K141564 · Zimmer Medizinsystems GmbH · ILY · Oct 24, 2014 · Physical Medicine
Device Facts
Record ID
K141564
Device Name
OPTON PRO
Applicant
Zimmer Medizinsystems GmbH
Product Code
ILY · Physical Medicine
Decision Date
Oct 24, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
temporary relief of minor muscle and joint aches, pains and stiffness · temporary relief of muscle spasm · temporary relief of minor pain and stiffness associated with arthritis · promoting relaxation of the muscle tissue · temporary increase in local blood circulation
Device Story
OptonPro is a Class IV laser-based topical heat lamp; emits infrared energy (810nm and 980nm) to elevate tissue temperature. System includes four diodes, flexible fiber optic cable, and 10mm output applicator. Aiming beam (650nm) assists targeting. Operated via touch screen interface; two modes: continuous wave and serial pulse. Safety features: software security key lock, footswitch, emergency off switch, hardware interlock, and protective glasses. Integrated power control unit tests laser performance; thermal threshold test assesses warmth sensitivity. SD card stores logs and updates. Used in clinical settings by healthcare providers to treat musculoskeletal conditions; provides therapeutic heating (40–45°C) to alleviate pain, stiffness, and spasms, and improve circulation.
Clinical Evidence
Bench testing only. Temperature measurements conducted on test candidates confirmed the device reaches a therapeutic range of 40–45°C within 5 minutes and maintains this range for a 10-minute treatment duration.
Technological Characteristics
Class IV laser system; 810nm/980nm infrared output; 650nm aiming beam. Max output 7W. Plastic housing. Touch screen interface. SD card for data/updates. Complies with IEC/EN 60601-1, IEC/EN 60601-1-2, IEC 60601-2-22, and IEC 60825-1. Applied part type B. Protection class II.
Indications for Use
Indicated for patients requiring temporary relief of minor muscle/joint aches, pains, and stiffness; muscle spasms; pain/stiffness associated with arthritis; muscle tissue relaxation; and increased local blood circulation.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol depicts a stylized human figure with three faces in profile, representing the department's focus on health and human well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 24, 2014
Zimmer MedizinSysteme GmbH % Ms. Kirsten Langen Tuv Sud America Incorporated 1775 Old Highway, 8 North West New Brighton, Minnesota 55112
Re: K141564
Trade/Device Name: OptonPro Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: ILY Dated: September 10, 2014 Received: September 29, 2014
Dear Ms. Langen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
for
Sincerely yours,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K141564
Device Name OptonPro
Indications for Use (Describe)
- · temporary relief of minor muscle and joint aches, pains and stiffness
- · temporary relief of muscle spasm
- · temporary relief of minor pain and stiffness associated with arthritis
- · promoting relaxation of the muscle tissue
- · temporary increase in local blood circulation
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo consists of the word "zimmer" in a bold, sans-serif font, with a horizontal line above the "z" and "i". Below the word "zimmer" is the phrase "MedizinSysteme" in a smaller, lighter font. The logo is simple and modern, and the use of two different fonts creates a visual hierarchy.
Tel. 0731. 9761-0 Fax 0731.9761-118 info@zimmer.de www.zimmer.de
# 5. 510(k) Summary
The following information is provided as required in 21CFR807.87 and the "Guidance Document for Heating and Cooling Devices 510(k)s".
### a.) General
| 510(k) Submitter: | Zimmer MedizinSysteme GmbH<br>Junkersstrasse 9<br>D-89231 Neu-Ulm<br>Germany |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Mr. Armin Petraschka<br>Project Manager<br>Phone: +49-731-9761-140<br>Fax: +49-731-9761-4475<br>E-mail: a.petraschka@zimmer.de |
| Establishment<br>Registration: | 8010720 |
| Submission Date: | April 25th, 2014 |
| Device Names: | |
| Device Name: | <i>Opton Pro</i> |
| Trade Names: | <i>Opton Pro</i> |
| Common Names: | Heating and Cooling Devices |
| Regulation Numbers<br>and Classification<br>Names: | 21 CFR 890.5500 – Infrared Lamp |
| Classification: | Class II |
| Product Codes: | ILY |
| Panel: | 89 - Physical Medicine |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The word "zimmer" is in bold, gray letters. Underneath the word "zimmer" is the word "MedizinSysteme" in a smaller, gray font.
Fax 0731.9761-11
info@zimmer.de www.zimmer.de
#### b.) Predicate Devices
| 1. | 2. |
|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Device<br>K-1200 | Vectra Genisys Laser System |
| Manufacturer<br>ELTECH s.r.l<br>Via Castagnole, 20/H<br>31100 Trevisio, Italy | CHATTANOOGA GROUP (Encore Medical)<br>4717 Adams Road. P.O. Box 489<br>Hixson, Tennessee 37343-0489 |
| 510(k) number<br>K091497 | K040662 |
### c.) Device Description
OptonPro is a topical heat lamp with laser light. It emits energy in the infrared spectrum to provide topical heating for the purpose of elevating the tissue temperature.
