K141520 · Alere Scarborough, Inc D/B/A Binax, Inc. · OCC · Jun 13, 2014 · Microbiology
Device Facts
Record ID
K141520
Device Name
ALERE I INFLUENZA A & B
Applicant
Alere Scarborough, Inc D/B/A Binax, Inc.
Product Code
OCC · Microbiology
Decision Date
Jun 13, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3980
Device Class
Class 2
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K141520 · Jun 13, 2014
ALERE I INFLUENZA A & B
Alere Scarborough, Inc D/B/A Binax, Inc.
Prospective clinical study of patients presenting with flu-like symptoms
The study evaluated the clinical performance of the Alere™ i Influenza A & B assay by testing nasal swab specimens collected from patients presenting with respiratory infection symptoms at eight U.S. clinical sites during the 2012-2013 influenza season.
Direct Nasal Swab Study; Multi-site prospective clinical study; Follow-up/Duration: 2012-2013 influenza season; Study Period: 2012-2013
Patients presenting at study centers with flu-like symptoms; Sample Size: 585 evaluable specimens; Number of Sites: 8
Viral culture
Sensitivity and specificity for detection of Influenza A and B
Indications for Use
The Alere™ i Influenza A & B assay performed on the Alere™ i Instrument is a rapid molecular in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology for the qualitative detection and discrimination of influenza A and B viral RNA in nasal swabs from patients with signs and symptoms of respiratory infection. It is intended for use as an aid in the differential diagnosis of influenza A and B viral infections in humans in conjunction with clinical and epidemiological risk factors. The assay is not intended to detect the presence of influenza C virus. Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions. Performance characteristics for influenza A were established during the 2012-2013 influenza season when influenza A/H3 and A/H1N1 pandemic were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary. If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
Device Story
Alere™ i Influenza A & B is a rapid, instrument-based isothermal nucleic acid amplification test for qualitative detection/differentiation of influenza A and B viral RNA. Input: nasal swab specimens. Process: sample eluted in buffer, transferred via cartridge to Test Base containing lyophilized reagents; Alere™ i Instrument performs automated heating, mixing, and fluorescence-based detection of amplified RNA using molecular beacons. Output: qualitative results for influenza A and B displayed on instrument screen; results stored in on-board archive; printable via USB-connected printer. Used in professional settings (medical labs, point-of-care) by healthcare providers. Benefits: rapid differential diagnosis of influenza A/B to aid clinical decision-making.
Clinical Evidence
Multi-site prospective study (n=585 evaluable specimens) compared Alere™ i Influenza A & B to viral culture. Influenza A sensitivity: 97.9% (95% CI: 92.6%-99.4%), specificity: 86.2% (95% CI: 82.8%-89.0%). Influenza B sensitivity: 92.5% (95% CI: 84.6%-96.5%), specificity: 96.5% (95% CI: 94.5%-97.8%). Discrepant results were resolved using an FDA-cleared RT-PCR assay.
Technological Characteristics
Isothermal nucleic acid amplification technology. System includes disposable Sample Receiver, Test Base (lyophilized reagents), Transfer Cartridge, and Alere™ i Instrument (reader). Detection via fluorescently labeled molecular beacons. Connectivity: USB port for printer, on-board data storage. Automated sample prep, amplification, and detection.
Indications for Use
Indicated for qualitative detection and discrimination of influenza A and B viral RNA in nasal swabs from patients of all ages presenting with signs and symptoms of respiratory infection. Not indicated for detection of influenza C.
Regulatory Classification
Identification
A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.
Special Controls
*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.
{0}------------------------------------------------
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K141520
#### SUBMITTER
Alere Scarborough, Inc. 10 Southgate Road Scarborough, ME 04074 Establishment Registration Number: 1221359
#### CONTACT PERSON
Angela Drysdale (207) 730-5737 (Office) (207) 730-5767 (FAX) Angela.drysdale@alere.com (email)
#### DATE PREPARED
6/10/2014
TRADE NAME Alere™ i Influenza A & B
COMMON NAME Alere™ i flu, Alere™ i, Alere™ Influenza A & B
#### CLASSIFICATION NAME
Respiratory Viral Panel Multiplex Nucleic Acid Assay (per 21 CFR 866.3980) Instrumentation for Clinical Multiplex Test Systems (per 21 CFR 862.2570)
CLASSIFICATION Class II
PRODUCT CODE OCC, OZE, OOI
PANEL Microbiology (83)
# PREDICATE DEVICE
IQuum Liat™ Influenza A/B Assay, K111387
{1}------------------------------------------------
#### DEVICE DESCRIPTION
Alere™ i Influenza A & B is a rapid, instrument-based isothermal test for the qualitative detection and differentiation of influenza A and influenza B from nasal swab specimens collected from patients presenting with signs and symptoms of respiratory infection. The Alere™ i Influenza A & B System utilizes isothermal nucleic acid amplification technology and is comprised of:
- Sample Receiver single use, disposable containing the elution buffer .
