K141555 · Lutronic Corporation · GEX · Nov 6, 2014 · General, Plastic Surgery
Device Facts
Record ID
K141555
Device Name
ADVANTAGE
Applicant
Lutronic Corporation
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 6, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ADVANTAGE Laser System is indicated for use in surgical, aesthetic, and cosmetic applications in the medical specialties of general and plastic surgery, and dermatology. The ADVANTAGE Laser System is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin. The ADVANTAGE Laser System with D1-800 Handpiece is indicated for the treatment of vascular lesions, including angiomas, hemangiomas, telangiectasia and other benign vascular lesions, and the treatment for pseudofolliculitis barbae. The ADVANTAGE Laser System with D1-800 Handpiece is also indicated for hair removal, permanent hair reduction defined as the long-term stable reduction in the number of hairs re-growing after the last treatment measured at 6, 9, and 12 months, and the treatment of benign pigmented lesions and leg veins. The ADVANTAGE Laser System with D3-800 Handpiece is indicated for hair removal and permanent hair reduction defined as the long-term stable reduction in the number of hairs re-growing after the last treatment measured at 6, 9, and 12 months. The ADVANTAGE Laser System with D1-1064 Handpiece is intended for treatment of vascular lesions, including angiomas, hemangiomas, telangiectasia, port wine stains, leg veins and other benign vascular lesions. The ADVANTAGE Laser System with D1-1064 Handpiece is intended for hair removal, permanent hair reduction, and the treatment of Pseudofolliculitis Barbae (PFB). Permanent hair reduction is defined as the long-term. stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The ADVANTAGE Laser System with D1-1064 Handpiece is intended for the treatment of benign pigmented lesions, including age spots, solar lentigines, café-au-lait spots, nevi of Ota/Ito, melasma, Becker's nevi and other benign pigmented lesions. The ADVANTAGE Laser System with D1-1064 Handpiece is also intended for treatment of wrinkles. The ADVANTAGE Laser System with D1-1064 Handpiece is intended for use on all skin types (Fitzpatrick skin types I-VI), including tanned skin.
Device Story
ADVANTAGE Laser System is a powered laser surgical instrument; delivers laser energy via interchangeable handpieces (D1-800, D3-800, D1-1064). D1-1064 handpiece utilizes 1064 nm wavelength; 10x10 mm tip; fluence up to 100 J/cm2; pulse duration 5-400 msec; repetition rate up to 3 Hz. System includes water-cooled active skin cooling. Operated by clinicians in medical settings. Laser energy targets vascular/pigmented lesions, hair follicles, or skin tissue for wrinkle treatment. Clinical benefit includes permanent hair reduction and lesion/wrinkle improvement. Output is controlled laser energy delivery; provider monitors treatment area visually to guide application.
Clinical Evidence
No clinical data presented; bench testing only.
Technological Characteristics
Powered laser surgical instrument; 1064 nm wavelength (D1-1064 handpiece); 10x10 mm spot size; 100 J/cm2 max fluence; 5-400 msec pulse duration; 3 Hz max repetition rate. Features active water-cooling system. Class II device.
Indications for Use
Indicated for surgical, aesthetic, and cosmetic applications in general/plastic surgery and dermatology. Patient population: all skin types (Fitzpatrick I-VI), including tanned skin. Specific indications: vascular lesions (angiomas, hemangiomas, telangiectasia, port wine stains, leg veins), pseudofolliculitis barbae, hair removal/permanent hair reduction, benign pigmented lesions (age spots, solar lentigines, café-au-lait spots, nevi of Ota/Ito, melasma, Becker's nevi), and wrinkles.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Lutronic Corporation ADVANTAGE Laser System (K113502)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing right, with flowing lines suggesting movement or connection. The profiles are stacked on top of each other, creating a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 6, 2014
Lutronic Corporation Dr. Jhung Won Vojir Global Regulatory Officer 6 Neshaminy Interplex, Suite 100 Trevose, Pennsylvania 19053
Re: K141555
Trade/Device Name: ADVANTAGE Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: October 10, 2014 Received: October 14, 2014
Dear Dr. Jhung Won Vojir:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known):
Device Name: ADVANTAGE Laser System
Indications for Use:
The ADVANTAGE Laser System is indicated for use in surgical, aesthetic, and cosmetic applications in the medical specialties of general and plastic surgery, and dermatology. The ADVANTAGE Laser System is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin.
The ADVANTAGE Laser System with D1-800 Handpiece is indicated for the treatment of vascular lesions, including angiomas, hemangiomas, telangiectasia and other benign vascular lesions, and the treatment for pseudofolliculitis barbae. The ADVANTAGE Laser System with D1-800 Handpiece is also indicated for hair removal, permanent hair reduction defined as the long-term stable reduction in the number of hairs re-growing after the last treatment measured at 6, 9, and 12 months, and the treatment of benign pigmented lesions and leg veins.
The ADVANTAGE Laser System with D3-800 Handpiece is indicated for hair removal and permanent hair reduction defined as the long-term stable reduction in the number of hairs re-growing after the last treatment measured at 6, 9, and 12 months.
The ADVANTAGE Laser System with D1-1064 Handpiece is intended for treatment of vascular lesions, including angiomas, hemangiomas, telangiectasia, port wine stains, leg veins and other benign vascular lesions.
