SOLID STATE X-RAY IMAGING DEVICE, DIGITAL FLAT PANEL X-RAY DETECTOR, 1417WGC
Applicant
Rayence Co., Ltd.
Product Code
MQB · Radiology
Decision Date
Jul 16, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
1417WGC Digital Flat Panel X-Ray Detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.
Device Story
1417WGC is a wired/wireless digital solid-state X-ray detector using a scintillator coupled to an a-Si TFT sensor. It captures X-ray images, digitizing them into RAW files for transmission via wireless LAN (IEEE 802.11a/g/n) or wired connection to a separate console system. The device is integrated into general radiographic systems and operated by healthcare professionals. It features a thinner, lighter design than its predecessor, with an internal battery and dedicated power supply (RP003A). The detector supports 14-bit and 16-bit A/D conversion. Output is processed by external DICOM-compatible software for diagnostic analysis. By providing digital images, it replaces traditional film/screen systems, potentially improving workflow efficiency and diagnostic clarity through enhanced spatial resolution and soft tissue contrast.
Clinical Evidence
Bench testing included MTF, DQE, and NPS analysis per IEC 6220-1, showing 1417WGC performance is equivalent or superior to the predicate (DQE 49% vs 46% at lowest spatial frequency). Clinical evidence consisted of a comparative review by a licensed US radiologist of sample radiographs taken with both 1417WGC and 1417WGA across similar patient age groups and anatomical structures. Results indicated 1417WGC images were comparable or superior in spatial resolution and soft tissue contrast.
Technological Characteristics
Digital flat-panel X-ray detector; scintillator coupled to a-Si TFT sensor. Dimensions: 460 x 384 x 15.4 mm; weight: 3.1 kg. Connectivity: Wireless LAN (IEEE 802.11a/g/n, 2.4/5 GHz, 2x2 MIMO) and wired Ethernet. A/D conversion: 14/16-bit. Power: Dedicated supply (RP003A, 100-240VAC input, 24VDC output). Safety/Performance standards: IEC 60601-1 (electrical/mechanical safety), IEC 60601-1-2 (EMC).
Indications for Use
Indicated for digital imaging in general radiographic systems for human anatomy. Intended to replace film/screen systems for general purpose diagnostic procedures. Contraindicated for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
510(k) Submission — 1417WGC
# 510(k) Summary
# JUL 16 2014
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date 510k summary prepared: June 11, 2014
Submitter's Name, address, telephone number, a contact person:
| Submitter's Name : | Rayence Co., Ltd. |
|-------------------------|---------------------------------------------------------|
| Submitter's Address: | 14, Samsung 1-ro 1-gil, Hwaseong-si, Gyeonggi-do, Korea |
| Submitter's Telephone: | +82-31-8015-6459 |
| Contact person: | Mr. Kee Dock Kim / Manager / +82-31-8015-6459 |
| Official Correspondent: | Dave Kim (davekim(@mtech-inc.net) |
| (U.S. Designated agent) | |
| Address: | 8310 Buffalo Speedway, Houston, TX 77025 |
| Telephone: | +713-467-2607 |
| Fax: | +713-583-8988 |
Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:
| Trade/proprietary name: | 1417WGC |
|-------------------------|-----------------------------------------|
| Common Name: | Digital Flat Panel X-ray Detector |
| Classification Name : | 21CFR892.1680 / Stationary x-ray system |
| Product Code: | MQB |
{1}------------------------------------------------
#### Predicate Device :
| Manufacturer | : Rayence Co., Ltd. |
|---------------|------------------------------------------------|
| Device | : 1417WGA |
| 510(k) Number | : K131114 (Decision Date - September 17, 2013) |
### Device Description :
1417WGC is a wired/wireless digital solid state X-ray detector that is based on flat-panel technology. The wireless LAN(IEEE 802.11a/g/n) communication signals images captured to the system and improves the user operability through high-speed processing. This radiographic image detector and processing unit consists of a scintillator coupled to an a-Si TFT sensor. This device needs to be integrated with a radiographic imaging system. It can be utilized to capture and digitalize X-ray images for radiographic diagnosis The RAW files can be further processed as DICOM compatible image files by separate console SW (not part of this 510k submission) for a radiographic diagnosis and analysis.
#### Indication for use :
1417WGC Digital Flat Panel X-Ray Detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.
{2}------------------------------------------------
#### 1. Summary of Design Control Risk management
The 1417WGC digital X-ray detector is a modification of 1417WGA (K131114). The new device is designed with a thinner and lighter body compared to its predecessor, 1417WGA. The battery of 1417WGA was attached on the back of the device but the newly designed 1417WGC device has its battery inserted inside the new device.
In addition, a dedicated power supply (model name: RP003A) for 1417WGC is developed and added to its component list. (Optional user selection).
The risks and the hazardous impact of the device modification were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design change were successfully mitigated and accepted.
# 2. Summary of the technological characteristics of the device compared to the predicate device:
The 1417WGC SSXI detector described in this 510(k) has the same indications for use and technical characteristics as its predicate device, 1417WGA flat panel detector, of Rayence Co., Ltd.
