K141515 · National Advanced Endoscopy Devices, Inc. · HIH · Jan 7, 2015 · Obstetrics/Gynecology
Device Facts
Record ID
K141515
Device Name
AED HYSTEROSCOPE AND ACCESSORIES
Applicant
National Advanced Endoscopy Devices, Inc.
Product Code
HIH · Obstetrics/Gynecology
Decision Date
Jan 7, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
AED Hysteroscope and Accessories are used to permit direct viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures.
Device Story
Reusable rod-lens hysteroscope system; includes various examination and operative sheaths. Device inputs: optical images of cervical canal/uterine cavity via rod-lens system; fiber optic light carrier provides illumination. Physician operates device in clinical setting; views anatomy directly through eyepiece or via connected imaging system. Sheaths facilitate media inflow/outflow for continuous flow visualization; operative sheaths provide instrument channels for surgical tools. Device enables diagnostic assessment and surgical intervention within uterine cavity. Benefits include direct visualization for improved procedural accuracy.
Clinical Evidence
Bench testing only. Performance testing conducted per IEC 60601-2-18, including thermal testing and simulated use testing. Device met all design and performance requirements.
Indicated for patients requiring direct visualization of the cervical canal and uterine cavity for diagnostic or surgical procedures.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
Predicate Devices
Karl Storz Fixed Magnification Telescope/KS Variable Magnification Telescope (K935716)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight, composed of three overlapping profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 7, 2015
National Advanced Endoscopy Devices. Inc. Gayle Butler Compliance Manager 22134 Sherman Way Canoga Park CA 91303
Re: K141515
Trade/Device Name: AED Hysteroscope And Accessories Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope and accessories Regulatory Class: II Product Code: HIH Dated: November 24, 2014 Received: December 4, 2014
Dear Gayle Butler,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -A
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K141515
Device Name AED Hysteroscope and Accessories
### Indications for Use (Describe)
AED Hysteroscope and Accessories are used to permit direct viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY CHILL CHILL CHE CHE CHARRE OF CHE
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
National Advanced Endoscopy Devices, Inc.
Premarket Notification: K141515
# 510(k) Summary of Safety and Effectiveness
| Date: | December 22, 2014 | |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | National Advanced Endoscopy Devices, Inc.<br>22134 Sherman Way<br>Canoga Park, CA 91303<br>Telephone: 818.227.2720<br>Fax: 818.227.2724<br>Contact Person: Gayle M. Butler<br>Compliance Manager | |
| Product: | Trade Name:<br>Classification:<br>Common Name:<br>Classification Name: | AED Hysteroscope and Accessories<br>Class II<br>Hysteroscope and Accessories<br>Hysteroscope (And Accessories)<br>(HIH, 21 CFR 884.1690) |
| Predicate Devices: | Karl Storz Fixed Magnification Telescope/KS Variable<br>Magnification Telescope K935716 | |
| Device<br>Description: | AED Hysteroscopes<br>5705B AED Hysteroscope<br>4.0mm diameter, 30° Angle of View, 36cm length, 30cm working length<br>5729B AED Hysteroscope<br>2.9mm diameter, 30° Angle of View, 36cm length, 30cm working length<br>AED Sheaths<br>Examination Sheaths | |
| | | |
| | | |
| | 18-2305 for use with a 4.0mm diameter, 30° Angle of View Hysteroscope<br>5.0mm diameter, 31cm length, 26cm working length, 1 stopcock<br>18-2306 for use with a 2.9mm diameter, 30° Angle of View Hysteroscope<br>4.0mm diameter, 31cm length, 26cm working length, 1 stopcock | |
| | | 18-2304 Continuous Flow for use with a 2.9mm diameter, 30° Angle of View<br>Hysteroscope<br>Inner Sheath: 3.8mm diameter, 27cm length, 25cm working length, 1 stopcock<br>Outer Sheath: 4.5mm diameter, 24cm length, 18cm working length, 1 stopcock |
{4}------------------------------------------------
## 18-2309 Continuous Flow for use with a 4.0mm diameter, 30° Angle of View Hysteroscope
Inner Sheath: 5.2mm diameter, 27cm length, 25cm working length, 1 stopcock Outer Sheath: 6.2mm diameter, 24cm length, 18cm working length, 1 stopcock