The ADVANTAGE Laser System with D1-1064 Handpiece is intended for hair removal, permanent hair reduction, and the treatment of Pseudofolliculitis Barbae (PFB). Permanent hair reduction is defined as the long-term. stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over The Counter Use (Part 21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
The ADVANTAGE Laser System with D1-1064 Handpiece is intended for the treatment of benign pigmented lesions, including age spots, solar lentigines, café-au-lait spots, nevi of Ota/Ito, melasma, Becker's nevi and other benign pigmented lesions. The ADVANTAGE Laser System with D1-1064 Handpiece is also intended for treatment of wrinkles.
The ADVANTAGE Laser System with D1-1064 Handpiece is intended for use on all skin types (Fitzpatrick skin types I-VI), including tanned skin.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over The Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
## 510(k) Summary for the Lutronic Corporation ADVANTAGE Laser System
This 510(k) Summary is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.
## 1. General Information
2.
| Applicant: | Lutronic Corporation<br>219, Sowon-ro<br>Deogyang-gu<br>Goyang-si, Gyeonggi-do 410-722<br>Republic of Korea |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jhung Won Vojir, Ph.D.<br>Global Regulatory Officer<br>Lutronic, Inc.<br>6 Neshaminy Interplex, Suite 100<br>Trevose, PA 19053<br>Telephone:<br>215-205-2219<br>Fax:<br>609-488-6958<br>Email: jvojir@lutronic.com) |
| Summary Preparation Date: | June 11, 2014 |
| Names | |
| Device Name: | ADVANTAGE Laser System |
| Classification Name: | Powered laser surgical instrument |
| Regulation Name: | Laser surgical instrument for use in general<br>and plastic surgery and in dermatology |
| Common/Usual Name: | Pulse Diode Array Laser |
| Product Code: | GEX |
| Regulation No.: | 878.4810 |
| Class: | II |
| Panel Identification: | General & Plastic Surgery |
{5}------------------------------------------------
#### 3. Predicate Devices
Lutronic Corporation ADVANTAGE Laser System, cleared in 510(k) K113502 and the Lumenis Ltd. ET LighSheer 1060, cleared in 510(k) K133319.
### 4. Device Description
The ADVANTAGE Laser System with D1-1064 Handpiece is a treatment handpiece, intended to be used with ADVANTAGE Laser System 510(k) cleared in K113502 on April 5, 2012. The ADVANTAGE Laser System with D1-1064 Handpiece delivers laser energy through a 10 x 10 mm tip with a fluence of up to 100 J/cm2. The settings for this handpiece are pulse duration from 5-400 msec, and a pulse repetition rate up to 3 Hz maximum. The ADVANTAGE Laser System with D1-1064 Handpiece is water-cooled to provide active skin cooling.
#### 5. Indications for Use
The ADVANTAGE Laser System is indicated for use in surgical, aesthetic, and cosmetic applications in the medical specialties of general and plastic surgery, and dermatology. The ADVANTAGE Laser System is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin.
The ADVANTAGE Laser System with D1-800 Handpiece is indicated for the treatment of vascular lesions, including angiomas, hemangiomas, telangiectasia and other benign vascular lesions, and the treatment for pseudofolliculitis barbae. The ADVANTAGE Laser System with D1-800 Handpiece is also indicated for hair removal, permanent hair reduction defined as the long-term stable reduction in the number of hairs re-growing after the last treatment measured at 6, 9, and 12 months, and the treatment of benign pigmented lesions and leg veins.
The ADVANTAGE Laser System with the D3-800 Handpiece is indicated for hair removal and permanent hair reduction defined as the long-term stable reduction in the number of hairs re-growing after the last treatment measured at 6, 9. and 12 months.
{6}------------------------------------------------
The ADVANTAGE Laser System with D1-1064 Handpiece is intended for treatment of vascular lesions, including angiomas, hemangiomas, telangiectasia, port wine stains, leg veins and other benign vascular lesions.
The ADVANTAGE Laser System with D1-1064 Handpiece is intended for hair removal, permanent hair reduction, and the treatment of Pseudofolliculitis Barbae (PFB). Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
The ADVANTAGE Laser System with D1-1064 Handpiece is intended for the treatment of benign pigmented lesions, including age spots, solar lentigines, caféau-lait spots, nevi of Ota/Ito, melasma, Becker's nevi and other benign pigmented lesions.
The ADVANTAGE Laser System with D1-1064 Handpiece is also intended for treatment of wrinkles.
The ADVANTAGE Laser System with D1-1064 Handpiece is intended for use on all skin types (Fitzpatrick skin types I-VI), including tanned skin.
#### 6. Substantial Equivalence
The ADVANTAGE Laser System with D1-800 Handpiece and D3-800 Handpiece is substantially equivalent to the previously cleared ADAVANTAGE Laser System, cleared in 510(k) K113502. The ADVANTAGE Laser System with D1-800 handpiece and the D3-800 handpiece has the same intended use as the predicate device.
The ADVANTAGE Laser System with D1-1064 Handpiece has the same principles of use and mechanisms of operation as Lutronic Corporation treatment handpieces cleared in 510(k) K113502, and has equivalent performance characteristics. The materials used for ADVANTAGE Laser System with D1-1064 nm and the manufacturing methods are identical to the devices cleared in 510(k) K113502. The difference resides in the wavelength of 1064 nm for the proposed device in comparison with 805 nm for the cleared treatment handpieces. The ADVANTAGE Laser System with D1-1064 Handpiece has the same intended use and have equivalent performance characteristics as the Lumenis Ltd. ET LightSheer cleared in 510(k) K13319.
#### 7. Performance Data
None presented.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.