#### 2.1 Detector
The mechanical design for each device is different as following.
Image /page/2/Figure/9 description: This image shows an exploded diagram of a product, with labels indicating the different components. The components are labeled 1 through 6, and include the carbon plate, carbon case, PC sheet assembly, middle block assembly, main board, and cap assembly. The image also includes the product names 1417WGC and 1417WGA, along with their dimensions and weights. The dimensions for 1417WGC are 460 x 384 x 15.4 mm with a weight of 3.1 kg, while the dimensions for 1417WGA are 460 x 417 x 15.9 mm with a weight of 3.9 kg.
{3}------------------------------------------------
Image /page/3/Figure/1 description: This image shows an exploded view of a device with its components labeled. The components are: 1. Carbon Plate, 2. Front cover, 3. Middle block & Panel, 4. Rear Cover, 5. Battery Pack, and 6. Handle. The exploded view shows how the components fit together to form the device.
- 2.2 Power supply
# 1417WGC requires power supply (Model Name: RP003A)
| | 1417WGC | 1417WGA |
|--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Model Name | RP003A | |
| | Image: RP003A device | |
| | Image: Connectors | The power supply<br>unit is not provided<br>from Rayence. |
| | ① Power plug connector: Connect with<br>AC power supply cord<br><br>② LAN connector : Ethernet port for<br>transmitting the image/command between<br>detector and PC<br><br>③ Link cable connector: Power connector<br>for detector operating<br><br>④ LED indicator: Four LEDs to display<br>the status<br>*Changeable part: Fuse T3.15 AL 250V | |
| Manufacturer | Rayence Co., Ltd. | |
{4}------------------------------------------------
#### 510(k) Submission - 1417WGC
| Dimension | 188 X 92 X 41.5 mm | | | |
|-----------------------|---------------------------------------------------------------------|---------------------------------|------|-----|
| Weight | 0.46 kg | | | |
| | | * Use CE or UL approved product | | |
| Power<br>requirements | Input rate : 100~240VAC (50/60Hz)<br>Output : Typ. 24VDC (Max 1.6A) | | Min | Max |
| | | Voltage | 18V | 24V |
| | | Current | 1.9A | |
The 1417WGC flat panel detector is equipped with a dedicated power supply (Model Name: RP003A). The electromagnetic compatibility test for the new device with the power supply has been conducted and the test report is included in this submission. The risk factors associated with the power supply such as power overload and overheating have been assessed and control measures to mitigate risks are discussed.
### 2.3 A/D conversion (analog-to-digital conversion)
The detector displays an image by digitizing and recording DN value of the analog electrical signal measured. The predicate device used 14 bit series to represent 16,384 values on the gray scale whereas the newly improved subject device is capable of processing both 14 and 16 bit (65,536 gray scale values) series based on the user needs and market requirements.
| Characteristic | Proposed<br>Rayence Co.,Ltd.<br>1417WGC | Predicate<br>Rayence Co.,Ltd.<br>1417WGA |
|-------------------|-----------------------------------------|------------------------------------------|
| A/D<br>conversion | 14 / 16 bit | 14 bit |
### 2.4 Recommended Generator Specification
| Model | Manufacture | Specification | | |
|---------|--------------------------------------|---------------|--------|--------|
| CMP 200 | Communications & Power<br>Industries | 32kW | 40kW | 50kW |
| | | kVp | 40-125 | 40-150 |
| | | mA | 10-400 | 10-500 |
{5}------------------------------------------------
# 510(k) Submission - 1417WGC
| EDITOR HFe 501 | Rontgenwerk Bochum | kVp | 40-150 |
|----------------|--------------------|-----|--------------------|
| | | mA | 10-630 |
| UD150L-40E/40F | Shimadzu | kVp | 40-150 |
| | | mA | @100 kVp- 500(320) |
| | | mA | @80 kVp- 630(400) |
| PXR-321B | Poskom Co.,Ltd. | kVp | 125/150 |
| | | mA | 500 |
ा
Note: To our best knowledge, the 1417WGC detector is compatible with the X-ray generators with the specifications described above. For any questions regarding the compatibility issue for other generators which are not listed above, please contact Rayence representative.
