← Product Code [OUR](/productcode/OUR) · K141549

# SImmetry Sacroiliac Joint Fusion System (K141549)

_Zyga Technology, Inc. · OUR · Jan 15, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K141549

## Device Facts

- **Applicant:** Zyga Technology, Inc.
- **Product Code:** [OUR](/productcode/OUR.md)
- **Decision Date:** Jan 15, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The SImmetry® Sacroiliac Joint Fusion System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.

## Device Story

SImmetry Sacroiliac Joint Fusion System consists of sterile, cannulated, self-tapping titanium alloy (Ti-6Al-4V ELI) lag or fully threaded implants; designed to transfix sacrum and ilium to provide stability for bony fusion. Implants available in various diameters (6.5mm–14.5mm) and lengths (30mm–110mm) to accommodate patient anatomy; some include pre-assembled or separate washers. Used by surgeons in clinical settings to treat sacroiliac joint disruptions and degenerative sacroiliitis. Implants provide mechanical stabilization of the joint, facilitating fusion. Performance verified via bench testing (insertion torque, axial pullout strength, torque to failure, bending yield strength, bending fatigue life) and cadaveric testing to ensure structural integrity under expected physiological loads.

## Clinical Evidence

Bench testing only. Testing included insertion torque, axial pullout strength, torque to failure, mean bending yield strength, and bending fatigue life. Cadaveric testing was performed to validate the surgical implantation technique.

## Technological Characteristics

Material: Titanium Alloy (Ti-6Al-4V ELI). Design: Cannulated, self-tapping, lag or fully threaded bone fixation fasteners. Dimensions: 6.5mm–14.5mm diameter, 30mm–110mm length. Features: Optional washers, pre-assembled or separate. Principle: Mechanical transfixation of the sacrum and ilium for stabilization. Sterilization: Sterile packaged.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- SImmetry Sacroiliac Joint Fusion System ([K130092](/device/K130092.md))
- SI-LOK Sacroiliac Joint Fixation System ([K112028](/device/K112028.md))
- Pioneer Cannulated Screw System ([K102903](/device/K102903.md))
- Synthes 6.5mm Cannulated Screw ([K021932](/device/K021932.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, resembling a bird-like shape.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 15, 2015

Zyga Technology, Incorporated Ms. Diane Brinza Director of Regulatory Affairs 5600 Rowland Road, Suite 200 Minnetonka, Minnesota 55343

Re: K141549

Trade/Device Name: SImmetry® Sacroiliac Joint Fusion System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: OUR Dated: December 12, 2014 Received: December 15, 2014

Dear Ms. Brinza:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Ms. Diane Brinza

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for Zyga. The logo is in orange and features the company name in a stylized font. The letters are connected and have a rounded appearance. The letters are all lowercase except for the Z, which is uppercase. The logo also includes the trademark symbol.

## INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K141549

Device Name: SImmetry® Sacroiliac Joint Fusion System

Indications For Use: The SImmetry® Sacroiliac Joint Fusion System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.

Prescription Use AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "zyga" in orange font. The font is sans-serif and the letters are connected. The background is white. The logo is simple and modern.

# 510(k) Summary

### [As required by 21 CFR 807.92(c)]

### Submitter / Contact Person / Date of Preparation

| Submitter           | Zyga Technology, Inc.<br>5600  Rowland Road, Suite 200<br>Minnetonka, MN 55343                              |
|---------------------|-------------------------------------------------------------------------------------------------------------|
| Contact Person      | Diane Brinza<br>Director of Regulatory Affairs<br>Ph. 952-698-9959<br>Fax. 952-698-9940<br>dbrinza@zyga.com |
| Date of Preparation | January 14, 2015                                                                                            |