- . Test Base - single use, disposable comprising two sealed reaction tubes, each containing a lyophilized pellet
- Transfer Cartridge single use, disposable for transfer of the eluted sample to the Test Base, and .
- Alere™ i Instrument repeat use reader .
The reaction tubes in the Test Base contain the reagents required for amplification of the target nucleic acid and an internal control. Alere™ i Influenza A & B utilizes a pair of templates (similar to primers) for the specific amplification of RNA from influenza A and B and a fluorescently labeled molecular beacon designed to specifically identify the amplified RNA targets. Alere™ i Influenza A & B is performed within the confinement of the Test Base, and no other part of the Alere™ i Instrument has contact with the sample during the amplification process. This minimizes the risk of instrument contamination and sample carryover between measurements.
To perform the assay, the Sample Receiver and Test Base are inserted into the Alere™ i Instrument and the elution buffer is automatically the instrument. The sample is added to the Sample Receiver and transferred via the Transfer Cartridge to the Test Base, resuspending the lyophilized pellet contained within the Test Base and initiating target amplification. Heating, mixing and detection by fluorescence is provided by the instrument, with results automatically reported.
Results are displayed by the Alere™ i Instrument separately for influenza A and influenza B. Results are also stored in an on-board archive and are assigned to a sample ID that has been entered into the Alere™ i Instrument by the operator, and the date/time the test was performed. Data can be retrieved and downloaded by the operator at any time after testing. An external Alere™ Universal Printer can be attached via USB to the Alere™ i Instrument to print test results.
#### INTENDED USE
The Alere™ i Influenza A & B assay performed on the Alere™ i Instrument is a rapid molecular in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology for the qualitative detection and discrimination of influenza A and B viral RNA in nasal swabs from patients with signs and symptoms of respiratory infection. It is intended for use as an aid in the differential diagnosis of influenza A and B viral infections in humans in conjunction with clinical and epidemiological risk factors. The assay is not intended to detect the presence of influenza C virus.
Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.
Performance characteristics for influenza A were established during the 2012-2013 influenza season when influenza A/H3 and A/H1N1 pandemic were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary.
If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local
{2}------------------------------------------------
health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
# TECHNOLOGICAL CHARACTERISTICS
Alere™ i Influenza A & B and the predicate device, IQuum Liat™ Influenza A/B Assay, have the same intended use, indications for use, and utilize similar basic principles of operation. They are both molecular tests for the qualitative detection of influenza A and B viral nucleic acid.
#### DEVICE COMPARISON
Alere™ i Influenza A & B was compared to the legally marketed predicate device, the IQuum Liat ™ Influenza A/B Assay.
| Parameter | Alere™ i Influenza A & B | IQuum Liat ™ Influenza A/B Assay (K111387) |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Product Code | OCC, OZE, OO1 | OCC, OOI |
| Assay Target | Influenza A, Influenza B | Same |
| Intended Use | Alere™ i Influenza A & B is a rapid, instrument-based, molecular <i>in vitro</i> diagnostic test utilizing isothermal nucleic acid amplification technology for the qualitative detection of influenza A and B viral nucleic acid in nasal swab specimens. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections in humans in conjunction with clinical and epidemiological risk factors. The assay is not intended to detect the presence of influenza C virus Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.<br><br>Performance characteristics for influenza A were established during the 2012-2013 influenza season when influenza A/H3 and A/H1N1 pandemic were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary. If infection with a novel influenza A virus is suspected based on current clinical and | The IQuum Liat™ Influenza A/B Assay performed on the Liat™ Analyzer is an automated multiplex real-time RT-PCR assay for the rapid <i>in vitro</i> qualitative detection and discrimination of influenza A virus and influenza B virus RNA in nasopharyngeal swab specimens from patients with signs and symptoms of respiratory infection in conjunction with clinical and epidemiological risk factors. The test is intended for use as an aid in the differential diagnosis of influenza A and influenza B in humans and is not intended to detect influenza C.<br><br>Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. |
| Parameter | Alere™ i Influenza A & B | IQuum Liat ™ Influenza A/B Assay (K111387) |
| | epidemiological screening criteria<br>recommended by public health<br>authorities, specimens should be<br>collected with appropriate<br>infection control precautions for<br>novel virulent Influenza viruses<br>and sent to state or local health<br>department for testing. Viral<br>culture should not be attempted in<br>these cases unless a BSL 3+ facility<br>is available to receive and culture<br>specimens. | |
| Intended Environment for<br>Use | Professional use, in a medical<br>laboratory or point-of-care | Professional use, in a medical laboratory |
