Alvimedica Tibbi Urunler Sanayi VE Dis Ticaret A.S
Product Code
DQY · Cardiovascular
Decision Date
Dec 3, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Alviguide™ Blue+ Interventional Cardiology Guiding Catheter is intended for use in the intravascular introduction of interventional/diagnostic devices into the coronary vascular system. The Alguide™ Endovascular Guiding Catheter is intended for use in the introduction of interventional/ diagnostic devices into the peripheral vascular system.
Device Story
Alviguide™ Blue+ and Alguide™ are sterile, single-lumen, single-use guide catheters. They facilitate the delivery of interventional or diagnostic devices into the coronary or peripheral vascular systems. Operated by physicians in clinical settings, these catheters provide a pathway for secondary devices. They feature a tri-layer construction (polymer inner layer, stainless steel wire braid, radiopaque outer layer) and a soft distal tip. Available in various French sizes (5-8F) and lengths (45-100cm) with multiple distal shape configurations. The device is inserted into the vasculature to guide other instruments to target sites, aiding in diagnostic or interventional procedures.
Clinical Evidence
Bench testing only. Testing included visual inspection, dimensional verification, withdrawal force, hub leakage, kink resistance, tensile strength, radiopacity, torque testing, packaging integrity, and EO sterilization validation. Biocompatibility testing per ISO 10993-1 for limited exposure (<24 hours) included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemocompatibility, and material-mediated pyrogenicity. Results demonstrate performance consistent with predicate devices.
Indicated for patients requiring intravascular introduction of interventional or diagnostic devices into the coronary vascular system (Alviguide™ Blue+) or peripheral vascular system (Alguide™).
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Medtronic 5F Sherpa Active NX Guide Catheter (K062420)
Merit Medical Systems Concierge Guiding Catheter [VasCon LLC Polaris Guiding Catheter (K043387)]
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with three profiles facing right, suggesting a sense of forward movement and progress. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 3, 2014
Alvimedica c/o Lori Adels Experien Group, LLC 755 N Mathilda Avenue. Suite #100 Sunnyvale, CA 94085
Re: K141529
Trade/Device Name: Alviguide™ Blue+ Interventional Cardiology Guiding Catheter and Alguide™ Endovascular Guiding Catheter Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter guide wire Regulatory Class: Class II Product Code: DQY Dated: October 31, 2014 Received: November 3, 2014
Dear Ms. Adels:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Melissa A. Torres -S
For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
Device Name
Alviguide™ Blue+ Interventional Cardiology Guiding Catheter, Alguide™ Endovascular Guiding Catheter
Indications for Use (Describe)
The Alviguide™ Blue+ Interventional Cardiology Guiding Catheter is intended for use in the intravascular introduction of interventional/diagnostic devices into the coronary vascular system.
The Alguide™ Endovascular Guiding Catheter is intended for use in the introduction of interventional/ diagnostic devices into the peripheral vascular system.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Notification K
## GENERAL INFORMATION
#### Applicant:
Alvimedica Tıbbi Ürünler Sanavi ve Dıs Ticaret A.S. İstanbul Trakya Serbest Bölgesi Ferhatpaşa Mahallesi Atatürk Bulvarı Manolya Sokak No:7 34540 Catalca, İstanbul Turkev Phone: +90 212 786 60 80 FAX: +90 212 786 60 90
#### Contact Person:
Lori E. Adels, Ph.D Regulatory Consultant for Alvimedica Experien Group, LLC 755 N. Mathilda Avenue, Suite 100 Sunnyvale, CA 94085 U.S.A. Phone: (408) 400-0856 ext. 112 FAX: (408) 400-0865
Date Prepared: June 6, 2014
## DEVICE INFORMATION
## Trade Name:
Alviguide™ Blue+ Interventional Cardiology Guiding Catheter Alguide™ Endovascular Guiding Catheter
## Classification Name:
Percutaneous catheter
## Generic/Common Name:
Guide catheter
## Classification:
Class II per 21 CFR§870.1250
Product Code: DOY
{4}------------------------------------------------
## PREDICATE DEVICES
- Medtronic 5F Sherpa Active NX Guide Catheter (K062420)
- Medtronic 6F Launcher Guide Catheter [Medtronic 6F Z3 Guide Catheter (K021256)]
- Medtronic 7F Launcher Guide Catheter (K022764) ●
- Medtronic 8F Launcher Guide Catheter (K023402) ●
- Merit Medical Systems Concierge Guiding Catheter ● [VasCon LLC Polaris Guiding Catheter (K043387)]
## INDICATIONS FOR USE
The Alviguide™ Blue+ Interventional Cardiology Guiding Catheter is intended for use in the intravascular introduction of interventional/diagnostic devices into the coronary vascular system.
The Alguide™ Endovascular Guiding Catheter is intended for use in the intravascular introduction of interventional/diagnostic devices into the peripheral vascular system.
## PRODUCT DESCRIPTION
The Alviguide™ Blue+ and Alguide™ Guide Catheters are sterile, non-pyrogenic, single lumen catheters with a tri-layer catheter body construction consisting of a polymer inner layer and a radiopaque outer layer surrounding a stainless steel wire braid, and a soft tip at the distal end. The catheters are for single-use only.
The Alviguide™ Blue+ and Alguide™ Guide Catheters are provided in a variety of distal shape configurations and are available in Fr Sizes 5, 6, 7 and 8 and lengths of 45, 60, 65 and 80cm (Alguide™), and 100cm (Alviguide™ Blue+ and Alguide™).
## TECHNOLOGICAL CHARACTERISTICS
The indication for use, design and materials used in the Alviguide Blue+ and Alguide Guide Catheters are similar to those of the predicate guide catheters.
# SUBSTANTIAL EQUIVALENCE
The indications for use for the predicate devices are substantially equivalent to the proposed indications for use for the Alviguide Blue+ and Alguide Guide Catheters. Anv differences in the technological characteristics between the devices do not raise any new issues of safety or effectiveness. Thus, the Alviguide Blue+ and Alguide Guide Catheters are substantially equivalent to the predicate devices.
{5}------------------------------------------------
## TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION
All necessary bench and clinical testing was conducted on the Alviguide Blue+ Guide Catheters to support a determination of substantial equivalence to the predicate devices.
## Nonclinical Testing Summary:
The nonclinical, bench testing included:
- Visual Inspection
- Dimensional Verification ●
- Catheter Sheath Introducer Withdrawal Force ●
- Hub Leakage
- Kink Diameter and Force ●
- Tensile Strength
- Radiopacity
- Torque Testing
- Packaging integrity (Visual inspection, seal strength test) ●
- Sterilization (EO sterilization evaluation, EO residuals, endotoxin)
## Biocompatibility Testing Summary:
Biocompatibility testing was conducted in compliance with ISO 10993-1, for externally communicating devices with limited exposure (<24 hours) to circulating blood, and included:
- Cytotoxicity ●
- Sensitization ●
- Intracutaneous Reactivity ●
- Acute Systemic Toxicity
- Hemocompatibility
- . Material Mediated Pyrogen
The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of the Alviguide Blue+ and Alguide Guide Catheters meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the Alviguide Blue+ and Alguide Guide Catheters do not raise new questions of safety or effectiveness when compared to the predicate devices.
## CONCLUSION
Based on the similar indication, design and materials, and the results of the bench testing, the Alviguide Blue+ Interventional Cardiology Guiding Catheter and Alguide Endovascular Guiding Catheter are considered substantially equivalent to the predicate devices listed above.
## SUMMARY
The Alviguide Blue+ Interventional Cardiology Guiding Catheter and the Alguide Endovascular Guiding Catheter are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.