### General Information

| Trade Name                                      | SImmetry® Sacroiliac Joint Fusion System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common / Usual<br>Name                          | Sacroiliac Joint Fixation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Classification                                  | 21 CFR Part 888.3040<br>Smooth or threaded metallic bone fixation fastener                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Regulatory Class                                | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Product Code                                    | OUR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Manufacturer                                    | Zyga Technology, Inc.<br>5600 Rowland Road, Suite 200<br>Minnetonka, MN 55343                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Identification of<br>Predicate Devices          | Zyga Technology, Inc. - Slmmetry Sacroiliac Joint Fusion<br>System (K130092)<br>Globus Medical, Inc. - SI-LOK Sacroiliac Joint Fixation<br>System (K112028)<br>Pioneer Surgical Technology, Inc. – Pioneer Cannulated<br>Screw System (K102903)<br>Synthes (USA) - Synthes 6.5mm Cannulated Screw<br>(K021932)                                                                                                                                                                                                                                                                                                                                                                                                           |
| Device Description                              | The SImmetry Sacroiliac Joint Fusion System consists of<br>sterile packaged lag or fully threaded, self-tapping<br>cannulated titanium implants designed to transfix the<br>sacrum and ilium, providing stability for bony fusion. The<br>surqical implants are available in various sizes to<br>accommodate patient anatomy. Implants have major                                                                                                                                                                                                                                                                                                                                                                        |
|                                                 | diameters ranging from 6.5mm - 14.5mm, in 2mm<br>increments. Lengths in 5mm increments range from 30mm-<br>110mm for fully threaded implants and 50mm to 110mm<br>for partially threaded implants. All partially threaded<br>implants and 6.5mm diameter fully threaded implants have<br>a pre-assembled washer. Individually sterile packaged<br>washers are available for fully threaded implants having<br>diameters ranging from 8.5mm - 14.5mm. All devices are<br>manufactured from Titanium Alloy (Ti-6Al-4V ELI).                                                                                                                                                                                                |
| Intended Use                                    | The SImmetry Sacroiliac Joint Fusion System is intended<br>for sacroiliac joint fusion for conditions including sacroiliac<br>joint disruptions and degenerative sacroiliitis.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Summary of Non-<br>clinical<br>Performance Data | Tests completed for insertion torque and axial pullout<br>strength demonstrate the subject devices are substantially<br>equivalent with respect to mechanical performance of the<br>K130092 predicate device. Cadaveric testing demonstrates<br>the subject devices can be safely implanted using the<br>SImmetry surgical technique. Additionally testing for torque<br>to failure, mean bending yield strength and bending fatigue<br>life provide evidence the device is capable of withstanding<br>expected loading without failure. This performance data<br>demonstrates that the devices included in the SImmetry<br>Sacroiliac Joint Fusion System do not raise any new<br>questions of safety or effectiveness. |
| Technological<br>Comparison                     | The principle of operation of the subject devices is identical<br>to that of the predicate devices cleared under the following<br>submissions:<br>K130092 Slmmetry Sacroiliac Joint Fusion System K112028 SI-LOK Sacroiliac Joint Fixation System K102903 Pioneer Cannulated Screw System K021932 Synthes 6.5mm Cannulated Screw          Additionally the subject devices are comprised of the same<br>Titanium Alloy (Ti-6Al-4V ELI) and manufactured using the<br>same manufacturing processes as the K130092 predicate<br>device.                                                                                                                                                                                    |
| Conclusion                                      | Equivalence for the SImmetry System is based on<br>similarities in indications for use, design features,<br>operational principles, and material composition when<br>compared to the predicate devices cleared under the<br>following submissions:<br>K130092 Slmmetry Sacroiliac Joint Fusion System K112028 SI-LOK Sacroiliac Joint Fixation System K102903 Pioneer Cannulated Screw System K021932 Synthes 6.5mm Cannulated Screw                                                                                                                                                                                                                                                                                     |

5600 Rowland Road, State 200, Minnetorka, MN 55343 Phone 952.698.9942 Fax 952.698.9940 zyga.com

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "zyga" in orange color. The font is sans-serif and the letters are connected. The background is white. 

5600 Rowland Road, Siate 200, Minnetorka, MN 55343 Phone 952.698.9942 Fax 952.698.9940 zyga.com

---

**Source:** [https://fda-staging.innolitics.com/device/K141549](https://fda-staging.innolitics.com/device/K141549)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).