| Instrumentation | Alere™ i Instrument | Liat ™ Analyzer |
| Self-Contained System | Integrated PC, Software, and Touch<br>Screen Display | Same |
| Automated Assay | Yes. Sample preparation,<br>amplification, detection, and result<br>interpretation. | Same |
| Assay Information | | |
| Sample Type | Nasal Swab | Nasopharyngeal Swab |
| Influenza A Viral Target | PB2 segment | Matrix Gene |
| Influenza B Viral Target | PA segment | Non-Structural Protein (NSP) Gene |
| Technology | Isothermal nucleic acid<br>amplification for detecting the<br>presence/absence of viral RNA in<br>clinical specimens | RT-PCR for detecting the presence/absence of<br>viral RNA in clinical specimens |
| Detection Method | Assay uses different reporter dyes<br>for each target | Multiplex assay using different reporter dyes for<br>each target |
| Internal Control | Yes | Yes |
| Result Interpretation | Automated | Same |
| Assay Result | Qualitative | Same |
| Time to Result | < 15 minutes | ~ 20 minutes |
{3}------------------------------------------------
#### PERFORMANCE SUMMARY
#### CLINICAL STUDY
Clinical performance characteristics of Alere™ i Influenza A & B were evaluated in a multi-site prospective study during the 2012-2013 flu season in the U.S. A total of eight investigational sites throughout the U.S. participated in the study. To be enrolled in the study, patients had to be presenting at the participating study centers with flu-like symptoms. Direct nasal swab specimens with flu-like symptoms were collected and tested using the Alere™ i Influenza A & B at the eight study sites. Viral culture performed according to standard virology culture procedures, was utilized as the reference method for this study.
Two nasal swabs were collected from one nostril from each subject using standard collection methods. At all sites, one nasal swab was tested directly on Alere™ i Influenza A & B, according to product instructions. The other nasal swab was eluted in 3-mLof viral transport media (VTM). Six of the eight sites (Site 1, Site 4, Site 8, Site 10, Site 11, and Site 12) shipped nasal swab samples in VTM to a central testing laboratory for viral
{4}------------------------------------------------
culture testing. This central testing laboratory was located at Site 1, which also participated as a sample collection and Alere™ i Influenza A & B testing site. The nasal swab samples in VTM from Site 2 and Site 9 were cultured on site by a local laboratory.
External control testing, using Alere™ i Influenza A & B Positive and Negative Controls, was performed prior to sample testing each day and on each Alere™ i instrument the testing was performed, at all study sites.
All specimens generating discrepant Alere™ i Influenza A & B and viral culture results were investigated by testing using an FDA-cleared Influenza RT-PCR assay at a central testing laboratory located at Site 1.
A total of 612 nasal swab specimens were enrolled in this study. Of those, 27 nasal swab specimens did not meet eligibility criteria. A total of 585 direct nasal swab specimens were considered evaluable. Patient age and gender distribution for all the evaluable specimens is presented in the table below.
| Age Group ( ) ================================================================================================================================================================ | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------|------|
| <1 year | ટેર્ણ | ਟੇਰੇ |
| 1 to 5 years | 108 | 117 |
| 6 to 10 years | દર્દ | 47 |
| 11 to 15 years | 21 | 20 |
| 16 to 21 years | 14 | 8 |
| >21 to 60 years | 51 | 20 |
| >60 years | 5 | पे |
| Total | 310 | 275 |
# Age and Gender Distribution - Direct Nasal Swab Study
Of the evaluable 585 specimens, Alere™ i Influenza A & B generated influenza A invalid results for 14 specimens and influenza B invalid results for 16 specimens, resulting in a total of 571 specimens for influenza A performance analysis and 569 specimens for influenza B performance analysis.
Compared to the viral culture reference method, the performance of Alere™ i Influenza A and influenza B are presented in the two tables below.
#### Alere™ i Influenza A & B Influenza A Direct Nasal Swab Performance against Viral Culture
| AlereTM Influenza A & B - Flu A | Positive | Negative | Total |
|---------------------------------|----------|----------------------|-------|
| Positive | 92 | 66* | 158 |
| Negative | 2b | 411 | 413 |
| Total | 94 | 477 | 571 |
| Sensitivity: 92/94 | 97.9% | (95%CI: 92.6%-99.4%) | |
| Specificity: 411/477 | 86.2% | (95%CI: 82.8%-89.0%) | |
ª Flu A nucleic acid was detected in 58/66 False Positive specimens using an FDA-cleared molecular test
b Flu A nucleic acid was not detected in 1/2 False Negative specimens using an FDA-cleared molecular test
Alere™ i Influenza A & B Influenza B Direct Nasal Swab Performance against Viral Culture
| Alere™ i<br>'Influenza A &<br>B - Flu B | Positive | Negative | Total |
|-----------------------------------------|----------|----------------------------|-------|
| Positive | 74 | 17 | 91 |
| Negative | 6 | 472 | 478 |
| Total | 80 | 489 | 569 |
| Sensitivity: 74/80 | 92.5% | (95%CI: 84.6%-96.5%) | |
| Specificity: 472//489 | | 96.5% (95%CI: 94.5%-97.8%) | |
{5}------------------------------------------------
ª Flu B nucleic acid was detected in 15/17 False Positive specimens using an FDA-cleared molecular test
6 Flu B nucleic acid was not detected in 4/6 False Negative specimens using an FDA-cleared molecular test
Performance of Alere™ i Influenza A & B for the detection of influenza B versus culture is presented in the table below stratified by patient age.