# 3. Proposed Labeling- Changes in comparison to the predicate device
| Item | Differenc<br>e | 1417WGC | 1417WGA |
|-------------------------|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Devi<br>ce<br>Labe<br>1 | Detector | Image: Detector label for 1417WGC | Image: Detector label for 1417WGA |
| | Power<br>Supply | Image: Power supply label for 1417WGC | - |
| Man<br>ual | Indicatio<br>ns for<br>Use | 1417WGC Digital Flat Panel<br>X-<br>Ray Detector is indicated for<br>digital imaging solution<br>designed for general<br>radiographic system for<br>human anatomy. It is intended<br>to replace film or screen based<br>radiographic systems in all<br>general purpose diagnostic | 1417WGA Digital Flat Panel X-<br>Ray Detector is indicated for<br>digital imaging solution designed<br>for general radiographic system<br>for human anatomy. It is<br>intended to replace film or screen<br>based radiographic systems in all<br>general purpose diagnostic<br>procedures. Not to be used for<br>mammography. |
{6}------------------------------------------------
| | | procedures. Not to be used for<br>mammography. | |
|----------------------|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| | Product<br>Specificat<br>ion | (page 25)<br>Pixel matrix: 3328 x 2816<br>Pixel area (active): 415 x 350<br>mm<br>Dimensions : 460 x 384 x<br>15.4 mm<br>Weight: 3.1 kg | (page 25)<br>Pixel matrix: 3328 x 2816<br>Pixel area (active): 415 x 350<br>mm<br>Dimensions : 460 x 417 x 15.9<br>mm<br>Weight: 3.9 kg |
| Wire<br>less<br>Spec | Standard | 802.11a/g/n compliance | 802.11a/g/n compliance |
| | Peak<br>Rate | 300Mbps | 300Mbps |
| | Frequenc<br>y | 2.4 GHz / 5 GHz | 2.4 GHz / 5 GHz |
| | Bandwidt<br>h | 20MHz/40MHz | 20MHz/40MHz |
| | MIMO | 2×2 | 2×2 |
### 4. Summary of Performance Testing
The 1417WGC flat panel detector is a modified version of 1417WGA (K131114), FDA cleared predicate device from Rayence. Indications for use, material, form factor, performance, and safety characteristics between 1417WGC and 1417WGA are identical.
The non-clinical test report and clinical consideration report were prepared and submitted to FDA separately to demonstrate the substantial equivalency between two similar detectors. The nonclinical test report contains the MTF, DQE and NPS test results of 1417WGC and 1417WGA by using the identical test equipment and same analysis method described by IEC 6220-1. The comparison of the MTF for 1417WGC and 1417WGA detector demonstrated that the MTF of the 1417WGA detector performed almost same with 1417WGC. Therefore, the overall resolution performance and sharpness of 1417WGC is almost same with 1417WGA. The DQE represents the ability to visualize object details of a certain size and contrast. 1417WGC demonstrated higher DOE performance than 1417WGA at various spatial frequencies and provides almost same Signalto Noise Ratio (SNR) transfer from the input to the output of a detector as a function of frequency. At the lowest spatial frequency, 1417WGC has a DQE of 49% and that of 1417WGA is 46%. 1417WGC also exhibited NPS which has almost same performance with 1417WGA. Therefore, the image quality of 1417WGC is found to be substantially equivalent to 1417WGA at the same patient exposure.
To further demonstrate the substantial equivalency of two devices, clinical images are taken from both devices and reviewed by a licensed US radiologist to render an expert opinion. Both the test subject (1417WGC) and control group (1417WGA) are evaluated and compared by taking sample radiographs of similar age groups and anatomical structures in accordance with the test protocol of diagnostic radiography evaluation procedure.
{7}------------------------------------------------
Based on the non-clinical and clinical consideration test and the outcome of a comparative review by an expert for both devices, we can claim the substantial equivalency between 1417WGC and its predicate device, 1417WGA in terms of image quality.
After comparing a broad review of plain radiographic images taken with the 1417WGC and the 1417WGA, the images obtained with the 1417WGC are comparable or superior to the same view obtained from a similar patient with the 1417WGA. In general, both the spatial resolution and soft tissue contrast are superior using the 1417WGC. Specifically, the soft tissues on extremity films were seen with better clarity. There is no difficulty in evaluating a wide range of anatomic structures necessary to provide a correct conclusion.
The manufacturing facility is in conformance with the design control procedure requirements and the relevant EPRC standards as specified in 21 CFR 802.30 and the records are available for review.
## 5. Summary for any testing in the submission:
Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1: 2005 + CORR.1(2006) + CORR(2007) (Medical electrical equipment Part 1:General requirements for basic safety and essential performance) was performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2:2007.
Non-clinical & Clinical considerations according to FDA Guidance "Guidance for the Submissions of 510(k)'s for Solid State X-ray Imaging Devices" was performed.
All test results were satisfactory.
#### 6. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Rayence Co., Ltd. concludes that 1417WGC is safe and effective and substantially equivalent in comparison with 1417WGA, the predicate device as described herein.
8
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines above two wavy lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 16, 2014
Rayence Co., Ltd. % Mr. Dave Kim Medical Device Regulatory Affairs 8310 Buffalo Speedway HOUSTON TX 77025
Re: K141563
Trade/Device Name: Solid State X-ray Imaging Device, Digital Flat Panel X-ray Detector,
1417wgc Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: June 12, 2014 Received: June 16, 2014
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic form in the quand of oversions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{9}------------------------------------------------
Page 2-Mr. Kim
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known) K141563
#### Device Name
1417WGC Digital Flat Panel X-Ray Detector
Indications for Use (Describe)
1417WGC Digital Flat Parel X-Ray Detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 807 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)
\$(m,p)\$
{11}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.