## 18-2313 5.1mm diameter, 1 luer lock adapter
5.1mm diameter, 31cm length, 26cm working length, 1 stopcock with luer adapter
#### 18-2314 5.1mm diameter. 1 luer lock adapter. rotatable
5.1mm diameter. 31cm length. 26cm working length. 1 rotatable stopcock with luer adapter
#### Examination/Operative Sheaths
## 18-2300 for use with a 4.0mm diameter, 30° Angle of View Hysteroscope
1 Instrument Channel and 1 Rotatable Stopcock
5.0mm diameter, 31cm length, 22cm working length, 2.0mm Channel Diameter
#### 18-2302 for use with a 2.9mm diameter, 30° Angle of View Hysteroscope 1 Instrument Channel and 1 Rotatable Stopcock
4.0mm diameter, 31cm length, 22cm working length, 2.0mm Channel Diameter
#### 18-2307 for use with a 4.0mm diameter, 30° Angle of View Hysteroscope
1 Instrument Channel and 2 Stopcocks
5.0mm diameter, 31cm length, 22cm working length, 2.0mm Channel Diameter , Continuous Flow
## 18-2308 for use with a 2.9mm diameter, 30° Angle of View Hysteroscope
1 Instrument Channel and 2 Stopcocks
4.0mm diameter, 31cm length, 22cm working length
2.0mm Channel Diameter, Continuous Flow
The AED Hysteroscope and Accessories is a reusable surgical device that incorporate operative and examination sheaths.
The AED Hysteroscope is a reusable rod lens Hysteroscope consisting of an eyepiece lens and a light post connection for fiber optic light cables. A shaft composed of surgical grade stainless steel encloses the glass rod-lens system and a built-in fiber optic light carrier made of glass/acrylic. The light post body is also comprised of surgical stainless steel. The eyepiece is made from Ultem.
Examination Sheaths are designed to be used with a hysteroscope alone. An operative sheath is an examination sheath with one or more instrument channels. The operative and examination sheaths are composed of surgical grade stainless steel with the exception of a Fixing Ring (where applicable) which is composed of Nylon.
{5}------------------------------------------------
There are two designs for Continuous Flow Sheaths. One design consists of a single sheath with two interior channels (18-2307 and 18-2308) for the inflow and outflow of media to be used with the hysteroscope alone. Alternatively, an inner and outer sheath may be used to perform the same function (18-2304 and 18-2309).
The device is reusable and provided non-sterile. It must be cleaned and sterilized before use.
# Indication for Use:
AED Hysteroscope and Accessories are used to permit direct viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures.
# Comparison of Technological Characteristics:
The AED Hysteroscope and Accessories has the same fundamental technological characteristics as the predicate devices and is substantially equivalent in design, materials and intended use as the predicate devices.
# Karl Storz Fixed Magnification Telescope/KS Variable Magnification Telescope (K935716):
- Same Indications for Use ●
- Utilizes the Same Operating Principle ●
- Same Basic Mechanical Design ●
- Manufactured Using the Same Materials
The AED Hysteroscope and Accessories differ from the predicate device Karl Storz Fixed Maqnification Telescope/KS Variable Magnification Telescope (K935716):
- The KS Variable Magnification Telescope has an ● adjustable ocular lens that can, using an exterior knob, be moved to adjust magnification to 80x. The AED Hysteroscope has a fixed nonadjustable ocular lens.
{6}------------------------------------------------
National Advanced Endoscopy Devices, Inc.
There is no difference between the Karl Storz Fixed Maqnification Telescope and the AED Hysteroscope.
## Performance Testing:
Testing was performed according to IEC 60601-2-18. Thermal testing and simulated use testing were performed to support safety and effectiveness and substantial equivalence to the predicate devices. The AED Hysteroscope and Accessories met all specified design and performance requirements.
# Performance
# Standards:
The device conforms to ISO 8600-4:2014, 8600-1:2013, 8600-3:1997 Amendment 1 2003, 8600-5:2005, 8600-6:2005 and IEC 60601-2-18 (1996) and IEC 60601-2-18 Edition 3.0 2009-08.
- Conclusion: Based on the technical information, intended use and performance information provided in this premarket notification, the AED Hysteroscopes and Accessories have been shown to be substantially equivalent to the current legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.