Nasal Swab Performance Obtained for Influenza A and Influenza B with the Alere™ i Influenza A & B in Comparison to Viral Culture - Stratified by Patient Age
| Influenza<br>Type | ≤ 5 Years of Age<br>(n = 332) | | 6 - ≤ 21 Years of Age<br>(n = 162) | | ≥ 22 Years of Age<br>(n = 77) | |
|-------------------|-----------------------------------|-------------------------------------|------------------------------------|-------------------------------------|---------------------------------|-----------------------------------|
| | Sensitivity<br>95% CI | Specificity<br>95% CI | Sensitivity<br>95% CI | Specificity<br>95% CI | Sensitivity<br>95% CI | Specificity<br>95% CI |
| Flu A | 98.3%<br>(58/59)<br>91.0% - 99.7% | 89.0%<br>(243/273)<br>84.7% - 92.2% | 100%<br>(31/31)<br>89.0% - 100% | 85.5%<br>(112/131)<br>78.5% - 90.5% | 75.0%<br>(3/4)<br>30.1% - 95.4% | 76.7%<br>(56/73)<br>65.8% - 84.9% |
| Flu B | 88.9%<br>(32/36)<br>74.7% - 95.6% | 98.0%<br>(288/294)<br>95.6% - 99.1% | 94.4%<br>(34/36)<br>81.9% - 98.5% | 96.8%<br>(122/126)<br>92.1% - 98.8% | 100%<br>(8/8)<br>67.6% - 100% | 89.9%<br>(62/69)<br>80.5% - 95.0% |
Alere™ i Influenza A & B detected one mixed influenza A and B infection in the prospective clinical ' evaluation. This sample tested positive for influenza B only by viral culture, but tested positive for influenza A only by an FDA cleared Influenza RT-PCR assay.
During the prospective clinical study, the initial invalid rate {before repeat testing per the product instructions) was 5.8% (34/585) (95% Cl: 4.2% to 8.0%) for Flu A, and 3.6% (21/585) (95% Cl: 2.4% to 5.4%) for Flu B. After repeat testing per the product instructions, the invalid rate was 2.4% (14/585) (95% CI: 1.4%, 4.0%) for Flu A, and 2.7% (16/585) (95% CI: 1.7%, 4.4%) for Flu B.
# ANALYTICAL STUDIES
### ANALYTICAL SENSITIVITY
Alere™ i Influenza A & B limit of detection (LoD) in natural nasal swab matrix was determined by evaluating different concentrations of 3 strains of influenza A and 2 strains of influenza B virus in Alere™ i Influenza A & B. Three strains of influenza A virus representing each of the three common currently or recently circulating influenza A subtypes (i.e., A/H1N1, A/H3N2 seasonal, and A/H1N1 pandemic (pdm)) and two strains of influenza B virus representing each of the two influenza B genetic lineages (i.e., Victoria and Yamagata) were included in this study.
Presumed negative natural nasal swab specimens were eluted in UTM. Swab elutes were combined and mixed thoroughly to create a clinical matrix pool to be used as the diluent. Each influenza virus strain was diluted in this natural nasal swab matrix pool to generate virus dilutions for testing. The vender provided virus strains were re-titered and the concentrations (in TCIDso/mL) were determined by standard virologic method. The concentration for each dilution (in genome equivalents/mL) was also assessed using laboratory developed and validated influenza A and influenza B quantitative real-time PCR assays.
Contrived nasal swab specimens were prepared by coating 10 microliters of each virus dilution onto the swab. The contrived swab samples were tested according to product instructions.
The LoD for each influenza strain tested was determined as the lowest virus concentration that was detected ≥ 95% of the time (i.e., concentration at which at least 19 out of 20 replicates tested positive).