The OptonPro contents a Class IV laser unit with four simultaneously operating diodes with wavelengths of 810nm and 980mm transmitting their energy to an applicator through a flexible fibre optic cable. The applicator has an output spot size of 10 mm.
The optimum ration of absorption and penetration depth quarantees efficient stimulation of tissue and pain receptors. The treatment effect of the OptonPro is achieved with 980nm and 810nm. Additionally the OptonPro contains an aiming beam with 650nm.
Skin type, indication and desired depth of treatment define the operating mode and the required energy. Two operating modes are possible: continuous wave and serial pulse. The duty cycle can be adapted additionally (1:1 – 1:10). Patient feedback and direct access to the treatment parameters enable to adjust the application for individual requirements and to achieve treatment times of a few minutes.
The OptonPro includes comprehensive safety features. In addition to multi-step activation of the laser by the software, with software security key lock and footswitch for starting treatment, protective glasses, an emergency off switch and a hardware interlock are included in the system. The performance of the laser unit can be tested with an integrated power control unit. A thermal threshold test offers a tool to assess the warmth sensitivity and the individual local absorption rate. Additionally the user is assisted with a special reminder function for a delivered particular amount of energy.
OptonPro is an intuitive therapy system with touch screen and menu-quided operation. The maximum output power is 7 W. The adjustable output power corresponds with the chosen duty cycle. The OptonPro has a large colour display with 12 inch. A SD card is used as a memory, as a logbook and for updates.
Individual adjustments can be made at any time and saved for easy access. A program menu with 3 fixed programs and a favourites and memory menu with 120 memory slots fur custom programs collects the most common settings and shows them when the OptonPro is started if required.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo consists of the word "zimmer" in a bold, sans-serif font, with a horizontal line above the "z" and "i". Below the word "zimmer" is the phrase "MedizinSysteme" in a smaller, sans-serif font. The logo is in grayscale.
Tel. 0731. 9761-0 Fax 0731.9761-118 info@zimmer.de www.zimmer.de
# d.) Statement of indications for use
The indications for use of the proposed device are the same as those for the predicate devices:
- temporary relief of minor muscle and joint aches, pains, and stiffness ●
- temporary relief of muscle spasm .
- temporary relief of minor pain and stiffness associated with arthritis .
- . promoting relaxation of the muscle tissue
- temporary increase of local blood circulation .
### e.) Comparison of Technological Characteristics
| No | Feature | OptonPro | K-1200 | Vectra Genisys Laser System |
|-----|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. | power source | 100 - 240 VAC, 50 / 60Hz | 100 - 240 VAC, 47 - 63 Hz | 120-240VAC, 50 / 60Hz |
| 2. | battery powered | - | + | + |
| 3. | dimensions<br>(h x w x d) [cm] | 30 x 35 x 20<br>(11.81 x 13.78 x 7.87 in) | 19 x 18 x 20<br>(7.5 x 7 x 8 in) | 16.3 x 28.8 x 32.8<br>(6.4 x 11.3 x 12.9 in) |
| 4. | weight [kg] | 3.8<br>(8.38 lb) | 1.3<br>(2.9 lb) | 2.3<br>(5.07 Ib) |
| 5. | housing materials | plastic | plastic | plastic |
| 6. | laser class | 4 | 4 | 3B |
| 7. | applied part | B | B | B |
| 8. | protection class | ll | ll | ll |
| 9. | laser system | laser diodes,<br>fibre-optical cable | laser diodes,<br>fibre-optical cable | laser diodes |
| 10. | software control | + | + | + |
| 11. | compliance with<br>voluntary<br>standards | Yes,<br>-IEC/EN 60601-1<br>-IEC/EN 60601-1-2<br>-IEC 60601-2-22<br>-IEC 60825-1 | Yes,<br>-IEC/EN 60601-1<br>-IEC 60825-1 | Yes,<br>-IEC/UL/EN 60601-1<br>-IEC/UL/EN 60601-1-2<br>-IEC/UL/EN 60601-2-22<br>-IEC 60825-1<br>-21CFR 1040.10 & 1040.11<br>-CAN/CSA C22.2 No.601.1-<br>M90 w/A2 |
| 12. | prescription use | + | + | + |
| 13. | intended use | emit energy in the infrared<br>spectrum to provide topical<br>heating for the purpose of<br>elevating tissue temperature<br>heating for the following<br>indications for use: | emit energy in the infrared<br>spectrum to provide topical<br>heating for the purpose of<br>elevating tissue temperature<br>heating for the following<br>indications for use: | indicated for topical heating for<br>the following indications for<br>use: |
