheral vasculature. The animal testing show the Cleaner Rotational Thrombectomy System successfully restores patency to occluded vessels. Bench testing included: infusion testing of the predicate device to the proposed device, Electrical Safety, and EMC testing.

# Conclusions:

Rex Medical considers the Cleaner™ Rotational Thrombectomy System to be substantially equivalent to the predicate devices listed above. The conclusions are based on performance testing and similarities in indications for use, materials, technological characteristics, principle of operation and design features.

---

**Source:** [https://fda-staging.innolitics.com/device/K141617](https://fda-staging.innolitics.com/device/K141617)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).