{6}------------------------------------------------
The confirmed LoDs in natural nasal swab matrix for each influenza strain tested are presented in the table below:
| Influenza Strain | Influenza A Subtype or Influenza B Genetic Lineage | LoD (TCID50/mL) | LoD (TCID50/Swab)* | LoD (Genome Equivalents/mL) | LoD (Genome Equivalents/Swab)* |
|------------------------|----------------------------------------------------|-----------------|--------------------|-----------------------------|--------------------------------|
| A/Puerto Rico/8/34 | A/H1N1 | $1.88 x 10^5$ | $1.88 x 10^3$ | $4.22 x 10^6$ | ( $4.22 x 10^4$ |
| A/Perth/16/2009 | A/H3N2 | $8.60 x 10^2$ | $8.60 x 10^0$ | $7.91 x 10^4$ | $7.91 x 10^2$ |
| A/California/7/2009 | A/2009<br>H1N1 pdm | $1.25 x 10^4$ | $1.25 x 10^2$ | $5.20 x 10^6$ | $5.20 x 10^4$ |
| B/Malaysia/2506/2004 | B Victoria<br>lineage | $1.90 x 10^3$ | $1.90 x 10^1$ | $7.24 x 10^4$ | $7.24 x 10^2$ |
| B/Bangladesh/3333/2007 | B Yamagata<br>lineage | $5.55 x 10^2$ | $5.55 x 10^0$ | $7.36 x 10^4$ | $7.36 x 10^2$ |
Limit of Detection (LoD) Study Results - Natural Nasal Swab Matrix
*Note: 10 ul of each virus dilution was coated onto a swab
#### REACTIVITY TESTING
An analytical reactivity (inclusivity) study was performed to determine whether the Alere™ i Influenza A & B assay is able to detect a variety of influenza A and B strains that represent temporal and geographic diversity.
Vender provided stocks of influenza A and B strains were diluted in UTM to generate virus dilutions for testing. The concentration (in TCIDs0/mL, or EID50/mL) for each strain was determined by standard virologic method. The concentration for each dilution (in genome equivalents/ml.) was also assessed using laboratory developed and validated influenza B quantitative real-time PCR assays.
Contrived swab specimens were prepared by coating 10 microliters of virus dilution onto each swab. The contrived swab samples were tested according to product instructions.
The starting dilution concentration selected for testing in this study was higher than the established LoDs in the Limit of Detection study. Each starting dilution per virus strain was tested in triplicates initially. If the initial testing concentration tested positive for all three replicates, the strain was further diluted 10-fold and tested in triplicates until at least one out three replicates generated a negative result. When a negative result was obtained, additional 2-fold dilutions were tested, starting from the highest dilution that produced 100% (3/3) positive results. A concentration level was considered "reactive/positive" in this study for all but one strain tested (i.e., B/Texas/06/2011 – see footnote "c" under the table below) if all three replicates generated a positive result for the expected influenza virus.
The Alere™ i Influenza A & B assay detected all strains tested at the concentrations indicated in the table below:
| Influenza Strain | Influenza A<br>Subtype or<br>Influenza B<br>Genetic<br>Lineage | Test Concentration (in TCID50 or Genome Equivalents,<br>unless indicated otherwise) | | | | Flu A<br>Result<br>(n=3,<br>unless<br>indicated<br>otherwise<br>) | Flu B<br>Result<br>(n=3,<br>unless<br>indicated<br>otherwise<br>) |
|--------------------------|----------------------------------------------------------------|-------------------------------------------------------------------------------------|------------------|------------------------------|-----------------------------|-------------------------------------------------------------------|-------------------------------------------------------------------|
| | | TCID50/mL | TCID50/Swab* | Genome<br>Equivalents/<br>mL | Genome<br>Equivalents/Swab* | | |
| A/New Caledonia/20/1999* | A/H1N1 | $9.19 x 10^5$ | $9.19 x 10^3$ | $4.09 x 10^6$ | $4.09 x 10^4$ | + | - |
| A/New Jersey/8/76* | A/H1N1 | $3.41 x 10^1$ | $3.41 x 10^{-1}$ | $1.52 x 10^5$ | $1.52 x 10^3$ | + | - |
