# ORTHOFIX ANKLE HINDFOOT NAILING SYSTEM (K141571)

_Orthofix Srl · HSB · Nov 12, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K141571

## Device Facts

- **Applicant:** Orthofix Srl
- **Product Code:** [HSB](/productcode/HSB.md)
- **Decision Date:** Nov 12, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3020
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Orthofix Ankle Hindfoot Nailing System is intended to facilitate tibiotalocalcaneal arthrodesis (fusion). Specific indications include: 1. Avascular necrosis of the talus 2. Failed total ankle arthroplasty 3. Trauma (malunited tibial pilon fracture) 4. Severe deformity or instability as a result of talipes equinovarus, cerebral vascular accident, paralysis or other neuromuscular disease 5. Revision ankle arthrodesis 6. Neuroarthropathy 7. Rheumatoid arthritis 8. Osteoarthritis 9. Pseudoarthrosis 10. Post-traumatic arthrosis 11. Previously infected arthrosis 12. Charcot foot 13. Severe endstage degenerative arthritis 14. Severe defects after tumor resection 15. Pantalar arthrodesis

## Device Story

System comprises implantable intramedullary nails, locking end caps, and locking screws; used to facilitate tibiotalocalcaneal arthrodesis (fusion). Implants manufactured from titanium alloy (Ti6Al4V). System includes instrumentation for implantation and explantation. Used in clinical settings by orthopedic surgeons. Device provides mechanical stabilization of the ankle and hindfoot to promote bone fusion. Benefits include correction of severe deformities, stabilization of trauma-related injuries, and treatment of degenerative or post-traumatic arthrosis.

## Clinical Evidence

Bench testing only. Mechanical testing included static four-point bend, static torsion, and bending fatigue testing conducted in accordance with ASTM F1264-03. Results demonstrate substantial equivalence to the predicate device.

## Technological Characteristics

Materials: Titanium alloy (Ti6Al4V). Components: Intramedullary nails, locking end caps, locking screws. Design features: Distal nail flare, variable fixation points, smaller inner diameter, threaded-head screw options. Testing standards: ASTM F1264-03 (Standard Specification and Test Methods for Intramedullary Fixation Devices).

## Regulatory Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

## Predicate Devices

- Biomet Phoenix Ankle Arthrodesis Nail System ([K091976](/device/K091976.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 12, 2014

Orthofix Srl % Mr. Troy Brooks Senior Regulatory Affairs Specialist 3451 Plano Parkway Lewisville, Texas 75056

Re: K141571

Trade/Device Name: Orthofix Ankle Hindfoot Nailing System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB Dated: October 10, 2014 Received: October 14, 2014

Dear Mr. Brooks,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, Lori A. Wiggins -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K141571

Device Name Orthofix Ankle Hindfoot Nailing System

Indications for Use (Describe)

The Orthofix Ankle Hindfoot Nailing System is intended to facilitate tibiotalocalcaneal arthrodesis (fusion). Specific indications include:

- 1. Avascular necrosis of the talus
- 2. Failed total ankle arthroplasty
- 3. Trauma (malunited tibial pilon fracture)
- 4. Severe deformity or instability as a result of talipes equinovarus, cerebral vascular accident, paralysis
- or other neuromuscular disease
- 5. Revision ankle arthrodesis
- 6. Neuroarthropathy
- 7. Rheumatoid arthritis
- 8. Osteoarthritis
- 9. Pseudoarthrosis
- 10. Post-traumatic arthrosis
- 11. Previously infected arthrosis
- 12. Charcot foot
- 13. Severe endstage degenerative arthritis
- 14. Severe defects after tumor resection
- 15. Pantalar arthrodesis

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

#### 5.0 510(k) Summary

| Date Prepared:                         | October 10, 2014                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Purpose for Submission:                | New product offering                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Sponsor:                               | Orthofix Srl<br>Via delle Nazioni 9<br>37012 Bussolengo (VR)<br>Italy                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Contact:                               | Orthofix<br>Troy Brooks, RAC<br>Tel: 214-937-2047<br>Fax: 214-937-3322                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Trade Name /<br>Proprietary Name:      | Orthofix Ankle Hindfoot Nailing System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Common Name:                           | Intramedullary Nail                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Product Code:                          | HSB - Rod, Fixation, Intramedullary And Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Classification:                        | Class II - 21 CFR § 888.3020 - Intramedullary Fixation Rod                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Predicate Device:                      | Biomet Phoenix Ankle Arthrodesis Nail System – K091976                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Device Description:                    | The Orthofix Ankle Hindfoot Nailing System comprises a variety of<br>implantable intramedullary nails with respective locking end caps and<br>locking screws manufactured from titanium alloy (Ti6Al4V) which are<br>intended to facilitate tibiotalocalcaneal arthrodesis (fusion). The system<br>also includes instrumentation required for implantation and explantation<br>of the implants.                                                                                                                                                                                                                                                                                                                                                                          |
| Intended Use /<br>Indications For Use: | The Orthofix Ankle Hindfoot Nailing System is intended to facilitate<br>tibiotalocalcaneal arthrodesis (fusion). Specific indications include:<br>1. Avascular necrosis of the talus<br>2. Failed total ankle arthroplasty<br>3. Trauma (malunited tibial pilon fracture)<br>4. Severe deformity or instability as a result of talipes equinovarus,<br>cerebral vascular accident, paralysis or other neuromuscular disease<br>5. Revision ankle arthrodesis<br>6. Neuroarthropathy<br>7. Rheumatoid arthritis<br>8. Osteoarthritis<br>9. Pseudoarthrosis<br>10. Post-traumatic arthrosis<br>11. Previously infected arthrosis<br>12. Charcot foot<br>13. Severe endstage degenerative arthritis<br>14. Severe defects after tumor resection<br>15. Pantalar arthrodesis |
| Technological<br>Characteristics       | The technological characteristics of the Orthofix Ankle Hindfoot Nailing<br>System are similar to the predicate device in terms of general design                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |

Summary:

dimensions, intended use, materials, and performance characteristics.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Orthofix. The logo consists of a blue, abstract shape on the left, followed by the word "ORTHOFIX" in bold, black letters. A small registration mark is present to the right of the word "ORTHOFIX".

The following technological differences exist between the Orthofix Ankle Hindfoot Nailing System and the predicate device:

- The intramedullary nails have a distal flare
- The intramedullary nails have additional fixation points in some lengths
- The intramedullary nails have a smaller inner diameter

System is substantially equivalent to the predicate device.

- The locking screws are also offered in a threaded-head version
- The locking screws are also offered in lengths of 115mm and 120mm

| Non-Clinical Testing: | Mechanical testing of the Orthofix Ankle Hindfoot Nailing System<br>consisting of Static Four-Point Bend testing, Static Torsion testing, and<br>Bending Fatigue testing was conducted in accordance with ASTM F1264-<br>03 “Standard Specification and Test Methods for Intramedullary Fixation<br>Devices.” Test results demonstrate that the Orthofix Ankle Hindfoot<br>Nailing System is substantially equivalent to the predicate device. |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial:          | Based upon similarities in design, materials, intended use, indications for                                                                                                                                                                                                                                                                                                                                                                    |
| Equivalence:          | use and the results of mechanical testing, the Orthofix Hindfoot Nailing                                                                                                                                                                                                                                                                                                                                                                       |

---

**Source:** [https://fda-staging.innolitics.com/device/K141571](https://fda-staging.innolitics.com/device/K141571)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).