| 14. | indications for<br>use | - temporary relief of minor<br>muscle and joint aches, pains<br>and stiffness<br>- temporary relief of muscle<br>spasm<br>- temporary relief of minor pain<br>and stiffness associated with<br>arthritis | - temporary relief of minor<br>muscle and joint pain<br>- temporary relief of muscle<br>spasm<br>- temporary relief of minor pain<br>and stiffness associated with<br>arthritis<br>- promoting relaxation of the | - temporary increase in local<br>blood circulation<br>- temporary relief of minor<br>muscle and joint aches, pains<br>and stiffness<br>- relaxation of muscles<br>- temporary relief of muscle<br>spasm |
| | | - promoting relaxation of the<br>muscle tissue<br>- temporary increase in local<br>blood circulation | muscle tissue<br>- temporary increase local<br>blood circulation | - temporary relief of minor pain<br>and stiffness associated with<br>arthritis |
| 15. | touch screen<br>interface | + | + | |
| 16. | color LCD screen | + | + | - |
| 17. | external memory | + | + | - |
| 18. | functional cart<br>option | + | + | + |
| 19. | number of output<br>channels | 1 | 1 | 1 |
| 20. | modifiable<br>waveform<br>parameters /<br>customize<br>treatment<br>parameters | + | + | + |
| 21. | possibility to<br>switch the laser<br>on and off | + | + | + |
| 22. | interlock | + | + | + |
| 23. | emergency off<br>switch | + | + | + |
| 24. | laser warning<br>signal | + | + | + |
| 25. | spacer | + (optional) | + | - |
| 26. | number of output<br>modes | 2 | 3 | 2 |
| 27. | output modes | continuous wave, serial pulse, | continuous wave, serial pulse,<br>intense super pulse | continuous wave, serial pulse |
| 28. | favorites / named<br>user defined<br>programs | + | + | + |
| 29. | laser<br>performance<br>testing | + | N/A | + |
| 30. | electronic key<br>lock | + | N/A | + |
| 31. | calculated and<br>displayed energy | + | + | + |
| 32. | displayed<br>treatment time | + | + | + |
| 33. | wavelength<br>aiming beam [nm] | 650 | 635 or 650 | 670 |
| 34. | wavelength laser<br>beam [nm] | 810 and 980 | 800 and / or 970 | 820 or 850 |
| 35. | output power [W] | 1 - 7 | 0.1 - 12 | 0.1 - 1.44 |
| 36. | repetition rate f<br>(Hz) | 1 - 50<br>CW | 1 - 20 000<br>CW<br>Intense Super Pulse (ISP) | 8 - 10 000<br>CW |
| 37. | area output spot<br>size [cm2] | 0.8 | 1 - 5 | 0.07 - 31.2 |
| 38. | duty cycle [%] | 10 - 50 | 50 | 90 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The word "zimmer" is in a bold, sans-serif font, with a horizontal line above the "z". Below the word "zimmer" is the word "MedizinSysteme" in a smaller, sans-serif font. The logo is in grayscale.
Tel. 0731. 9761-0
Fax 0731. 9761-118
info@zimmer.de
www.zimmer.de
5
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo is in a dark gray color. The word "zimmer" is in a bold, sans-serif font, and the words "MedizinSysteme" are in a smaller, lighter font below the word "zimmer."
Tel 0731 9761-0 Fax 0731.9761-118
info@zimmer.de www.zimmer.de
# f.) Summary Comparison with predicate devices
The OptonPro shares the same intended use and the same or similar basic characteristics and features as the predicate devices. In addition, any differences in their technological characteristics are explained to demonstrate in this submission that these differences do not raise any new questions of safety and effectiveness.
# q.) Non-clinical Tests Performed
Validation documentation, product testing and a comparison of the technical characteristics and features according to relevant standards were provided to demonstrate that OptonPro is as safe and effective in its intended use.
The OptonPro achieves a therapeutic temperature range of 40 – 45°C as accepted by the FDA.
The maximal transition time for reaching this therapeutic temperature range is 5 min.
The therapeutic temperature range was maintained for a recommended treatment time of 10 min.
The temperature measurements were conducted on a few test candidates on different physical locations.
# h.) Conclusion Substantial Equivalence
Drawn from the comparison between the predicate devices and the OptonPro devices it demonstrates that the OptonPro device is as safe and effective as the predicate devices and therefore are substantially equivalent to the compared devices on the basis of similarities in operating principles, intended use and functional performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.