#### Analytical Reactivity Study Results
{7}------------------------------------------------
| A/Brisbane/59/2007 * | A/H1N1 | $2.11 \times 10^4$ | $2.11 \times 10^2$ | $3.39 \times 10^5$ | $3.39 \times 10^3$ | + | - |
|----------------------------------|------------------------------------------|---------------------------------|--------------------------------------|--------------------|--------------------|---|---|
| A/WSN/33 * | A/H1N1 | $2.11 \times 10^2$ | $2.11 \times 10^0$ | $2.43 \times 10^5$ | $2.43 \times 10^3$ | + | - |
| A/Port Chalmers/1/73 | A/H3N2 | $4.22 \times 10^4$ | $4.22 \times 10^2$ | $1.31 \times 10^6$ | $1.31 \times 10^4$ | + | - |
| A/Hong Kong/8/68 | A/H3N2 | $7.03 \times 10^0$ | $7.03 \times 10^{-2}$ | $2.70 \times 10^5$ | $2.70 \times 10^3$ | + | - |
| A/Aichi/2/68 | A/H3N2 | $2.08 \times 10^5$ | $2.08 \times 10^3$ | $7.47 \times 10^6$ | $7.47 \times 10^4$ | + | - |
| A/Victoria/3/75 | A/H3N2 | $3.68 \times 10^5$ | $3.68 \times 10^3$ | $3.39 \times 10^6$ | $3.39 \times 10^4$ | + | - |
| A/Wisconsin/67/2005 | A/H3N2 | $6.81 \times 10^4$ | $6.81 \times 10^2$ | $2.57 \times 10^6$ | $2.57 \times 10^4$ | + | - |
| A/Brisbane/10/2007 | A/H3N2 | $3.16 \times 10^2$ | $3.16 \times 10^0$ | $3.37 \times 10^5$ | $3.37 \times 10^3$ | + | - |
| A/Texas/50/2012 | A/H3N2 | $2.50 \times 10^0$ | $2.50 \times 10^{-2}$ | $6.35 \times 10^3$ | $6.35 \times 10^1$ | + | - |
| A/Victoria/361/2011 | A/H3N2 | $1.56 \times 10^1$ | $1.56 \times 10^{-1}$ | $3.53 \times 10^5$ | $3.53 \times 10^3$ | + | - |
| A/California/4/2009 | A/H1N1 (pdm) | $1.47 \times 10^4$ | $1.47 \times 10^2$ | $1.07 \times 10^6$ | $1.07 \times 10^4$ | + | - |
| A/Maryland/04/2011 | A/H1N1 (pdm) | $7.88 \times 10^4$ | $7.88 \times 10^2$ | $3.81 \times 10^6$ | $3.81 \times 10^4$ | + | - |
| A/New York/18/2009 | A/H1N1 (pdm) | $1.25 \times 10^2$ | $1.25 \times 10^0$ | $9.16 \times 10^5$ | $9.16 \times 10^3$ | + | - |
| A/Anhui/1/2013<br>(Inactivated)* | A/H7N9<br>(Detected in<br>China in 2013) | $4.00 \times 10^6$<br>EID50/mL | $4.00 \times 10^4$<br>EID50/Swab | $1.72 \times 10^6$ | $1.72 \times 10^4$ | + | - |
| A/Indiana/10/2011* | A/H3N2v | $2.00 \times 10^8$<br>EID50/mL | $2.00 \times 10^6$<br>EID50/Swab | $2.00 \times 10^6$ | $2.00 \times 10^4$ | + | - |
| B/Lee/40 | Victoria Lineage | $5.00 \times 10^1$<br>CEID50/mL | $5.00 \times 10^{-1}$<br>CEID50/Swab | $5.40 \times 10^4$ | $5.40 \times 10^2$ | - | + |
| B/Victoria/504/2000 | Victoria Lineage | $1.19 \times 10^3$ | $1.19 \times 10^1$ | $8.29 \times 10^4$ | $8.29 \times 10^2$ | - | + |
| B/Nevada/03/2011 | Victoria Lineage | $1.75 \times 10^3$ | $1.75 \times 10^1$ | $1.13 \times 10^5$ | $1.13 \times 10^3$ | - | + |
| B/Montana/05/2012 | Victoria Lineage | $9.00 \times 10^1$ | $9.00 \times 10^{-1}$ | $2.55 \times 10^4$ | $2.55 \times 10^2$ | - | + |
| B/Maryland/1/59 | Yamagata<br>Lineage | $8.51 \times 10^2$ | $8.51 \times 10^0$ | $1.13 \times 10^5$ | $1.13 \times 10^3$ | - | + |
| B/Russia/69b | Yamagata<br>Lineage | $4.44 \times 10^1$ | $4.44 \times 10^{-1}$ | $2.96 \times 10^6$ | $2.96 \times 10^4$ | - | + |
| B/Wisconsin/01/2010c | Yamagata<br>Lineage | $3.68 \times 10^4$ | $3.68 \times 10^2$ | $1.16 \times 10^6$ | $1.16 \times 10^4$ | - | + |
| B/Massachusetts/2/2012 | Yamagata<br>Lineage | $6.25 \times 10^1$ | $6.25 \times 10^{-1}$ | $2.28 \times 10^5$ | $2.28 \times 10^3$ | - | + |
| B/Texas/06/2011c | Yamagata | $2.89 \times 10^5$ | $6.25 \times 10^3$ | $2.00 \times 10^6$ | $2.00 \times 10^4$ | - | + |
* Note: 10 ul of each virus dilution was coated onto a swab
" Although this test has been shown to detect A/H/N9 (detected in China in 2013) and A/H3N2v viruses cultured from positive human respiratory specimens tharacterstics of this device with clinical specimens that are positive for the A/H1N1 (pre-2009 pandemic), A/H7N9 (detected in China in 2013) and A/H3N2v viruses have not been established.
* Influenza B/Russia/69 lowest level in which 3/3 replicates were positive is approximately 40 to 150 x the LoD (as comparing to the Genome Equivalents/Syab values generated in the LoD with simulated clinical matrix study testing B/Malaysia/2008, respectively). A polymorphism within segment PA of the Influenza B genome was identified at a position which is 4 nucleotides from the 3 -nd of template 2. This G to A polynorphism results in a C/C (product/template) match to an A/C (product/template) mismatch is determined to be moderately destabilizing, and coupled to its position only 4 nucleotides from the 3 recognition region, its impact on annealing is potentially great. The frequency of this G to A polymorphism is determined to be very low. In analyzing the strains present in the NCBI Influenza Virus Resource database from 2/2005 to 3/2014 (N=986), no strains contained this polymorphism, suggesting that it has not been circulating for an extended period of time.
^ Influenza B/Wisconsin/01/2010 lowest level in which 3/3 replicates were positive is approximately 15 to 60 x the LoD, and Influenza B/Texas/06/2011 lowest level in which at least 1/3 replicates were positive is approximately 25 to 100 x the LoD (as comparing to the Genome Equivalents/Swab values generated in the LoD with simulated clinical matrix study testing B/Malaysia/2008, respectively). A single G to A polymorphism within segment PA of the Influenza B genome was identified at a position which is 5 nucleotides from the 3'-end of the molecular beacon annealing region. The G to A polymorphism results in a C/G match to a C/A mismatch between the molecular beacon and product 1. The C/A mismatch is detabilizing that can significantly reduce assay sensitivity. An assessment of what impact this polymorphism would have on the melting temperature (Tm) of the molecular beacon (product 1 annealing was performed and the results showed a Tm drop from 62.3℃ to 55.6℃, just below the assay running temperature. This suggests that annealing would occur, but at a greatly reduced level, with a concomitivity. The frequency of this G to A polymorphism is found at a frequency of approximately 5% within the NCBI Influenza Virus Resource database covering the time from 2/2005 through 3/2014.
#### ANALYTICAL SPECIFICITY (CROSS-REACTIVITY)
To determine the analytical specificity of Alere™ i Influenza A & B, 53 commensal and pathogenic microorganisms (37 bacteria, 15 viruses and 1 yeast) that may be present in the nasal cavity or nasopharynx were tested. All of the following microorganisms were negative when tested at concentrations.ranging from
{8}------------------------------------------------
108 to 10ª cells/mL, CFU/mL (bacteria), 104 to 10º TCIDs0/mL or CEID50/mL (viruses), and 10º cells/mL (yeast).
#### Bacteria
Acinetobacter calcoaceticus Bacteroides fragilis Bordetella pertussis Chlamydia pneumoniae Corynebacterium diphtheriae Enterococcus faecalis Escherichia coli Gardnerella vaginalis Haemophilus influenzae Klebsiella pneumoniae Lactobacillus casei Lactobacillus plantarum Leaionella pneumophila Listeria monocytogenes Moraxella/Branhamella catarrhalis Mycobacterium avium Mycobacterium intracellulare Mycobacterium tuberculosis Mycoplasma pneumoniae Neisseria gonorrhoeae Neisseria meningitidis Neisseria sicca Neisseria subflava Proteus vulgaris Pseudomonas aeruginosa Serratia marcescens Staphylococcus aureus Staphylococcus epidermidis Streptococcus, Group A Streptococcus, Group B Streptococcus, Group C Streptococcus, Group F Streptococcus, Group G Streptococcus mutans Streptococcus pneumoniae Streptococcus salivarius Streptococcus sanguinis
#### Viruses
- Adenovirus type 1 · Candida albicans Adenovirus type 7 Human Coronavirus OC43 Human Coronavirus 229E Enterovirus/Coxsackievirus B4 Human Cytomegalovirus (CMV) (Herpes V) Epstein Barr Virus Human metapneumovirus Measles (Edmonston) Mumps (Enders) Parainfluenza 1 Parainfluenza 2 Parainfluenza 3 Respiratory Syncytial Virus type B Rhinovirus type 1A
Yeast
#### INTERFERING SUBSTANCES
The following substances, naturally present in respiratory specimens or that may be artificially introduced into the nasal cavity or nasopharynx, were evaluated with Alere™ i Influenza A & B at the concentrations listed below and were found not to affect test performance.
#### Substance
Mucin Whole Blood Sinus Buster Nasal Spray NeoSynephrine Cold & Sinus Extra Strength Spray
# Concentration
20 µg/mL 50 µl/mL 200 µl/mL 200 µl/mL
{9}------------------------------------------------
Zicam Extreme Congestion Relief 4-acetamidophenol Acetylsalicylic acid Albuterol Chlopheniramine Dexamethasone Dextromethorphan Diphenhydramine› Doxylamine Succinate Ephedrine Flunisolide Guaiacol glycerol ether Mupirocin Oxymetazoline Phenylephrine Rebetol Relenza Rimatadine Tamiflu Tobryamycin Triamcinolone
200 µl/mL 200 µg/mL 650 ug/mL 400 ng/mL 145 ng/mL 0.80 mg/mL 1 µl/mL 5 µg/mL 236 ng/mL 237 ng/mL 6.8 ng/mL 3.5 ng/mL 12 mg/mL 0.6 mg/mL 12 mg/mL 4.5 µg/mL 282 ng/mL 282 ng/mL 1.1 µg/mL 2.43 mg/mL 40 µg/mL
{10}------------------------------------------------
#### REPRODUCIBILITY
A reproducibility study of Alere™ i Influenza A & B was conducted by operators from three sites using panels of blind coded specimens containing negative (below the limit of detection), low positive (at the limit of detection), and moderate positive (above the limit of detection) influenza A and B viral samples.
Virus dilutions were prepared using one influenza A strain and one influenza B strain in Universal Transport Medium (UTM). The concentrations of the viral stocks (in TCID50/mL) were determined by standard virologic method prior to inactivation by the venders. The concentration for each dilution (in genome equivalents/mL] was also assessed using laboratory developed and validated influenza B quantitative real-time PCR assays.
Contrived nasal swab specimens were prepared by coating 10 microliters of each virus dilution onto the swab. The contrived swab samples were tested according to product instructions.
Participants tested each sample multiple times on five different days. The percent agreement with expected results for the influenza A moderate positive, and high negative samples were 100% (90/90), 100 % (90/90) and 70% (63/90), respectfully. The percent agreement with expected result for the influenza B moderate positive, low positive, and high negative samples were 100% (90/90), 92% (83/90) and 90% (81/90), respectfully. All of the true negative samples (90) generated negative test results. There were no significant differences observed within run (replicates tested by one operator), between run (five different days), between sites (three sites), or between operators (six operators).
The Reproducibility Study site-to-site qualitative results (agreements with expected results) are presented in the table below:
| and and the first to the simple simple simments of the states of the light | | | | | | | The December 2017 - Overall Percent | | |
|----------------------------------------------------------------------------|-----------------------|-------|-----------------------|-------|--------------------------|-------|-------------------------------------|-----------------------------|--|
| Sample f | Site 1 | | * Site 2 | | Site 3 Site 3 | | | | |
| Category . | Percent.<br>Agreement | | Percent<br>oreement . | | Percent War<br>Agreement | | Agreement an | 1. 17 19 11. 11. 11. 11.11. | |
| HN1<br>Influenza A | 66.7% | 20/30 | 80.0% | 24/30 | 63.3% | 19/30 | 70.0%<br>(63/90) | (59.9%, 78.5%) | |
| LP<br>Influenza A | 100% | 30/30 | 100% | 30/30 | 100% | 30/30 | 100%<br>(90/90) | (95.9%, 100%) | |
| MP<br>Influenza A | 100% | 30/30 | 100% | 30/30 | 100% | 30/30 | 100%<br>(90/90) | (95.9%, 100%) | |
| HNI<br>Influenza B | 86.7% | 26/30 | 100% | 30/30 | 83.3% | 25/30 | 90.0%<br>(81/90) | (82.1%, 94.6%) | |
| LP<br>Influenza B | 93.3% | 28/30 | 86.7% | 26/30 | 96.7% | 29/30 | 92.2%<br>(83/90) | (84.8%, 96.2%) | |
| MP<br>Influenza B | 100% | 30/30 | 100% | 30/30 | 100% | 30/30 | 100%<br>(90/90) | (95.9%,<br>100.0%) | |
| TN | 100% | 30/30 | 100% | 30/30 | 100% | 30/30 | 100%<br>(90/90) | (95.9%, 100%) | |
#### Reproducibility Study Site-To-Site Qualitative Results
1Percent Agreement correlates to the percent of negative results.
Signed _
Date
Angela Drysdale VP. Regulatory & Clinical Affairs – Infectious Disease Alere Scarborough, Inc.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its wings and tail feathers. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002
June 13, 2014
Alere Scarborough, Inc. Angela Drysdale Vice President, Regulatory and Clinical Affairs - Infectious Disease 10 Southgate Road Scarborough, ME 04074
Re: K141520
Trade/Device Name: Alere™ i Influenza A&B Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory Viral Panel Multiplex Nucleic Acid Assay Regulatory Class: II Product Code: OCC. OZE. OOI Dated: June 06, 2014 Received: June 09, 2014
Dear Ms. Drysdale:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{12}--…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.